The Left Bundle Cardiac Resynchronization Therapy Trial
NCT ID: NCT05434962
Last Updated: 2025-07-17
Study Results
The study team has not published outcome measurements, participant flow, or safety data for this trial yet. Check back later for updates.
Basic Information
Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.
ACTIVE_NOT_RECRUITING
NA
176 participants
INTERVENTIONAL
2022-06-27
2025-08-31
Brief Summary
Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.
Related Clinical Trials
Explore similar clinical trials based on study characteristics and research focus.
Left Bundle Branch Area Pacing Using Conventional Stylet-driven Pacemaker Leads for Cardiac Resynchronization Therapy
NCT05365568
Left vs Left Randomized Clinical Trial
NCT05650658
Left Bundle Branch Pacing Versus Right Ventricular Pacing in Patients With Atrioventricular Block
NCT05722379
Acute Effects of LBBP Versus BVP for CRT
NCT04505384
Safety and Effectiveness of Left Bundle Branch Pacing in Patients With Cardiac Dysfunction and AV Block
NCT05553626
Detailed Description
Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.
Indication: Pacing and sensing in the right atrium or ventricle. It is also designed for pacing and sensing in the bundle as an alternative to pacing in the right ventricle in a single or dual chamber pacing system.
Intended use: Patients with an indication for cardiac resynchronization therapy.
Duration of the study: Approximate inclusion period of 18 months
Follow-up: Patients will be followed for 12 months ± 4 weeks after implant
Planned study period: The study is planned to start in Q1 2022
No. of sites: Approximately 11 sites in Spain.
No. of subjects: 176 patients will be enrolled in the study
Study type: Post-market study
Study rationale: Previous studies show 85% implantation success, significant reduction in QRS duration, and significant improvement in NYHA functional class, LVEF, and LV end-diastolic diameter. Left bundle branch pacing could be an alternative to conventional biventricular pacing.
Study design: Multicenter, prospective, randomized, non-inferiority, open.
Objectives of the study:
* Primary objective: The primary objective of the study is to demonstrate that LBBAP is associated with a non-inferior percentage of CRT responders at 6 months follow-up than conventional BiV pacing using a conventional CS lead in patients with CRT indications according to current guidelines. Positive CRT response will be defined either by an improved Clinical Composite Score or ≥15% reduction in left ventricular end-systolic volume at 6 months follow-up
* Secondary objectives: To evaluate reverse ventricular remodeling response and dyssynchrony parameters based on echocardiographic data at 6 and 12 months follow-up. To asess clinical outcome using the Clinical Composite Score, 6 minute walking test, quality of life measurements (EuroQuol EQ 5-D) at 6 and 12 months follow-up. To evaluate HF related hospitalizations at 6 and 12 months follow-up. To evaluate mortality (global and cardiovascular) and need for cardiac transplantation at 6 and 12 month follow-up. To evaluate incidence of ventricular arrhythmias at 6 and 12 month follow-up. To evaluate differences in CRT response depending on the final LBBAP modality (LBBP vs deep septal pacing). To evaluate device related complications at 6 and 12 month follow-up.
Primary Outcome Measures: Improved Clinical Composite Score or ≥15% reduction in left ventricular end-systolic volume at 6 months follow-up.
Participating sites:
* Hospital Universitario y Politécnico La Fe, Valencia
* Hospital Clinic de Barcelona
* Hospital Virgen de las Nieves, Granada
* Hospital Virgen de la Arrixaca, Murcia
* Hospital Puerta de Hierro, Madrid
* Hospital General de Alicante
* Hospital Bellvitge, Barcelona
* Hospital 12 de Octubre, Madrid
* Hospital Universitario Lozano Blesa, Zaragoza
* Hospital Juan Ramón Jiménez, Huelva
* Hospital Universitario San Cecilio, Granada
Study population: Subjects 18 years of age or older will be included in the study. Both sexes are eligible to participate. Subjects must present a left bundle branch block (according to Strauss criteria) with indication of CRT class I or IIa according to clinical practice guidelines. A selection of the subjects will be made before inclusion following the selection clinical investigation plan. After confirming that the subject meets the inclusion criteria and does not meet the exclusion criteria, they will be included in the clinical study.
Study population criteria:
* Inclusion criteria: age \> 18 years; Class I or IIa indication for CRT according to current ESC or ACC/AHA/HRS guidelines; left bundle branch block according to Strauss criteria; Signed and dated informed consent.
* Exclusion criteria: Pregnant woman; Permanent or long-lasting atrial fibrillation; Previous CRT device; Previous PM/ICD with ventricular pacing \> 10%; Conventional pacemaker indication; Nonspecific IVCD (intraventricular conduction delay); Unstable angina, acute Myocardial Infarction (MI), Coronary Artery Bypass Graft (CABG), Percutaneous Coronary Angioplasty (PCI), valve repair or replacement within 90 days prior enrollment; Indication for valve repair or replacement; Already included in another clinical study that could confoud the results of the present study; Life expectancy \< 12 months.
Device Implantation Procedure: CRT obtained by biventricular pacing vs. CRT obtained by stimulating the left branch area.
Follow-up: Follow-up should be carried out with pre-established periodic clinical appointments 3, 6, and 12 months after hospital discharge. In the follow-up, routine examinations such as vital signs, electrocardiogram and echocardiogram (6 and 12 months) are recorded to check the evaluation of the reverse ventricular remodeling response and asynchrony parameters based on echocardiographic data.
Statistical considerations: Once patients have signed the informed consent at baseline visit, patients will be considered as "Enrolled population" in the statistical analysis. The sample size is based on the primary endpoint: Improved Clinical Composite Score or ≥15% reduction in left ventricular end-systolic volume at 6 months follow-up, showing that LBBAP is non-inferior to BiV-CRT. Accounting for a global attrition rate of 10% and considering the block size for randomization, 176 patients shall be enrolled.
Conditions
See the medical conditions and disease areas that this research is targeting or investigating.
Study Design
Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.
RANDOMIZED
PARALLEL
TREATMENT
NONE
Study Groups
Review each arm or cohort in the study, along with the interventions and objectives associated with them.
Cardiac resynchronization therapy (CRT) obtained by stimulating the left branch area
The CRT-P/CRT-D device and leads should be implanted according to the physician's standard practice. Only locally approved Medtronic CRT-P/CRT-D generators will be used during the study. It is strongly encouraged the designation of a single and experienced implanting physician at each center for the LBBAP implant procedures.
During the implant procedure lead impedances, pacing and sensing parameters will be measured using a pacing system analyzer (PSA).
For LBBAP, the Medtronic 3830 lead will be used. Acceptable LBBAP threshold should be \<2,5V@0,5ms.
Cardiac resynchronization therapy
Standard BiV-CRT The CRT-P/CRT-D device and leads should be implanted according to the physician's manual provided with the devices. Only locally approved Medtronic CRT-P/CRT-D generators will be used during the study. The LV lead should be implanted at a lateral or posterolateral CS branch in a basal or mid ventricular position confirmed by ortogonal fluoroscopic views (RAO and LAO 40º). Acceptable LV pacing threshold should be \<3V@0,5ms and phrenic nerve stimulation (PNS) margin should be \>1V with respect to the pacing threshold.
LBBAP-CRT The CRT-P/CRT-D device and leads should be implanted according to the physician's standard practice. Only locally approved Medtronic CRT-P/CRT-D generators will be used during the study. During the implant procedure lead impedances, pacing and sensing parameters will be measured using a pacing system analyzer (PSA). For LBBAP, the Medtronic 3830 lead will be used. Acceptable LBBAP threshold should be \<2,5V@0,5ms
Cardiac resynchronization therapy (CRT) obtained by biventricular pacing
The CRT-P/CRT-D device and leads should be implanted according to the physician's manual provided with the devices. Only locally approved Medtronic CRT-P/CRT-D generators will be used during the study.
Investigators may use any market approved right atrial (RA) pace/sense lead, right ventricular (RV) pacing/defibrillator lead with pace/sense capabilities and any market approved unipolar/bipolar/quadripolar LV pacing lead.
During the implant procedure lead impedances, pacing and sensing parameters will be measured using a pacing system analyzer (PSA).
The LV lead should be implanted at a lateral or posterolateral CS branch in a basal or mid ventricular position confirmed by ortogonal fluoroscopic views (RAO and LAO 40º).
Acceptable LV pacing threshold should be \<3V@0,5ms and phrenic nerve stimulation (PNS) margin should be \>1V with respect to the pacing threshold
Cardiac resynchronization therapy
Standard BiV-CRT The CRT-P/CRT-D device and leads should be implanted according to the physician's manual provided with the devices. Only locally approved Medtronic CRT-P/CRT-D generators will be used during the study. The LV lead should be implanted at a lateral or posterolateral CS branch in a basal or mid ventricular position confirmed by ortogonal fluoroscopic views (RAO and LAO 40º). Acceptable LV pacing threshold should be \<3V@0,5ms and phrenic nerve stimulation (PNS) margin should be \>1V with respect to the pacing threshold.
LBBAP-CRT The CRT-P/CRT-D device and leads should be implanted according to the physician's standard practice. Only locally approved Medtronic CRT-P/CRT-D generators will be used during the study. During the implant procedure lead impedances, pacing and sensing parameters will be measured using a pacing system analyzer (PSA). For LBBAP, the Medtronic 3830 lead will be used. Acceptable LBBAP threshold should be \<2,5V@0,5ms
Interventions
Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.
Cardiac resynchronization therapy
Standard BiV-CRT The CRT-P/CRT-D device and leads should be implanted according to the physician's manual provided with the devices. Only locally approved Medtronic CRT-P/CRT-D generators will be used during the study. The LV lead should be implanted at a lateral or posterolateral CS branch in a basal or mid ventricular position confirmed by ortogonal fluoroscopic views (RAO and LAO 40º). Acceptable LV pacing threshold should be \<3V@0,5ms and phrenic nerve stimulation (PNS) margin should be \>1V with respect to the pacing threshold.
LBBAP-CRT The CRT-P/CRT-D device and leads should be implanted according to the physician's standard practice. Only locally approved Medtronic CRT-P/CRT-D generators will be used during the study. During the implant procedure lead impedances, pacing and sensing parameters will be measured using a pacing system analyzer (PSA). For LBBAP, the Medtronic 3830 lead will be used. Acceptable LBBAP threshold should be \<2,5V@0,5ms
Eligibility Criteria
Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.
Inclusion Criteria
* Class I or IIa indication for CRT according to current ESC or ACC/AHA/HRS guidelines
* left bundle branch block according to Strauss criteria
* Signed and dated informed consent
Exclusion Criteria
* Permanent or long-lasting atrial fibrillation
* Previous CRT device
* Previous PM/ICD with ventricular pacing \> 10%
* Conventional pacemaker indication
* Nonspecific IVCD (intraventricular conduction delay)
* Unstable angina, acute Myocardial Infarction (MI), Coronary Artery Bypass Graft (CABG), Percutaneous Coronary Angioplasty (PCI), valve repair or replacement within 90 days prior enrollment
* Indication for valve repair or replacement
* Already included in another clinical study that could confoud the results of the present study
* Life expectancy \< 12 months
18 Years
ALL
No
Sponsors
Meet the organizations funding or collaborating on the study and learn about their roles.
Spanish Society of Cardiology
OTHER
Óscar Cano Pérez
OTHER
Responsible Party
Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.
Óscar Cano Pérez
Cardiologist
Locations
Explore where the study is taking place and check the recruitment status at each participating site.
Hospital General Universitario de Alicante Doctor Balmis
Alicante, Alicante, Spain
Hospital Universitario de Bellvitge
L'Hospitalet de Llobregat, Barcelona, Spain
Hospital Universitario Virgen de las Nieves
Granada, Granada, Spain
Hospital Universitario San Cecilio
Granada, Granada, Spain
Hospital Universitario Juan Ramón Jiménez
Huelva, Huelva, Spain
Hospital 12 Octubre
Madrid, Madrid, Spain
Hospital Universitario Puerta De Hierro
Majadahonda, Madrid, Spain
Hospital Universitario Virgen de la Arrixaca
Murcia, Murcia, Spain
Hospital Universitari i Politècnic La Fe
Valencia, Valencia, Spain
Hospital Clínico Universitario Lozano Blesa
Zaragoza, Zaragoza, Spain
Countries
Review the countries where the study has at least one active or historical site.
Other Identifiers
Review additional registry numbers or institutional identifiers associated with this trial.
LB-CRT-001