How Differences in Oximeter Performance May Affect Clinical Decision

NCT ID: NCT04474574

Last Updated: 2025-01-22

Study Results

Results pending

The study team has not published outcome measurements, participant flow, or safety data for this trial yet. Check back later for updates.

Basic Information

Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.

Recruitment Status

COMPLETED

Total Enrollment

50 participants

Study Classification

OBSERVATIONAL

Study Start Date

2023-07-01

Study Completion Date

2023-10-30

Brief Summary

Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.

In clinical practice discrepancies between overnight SpO2 recordings performed by 2 devices used simultaneously are regularly observed. However, this has not been systematically studied or quantified. It is therefore important to determine if these discrepancies are anecdotic, or frequent, and to what extent this may affect decisions in clinical practice such as implementing (or withdrawing) oxygen in subjects under noninvasive ventilation (NIV) or continuous positive airway pressure (CPAP), or adjusting NIV settings.

Detailed Description

Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.

Pulse oximetry estimates oxygen saturation in the arterial blood by trans-illuminating a translucent tissue (usually a fingertip, or an ear lobe) with light-emitting diodes at 2 specific wavelengths. Absorption of light at these different wavelengths (660 nm, red, and 940 nm, infrared respectively) differs between oxygenated and deoxygenated hemoglobin. The amount of light which is transmitted at both wave lengths is then quantified and processed by an algorithm that displays a saturation value. The signals are corrected for the pulsatile nature of arterial oxygen flow. Devices also provide pulse rate measured by plethysmography (pulse-related changes in volume of fingertip or earlobe). Since its introduction in the early 1980s, pulse oximetry has proven to be an essential tool for the non-invasive assessment of blood oxygen saturation (SpO2). The use of pulse oximeter is now widespread and interests acute care settings as well as primary care settings. Although there have been recent improvements in signal analysis such as increasing sampling frequency and improving reflectance technology \[4\], the accuracy of commercially available oximeters differs. First of all, there is variability in the way accuracy of pulse oximetry devices is reported within 2% (± 1 SD) or within 5% (± 2 SD) of reference measurements obtained by blood gases analysis \[5\]. Secondly, there is a variability between commercially available devices, especially below a SpO2 of 90%. This is partly explained by the fact that calibration of the different algorithms employed in signal processing is limited by the range of saturations that can be safely obtained in healthy volunteers. In respiratory medicine, it has been shown that such variability of accuracy could affect the diagnosis of Obstructive Sleep Apnea Hypopnea Syndrome (OSAHS) and impact on clinical decisions as the recorded number of apneas/hypopneas varied between devices during nocturnal sleep studies \[8\] Nocturnal hypoxemia (NH) is considered as one of the major determinants of adverse cardiovascular complications and neurocognitive impairment in patients suffering of chronic respiratory failure (CRF). Because of its simplicity, short set-up time and short time response, pulse oximetry is a valuable screening tool for nocturnal hypoxemia despite its disadvantages such as motion artefacts or sensitivity to perfusion. Therefore, definitions of NH rely solely on nocturnal oximetry recording: for instance, in a consensus statement on noninvasive ventilation, spending \> 5% of sleep time under 88% of SpO2 was considered as a relevant threshold. Definitions of NH remain arbitrary and different expert-based thresholds have been suggested in the medical literature \[10\]. Patients suffering from nocturnal hypoventilation, especially those with an average daytime SpO2 close to the steep portion of the hemoglobin dissociation curve (SpO2 between 90-94%), are at higher risk for NH. Therefore, in these patients, device imprecision could have a significant impact on medical decisions, such as deciding to adjust ventilator settings in patients under noninvasive ventilation (NIV) and/or implementing nocturnal oxygen supplementation.

In patients with CRF and NIV, after optimal adjustment of ventilator settings, prescription of nocturnal oxygen supplementation is common practice although impact of nocturnal oxygen supplementation on survival, patient comfort, or prevention of cor pulmonale has yet to be demonstrated. Practically speaking, it increases considerably the burden of the treatment for the patient and care givers (additional connections and tubings, noise of the oxygen concentrator etc.). To our knowledge, no study has evaluated how the use of different pulse oximeters could impact on this decision. Three types of devices are used in clinical practice: pulse oximetry using a probe connected to the home ventilator device, pulse oximetry combined with transcutaneous capnography, or using a "wrist watch" type of device.

Conditions

See the medical conditions and disease areas that this research is targeting or investigating.

Sleep Apnea Alveolar Hypoventilation Nocturnal Hypoxemia

Study Design

Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.

Observational Model Type

CASE_ONLY

Study Time Perspective

PROSPECTIVE

Interventions

Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.

Simultaneous recording of nocturnal SpO2

The intervention will consist in performing a simultaneous recording of nocturnal SpO2 and pulse rate for at least 4 hours with 3 different devices.

Intervention Type DEVICE

Eligibility Criteria

Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.

Inclusion Criteria

* \> 18 years old
* Awake room air SpO2 between 90 and 94%
* NIV or CPAP therapy
* Ambulatory or hospitalized patient in a clinically stable respiratory condition without any vasopressor treatment.

Exclusion Criteria

* Hospitalization in an acute care setting (e.g. emergency room, Intensive Care Unit, Intermediate Care Unit)
* Any vasopressor treatment
* Peripheral vascular pathologies that can affect digital perfusion (e.g. history of ischemia, Raynaud's phenomenon, any type of vasculitis).
* Mechanical obstacles that may limit quality of signal (e.g. nail polish, bandage, splint, plaster).
* Patient already treated by long term nocturnal oxygen therapy (LTOT)
Minimum Eligible Age

18 Years

Eligible Sex

ALL

Accepts Healthy Volunteers

No

Sponsors

Meet the organizations funding or collaborating on the study and learn about their roles.

University Hospital, Geneva

OTHER

Sponsor Role lead

Responsible Party

Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.

Jean-Paul Janssens

Professor

Responsibility Role PRINCIPAL_INVESTIGATOR

Locations

Explore where the study is taking place and check the recruitment status at each participating site.

Geneva University Hospital

Geneva, , Switzerland

Site Status

Countries

Review the countries where the study has at least one active or historical site.

Switzerland

References

Explore related publications, articles, or registry entries linked to this study.

Karege G, Egger B, Lawi D, Bergeron A, Janssens JP. How Using Different Oximeters May Affect Clinical Decision-Making: A Method Comparison Study in Patients Receiving CPAP or Noninvasive Ventilation. Chest. 2025 Sep 9:S0012-3692(25)05158-X. doi: 10.1016/j.chest.2025.08.018. Online ahead of print.

Reference Type DERIVED
PMID: 40930342 (View on PubMed)

Other Identifiers

Review additional registry numbers or institutional identifiers associated with this trial.

2020-01813

Identifier Type: -

Identifier Source: org_study_id