Qualitative and Quantitative Endothelium Changes After Cataract Surgery: Ultrasound Phacoemulsification vs Nanolaser Technique
NCT ID: NCT05886283
Last Updated: 2023-06-05
Study Results
The study team has not published outcome measurements, participant flow, or safety data for this trial yet. Check back later for updates.
Basic Information
Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.
COMPLETED
NA
74 participants
INTERVENTIONAL
2017-03-01
2020-12-30
Brief Summary
Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.
Setting: Department of ophthalmology, Nabeul, Tunisia. Design: Prospective cohort study. Methods: The study included eyes with nuclear cataract density grade 1, 2, 3, or 4 according to LOCS III, divided into two groups; Group 1 had conventional US, and group 2 had nanosecond laser. The Endothelial Cell Density (ECD), coefficient of variation (CoV) in cell size, percentage of hexagonal cells, central corneal thickness and best visual acuity were evaluated over 24 months.
Related Clinical Trials
Explore similar clinical trials based on study characteristics and research focus.
Endothelial Cell Loss After Cataract Surgery
NCT00932191
Effect of Minimising Ultrasound Power to 1% During Cataract Surgery on Corneal Endothelium.
NCT01259349
Ultrasonic System for Cataract Characterization and Optimal Phacoemulsification Energy Evaluation
NCT04461912
Correlation Between Postoperative Central Corneal Thickness and Endothelial Damage After Cataract Surgery
NCT03972670
Correlation Analysis of Lens Nucleus Density and Surgical Parameters Based on Ss-oct and Scheimpflug Technology
NCT05028530
Detailed Description
Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.
A randomized prospective, uncontrolled cohort study was conducted at the Department of Ophthalmology of the Mohamed Taher Maamouri Hospital (MTMH) from March 2017 to May 2020. The MTMH ethics committee approved this study. All patients provided informed consent for the use of their data. The study adhered to all tenets of the declaration of Helsinki.
All patients underwent complete ocular examination including visual acuity, clinical corneal assessment, intraocular pressure (IOP) measurement, and nuclear hardness grading based on the lens opacity classification system III (LOCS III), commonly referred to in most studies. Only the nuclear component of the cataract (NO1 to NO4) was taken into account in the evaluation.
The endothelium of the central cornea was examined with a non-contact specular microscope (SM) (TOMEY CORP EM-4000).
Patients were randomly divided into 2 groups. Group 1 had US PHACO using the phaco-chop technique (Stellaris PC: Bausch + Lomb®) and group 2 underwent an NL PHACO (Cetus A.R.C. Laser system®).
The Cetus A.R.C. Laser system is composed of a base that generates a 4 to 5 ns pulsed Nd: YAG of 1064 nm wavelength laser with a pulse frequency up to 10 Hz and an optic fiber that transmits the laser pulses to the disposable single-use coaxial handpiece. The probe used has a total diameter of 720 µm, occupied in its center by a 320 µm quartz optical fiber that transferred the laser pulse towards a titanium plate in the phaco probe. The individual pulse energy ranged from 30 to 50 % of maximum power and the pulse frequency was 1-2 Hz.
The base unit is connected to the same phaco aspiration/irrigation system and is controlled by the pedal of that system.
In both groups, surgery started with a clear corneal incision of 2.2 mm at 9 o'clock and a port incision of 1 to 1.5 mm at 2 o'clock using the same knife. Continuous Circular Capsulorhexis (CCC) of 6 mm was performed under the same ophthalmic viscosurgical device (3% Sodium hyaluronate, 4% chondroitin sulfate, DuoVisc® Alcon ). A hydro dissection was performed, then followed by phacoemulsification.
In-group 1 (horizontal phaco chop): after cortex aspiration, the nucleus was held with the phaco tip at a high vacuum. The phaco chopper was then introduced from the side port incision to engage, under the lower edge of CCC. It was then drawn towards the phaco tip to cleave it by a manual separation between the two instruments. The same process continued for the two nuclear halves and the fragments were then aspirated with phaco power. The standard parameters used during phacoemulsification were a vacuum level of 500 mmHg, pressurized irrigation of 90 mmHg, and 40% of phaco power. Energy was expressed in % on the PHACO machine and initially converted into Watts according to the following curve "Figure 1" then into Joules based on the formula: Energy (Joules) = Power (Watts) × time (seconds).
In-group 2: The cortex and epinucleus were aspirated. The nucleus was fragmented using the shockwaves emanating from the phaco tip and then aspirated. The phaco chopper was used to accelerate the mechanical fragmentation of the nucleus and minimize the dissipated energy. The energy used was calculated automatically by the ARC Laser machine and was displayed on the screen.
In both groups, after bimanual infusion/aspiration cortex removal and Visco expansion of the capsular bag, a hydrophobic single-piece: AcrySof acrylic (SA60AT, Alcon) IOL was implanted. The duration of the whole procedure (in minutes) was noted.
Uncorrected Visual acuity (UVA) at 1 day postoperative and BCVA at 1 week, 1, 3, 6, and 24 months after surgery as well as corneal edema and anterior chamber (AC ) reaction were recorded. A visual outcome of 0.5 logarithm of the minimal angle of resolution (logMar) without correction was considered a successful result based on the World Health Organization definition (1, 15, 16). The SM was performed at 1 week, 1, 3, 6, and 24 months after surgery. The investigators reported endothelial cell density (EDC) (cell/ mm²), the percentage of hexagonal cells, coefficient of variation (CoV) in cell size, and central corneal thickness (CCT) at each follow-up visit.
All the data were analyzed using SPSS software (version 21.0; SPSS, Inc., Chicago, IL, USA). BCVA data was converted into log MAR for statistical analyses. Quantitative variables were presented as medians, means, and standard deviations. The investigators used Student's t-test for the independent series. Pearson's correlation coefficient was used for searching the statistical relationship or association between variables. A p-value \< 0.05 was accepted as statistically significant.
Conditions
See the medical conditions and disease areas that this research is targeting or investigating.
Study Design
Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.
RANDOMIZED
PARALLEL
TREATMENT
SINGLE
Study Groups
Review each arm or cohort in the study, along with the interventions and objectives associated with them.
ultrasound phaco groupe
this group had US PHACO using the phaco-chop technique (Stellaris PC: Bausch + Lomb®)
phacoemulsification
we started with a clear corneal incision of 2.2 mm at 9 o'clock and a port incision of 1 at 2 o'clock. Continuous Circular Capsulorhexis of 6 mm was performed under the same ophthalmic viscosurgical device. A hydro dissection was performed, then followed by phacoemulsification.
the Nucleus has been treated using 2 different techniques After bimanual infusion/aspiration cortex removal and Visco expansion of the capsular bag, a hydrophobic single-piece: AcrySof acrylic IOL was implanted.
utrasound phacoemulsification
US group:The nucleus was held with the phaco tip at a high vacuum. The phaco choppe was drawn towards the phaco tip to cleave it by a manual separation between the two instruments. The fragments were then aspirated with phaco power. The standard parameters used during phacoemulsification were a vacuum level of 500 mmHg, pressurized irrigation of 90 mmHg, and 40% of phaco power
nanolaser phaco group
this group underwent an NL PHACO (Cetus A.R.C. Laser system®)
phacoemulsification
we started with a clear corneal incision of 2.2 mm at 9 o'clock and a port incision of 1 at 2 o'clock. Continuous Circular Capsulorhexis of 6 mm was performed under the same ophthalmic viscosurgical device. A hydro dissection was performed, then followed by phacoemulsification.
the Nucleus has been treated using 2 different techniques After bimanual infusion/aspiration cortex removal and Visco expansion of the capsular bag, a hydrophobic single-piece: AcrySof acrylic IOL was implanted.
nanolaser phacoemulsification
Laser group: The nucleus was fragmented using the shockwaves emanating from the phaco tip and then aspirated. The phaco chopper was used to accelerate the mechanical fragmentation of the nucleus and minimize the dissipated energy.
Interventions
Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.
phacoemulsification
we started with a clear corneal incision of 2.2 mm at 9 o'clock and a port incision of 1 at 2 o'clock. Continuous Circular Capsulorhexis of 6 mm was performed under the same ophthalmic viscosurgical device. A hydro dissection was performed, then followed by phacoemulsification.
the Nucleus has been treated using 2 different techniques After bimanual infusion/aspiration cortex removal and Visco expansion of the capsular bag, a hydrophobic single-piece: AcrySof acrylic IOL was implanted.
utrasound phacoemulsification
US group:The nucleus was held with the phaco tip at a high vacuum. The phaco choppe was drawn towards the phaco tip to cleave it by a manual separation between the two instruments. The fragments were then aspirated with phaco power. The standard parameters used during phacoemulsification were a vacuum level of 500 mmHg, pressurized irrigation of 90 mmHg, and 40% of phaco power
nanolaser phacoemulsification
Laser group: The nucleus was fragmented using the shockwaves emanating from the phaco tip and then aspirated. The phaco chopper was used to accelerate the mechanical fragmentation of the nucleus and minimize the dissipated energy.
Eligibility Criteria
Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.
Inclusion Criteria
* Preoperative endothelial cell count (ECC) less than 1500 cells/mm2.
* Pathological alterations of the anterior segment such as corneal opacities, cornea Guttata, uveitis, pseudoexfoliative syndrome, glaucoma, high myopia axil length (≥ 26 mm), or suffering from chronic pathologies that may affect corneal endothelium
Exclusion Criteria
* Surgery was complicated with capsular rent and vitreous loss.
50 Years
90 Years
ALL
Yes
Sponsors
Meet the organizations funding or collaborating on the study and learn about their roles.
University Tunis El Manar
OTHER
Responsible Party
Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.
Walid Zbiba
professor
Locations
Explore where the study is taking place and check the recruitment status at each participating site.
Mohamed Taher Maamouri Hospital
Nabeul, , Tunisia
Countries
Review the countries where the study has at least one active or historical site.
References
Explore related publications, articles, or registry entries linked to this study.
Zbiba W, Kharrat M, Sayadi S, Kallel Z, Marzouk G. Qualitative and quantitative endothelium changes after cataract surgery: ultrasound phacoemulsification vs. nanolaser technique. Front Med (Lausanne). 2023 Sep 21;10:1097404. doi: 10.3389/fmed.2023.1097404. eCollection 2023.
Other Identifiers
Review additional registry numbers or institutional identifiers associated with this trial.
14-03-2023
Identifier Type: -
Identifier Source: org_study_id
More Related Trials