Fetal Life: Smart-Device Based Uterine Activity Monitoring
NCT ID: NCT03940365
Last Updated: 2022-01-26
Study Results
The study team has not published outcome measurements, participant flow, or safety data for this trial yet. Check back later for updates.
Basic Information
Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.
COMPLETED
NA
17 participants
INTERVENTIONAL
2019-05-15
2021-03-31
Brief Summary
Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.
The study aims to test a uterine contraction monitor that can be applied by the patient to her abdomen and connected to her smart device to detect the frequency and duration of uterine contractions. Phase 1 will collect data for calibration of the monitor and for a noninferiority study comparing the Fetal Life uterine contraction monitor (tocometer) with standard tocometry devices. Phase 2 will be a feasibility study in which patients will position the monitor, connect to the corresponding Fetal Life app and measure their own uterine contraction activity.
The study's hypothesis is that after calibration, the new monitor will reliably detect at least 95% of the number and duration of contractions detected by the standard monitor, and that the device will perform comparably whether applied by the patient or by research staff.
Related Clinical Trials
Explore similar clinical trials based on study characteristics and research focus.
Electrode-Based Sensor for Non-Invasive Fetal Heart Rate Monitoring With Improved Reliability
NCT02024165
Dual System Study: Transvaginal Fetal Pulse Oximetry Measurement Systems
NCT07296419
The Fetal EKG Study
NCT02320279
Trans-abdominal Fetal Pulse Oximetry; Tissue Light Scattering and Signal Integrity
NCT04876846
Accuracy and Reliability of Novii: Fetal Heart Rate (FHR), Maternal Heart Rate (MHR) and Uterine Contraction (UA) Compared With Doppler, Scalp Fetal Scalp Electrode (FSE), Tocodynamometer (TOCO) and Intra Uterine Pressure Catheter (IUPC)
NCT03409146
Detailed Description
Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.
Patients are educated during prenatal care that contractions should be timed from the start of one contraction to the start of the next. These instructions can be difficult for patients to remember, and patients who are having painful contractions may prefer a more automated method for timing contractions. All methods currently available for home contraction tracking (on paper, using a smart device, or using one of many available smart device applications) require manual data entry either by the patient or her support partners.
Patients are cautioned during pregnancy to maintain awareness of the development of uterine contractions. Early in pregnancy, patients are frequently counseled that infrequent, irregular contractions can be normal, but that frequent contractions occurring in a pattern can be a sign of preterm labor. As dehydration can cause the uterus to contract, preterm patients who notice contractions occurring in a pattern are often instructed to drink water, observe their contraction pattern, and come to the hospital if their contractions continue to occur every 15 minutes or more frequently following oral hydration. At term, patients are frequently counseled to come to the hospital when they are experiencing contractions that last about a minute, are occurring every five minutes, and have been at this duration and frequency for about an hour. Both preterm and at term, patients may have runs of contractions which resolve with time and hydration, but patients in both situations need to recognize when their contractions have reached the parameters recommended by their health care providers to seek medical attention.
The new tocometer device tested in this study has significant cost savings in comparison to existing devices, and connects to a smart phone or tablet rather than a bulky receiver. The device will allow the patient to apply a monitor belt, connect it to her smart phone, and view a graphical representation of her contraction pattern with verbal descriptors that she can share with her health care team. Many patients have at least one visit to the hospital for evaluation of a contraction pattern which did not warrant this time and expense. This device may help patients to share a more accurate report of their symptoms with their prenatal care providers, potentially decreasing the stress of manual contraction tracking, improving the understanding of progress in early labor and decreasing unnecessary hospital visits. The device and app will not give the patient instructions on whether to come to the hospital, but will be used to help the patient understand if her uterine activity meets the parameters her prenatal care provider has given her to come in for evaluation. The Fetal Life device is constructed of smooth, latex-free plastic, is powered by a Lithium battery similar to a watch battery, and is held in place using a standard tocometry belt manufactured by General Electric. The investigators consider this to be a nonsignificant risk device.
Phase 1 of this initial trial will allow a noninferiority comparison of the performance of this new contraction monitoring device with standard devices currently used in Labor and Delivery settings for monitoring duration and frequency of contractions. Phase 2 will assess feasibility for patient use.
Procedure for Calibration and Noninferiority Phase (Phase 1):
* Patients meeting inclusion criteria will be offered participation in the study by the Clinical Trials Unit (CTU) Staff
* If the patient agrees to participate, CTU staff will obtain consent.
* The patient will be assigned a participant number.
* The Fetal Life tocometer will be applied to the patient's lower abdomen in addition to the standard tocometer used on Labor and Delivery and secured with the monitor belt by CTU staff
* The Fetal Life tocometer will be paired the study iPad and output will be captured by CTU staff
* The patient will be asked to cough in order to create a peak that synchronizes the graphic output of the Fetal Life tocometer and the standard tocometer
* The patient will be monitored with the both the standard tocometer and the Fetal Life tocometer for 45-60 minutes
* If the maternal or fetal status requires changes of maternal position during the trial, the Fetal Life monitor may be left in place, disconnected from the iPad or removed from the patient, whichever is least disruptive, so as not to interfere with the patient's obstetrical care in any way
* CTU staff will document patient's age, parity, estimated gestational age and body mass index on a data collection sheet identified by participant number. No personal health information will be maintained on the data sheet.
* If 45-60 minutes of monitoring is not completed, CTU staff will document the reason why
* Patient will complete a brief survey on monitor comfort in comparison to the standard monitor
* Image files from the standard tocometry fetal monitoring system will be exported for comparison with the output of the Fetal Life tocometer.
* Fetal Life device will be sanitized with the antiviral, antibacterial wipes used routinely for sanitation of the standard tocometer devices before use with another patient. A new belt will be attached to the device for each subject.
* For calibration, deidentified data from standard tocometer and the Fetal Life device will be reviewed by the team in the University of Louisville School of Engineering so that any needed adjustments can be made to the pressure transducer's sensitivity or graphic output
* After calibration is completed (estimate 3 subjects needed), subjects will then be enrolled using the same protocol to collect data required for statistical analysis of non-inferiority
Procedure for Feasibility Phase (Phase 2):
* Patients meeting the inclusion criteria will be offered participation in the study by the Clinical Trials Unit (CTU) Staff
* If the patient agrees to participate, CTU staff will obtain consent.
* The patient will be assigned a participant number.
* The patient will read the Fetal Life monitor instructions
* Standard tocometry and fetal heart rate devices will be removed
* The patient will apply the Fetal Life monitor using the positioning belt and will connect to the study iPad or lap top computer via Bluetooth
* Fifteen minutes of monitoring will be captured via the iPad or lap top computer by CTU staff
* During this time, the patient will note when she perceives the beginning and the end of her contractions; this will be documented the Clinical Trials Unit staff with time notations
* Fetal Life monitor will be removed and the standard external fetal monitoring will resume
* Patient will complete the patient survey
Conditions
See the medical conditions and disease areas that this research is targeting or investigating.
Study Design
Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.
NA
SINGLE_GROUP
SUPPORTIVE_CARE
NONE
Study Groups
Review each arm or cohort in the study, along with the interventions and objectives associated with them.
Study Group
As detailed above, a single group will be used for the study to compare the output of the study device with the output of the standard device in each patient.
Fetal Life tocometer device
The fetal life tocometer device is a uterine contraction monitor which is less expensive to manufacture than existing contraction monitor devices and which pairs with a smart device rather than a bulky output device.
Interventions
Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.
Fetal Life tocometer device
The fetal life tocometer device is a uterine contraction monitor which is less expensive to manufacture than existing contraction monitor devices and which pairs with a smart device rather than a bulky output device.
Eligibility Criteria
Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.
Inclusion Criteria
* Contracting every five minutes or more frequently by standard tocometry monitor
* Category 1 fetal heart tracing (reassuring tracing with moderate variability and no decelerations)
* Cervical examination completed and awaiting repeat examination in 2-4 hours
* 37 weeks or greater
* Singleton pregnancy (not multiple gestation)
* Live fetus
* Undergoing tocometry monitoring with standard devices
* English speaking
* Age between 20 years and 40 years
Exclusion Criteria
* Any active skin lesions or infections
20 Years
40 Years
FEMALE
Yes
Sponsors
Meet the organizations funding or collaborating on the study and learn about their roles.
University of Louisville
OTHER
Fetal Life LLC
INDUSTRY
Responsible Party
Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.
Principal Investigators
Learn about the lead researchers overseeing the trial and their institutional affiliations.
Sara Petruska, MD
Role: PRINCIPAL_INVESTIGATOR
University of Louisville
Locations
Explore where the study is taking place and check the recruitment status at each participating site.
University Hospital, University of Louisville
Louisville, Kentucky, United States
Countries
Review the countries where the study has at least one active or historical site.
References
Explore related publications, articles, or registry entries linked to this study.
Lewis D, Downe S; FIGO Intrapartum Fetal Monitoring Expert Consensus Panel. FIGO consensus guidelines on intrapartum fetal monitoring: Intermittent auscultation. Int J Gynaecol Obstet. 2015 Oct;131(1):9-12. doi: 10.1016/j.ijgo.2015.06.019. No abstract available.
Hoffman E.B., Sen, P.K., Weinberg, C.R. Within-cluster resampling. Biometrika (2001), 88(4): 1121-1134.
Lorenz DJ, Levy S, Datta S. Inferring marginal association with paired and unpaired clustered data. Stat Methods Med Res. 2018 Jun;27(6):1806-1817. doi: 10.1177/0962280216669184. Epub 2016 Sep 20.
Other Identifiers
Review additional registry numbers or institutional identifiers associated with this trial.
19.0011
Identifier Type: -
Identifier Source: org_study_id
More Related Trials
Additional clinical trials that may be relevant based on similarity analysis.