Servo Controlled Oxygen Targeting (SCO2T) Study: Masimo vs. Nellcor
NCT ID: NCT04866342
Last Updated: 2024-09-25
Study Results
The study team has not published outcome measurements, participant flow, or safety data for this trial yet. Check back later for updates.
Basic Information
Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.
COMPLETED
NA
25 participants
INTERVENTIONAL
2020-11-27
2023-08-28
Brief Summary
Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.
Keeping oxygen levels stable is usually done by nurses adjusting the oxygen levels by hand (manual control). There is also equipment available that can do this automatically (servo control). It is not known which is best.
Research suggests that different automated devices control oxygen effectively as measured by the readings from their internal oxygen saturation monitoring systems. When compared to free-standing saturation monitors there appears to be variations in measured oxygen levels between devices. This could have important clinical implications.
This study aims to show the different achieved oxygen levels when babies are targeted to a set target range. Babies in the study will have both a saturation monitor and a transcutaneous oxygen monitor at the same time. Both types of monitor have been in long term use in neonatal units.
For a period of 12 hours, each baby will have their oxygen adjusted automatically using two different internal oxygen monitoring technologies (6 hours respectively). The investigators will compare the range of oxygen levels that are seen between the two oxygen saturation monitoring technologies.
The investigators will study babies born at less than 30 weeks gestation, who are at least 2 days old, on nasal high flow and still require added oxygen.
Related Clinical Trials
Explore similar clinical trials based on study characteristics and research focus.
SCO2T (Servo vs. Servo)
NCT04546568
Masimo O3 Regional Oximetry Device in Pediatric Patients Undergoing Cardiac Surgery
NCT04208906
Accuracy of Owlet Smart Sock Sensor 3rd Generation With Profound Hypoxia
NCT04446247
End-Tidal Oxygen for Intubation in the Emergency Department
NCT06578468
Pulse Contour Analysis and Tissue Oxymetry in Changing Vascular Tone
NCT00583687
Detailed Description
Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.
Research suggests that individual servo control devices control oxygen effectively as measured by the readings obtained from their internal SpO2 monitoring system. The device the investigators intend to study is available with two different oximeter monitoring systems. When compared to a separate free-standing SpO2 monitor the devices have a systematic bias in the clinically targeted range. Consequently, this could allow variations between infants in oxygenation large enough to influence important clinical outcome to go unrecognised.
There is a need to determine the achieved SpO2 and TcPO2 distributions associated with the use of different automated control systems as a first step in planning future trials. When this is measured over a small number of hours it is not anticipated that this would have an influence on clinical outcome.
This study is a prospective, single centre, randomised crossover trial of two different internal oximeter monitoring systems in an automated (servo) control device - IntellO2 (Vapotherm, USA) - delivering nasal high flow employing automated oxygen titration. Each infant will act as their own control. Infants born at less than 30 weeks gestation, greater than 48 hour of age and receiving supplementary oxygen will be eligible for inclusion.
The study will be undertaken in the Neonatal Unit at the Simpson Centre for Reproductive Health at the Royal Infirmary of Edinburgh.
Total study time is 12 hours for each infant. Infants will be randomised to commence on either Masimo oximetry or Nellcor oximetry using the Oxygen Assist Module (OAM), IntellO2 Vapotherm device. SpO2 (range 90-95%) will be continuously monitored on a second pulse oximetry probe connected to a bedside multiparameter monitor as per normal standard of care.
Additional monitoring will be carried out as shown below:
1. TcPO2 monitoring
2. FiO2 monitoring
3. Heart rate monitoring (used to validate SpO2 readings)
4. Arterial gas sampling (only if conducted by the direct care team as part of the routine care of the infant; no extra blood samples will be taken as part of the study)
FiO2 will be adjusted by the respiratory support device which has integrated automated oxygen control, set to maintain a SpO2 target range of 90-95%.The IntellO2 device uses Precision Flow technology (IntellO2, Vapotherm, USA). By means of a modified closed-loop algorithm, the devise uses MasimoSET or Nellcor pulse oximetry to target a user-set SpO2 value.
SpO2 readings will be downloaded directly from the multiparameter patient monitor. SpO2 will be measured using a Phillips MX500 multiparameter monitor (Phillips, Germany, CE 0366).TcPO2 will be measured using a SenTec Digital Monitoring System with OxiVent sensor (SenTec AG, Switzerland, European patent No. 1535055, CE 0120). Both monitors are routinely used in clinical practice. Transcutaneous data will be recorded contemporaneously and the site of the transcutaneous probe will be rotated on each infant every 2 hours. Control of sensor temperature and application duration are designed to meet all applicable standards and this monitoring device is used routinely in many neonatal units.
Conditions
See the medical conditions and disease areas that this research is targeting or investigating.
Study Design
Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.
RANDOMIZED
CROSSOVER
TREATMENT
NONE
To remove selection bias the 20 infants will be randomised using sealed, windowless, envelopes. Half will instruct to commence on automated control (Precision Flow technology IntellO2, Vapotherm, USA) using Masimo oximetry technology to target SpO2 to 90-95%. Half will have instructions to commence on automated control (Precision Flow technology IntellO2, Vapotherm, USA) using Nellcor oximetry technology to target SpO2 to 90-95%.
Study Groups
Review each arm or cohort in the study, along with the interventions and objectives associated with them.
Servo control - Masimo oximetry technology (Oxygen Assist Module, IntellO2, Vapotherm)
Automated control of oxygen. The oxygen saturation target range will be set to 90-95% (set to maintain an integral value of 93%) as per standard clinical practice.
Automated oxygen control can be overridden by manual adjustment of oxygen at any time if this is considered necessary to optimise control of oxygenation according to current clinical targets.
Servo control (closed-loop automatic control of the inspiratory fraction of oxygen (FiO2)) - IntellO2 OAM
FiO2 adjustments will be made by the IntellO2 Oxygen Assist Module (OAM) for Precision Flow (IntellO2, Vapotherm, USA). By means of a modified closed-loop algorithm, the devise uses MasimoSET or Nellcor pulse oximetry to target a user-set SpO2 value.
Manual adjustments of the inspired oxygen fraction can additionally be made as per standard care.
Servo control - Nellcor oximetry technology (Oxygen Assist Module, IntellO2, Vapotherm)
Automated control of oxygen. The oxygen saturation target range will be set to 90-95% (set to maintain an integral value of 93%) as per standard clinical practice.
Automated oxygen control can be overridden by manual adjustment of oxygen at any time if this is considered necessary to optimise control of oxygenation according to current clinical targets.
Servo control (closed-loop automatic control of the inspiratory fraction of oxygen (FiO2)) - IntellO2 OAM
FiO2 adjustments will be made by the IntellO2 Oxygen Assist Module (OAM) for Precision Flow (IntellO2, Vapotherm, USA). By means of a modified closed-loop algorithm, the devise uses MasimoSET or Nellcor pulse oximetry to target a user-set SpO2 value.
Manual adjustments of the inspired oxygen fraction can additionally be made as per standard care.
Interventions
Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.
Servo control (closed-loop automatic control of the inspiratory fraction of oxygen (FiO2)) - IntellO2 OAM
FiO2 adjustments will be made by the IntellO2 Oxygen Assist Module (OAM) for Precision Flow (IntellO2, Vapotherm, USA). By means of a modified closed-loop algorithm, the devise uses MasimoSET or Nellcor pulse oximetry to target a user-set SpO2 value.
Manual adjustments of the inspired oxygen fraction can additionally be made as per standard care.
Eligibility Criteria
Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.
Inclusion Criteria
2. Infants greater than 48 hours of age
3. Infants who are receiving supplementary oxygen
4. Person with parental responsibility able to give consent
Exclusion Criteria
2. Clinical condition of an infant would impair accurateTcPO2 measurement (e.g. impaired perfusion or requirement of inotropic or vasopressor support)
3. Parent/person with parental responsibility unable to give informed consent on behalf of the infant
4. Infants born less than 22 weeks gestation
2 Days
ALL
Yes
Sponsors
Meet the organizations funding or collaborating on the study and learn about their roles.
NHS Lothian
OTHER_GOV
University of Edinburgh
OTHER
Responsible Party
Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.
Locations
Explore where the study is taking place and check the recruitment status at each participating site.
The Simpson Centre for Reproductive Health, Royal Infirmary Edinburgh
Edinburgh, City Of Edinburgh, United Kingdom
Countries
Review the countries where the study has at least one active or historical site.
Other Identifiers
Review additional registry numbers or institutional identifiers associated with this trial.
AC20114
Identifier Type: -
Identifier Source: org_study_id