Visceral Manifold Study for the Repair of Thoracoabdominal Aortic Aneurysms

NCT ID: NCT03113383

Last Updated: 2022-11-08

Study Results

Results available

Outcome measurements, participant flow, baseline characteristics, and adverse events have been published for this study.

View full results

Basic Information

Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.

Recruitment Status

TERMINATED

Clinical Phase

NA

Total Enrollment

4 participants

Study Classification

INTERVENTIONAL

Study Start Date

2017-08-18

Study Completion Date

2021-10-06

Brief Summary

Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.

The primary objective of the clinical investigation is to assess the use of the the Visceral Manifold Thoracoabdominal Aortic Aneurysms (VTAAA) stent graft system to repair thoracoabdominal aortic aneurysms in patients having appropriate anatomy. The primary intent of the study is to assess safety (i.e. freedom from major adverse events (MAE) at 30 days) and preliminary effectiveness (i.e., treatment success and technical success) of the device (i.e., the proportion of treatment group subjects that achieve and maintain treatment success at one year).

Detailed Description

Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.

The primary purpose of this study is to evaluate the safety of this device as there are no or very limited devices and clinical options available for this patient population. The primary safety endpoint of this study is safety freedom from major adverse events (MAE) at 30 days or during hospitalization if this exceeds 30 days. Major adverse events include death, bowel ischemia, myocardial infarction, paraplegia, renal failure, respiratory failure, and stroke. The primary safety endpoint will be analyzed to determine statistical significance when compared to a target performance goal. A literature review of outcomes of open surgical repair was used to create the performance goal as there is not a comparable endovascular option to use for analysis. The performance goal was selected based on the range of subjects experiencing a major adverse event at 30 days. The range was calculated based on assumptions of the minimum and maximum number of subjects experiencing at least one MAE in the historical open surgical repair group . Based on the literature reviewed and the above assumptions the range of subjects experiencing at least one MAE in the open surgical repair group is 30.5% to 77.4%.

The primary effectiveness endpoint is the proportion of the study subjects with treatment success at 1 year. The data will be presented as quality outcomes with the number of study subjects with treatment success compared to the overall patient population.

Additionally, data outcomes from this study will be entered into a common vascular database so that data can be pooled with other PS-IDEs. This would provide consistent reporting across the PS-IDEs. Additionally, the PS-IDEs will be evaluating the same device and endpoints to allow for a pool-ability of data across the sites.

The pooled data will be separated into two separate study arms: primary study arm and expanded selection arm.

Conditions

See the medical conditions and disease areas that this research is targeting or investigating.

Thoracoabdominal Aortic Aneurysm

Study Design

Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.

Allocation Method

NON_RANDOMIZED

Intervention Model

SINGLE_GROUP

Primary Study Purpose

TREATMENT

Blinding Strategy

NONE

Study Groups

Review each arm or cohort in the study, along with the interventions and objectives associated with them.

Primary Arm

Use of the thoracic bifurcation and the visceral manifold to repair thoracoabdominal aortic aneurysms in patients having appropriate anatomy.

Group Type EXPERIMENTAL

Thoracoabdominal Aortic Aneurysm Repair

Intervention Type DEVICE

Thoracoabdominal aortic aneurysm repair

Expanded Selection Arm

Use of the thoracic bifurcation and the visceral manifold to repair thoracoabdominal aortic aneurysms in patients having appropriate anatomy.

Group Type EXPERIMENTAL

Thoracoabdominal Aortic Aneurysm Repair

Intervention Type DEVICE

Thoracoabdominal aortic aneurysm repair

Interventions

Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.

Thoracoabdominal Aortic Aneurysm Repair

Thoracoabdominal aortic aneurysm repair

Intervention Type DEVICE

Eligibility Criteria

Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.

Inclusion Criteria

* A patient may be entered into the study if the patient has at least one of the following:

* An aneurysm with a maximum diameter of greater than or equal to 5.5 cm or 2 times the normal diameter just proximal to the aneurysm using orthogonal (i.e., perpendicular to the centerline) measurements
* Aneurysm with a history of growth greater than or equal to 0.5 cm in 6 months
* Saccular aneurysm deemed at significant risk for rupture
* Symptomatic aneurysm greater than 4.5 cm
* Axillary or brachial and iliac or femoral access vessel morphology that is compatible with vascular access techniques, devices or accessories, with or without use of a surgical conduit
* Proximal landing zone for the thoracic bifurcation stent graft that has:

* Greater than or equal to 2.5 cm of nonaneurysmal aortic segment including previously placed graft material (neck) distal to the left subclavian artery (LSA) diameter in the range of 26-42 mm
* Adequate distance from the celiac artery, in order to accommodate cannulation from the antegrade access point when considering the total deployed length of the thoracic bifurcation and visceral manifold
* Minimum branch vessel diameter greater than 5 mm
* Iliac artery or aortic distal fixation site, including both native tissue and previously placed graft, greater than or equal to 15 mm in length and diameter in the range of 8 - 25 mm
* Age: greater than or equal to 18 years old
* Life expectancy: greater than 1 year


* Patient that meets the criteria for inclusion in the primary study arm but has one or more of the following criteria which would exclude them from the primary study arm:
* Minimum branch vessel diameters \<less than 5mm
* Urgent or emergent presentation
* Patient has a contained rupture
* Patient has a ruptured aneurysm
* Patient has a type B dissection (subacute or chronic) in the portion of the aorta intended to be treated
* Poor performance status including two major system failures (including but not limited to cardiovascular, pulmonary, renal, hepatobiliary, and neuromuscular)
* Baseline creatinine greater than greater than or equal to 2.0 mg/dL
* Anatomy that does not allow for maintenance of at least one hypogastric artery
* Anatomy that does not allow primary or assisted patency of the left subclavian artery
* Prior aneurysm repair that would involve relining of the previously placed graft material requiring placement of the investigational system in a landing zone that expands beyond any limits of the previously placed graft material
* Obstructive stenting of any or all of the visceral vessels

Or

* A patient that meets the criteria for inclusion into the primary study arm and:
* Would not be eligible for the primary study arm per a documented reason other than those outlined above, and
* Per the opinion of the Principal Investigator, with concurrence of the IRB, alternatives therapies are unsatisfactory and the probable risk of using the investigational device is no greater than the probable risk from the disease or condition.

These will be determined by standard of care assessments of their disease/aneurysm morphology by the treating investigator.

Exclusion Criteria

* Patient is a good candidate for and elects for open surgical repair
* Can be treated in accordance with the instructions for use with a legally marketed endovascular device
* Is eligible for enrollment in a manufacturer-sponsored IDE at the investigational site
* Unwilling to comply with the follow-up schedule
* Inability or refusal to give informed consent
* Urgent or emergent presentation
* Patient is pregnant or breastfeeding
* Patient has a contained rupture
* Patient has a ruptured aneurysm
* Patient has a dissection in the portion of the aorta intended to be treated
* Obstructive stenting of any or all of the visceral vessels
* Poor performance status including two major system failures (including but not limited to cardiovascular, pulmonary, renal, hepatobiliary, and neuromuscular)


* Known sensitivities or allergies to the materials of construction of the devices, including nitinol (Nickel: Titanium) polyester, platinum-iridium, polytetrafluoroethylene (PTFE), platinum, gold, polyethylene, or stainless steel.
* Known hypersensitivity or contraindication to anticoagulation or contrast media that cannot be adequately medically managed
* Uncorrectable coagulopathy
* Body habitus that would inhibit x-ray visualization of the aorta or exceeds the safe capacity of the equipment
* Patient has had a major surgical or interventional procedure unrelated to the treatment of the aneurysm planned less than 30 days of the endovascular repair
* Unstable angina (defined as angina with a progressive increase in symptoms, new onset at rest or nocturnal angina)
* Systemic or local infection that may increase the risk of endovascular graft infection
* Baseline creatinine greater than or equal to 2.0 mg/dL
* History of connective tissue disorders (e.g., Marfan Syndrome in the primary arm only, not applicable for patients in the the expanded access arm, Ehler's Danlos Syndrome)
* Prior aneurysm repair that would involve relining of the previously placed graft material requiring placement of the investigational system in a landing zone that expands beyond any limits of the previously placed graft material


* Minimum branch vessel diameter less than 5 mm
* Thrombus or excessive calcification in the proximal aortic neck
* Anatomy that would not allow maintenance of at least one patent hypogastric artery
* Anatomy that would not allow primary or assisted patency of the left subclavian artery
Minimum Eligible Age

18 Years

Eligible Sex

ALL

Accepts Healthy Volunteers

No

Sponsors

Meet the organizations funding or collaborating on the study and learn about their roles.

Medtronic

INDUSTRY

Sponsor Role collaborator

Vanderbilt University Medical Center

OTHER

Sponsor Role lead

Responsible Party

Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.

Thomas Naslund

Professor Vascular Surgery

Responsibility Role PRINCIPAL_INVESTIGATOR

Principal Investigators

Learn about the lead researchers overseeing the trial and their institutional affiliations.

Thomas C Naslund, MD

Role: PRINCIPAL_INVESTIGATOR

Chief, Division of Vascular Surgery

Locations

Explore where the study is taking place and check the recruitment status at each participating site.

Vanderbilt University Medical Center

Nashville, Tennessee, United States

Site Status

Countries

Review the countries where the study has at least one active or historical site.

United States

Provided Documents

Download supplemental materials such as informed consent forms, study protocols, or participant manuals.

Document Type: Study Protocol and Statistical Analysis Plan

View Document

Other Identifiers

Review additional registry numbers or institutional identifiers associated with this trial.

161627

Identifier Type: -

Identifier Source: org_study_id

More Related Trials

Additional clinical trials that may be relevant based on similarity analysis.

The (PIVOTAL) Study
NCT00444821 TERMINATED NA
Endovascular Treatment of Thoracic Aortic Disease
NCT00583817 ENROLLING_BY_INVITATION NA