Different Cryocompression Devices and Skin Temperature of the Knee
NCT ID: NCT05355116
Last Updated: 2024-11-06
Study Results
Outcome measurements, participant flow, baseline characteristics, and adverse events have been published for this study.
View full resultsBasic Information
Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.
COMPLETED
NA
32 participants
INTERVENTIONAL
2022-04-20
2022-07-27
Brief Summary
Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.
The aim of this study is to determine the ability of five different cryocompression.devices to effectively lower the skin temperature of the treatment area to within the therapeutic range.
Related Clinical Trials
Explore similar clinical trials based on study characteristics and research focus.
Skin Temperature Changes When Using a Cryocompression Device
NCT05136482
Effect of Pressure on Skin Temperature When Using a Cryocompression Device
NCT05454982
Skin Temperature Reduction With Hilotherm Device
NCT05789290
Prolonged Cryocompression and Skin Temperature: a Safety and Feasibility Pilot
NCT06673277
Combined Cryotherapy With Compression Versus Cryotherapy Alone After Orthopaedic Surgery
NCT05011084
Detailed Description
Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.
Protocol All eligible participants will first be required to provide informed consent prior to their participation in the study. Once this has been obtained, participants will each attend 5 individual testing sessions, each lasting no longer than 60 minutes. The height and mass of the participants will be measured during the first testing session in order to calculate BMI, and participants' age and sex will be recorded and included in the later data analysis.
Participants will be required to remain in a seated position with their leg in full extension and elevated, parallel to the floor, for the duration of each test. The leg that will be used for the study will be randomly selected for each participant, and will remain the same for each condition. Skin temperature will be measured using a thermocouple, which will be attached with tape 20 mm distal to the patella. One of the cryocompression devices (Physiolab S1; Physiolab ice pack; GameReady; Cryo/Cuff; or VPulse) will then be attached to the leg of the participant and the test will begin. Depending on the condition to which a participant has been randomly assigned for a given test, a different device will be used. The temperature and pressure settings for each test will be set according to the maximum capability of each device, which varies between devices. The lowest temperature setting that will be used throughout the study will be 1℃ (GameReady device) and the highest pressure setting that will be used is 50 mmHg (GameReady and Physiolab S1 devices). None of the temperature or pressure settings used are deemed to be unsafe or have a high risk of adverse reactions, as long as participants are not contraindicated according to the exclusion criteria for this study.
The temperature of the skin will be measured using a k-type thermocouple attached to an infrared thermometer immediately prior to the application of a device, and then every 5 minutes for the duration of the test. The length of a test will be 30 minutes, in accordance with the manufacturers' instructions/recommendations for each device.
Upon completion of a test, the cuff will be removed. If the temperature of the skin has been reduced to within 10-15℃, skin temperature will continue to be monitored every 5 minutes after the cuff has been removed until it rises above 15℃, at which point the test will end. This will allow for the total time to be measured that skin temperature remains within the target therapeutic range as a result of a single application of the cryocompression device. If the skin temperature is not reduced to within 10-15℃ while the cuff is worn, then the test will end after the cuff is initially removed. Upon completion of the test, the temperature sensor will be removed from the leg of the participant; thus completing their participation in the test. Once each participant has taken part in all conditions, they will have completed their participation in the study.
The following measures will be recorded: skin temperature at each time point, minimum skin temperature achieved, time to achieve minimum skin temperature, length of time with skin temperature within therapeutic range of 10-15℃. After each test, participants will also be asked to respond to the following question using a 5-point Likert scale: "How comfortable did you find the treatment you just experienced?" The possible answers to this question will be: "Very comfortable, Comfortable, Neutral, Uncomfortable, Very uncomfortable". This will provide insight into differences in perception of the treatment, which could be relevant for the likelihood of protocol adherence by patients in clinical settings.
Analysis The data will be analysed to detect any differences during testing compared to baseline measures, and between groups. All data will be first subjected to a Kolmogorov-Smirnov test to assess whether they are normally distributed. A repeated measures analysis of variance will be performed on all normally distributed data. A Friedman test will be performed on any data that are not normally distributed.
Ethical issues There is a low risk of cold injury to participants if their skin temperature is reduced to \<10℃ for any prolonged period of time (\>60 mins). The coldest temperature setting for this study is 1℃ (GameReady), which will be applied for 30 mins. However, device temperature settings do not equal the actual skin temperatures achieved, and it is not expected that skin temperatures will be reduced to lower than 5℃ for more than 15 minutes during any test. Therefore, the risk of harm to the participant is deemed to be minimal for this study.
Participants have the right to withdraw themselves and any collected data from the study at any time during, and for the first 7 days after, their participation without having to give a reason. They also have the right to terminate a test session at any time. Testing shall also be terminated as a result of any adverse reactions that emerge. Adverse reactions (e.g. pain) to the testing protocol are not expected due to the low-risk nature of the study, however any that occur shall be recorded and monitored until things return to baseline/normal. In order to monitor any pain/discomfort that might occur, a Numerical Pain Rating Scale (0-10 scale) will be within sight of the participant at all times during testing: a test will be terminated if reported pain/discomfort exceeds 5/10, though none is expected. Any reported pain (or other adverse reactions) shall be stored along with the data collected for that participant. If a test is terminated for a reason that could be mitigated in future, participants will be offered the opportunity to repeat the test another day, should they wish to continue taking part in the study. If a test is terminated due to an adverse reaction that can not be mitigated in future, the participant (and any collect-ed data) will be thanked for their time and withdrawn from the study.
Conditions
See the medical conditions and disease areas that this research is targeting or investigating.
Study Design
Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.
RANDOMIZED
CROSSOVER
TREATMENT
NONE
Study Groups
Review each arm or cohort in the study, along with the interventions and objectives associated with them.
Group A: Physiolab S1
The cryocompression device will be attached to the lower limb of participants using a cuff spanning from the mid-thigh to mid-calf. The device will exert an intermittent pressure of 25-50 mmHg throughout each test session. The device will pump temperature-controlled cold water through the cuff in order to reduce the skin temperature around the knee (and intra-articular temperature). The temperature of the water being pumped through the device will be 8℃. Each test session will last for 30 minutes. Participants will take part in all conditions, with at least 24 hours rest in between test sessions. Skin temperature around the knee will be measured prior to the cryocompression device being applied; every 5 minutes during the 30 minute test; and every 5 minutes after the test until skin temperature \>15℃.
Physiolab S1
A cryocompression device capable of circulating ice-water through a cuff at 6-12℃ with either a static pressure of 25mmHg or dynamic pressure of 25-50 mmHg or 25-75mmHg. The lowest temperature setting, and highest pressure setting will be used for this study.
Group B: Breg Vpulse
The cryocompression device will be attached to the lower limb of participants using a cuff spanning from the mid-thigh to mid-calf. The device will exert a dynamic peak pressure of 50 mmHg throughout each test session. The device will pump temperature-controlled cold water through the cuff in order to reduce the skin temperature around the knee (and intra-articular temperature). The temperature of the water being pumped through the device will not be lower than 5.5℃. The selected pressure and temperature represents the maximum capability of this device. Each test session will last for 30 minutes. Participants will take part in all conditions, with at least 24 hours rest in between test sessions. Skin temperature around the knee will be measured prior to the cryocompression device being applied; every 5 minutes during the 30 minute test; and every 5 minutes after the test until skin temperature \>15℃.
Breg Vpulse
A cryocompression device capable of circulating ice-water through a cuff at no lower than 5.5℃ with a dynamic peak pressure of 50 mmHg. The temperature of the ice-water, and the applied pressure, is not modifiable by the user.
Group C: Cryo/Cuff
The cryocompression device will be attached to the lower limb of participants using a cuff spanning from the mid-thigh to mid-calf. The pressure and temperature that the device applies to the treatment area is non-modifiable and undefined. The device will be used according to the manufacturer's recommendations, which will represent the maximum capability of this device. Each test session will last for 30 minutes. Participants will take part in all conditions, with at least 24 hours rest in between test sessions. Skin temperature around the knee will be measured prior to the cryocompression device being applied; every 5 minutes during the 30 minute test; and every 5 minutes after the test until skin temperature \>15℃.
Aircast Cryo/Cuff
A cryocompression device capable of circulating ice-water through a cuff at an undefined, non-modifiable temperature with a static pressure that is also undefined and non-modifiable. The device will be applied as per the manufacturer's instructions, which will represent its maximum capability.
Group D: GameReady
The cryocompression device will be attached to the lower limb of participants using a cuff spanning from the mid-thigh to mid-calf. The device will exert an intermittent pressure of 5-50 mmHg throughout each test session. The device will pump temperature-controlled cold water through the cuff in order to reduce the skin temperature around the knee (and intra-articular temperature). The temperature of the water being pumped through the device will be 1℃. Each test session will last for 30 minutes. Participants will take part in all conditions, with at least 24 hours rest in between test sessions. Skin temperature around the knee will be measured prior to the cryocompression device being applied; every 5 minutes during the 30 minute test; and every 5 minutes after the test until skin temperature \>15℃.
GameReady
A cryocompression device capable of circulating ice-water through a cuff at 1-12℃ with a dynamic pressure of 5-15 mmHg, 5-50 mmHg, or 5-75 mmHg. The lowest temperature setting, and highest pressure setting will be used for this study.
Group E: Physiolab Gel Therapy Wrap
The cryocompression device will be attached to the lower limb of participants using a cuff spanning from the mid-thigh to mid-calf. The pressure and temperature that the device applies to the treatment area is non-modifiable and undefined. The device will be used according to the manufacturer's recommendations, which will represent the maximum capability of this device. Each test session will last for 30 minutes. Participants will take part in all conditions, with at least 24 hours rest in between test sessions. Skin temperature around the knee will be measured prior to the cryocompression device being applied; every 5 minutes during the 30 minute test; and every 5 minutes after the test until skin temperature \>15℃.
Physiolab Gel Therapy Wrap
A cryocompression cuff that applies an undefined, non-modifiable temperature with a static pressure that is also undefined and non-modifiable. The device will be applied as per the manufacturer's instructions, which will represent its maximum capability.
Interventions
Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.
Physiolab S1
A cryocompression device capable of circulating ice-water through a cuff at 6-12℃ with either a static pressure of 25mmHg or dynamic pressure of 25-50 mmHg or 25-75mmHg. The lowest temperature setting, and highest pressure setting will be used for this study.
Breg Vpulse
A cryocompression device capable of circulating ice-water through a cuff at no lower than 5.5℃ with a dynamic peak pressure of 50 mmHg. The temperature of the ice-water, and the applied pressure, is not modifiable by the user.
Aircast Cryo/Cuff
A cryocompression device capable of circulating ice-water through a cuff at an undefined, non-modifiable temperature with a static pressure that is also undefined and non-modifiable. The device will be applied as per the manufacturer's instructions, which will represent its maximum capability.
GameReady
A cryocompression device capable of circulating ice-water through a cuff at 1-12℃ with a dynamic pressure of 5-15 mmHg, 5-50 mmHg, or 5-75 mmHg. The lowest temperature setting, and highest pressure setting will be used for this study.
Physiolab Gel Therapy Wrap
A cryocompression cuff that applies an undefined, non-modifiable temperature with a static pressure that is also undefined and non-modifiable. The device will be applied as per the manufacturer's instructions, which will represent its maximum capability.
Eligibility Criteria
Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.
Inclusion Criteria
Exclusion Criteria
* History of nerve damage or sensory deficit in the lower limbs (including frostbite)
* Hypersensitivity to cold, including hives
* Active inflammation or pain of the knee
* History of thrombosis, embolism, or other conditions related to impaired peripheral circulation
* Suffering from diagnosed diabetes, multiple sclerosis, rheumatoid arthritis, spinal cord injury, cardio-vascular disease, hypertension, Raynaud disease, cryoglobulinemia, or haemoglobinuria
* Confirmed or suspected tissue infection, an unstable fracture, a skin condition, or a tumour in the treatment area
* Cognitive impairment or communication barriers where adjustments can not be reasonably made
18 Years
ALL
Yes
Sponsors
Meet the organizations funding or collaborating on the study and learn about their roles.
Physiolab Technologies Ltd
INDUSTRY
University of Winchester
OTHER
Responsible Party
Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.
Principal Investigators
Learn about the lead researchers overseeing the trial and their institutional affiliations.
James Faulkner, PhD
Role: PRINCIPAL_INVESTIGATOR
University of Winchester
Locations
Explore where the study is taking place and check the recruitment status at each participating site.
Physiology Laboratory
Winchester, , United Kingdom
Countries
Review the countries where the study has at least one active or historical site.
References
Explore related publications, articles, or registry entries linked to this study.
Wilke B, Weiner RD. Postoperative cryotherapy: risks versus benefits of continuous-flow cryotherapy units. Clin Podiatr Med Surg. 2003 Apr;20(2):307-22. doi: 10.1016/S0891-8422(03)00009-0.
Bleakley C, McDonough S, MacAuley D. The use of ice in the treatment of acute soft-tissue injury: a systematic review of randomized controlled trials. Am J Sports Med. 2004 Jan-Feb;32(1):251-61. doi: 10.1177/0363546503260757.
Fang L, Hung CH, Wu SL, Fang SH, Stocker J. The effects of cryotherapy in relieving postarthroscopy pain. J Clin Nurs. 2012 Mar;21(5-6):636-43. doi: 10.1111/j.1365-2702.2010.03531.x. Epub 2011 Feb 20.
Waterman B, Walker JJ, Swaims C, Shortt M, Todd MS, Machen SM, Owens BD. The efficacy of combined cryotherapy and compression compared with cryotherapy alone following anterior cruciate ligament reconstruction. J Knee Surg. 2012 May;25(2):155-60. doi: 10.1055/s-0031-1299650.
Bellon M, Michelet D, Carrara N, Vacher T, Gafsou B, Ilhareborde B, Mazda K, Ferrero E, Simon AL, Julien-Marsollier F, Dahmani S. Efficacy of the Game Ready(R) cooling device on postoperative analgesia after scoliosis surgery in children. Eur Spine J. 2019 Jun;28(6):1257-1264. doi: 10.1007/s00586-019-05886-6. Epub 2019 Jan 31.
Selfe J, Hardaker N, Whittaker J and Hayes C. An investigation into the effect on skin surface temperature of three cryotherapy modalities. Thermology International. 2009; 19(4): 121-126.
Provided Documents
Download supplemental materials such as informed consent forms, study protocols, or participant manuals.
Document Type: Study Protocol and Statistical Analysis Plan
Other Identifiers
Review additional registry numbers or institutional identifiers associated with this trial.
HWB_REC_ 220228_Faulkner
Identifier Type: -
Identifier Source: org_study_id
More Related Trials
Additional clinical trials that may be relevant based on similarity analysis.