Hypothermic Oxygenated Perfusion Versus Static Cold Storage for Marginal Graft
NCT ID: NCT03031067
Last Updated: 2018-02-13
Study Results
The study team has not published outcome measurements, participant flow, or safety data for this trial yet. Check back later for updates.
Basic Information
Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.
COMPLETED
NA
20 participants
INTERVENTIONAL
2016-10-31
2018-02-28
Brief Summary
Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.
The addition of oxygen during the perfusion represents an innovation in the methods of preservation in approved clinical setting seems to add further improvements of the graft. The present study was designed in order to assess the impact of hypothermic oxygenated perfusion (PIO) of marginal human kidney and liver compared with SCS.
Related Clinical Trials
Explore similar clinical trials based on study characteristics and research focus.
Clinical Trial of New Hypothermic Oxygenated Perfusion System Versus Static Cold Storage
NCT03837197
Normothermic Machine Perfusion (NMP) Versus Hypothermic Machine Perfusion (HMP) in Human Kidney Transplantation
NCT05743751
Clinical Impact of Hypothermic Machine Perfusion in Renal Transplant Recipients
NCT02621281
Normothermic Machine Perfusion (NMP) Versus Static Cold Storage (SCS) in Human Kidney Transplantation
NCT05031052
Immunometabolism of Machine Perfusion Strategies
NCT06371924
Detailed Description
Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.
The marginal graft will be perfused with oxygenated solution of preservation at 4°C for two hours with Exiper, Bologna Machine Perfusion, developed by Medica s.p.a and our research group.
Flow and pressure values will be set up for the kidney and liver perfusion, otherwise. The oxygenation of solution will be performed by an oxygenator and a filter for decapneization / oxygenation. During the perfusion the oxygen pressure will be required between 600-750 mmHg (pO2 80-100 Kpa), as reported in the scientific literature. The pH, lactate concentration, and oxygen (PO2) and carbon dioxide (PCO2) partial pressure were measured in the preservation solution at T0 and T1 by means of a standard haemogasanalyzer.
After transplantation, the recipients will be monitored clinically to assess the graft function. The liver's data will be compared with similar case treated with SCS in our centre of transplant, retrospectively; instead, the kidney's data will be compared with the "twin" organ transplanted without PIO, prospectively.
Conditions
See the medical conditions and disease areas that this research is targeting or investigating.
Study Design
Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.
NON_RANDOMIZED
PARALLEL
TREATMENT
NONE
Study Groups
Review each arm or cohort in the study, along with the interventions and objectives associated with them.
Machine perfusion - Kidney
The marginal kidney will be perfused with oxygenated solution of preservation at 4°C for two hours with Exiper, Bologna Machine Perfusion.
Machine perfusion
The graft preservation will be performed perfusing with a oxygenated solution of preservation at 4°C for two hours with Exiper, Bologna Machine Perfusion, developed by Medica s.p.a and our research group.
Flow and pressure values will be set up for the kidney and liver perfusion, differently. The oxygenation of solution will be performed by an oxygenator and a filter for decapneization / oxygenation. During the perfusion the oxygen pressure will be required between 600-750 mmHg (pO2 80-100 Kpa), as reported in the scientific literature. The pH, lactate concentration, and oxygen (PO2) and carbon dioxide (PCO2) partial pressure were measured in the preservation solution at T0 and T1 by means of a standard haemogasanalyzer.
Static cold storage - Kidney
The marginal kidney that was stored to cold (SCS), previously.
No interventions assigned to this group
Machine perfusion - Liver
The marginal liver will be perfused with oxygenated solution of preservation at 4°C for two hours with Exiper, Bologna Machine Perfusion.
Machine perfusion
The graft preservation will be performed perfusing with a oxygenated solution of preservation at 4°C for two hours with Exiper, Bologna Machine Perfusion, developed by Medica s.p.a and our research group.
Flow and pressure values will be set up for the kidney and liver perfusion, differently. The oxygenation of solution will be performed by an oxygenator and a filter for decapneization / oxygenation. During the perfusion the oxygen pressure will be required between 600-750 mmHg (pO2 80-100 Kpa), as reported in the scientific literature. The pH, lactate concentration, and oxygen (PO2) and carbon dioxide (PCO2) partial pressure were measured in the preservation solution at T0 and T1 by means of a standard haemogasanalyzer.
Static cold storage - Liver
The marginal liver that was stored to cold (SCS), previously.
No interventions assigned to this group
Interventions
Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.
Machine perfusion
The graft preservation will be performed perfusing with a oxygenated solution of preservation at 4°C for two hours with Exiper, Bologna Machine Perfusion, developed by Medica s.p.a and our research group.
Flow and pressure values will be set up for the kidney and liver perfusion, differently. The oxygenation of solution will be performed by an oxygenator and a filter for decapneization / oxygenation. During the perfusion the oxygen pressure will be required between 600-750 mmHg (pO2 80-100 Kpa), as reported in the scientific literature. The pH, lactate concentration, and oxygen (PO2) and carbon dioxide (PCO2) partial pressure were measured in the preservation solution at T0 and T1 by means of a standard haemogasanalyzer.
Other Intervention Names
Discover alternative or legacy names that may be used to describe the listed interventions across different sources.
Eligibility Criteria
Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.
Inclusion Criteria
* Kidney recipient: single or dual kidney transplant recipients at our center, provided consent.
* Liver donor: ECD
* Liver recipient: liver transplant recipients at our center, provided consent.
Exclusion Criteria
* Kidney recipient: patients with severe atherosclerotic vascular disease, pathologies predisposing to the onset of intra-operative surgical complications, such as thrombophilia and hemophilia, antigenic incompatibility with donor determining the acute rejection of the organ; patients waiting for kidney transplant pre-emptive, multi-organ transplant, retransplantation.
* Liver donor: no ECD, donor with vascular anatomical abnormalities, donor with a risk infection, to increase cold ischemia time can not be avoided.
* Liver recipient: patients with acute liver disease and with vascular abnormalities and/or biliary tract requiring non-conventional reconstructive techniques, patients waiting for multi-organ transplant, retransplantation, urgency transplant ( or with MELD\>30).
18 Years
60 Years
ALL
No
Sponsors
Meet the organizations funding or collaborating on the study and learn about their roles.
Matteo Ravaioli
OTHER
Responsible Party
Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.
Matteo Ravaioli
Principal Investigator
Principal Investigators
Learn about the lead researchers overseeing the trial and their institutional affiliations.
Matteo Ravaioli, PhD
Role: PRINCIPAL_INVESTIGATOR
University of Bologna S. Orsola-Malpighi Hospital, Transplantation and General Surgery Unit
Antonio Daniele Pinna, Professor
Role: STUDY_DIRECTOR
University of Bologna S. Orsola-Malpighi Hospital, Transplantation and General Surgery Unit
Locations
Explore where the study is taking place and check the recruitment status at each participating site.
Azienda Ospedaliera di Bologna - Policlinico S. Orsola Malpighi
Bologna, , Italy
Countries
Review the countries where the study has at least one active or historical site.
References
Explore related publications, articles, or registry entries linked to this study.
O'Callaghan JM, Morgan RD, Knight SR, Morris PJ. Systematic review and meta-analysis of hypothermic machine perfusion versus static cold storage of kidney allografts on transplant outcomes. Br J Surg. 2013 Jul;100(8):991-1001. doi: 10.1002/bjs.9169.
Thuillier R, Allain G, Celhay O, Hebrard W, Barrou B, Badet L, Leuvenink H, Hauet T. Benefits of active oxygenation during hypothermic machine perfusion of kidneys in a preclinical model of deceased after cardiac death donors. J Surg Res. 2013 Oct;184(2):1174-81. doi: 10.1016/j.jss.2013.04.071. Epub 2013 May 21.
Hosgood SA, Nicholson HF, Nicholson ML. Oxygenated kidney preservation techniques. Transplantation. 2012 Mar 15;93(5):455-9. doi: 10.1097/TP.0b013e3182412b34.
Dutkowski P, Polak WG, Muiesan P, Schlegel A, Verhoeven CJ, Scalera I, DeOliveira ML, Kron P, Clavien PA. First Comparison of Hypothermic Oxygenated PErfusion Versus Static Cold Storage of Human Donation After Cardiac Death Liver Transplants: An International-matched Case Analysis. Ann Surg. 2015 Nov;262(5):764-70; discussion 770-1. doi: 10.1097/SLA.0000000000001473.
Other Identifiers
Review additional registry numbers or institutional identifiers associated with this trial.
159/2015/O/Disp
Identifier Type: REGISTRY
Identifier Source: secondary_id
PIO-700
Identifier Type: -
Identifier Source: org_study_id