Bone Bioactive Liquid Efficiency in Improving Dental Implant Osteointegration Oral Soft Tissue Hellingand Oral Surgery

NCT ID: NCT06371430

Last Updated: 2024-06-04

Study Results

Results pending

The study team has not published outcome measurements, participant flow, or safety data for this trial yet. Check back later for updates.

Basic Information

Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.

Recruitment Status

COMPLETED

Clinical Phase

NA

Total Enrollment

33 participants

Study Classification

INTERVENTIONAL

Study Start Date

2021-04-14

Study Completion Date

2024-04-14

Brief Summary

Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.

The investigators implanted 264 Galaxy TS implants (with or without BBL) in 33 patients across both jaws. Implant stability and osseointegration were assessed using Initial Stability Quotient (ISQ) measurements, computed tomography (CBCT) scans and pain evaluations at various intervals post-surgery. Further, implant surfaces were examined using scanning electron microscopy (SEM) and atomic force microscopy (AFM). In vitro studies evaluated the efficacy of BBL on dental pulp pluripotent stem cells (DPPSCs) osteogenesis, and inflammatory factor modulation in human macrophages.

Detailed Description

Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.

Ensuring implant stability and longevity is pivotal for patient satisfaction and quality of life. Our recent discovery, BBL, novel bone bioactive liquid, demonstrated promising properties in improving implant surfaces, alleviating pain, and expediting oral healing. In this multi-center randomized, double-blind clinical trial, investigators sought to evaluate the impact of BBL on the clinical performance of Galaxy TS implants.

Methods Investigators implanted 264 Galaxy TS implants (with or without BBL) in 33 patients across both jaws. Implant stability and osseointegration were assessed using Initial Stability Quotient (ISQ) measurements, computed tomography (CBCT) scans and pain evaluations at various intervals post-surgery. Further, implant surfaces were examined using scanning electron microscopy (SEM) and atomic force microscopy (AFM). In vitro studies evaluated the efficacy of BBL on dental pulp pluripotent stem cells (DPPSCs) osteogenesis, and inflammatory factor modulation in human macrophages.

Conditions

See the medical conditions and disease areas that this research is targeting or investigating.

Wound Healing Disorder Wounds and Injuries Wound Surgical

Study Design

Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.

Allocation Method

RANDOMIZED

Intervention Model

FACTORIAL

Primary Study Purpose

TREATMENT

Blinding Strategy

QUADRUPLE

Participants Caregivers Investigators Outcome Assessors

Study Groups

Review each arm or cohort in the study, along with the interventions and objectives associated with them.

Pain Scale

To assess the efficacy of the surgical procedure, pain scale evaluations were conducted on day 1,4 and 7, following the surgery. This evaluation involved a direct contact with the patients to gauge their subjective pain experience. A modified visual analog scale (VAS) was utilized, as described in reference \[34\]. No pain scale = 0, moderate pain scale = 5 and maximum pain scale = 10 \[21\]. In addition to efficacy measurements, safety measurements were also considered. The occurrence of adverse events (AEs) and serious adverse events (SAEs) was monitored throughout the duration of the study. These events were either detected by the investigator or reported by the patients themselves. The aim was to identify any potential complications or issues that arose because of the surgical procedure.

Group Type PLACEBO_COMPARATOR

Dental Implant insertion and tooth extraction

Intervention Type DEVICE

The osteotomies were carried out at a speed of 800 rpm, using sequential burs of similar diameter for both the maxilar and the mandibular side. The drilling protocols followed the recommendations provided by the manufacturer, the Galaxy-Ziacom implant system, for standard drilling.

The insertion of the implants was initiated using a motor handpiece at a speed of 20-50 rpm, without irrigation. The installation process was completed using a manual surgical torque wrench indicator. The maximum torque value (Ncm) reached at the end of implant insertion was recorded as the insertion torque (IT). A total of 264 implants were inserted in 33 patients, 160 implants were post-extraction and 104 implants were inserted after at least one-month post-extraction. The sutures were typically removed 1 week after the surgery, during the postoperative follow-up appointment.

RFA and ISQ measurements

After the implants were fully seated, a Smartpeg specific to the implant system and restorative platform diameter was used for each implant. RFA was then performed using an OsstellMentor device from Ostell/Integration Diagnostics in Gothenburg, Sweden. This analysis was used to measure the ISQ values for all implant surfaces at days 0, 7 and 30 post-surgeries. For each implant, 4 readings were taken per lingual, mesial, distal, and vestibular direction. The average ISQ values of all the readings were recorded. Following the RFA measurements, new sterile healing abutments were inserted, and the incisions were sutured to close the wounds.

Group Type PLACEBO_COMPARATOR

Dental Implant insertion and tooth extraction

Intervention Type DEVICE

The osteotomies were carried out at a speed of 800 rpm, using sequential burs of similar diameter for both the maxilar and the mandibular side. The drilling protocols followed the recommendations provided by the manufacturer, the Galaxy-Ziacom implant system, for standard drilling.

The insertion of the implants was initiated using a motor handpiece at a speed of 20-50 rpm, without irrigation. The installation process was completed using a manual surgical torque wrench indicator. The maximum torque value (Ncm) reached at the end of implant insertion was recorded as the insertion torque (IT). A total of 264 implants were inserted in 33 patients, 160 implants were post-extraction and 104 implants were inserted after at least one-month post-extraction. The sutures were typically removed 1 week after the surgery, during the postoperative follow-up appointment.

CBCT analysis

CBCTs were captured for implants from six patients with total of 24 implants, with 12 implants Galaxy TS (non-treated) on one side of the jaw and 12 implants Galaxy TSA (treated with BBL) on the other side of the jaw. All patients in this group had implants inserted in the upper jaw and each patient had 4 implants inserted (All on 4 technique). The CBCTs scans were conducted at days 1 and 60 post-surgeries using a 3D Accuitomo 170 scanner (J. MORITA EUROPE GMBH). The scans were performed with the following technical parameters: 90 kV, 5 mA, 87.5 mAs, voxel size ranging from 125-250 µm, 360° rotation, and scanning time of 17.5 s as previously described \[32\]. To ensure accuracy, a bite mark was utilized during CBCT scanning to correspond to the patients' midline. Consistency in bite mark placement was maintained to ensure that the same slice was obtained for each scan performed on the patient

Group Type EXPERIMENTAL

Dental Implant insertion and tooth extraction

Intervention Type DEVICE

The osteotomies were carried out at a speed of 800 rpm, using sequential burs of similar diameter for both the maxilar and the mandibular side. The drilling protocols followed the recommendations provided by the manufacturer, the Galaxy-Ziacom implant system, for standard drilling.

The insertion of the implants was initiated using a motor handpiece at a speed of 20-50 rpm, without irrigation. The installation process was completed using a manual surgical torque wrench indicator. The maximum torque value (Ncm) reached at the end of implant insertion was recorded as the insertion torque (IT). A total of 264 implants were inserted in 33 patients, 160 implants were post-extraction and 104 implants were inserted after at least one-month post-extraction. The sutures were typically removed 1 week after the surgery, during the postoperative follow-up appointment.

Interventions

Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.

Dental Implant insertion and tooth extraction

The osteotomies were carried out at a speed of 800 rpm, using sequential burs of similar diameter for both the maxilar and the mandibular side. The drilling protocols followed the recommendations provided by the manufacturer, the Galaxy-Ziacom implant system, for standard drilling.

The insertion of the implants was initiated using a motor handpiece at a speed of 20-50 rpm, without irrigation. The installation process was completed using a manual surgical torque wrench indicator. The maximum torque value (Ncm) reached at the end of implant insertion was recorded as the insertion torque (IT). A total of 264 implants were inserted in 33 patients, 160 implants were post-extraction and 104 implants were inserted after at least one-month post-extraction. The sutures were typically removed 1 week after the surgery, during the postoperative follow-up appointment.

Intervention Type DEVICE

Eligibility Criteria

Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.

Inclusion Criteria

* Age 18 years or older.
* sufficient residual bone volume to accommodate implant placement without requiring bone augmentation, minimum ridge height and width of ≥ 9mm and ≥ 6mm, respectively, and healed bone sites with a minimum of 1 months of post-extraction healing.

Exclusion Criteria

* Patients with alcoholism.
* Smoking habits.
* History of illicit drug use were excluded.
* Patients with heart diseases.
* Diabetes.
* Previous bone regenerative or augmentation procedures.
* Bleeding disorders.
* Compromised immune systems.
* History of radiation therapy
* Treatment with steroids or bisphosphonates
Minimum Eligible Age

18 Years

Eligible Sex

ALL

Accepts Healthy Volunteers

Yes

Sponsors

Meet the organizations funding or collaborating on the study and learn about their roles.

Universidad Pública de Navarra

OTHER

Sponsor Role collaborator

Dasman Diabetes Institute

OTHER

Sponsor Role collaborator

Biointelligent Technology Systems SL

INDUSTRY

Sponsor Role lead

Responsible Party

Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.

Maher Atari Abouasi

Director

Responsibility Role PRINCIPAL_INVESTIGATOR

Locations

Explore where the study is taking place and check the recruitment status at each participating site.

Biointelligent Technology Syst

Barcelona, , Spain

Site Status

Countries

Review the countries where the study has at least one active or historical site.

Spain

Other Identifiers

Review additional registry numbers or institutional identifiers associated with this trial.

Biointelligent Technology Syst

Identifier Type: -

Identifier Source: org_study_id

More Related Trials

Additional clinical trials that may be relevant based on similarity analysis.

Osteoimplant and Dental Implants Stability
NCT06657521 NOT_YET_RECRUITING EARLY_PHASE1