Version Testing of EnzySystem Version A for Hemophilia A
NCT ID: NCT06273865
Last Updated: 2025-08-15
Study Results
The study team has not published outcome measurements, participant flow, or safety data for this trial yet. Check back later for updates.
Basic Information
Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.
SUSPENDED
40 participants
OBSERVATIONAL
2024-05-29
2026-05-01
Brief Summary
Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.
To measure blood clotting, blood is taken from a vein. This blood is processed in the laboratory and then tested. A new device has been developed that requires only a very small volume of blood (5-10 drops of blood) to perform the laboratory tests. The long-term goal is that this device can be used by a doctor or at home to quickly measure blood clotting. In this research we want to compare this new system with the standard methods - measurement in the laboratory - and evaluate whether the correct value is determined.
Objective of the study:
The primary objective of this study is to demonstrate that the EnzySystem HemA version A can record thrombin generation TG and quantify FVIII activity levels within a time frame of 60 min in fresh whole blood samples of healthy volunteers and patients with hemophilia A in the Enzyre laboratory (for healthy volunteers) and the Radboudumc (for patients with hemophilia A).
Study design:
This is a cross-sectional observational study. All participants are asked to fill a questionnaire prior to blood collection. The blood of healthy volunteers will be collected in an office of Enzyre BV, the blood of patients will be collected in the Radboudumc. Blood collection, by venepuncture, will be conducted by a Radboudumc research nurse or physician of the research team in both locations. In total, four blood tubes with citrate as anticoagulant will be drawn (a total of around 11 mL).
Study population:
The study population consists of 20 healthy volunteers: evenly distributed between male and female; ages spread over the range from 20 to 70 years old; recruited by Enzyre via advertisement. 20 Patients: 5 severe hemophilia A; 5 moderate hemophilia A; 10 mild hemophilia A; recruited from the Hemophilia Treatment Center (HTC) Nijmegen-Eindhoven-Maastricht (NEM) (location Radboudumc).
Primary study parameters/outcome of the study:
Demonstrate that the EnzySystem HemA version A can record TG and quantitative FVIII activity levels within a time frame of 60 min in fresh blood samples of healthy volunteers and patients with hemophilia A.
Secondary study parameters/outcome of the study (if applicable):
Secondary study parameters are composed whether the measured values comply with the desired assay specificity and accuracy.
Outcomes are analysed for equivalence compared to one-stage FVIII assay, FVIII chromogenic assay, thrombin generation via the Nijmegen Hemostasis Assay, and possibly via the Technoclone assay.
Is it possible to measure FVIII activity with the EnzySystem HemA version A in fresh blood samples of healthy volunteers and patients with hemophilia A, compared to the gold standard with;
* Precision in the normal range (60-140%): min. 30%
* Precision in the low range (3-10%): min. 50%
* Limit of Detection range min. 100 % FVIII activity
* Limit of Detection low range min. 3 % FVIII activity Is it possible to measure TG with the EnzySystem HemA version A in fresh blood samples of healthy volunteers and patients with hemophilia A, compared to the gold standard with;
* Precision in the normal range (60-140%) of control samples: min. 30%
* Precision in patient with hemophilia A: min. 50%
* Limit of Detection, high range \> 400 nM thrombin activity
* Limit of Detection measured with Plasma, low range \< 50 nM thrombin
Other study parameters All samples will also be tested for other hemostasis specific parameters as these parameters may affect a proper measurement of both FVIII activity and Thrombin Generation. The following parameters will be measured in plasma obtained from the whole blood vacutainers. Moreover, left over samples (plasma) will eventually be used to develop other coagulation related parameters.
* von Willebrand Factor antigen levels
* von Willebrand Factor ristocetin activity levels
* Prothrombin Fragment 1+2 levels
* ADAMTS13 activity
* FVIII antigen levels
* blood group
Related Clinical Trials
Explore similar clinical trials based on study characteristics and research focus.
Real-life Clinical FEIBA Samples Measured Using the Version A of the HemA EnzySystem
NCT06357572
Home Monitoring in Hemophilia a
NCT06702579
Case and Use Scenario Study to Gain Knowledge on the User Needs for a Medical Device for Hemophilia a Monitoring
NCT06369740
Pharmacokinetic-guided Dosing of Emicizumab
NCT06320626
Emicizumab in Acquired Hemophilia A
NCT04188639
Detailed Description
Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.
Conditions
See the medical conditions and disease areas that this research is targeting or investigating.
Study Design
Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.
COHORT
CROSS_SECTIONAL
Study Groups
Review each arm or cohort in the study, along with the interventions and objectives associated with them.
Healthy volunteers
Several assays
Standard blood draw by venipuncture is performed in all study participants to collect blood samples in four 2.7 mL citrated blood tubes. Tube one is a dummy tube and is discarded. Tube two is used for the FVIII activity and thrombin generation assay on the EnzySystem. Tube three is used to perform conventional FVIII activity and thrombin generation assays. Tube four is used for duplicate measurements of the conventional assays and measurement of prothrombin F1+2 antigen levels, ADAMTS13 activity, FVIII antigen, VWF (von Willebrand Factor) antigen and von Willebrand ristocetin cofactor activity levels.
Hemophilia A patient
Several assays
Standard blood draw by venipuncture is performed in all study participants to collect blood samples in four 2.7 mL citrated blood tubes. Tube one is a dummy tube and is discarded. Tube two is used for the FVIII activity and thrombin generation assay on the EnzySystem. Tube three is used to perform conventional FVIII activity and thrombin generation assays. Tube four is used for duplicate measurements of the conventional assays and measurement of prothrombin F1+2 antigen levels, ADAMTS13 activity, FVIII antigen, VWF (von Willebrand Factor) antigen and von Willebrand ristocetin cofactor activity levels.
Interventions
Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.
Several assays
Standard blood draw by venipuncture is performed in all study participants to collect blood samples in four 2.7 mL citrated blood tubes. Tube one is a dummy tube and is discarded. Tube two is used for the FVIII activity and thrombin generation assay on the EnzySystem. Tube three is used to perform conventional FVIII activity and thrombin generation assays. Tube four is used for duplicate measurements of the conventional assays and measurement of prothrombin F1+2 antigen levels, ADAMTS13 activity, FVIII antigen, VWF (von Willebrand Factor) antigen and von Willebrand ristocetin cofactor activity levels.
Eligibility Criteria
Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.
Inclusion Criteria
* Age between 20 to 70 years old (equally distributed over the age range)
* 7 volunteers 20-40 years old
* 7 volunteers 40-60 years old
* 6 volunteers 60+ years old
Hemophilia A patients:
* Diagnosed with mild (FVIII activity levels 5-40%), moderate (FVIII activity levels 1-5%) or severe hemophilia A (FVIII activity levels \<1%)
* Medication
1. On demand treatment
2. Washout of medication of at least 24 hours after treatment with short half life (SHL) replacement therapy
3. Washout of medication of at least 72 hours after treatment with extended half life (EHL) replacement therapy
* Age 20-70 years old
Exclusion Criteria
* use of anticoagulants or platelet antagonists (aspirin or any TAR);
* known allergy to stainless steel;
* trauma or surgery within the last two weeks;
* pregnancy;
* use of:
* NSAIDs;
* antimicrobial medication;
* thyroid inhibitors; or SSRI's. A hemophilia A patient who meets any of the following criteria will be excluded from participation in this study:
* use of anticoagulants or platelet antagonists (aspirin or any TAR);
* known allergy to stainless steel;
* trauma or surgery within the last two weeks;
* a bleeding episode within the last two weeks;
* clinical indication of liver cirrhosis (echographic indication, enlarged spleen, decreased platelet count);
* pregnancy;
* FVIII inhibitors;
* Signs of inflammation or infection
* use of:
* NSAIDs;
* antimicrobial medication;
* thyroid inhibitors or SSRI's;
* Emicizumab.
20 Years
70 Years
ALL
Yes
Sponsors
Meet the organizations funding or collaborating on the study and learn about their roles.
Enzyre B.V.
INDUSTRY
Radboud University Medical Center
OTHER
Responsible Party
Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.
Principal Investigators
Learn about the lead researchers overseeing the trial and their institutional affiliations.
Saskia Schols, PhD
Role: PRINCIPAL_INVESTIGATOR
Radboud University Medical Center
Locations
Explore where the study is taking place and check the recruitment status at each participating site.
Enzyre BV
Nijmegen, Gelderland, Netherlands
Radboud university medical center
Nijmegen, , Netherlands
Countries
Review the countries where the study has at least one active or historical site.
Other Identifiers
Review additional registry numbers or institutional identifiers associated with this trial.
HEMSTOL77
Identifier Type: -
Identifier Source: org_study_id
More Related Trials
Additional clinical trials that may be relevant based on similarity analysis.