Performance Analysis of Hermetic Closed-loop Anesthesia Delivery System
NCT ID: NCT05967403
Last Updated: 2025-03-27
Study Results
The study team has not published outcome measurements, participant flow, or safety data for this trial yet. Check back later for updates.
Basic Information
Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.
RECRUITING
NA
150 participants
INTERVENTIONAL
2023-12-09
2026-11-30
Brief Summary
Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.
Closed loop anaesthesia delivery system (CLADS) is an indigenously developed patented (Patent no.502/DEL/2003 \& US 9,108,013 B2) computer-controlled anaesthesia delivery system which works with feedback loop information elicited by BIS monitoring and delivers propofol TIVA to the patient via a non-TCI automated infusion pump. It has been an extensively used and validated in patients undergoing both cardiac and non-cardiac surgical procedures.
A new compact and upgraded version of CLADS is now available. This new version incorporates the anesthetic depth monitor, hemodynamic monitor, controller, user interface and actuator syringe pump into a single, compact and user-friendly module.
The investigators aim to conduct a prospective randomized pilot study comparing the new CLADS and older CLADS version with respect to: adequacy of anaesthesia depth maintenance, performance characteristic of propofol delivery system, propofol requirement, haemodynamics stability, recovery from anesthesia and postoperative sedation.
Related Clinical Trials
Explore similar clinical trials based on study characteristics and research focus.
Comparison of TIVA by Closed Loop Anaesthesia Delivery System Versus Target Controlled Infusion Device
NCT04731480
Effect of Subanaesthetic Dose of Ketamine on Depth of Anaesthesia Consistency
NCT06986109
Effect of Gender on Propofol Requirement
NCT05725135
Evaluation of Propofol Dosing Based on Total Body Weight Using Closes-loop Anaesthesia Delivery System
NCT05305313
Total Intravenous Anesthesia in Percutaneous Transhepatic Cholangiography
NCT03103672
Detailed Description
Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.
The introduction of target-controlled infusion (TCI) pumps has allowed precision control in propofol delivery as compared to the conventional manually operated infusion pumps. In manually operated infusion pumps the clinician regulates the propofol infusion rate to achieve the desired anaesthesia depth, whereas TCI-pumps deliver propofol using inbuilt algorithms based on the pharmacokinetic (PK) and pharmacodynamic (PD) profile of propofol. Two of the most used PK-PD models for propofol delivery are the Marsh model and the Schneider model. Whereas in the Marsh model the user can set the desired target plasma concentration, Schneider model allows the user to set the desired target effect site concentration for achieving adequate depth of GA.
Over the last two decades the evolution and advancement in automated anaesthesia delivery systems, particularly for propofol administration, has made propofol-TIVA delivery more efficient by removing the human interface required for adjusting real-time propofol delivery, both rate and concentration of propofol. Automated anaesthesia delivery systems deliver propofol based on ascertaining frontal cortex electrical activity as determined by the processed electroencephalogram, the bispectral index (BIS) score. These devices regulate propofol delivery based on a feedback loop involving the BIS score (control variable) generated by the patient and the propofol infusion pump (actuator) and attempts to keep the values within a pre-assigned range, consistent with robust GA depth.
Closed loop anaesthesia delivery system (CLADS) is an indigenously developed patented (Patent no.502/DEL/2003 \& US 9,108,013 B2) computer-controlled anaesthesia delivery system which works with feedback loop information elicited by BIS monitoring and delivers propofol TIVA to the patient via a non-TCI automated infusion pump. The basic control algorithm is based on the relationship between the infusion rates of propofol and BIS values, taking into consideration the pharmacokinetic variables, such as, drug distribution and clearance. The system updates the EEG data every 5-seconds and calculates the BIS error, a difference between the target BIS and the actual BIS value using the proportional-integral-derivative (PID) controller. During anaesthesia induction the target concentration is achieved in a stepwise manner with BIS feedback received every 5-seconds. During maintenance phase of anaesthesia, the propofol delivery is modified every 1-epoch of 30-seconds duration. In each epoch an average of initial 3-BIS values (of every 5-seconds) and average of last 3-BIS values (of every 5-seconds) are compared, and a trend assessment is made. If the trend indicates increasing BIS values, then higher propofol rate is delivered by the infusion pump (actuator) and vice-versa. The control algorithm is implemented using a personal computer (PC) with a Pentium 4 processor. The PC controls communication with the infusion pump (Pilot-C, Fresenius, Paris, France) and the vital sign monitor (AS5, Datex Ohmeda Division, GE Healthcare, Singapore) through RS 232 serial ports.
CLADS has been extensively used and validated for administering propofol TIVA in patients undergoing both cardiac and non-cardiac surgical procedures. In a multi-centric study on evaluation of anaesthesia delivery by CLADS, it was shown that CLADS maintains depth of anaesthesia with far more precision as compared to manual administration.
A new compact and upgraded version of CLADS (Clarity Medical Private Ltd., Mohali, Punjab Indiais now available. The new version integrates the anaesthetic depth as well as the hemodynamic monitor, the controller, the user interface and the actuator syringe pump into a single, compact and user-friendly module.
The investigators aim to conduct a prospective randomised pilot to compare the multiple connected-unit conventional CLADS with the hermetically unit-integrated CLADS version, with respect to, the adequacy of anaesthesia depth (primary objective); and, performance characteristic, propofol requirement, haemodynamic stability, recovery from anaesthesia and postoperative sedation of the delivery systems (secondary objectives).
Conditions
See the medical conditions and disease areas that this research is targeting or investigating.
Study Design
Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.
RANDOMIZED
PARALLEL
Group-1 \[Conventional CLADS (CC) group, n= 75\]: Anaesthesia will be induced and maintained with propofol administered using the automated conventional CLADS.
Group-2 \[Hermetic CLADS (HC) group, n= 75\]: Anaesthesia will be induced and maintained with propofol administered using the hermetic CLADS.
OTHER
DOUBLE
Study Groups
Review each arm or cohort in the study, along with the interventions and objectives associated with them.
Conventional CLADS group
Propofol administration rate will be controlled by a feedback loop facilitated by BIS monitoring using the conventional (prototype) closed-loop anaesthesia delivery system (CLADS). A BIS value of 50 will be used as the target point for induction and maintenance of anesthesia
Propofol
Propofol administration rate will be controlled by a feedback loop facilitated by BIS monitoring using the conventional closed loop anaesthesia delivery system (CLADS). A BIS value of 50 will be used as the target point for induction and maintenance of anaesthesia.
Hermetic CLADS group
Propofol administration rate will be controlled by a feedback loop facilitated by BIS monitoring using the hermetic closed-loop anaesthesia delivery system (CLADS).\[(Clarity Medical Private Ltd., Mohali, Punjab India\]. A BIS value of 50 will be used as the target point for induction and maintenance of anesthesia
Propofol
Propofol administration rate will be controlled by a feedback loop facilitated by BIS monitoring using the hermetic closed loop anaesthesia delivery system (CLADS). A BIS value of 50 will be used as the target point for induction and maintenance of anaesthesia.
Interventions
Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.
Propofol
Propofol administration rate will be controlled by a feedback loop facilitated by BIS monitoring using the conventional closed loop anaesthesia delivery system (CLADS). A BIS value of 50 will be used as the target point for induction and maintenance of anaesthesia.
Propofol
Propofol administration rate will be controlled by a feedback loop facilitated by BIS monitoring using the hermetic closed loop anaesthesia delivery system (CLADS). A BIS value of 50 will be used as the target point for induction and maintenance of anaesthesia.
Eligibility Criteria
Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.
Inclusion Criteria
* ASA physical status I-II
* undergoing elective non-cardiac surgery of minimum 60-minutes duration
Exclusion Criteria
* Hepato-renal insufficiency.
* Any history of neurological disorder (e.g., epilepsy) or brain trauma
* Uncontrolled endocrine diseases (e.g., diabetes mellitus, hypothyroidism).
* Known allergy/hypersensitivity to the study drug.
* History of any psychiatric disorder and/or drug dependence/substance abuse.
* Requirement of postoperative ventilation.
* Refusal to informed consent.
18 Years
65 Years
ALL
No
Sponsors
Meet the organizations funding or collaborating on the study and learn about their roles.
Sir Ganga Ram Hospital
OTHER
Responsible Party
Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.
Nitin Sethi, DNB
Senior Consultant, Institute of Anaesthesiology, Pain & Perioperative Medicine
Principal Investigators
Learn about the lead researchers overseeing the trial and their institutional affiliations.
Nitin Sethi, DNB
Role: STUDY_DIRECTOR
Sir Ganga Ram Hospital, New Delhi, INDIA
Amitabh Dutta, MD, PGDHR
Role: PRINCIPAL_INVESTIGATOR
Sir Ganga Ram Hospital, New Delhi, INDIA
Goverdhan D Puri, MD, PhD
Role: STUDY_CHAIR
Post Graduate Institute of Medical Education & Research, Chandigarh, India
Locations
Explore where the study is taking place and check the recruitment status at each participating site.
Nitin Sethi
New Delhi, National Capital Territory of Delhi, India
Countries
Review the countries where the study has at least one active or historical site.
Central Contacts
Reach out to these primary contacts for questions about participation or study logistics.
Facility Contacts
Find local site contact details for specific facilities participating in the trial.
Other Identifiers
Review additional registry numbers or institutional identifiers associated with this trial.
EC/05/23/2282
Identifier Type: -
Identifier Source: org_study_id
More Related Trials
Additional clinical trials that may be relevant based on similarity analysis.