Neonatal Wireless Monitoring System for Intensive Care

NCT ID: NCT04241263

Last Updated: 2021-01-12

Study Results

Results pending

The study team has not published outcome measurements, participant flow, or safety data for this trial yet. Check back later for updates.

Basic Information

Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.

Recruitment Status

UNKNOWN

Clinical Phase

NA

Total Enrollment

20 participants

Study Classification

INTERVENTIONAL

Study Start Date

2020-03-01

Study Completion Date

2021-07-31

Brief Summary

Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.

Babies who require intensive care need to have their heart rate, oxygen levels and temperature measured continuously using specially designed monitors. These monitors allow the clinical team to constantly assess the baby and see if they are becoming unwell. However, these monitors each have separate wires to connect them to the baby. Another important part of care in these babies is being able to allow parents to hold their babies, getting them out of their incubators to be cuddled. This is called Kangaroo care and has been shown to impact on long term health for both babies and parents.

We have designed a system to free babies of the wires that tether them to the monitors. This system has been tested in adults and we now want to test them on the babies.

There are two study parts with two different questions:

1. How good is the data? This will compare the new wireless system with the existing system. Babies will have both standard wired monitors and the new wireless monitor on at the same time so we can compare the data to see how good is the quality of the data collected on the new system.
2. What do parents and staff think about the wireless system during Kangaroo care? Babies who no longer need monitoring (but who previously had been monitored with the wired system), will have the wireless system attached and the parent will take their baby from the cot for Kangaroo Care. After the baby has been placed back in the cot, we will conduct interviews with the parent and the nurse to find out what they thought of the new system.

We hope this will help the investigators to understand about how good the data is we collect and how we might be able to improve the system.

Detailed Description

Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.

Summary of study design

This is a single-center study recruiting from the neonatal intensive care unit (NICU) at Addenbrooke's Hospital, Cambridge. There are 2 study parts:

1. Quantitative assessment of the accuracy of the new system in comparison with standard clinical monitoring systems
2. Qualitative Assessment of parent and staff views of caring for babies with the new system in comparison with conventional clinical monitoring systems

Quantitative study arm Study population Up to 24 participants will be recruited from the Addenbrooke's Hospital NICU. The participants will be recruited from three groups, defined by the BAPM Categories of Care The population size has been selected to include sufficient babies to collect a reasonable amount of data to ascertain, with confidence, our defined outcomes. Furthermore, the data should come from across the range of variables seen in babies requiring intensive care while limiting exposure of the infants at most risk. The study will be performed in a stepwise fashion, starting with those infants requiring lower dependency care.

Inclusion criteria

* Babies being cared for on the NICU at Addenbrookes Hospital
* Written, informed parental consent Exclusion criteria
* Known to have an intolerance to silicone or silicone gel adhesives.
* Has a pacemaker or other implanted electronic device.

Planned procedures Participants will have the newly developed wireless sensors for ECG, pulse oximetry, and temperature attached adjacent to the current clinical, wired sensors. Once the application sites have been identified and prepared, the sensors will be placed on the baby by an appropriately trained individual. The wireless receiver will connect to a separate patient monitor to log the comparative data. This study monitor will be covered and silenced such that data will not be visible to the clinical team, and it will not be possible to confuse the study monitor from the 'clinical' monitor. The sensors will be in place for a period of up to 24 hours. After this period, the study sensors will be removed, and the site inspected by a member of the research team and scored using the standard tissue viability scoring system already in use on the unit. The sensors will be disposed of in keeping with the unit and Trust policy.

After the study intervention, a questionnaire will be given to the nurse caring for the baby to determine their views on the wireless system.

Data collection Data will be collected from both the study monitor and the 'clinical' monitor. The data are formed of two types: parameter data and waveform data. Parameter data has a two-second interval. The waveform data has a period of approximately 4ms.

Qualitative study arm

Study population Up to 10 participants will be recruited from the Addenbrooke's Hospital. This is in line with other studies with similar cohorts and balances the need to obtain a range of views and experiences, with the imposition on parents' time in such an environment.

Inclusion criteria

* BAPM level transitional care babies who do not clinically require vital sign monitoring.
* The baby has had vital sign monitoring at some point during their stay on the unit.
* Written, informed parental consent Exclusion criteria
* Known to have an intolerance to silicone or silicone gel adhesives.
* Has a pacemaker or other implanted electronic medical device.

Planned procedure Babies who are not indicated for vital sign monitoring will have the wireless monitoring system attached. Once the application sites have been identified and prepared, the sensors will be placed on the baby by an appropriately trained individual. The wireless receiver will connect to the study monitor to log the vital sign data, and all alarms will be silenced. The monitor will be covered and silenced such that it will not impact on clinical care.

The baby will be transferred from the cot/incubator to the parent for kangaroo care as per standard NICU procedure. The parent and baby will remain in contact for as long as the parent is comfortable, and the baby remains stable and in keeping with unit policy and procedures. The baby will then be transferred back to the cot, and monitoring will be removed.

Semi-structured interviews lasting approximately 30 minutes will then be conducted with, (i) the nurse, and (ii) the parent(s). These interviews will focus on previous experiences with vital sign monitoring systems, and their experience of using the new wireless system.

Conditions

See the medical conditions and disease areas that this research is targeting or investigating.

Preterm Infant Infant, Newborn Low; Birthweight, Extremely (999 Grams or Less)

Study Design

Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.

Allocation Method

NON_RANDOMIZED

Intervention Model

PARALLEL

Primary Study Purpose

DEVICE_FEASIBILITY

Blinding Strategy

NONE

This is a feasibility study but there are two arms and the babies, parents and staff will all be considered participants -it will be approximately 20 families recruited

Study Groups

Review each arm or cohort in the study, along with the interventions and objectives associated with them.

Comparative assessment of data quality

Quantitative comparison between the current wired system and the new wireless system

Group Type OTHER

Wireless vital sign monitoring - comparative

Intervention Type DEVICE

Participants who currently have vital sign monitoring will, in addition, have the wireless system attached. We will then collect up to 24 hours of vital sign data from both systems, which will allow us to make direct comparisons between them.

Usability

Qualitative assessment of the new wireless system

Group Type OTHER

Wireless vital sign monitoring - usability

Intervention Type DEVICE

Babies will be recruited who do not currently require vital sign monitoring. The wireless system will be attached to these babies by nurses who have been trained in its use. A parent will then be asked to perform Kangaroo Care for as long as they feel comfortable. After the baby is placed back in its cot, and the sensors have been removed, we will conduct (separate) interviews with the nurse and the parent on the neonatal intensive care unit (NICU).

Interventions

Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.

Wireless vital sign monitoring - comparative

Participants who currently have vital sign monitoring will, in addition, have the wireless system attached. We will then collect up to 24 hours of vital sign data from both systems, which will allow us to make direct comparisons between them.

Intervention Type DEVICE

Wireless vital sign monitoring - usability

Babies will be recruited who do not currently require vital sign monitoring. The wireless system will be attached to these babies by nurses who have been trained in its use. A parent will then be asked to perform Kangaroo Care for as long as they feel comfortable. After the baby is placed back in its cot, and the sensors have been removed, we will conduct (separate) interviews with the nurse and the parent on the neonatal intensive care unit (NICU).

Intervention Type DEVICE

Eligibility Criteria

Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.

Inclusion Criteria

(i) Babies being cared for on the NICU at Addenbrooke's Hospital (ii) Written, informed parental consent

Exclusion Criteria

(i) Known to have an intolerance to silicone or silicone gel adhesives. (ii) Has a pacemaker, or other implanted electronic device
Maximum Eligible Age

6 Months

Eligible Sex

ALL

Accepts Healthy Volunteers

Yes

Sponsors

Meet the organizations funding or collaborating on the study and learn about their roles.

University of Cambridge

OTHER

Sponsor Role collaborator

Cambridge University Hospitals NHS Foundation Trust

OTHER

Sponsor Role lead

Responsible Party

Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.

Kathryn Beardsall

University Lecturer, Consultant Neonatologist

Responsibility Role PRINCIPAL_INVESTIGATOR

Principal Investigators

Learn about the lead researchers overseeing the trial and their institutional affiliations.

kathryn beardsall

Role: PRINCIPAL_INVESTIGATOR

University of Cambridge

Locations

Explore where the study is taking place and check the recruitment status at each participating site.

Cambridge University Hospitals NHS Foundation Trust

Cambridge, Cambridgeshire, United Kingdom

Site Status RECRUITING

Countries

Review the countries where the study has at least one active or historical site.

United Kingdom

Central Contacts

Reach out to these primary contacts for questions about participation or study logistics.

Kathryn Beardsall, MRCP

Role: CONTACT

01223746791

Facility Contacts

Find local site contact details for specific facilities participating in the trial.

Kathryn Beardsall

Role: primary

Other Identifiers

Review additional registry numbers or institutional identifiers associated with this trial.

A095011

Identifier Type: -

Identifier Source: org_study_id

More Related Trials