Clinical Trial Testing the BioSerenity Pulse Oximeter (EOXY) Efficacy, Performance and Safety in Healthy Volunteer
NCT ID: NCT03614416
Last Updated: 2021-01-12
Study Results
Outcome measurements, participant flow, baseline characteristics, and adverse events have been published for this study.
View full resultsBasic Information
Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.
COMPLETED
NA
13 participants
INTERVENTIONAL
2017-12-08
2018-11-06
Brief Summary
Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.
Thirteen healthy subjects will participate in the study in total. Three subjects will be enrolled during a primary phase .They will participate at a secondary phase as well as ten others subjects. The purpose of the primary phase is to test the body area where the measurement can be done. The purpose of the secondary phase is to measure SpO2 and HR in desaturation conditions.
Related Clinical Trials
Explore similar clinical trials based on study characteristics and research focus.
Evaluation of the Signals Electrophysiological Measured by Intelligent Textiles
NCT03805243
Mobile Biosensor for Measuring Vital Signs in Healthcare and Home Settings
NCT04233827
SpO2 System Accuracy Testing With Different Sensors
NCT01613222
BiOSENCY BORA Band SpO2 Validation Study
NCT03918018
A Clinical Trial to Verify the Accuracy of Seers Technology's Pulse Oximeter (mobiCARE +Pulse, MP100W) in Healthy Volunteers
NCT05895682
Detailed Description
Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.
Subject safety during desaturation and saturation period, will be evaluated by reviewing vital sign: heart rate (HR) continuously and blood pressure/temperature between each desaturation plateau.
There are two secondaries objectives, the measurement of the variation of the fraction of inspired oxygen (FiO2), and the validation of the algorithm for data review and analysis.
For the study primary phase, the purpose of this phase is to test the body area where the SpO2 is better detected and define 3 measurement areas. For this, three healthy volunteers will be placed in normal O2 saturation, the SpO2 will be measured using the Bioserenity oximeter (noninvasive method) and will be compared to a gold-standard oximeter CE-marked. The investigator will determine the best area in the body where the SpO2 could be taken for the following subjects.
During a second phase, the measurement of Sp02 and HR with EOXY, will be performed in O2 saturation conditions up to 100% and then in desaturation by plateau from 100 % to 70%. 3 EOXY pulse oximeter recorded the SpO2 at the same time in 3 different areas of the body. As required by the standard, Sp02 measures will be simultaneously compared to SaO2 values measured by blood sampling taken from an indwelling arterial catheter. The HR obtained with EOXY will be compared simultaneously to a gold standard oximeter CE-marked.
As the example in the standard, it was choosen to do the test on the 70-100% range, with 5 plateaus (100%-97%, 97%-92%, 92%-85%, 84%-78%, 77%-70%) and to collect five Sp02 and Sa02 values (blood samples) and five HR measures, per plateau, for each of 13 subjects.
The desaturation will be controlled thanks' the FiO2 measures, that should allow to bring subjects near target levels of desaturation plateau.
Conditions
See the medical conditions and disease areas that this research is targeting or investigating.
Study Design
Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.
NA
SINGLE_GROUP
OTHER
NONE
Study Groups
Review each arm or cohort in the study, along with the interventions and objectives associated with them.
EOXY device and Gold standard oximter and SaO2 measures
Heart rate and SPO2 measures provided from EOXY device Heart rate measures provided from gold standard oximeter. SaO2 measures provided from blood sampling There is only one arm: all subjects have simultaneously three interventions (as required in the European standard ISO 80601-2-61), to qualify a pulse oximeter
EOXY device
Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
Gold standard oximeter
Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
SaO2 sampling
Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
Interventions
Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.
EOXY device
Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
Gold standard oximeter
Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
SaO2 sampling
Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard
Eligibility Criteria
Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.
Inclusion Criteria
* Subject signed informed consent prior to any screening procedure
* Subject in good health (COHb \< 3% , MetHb \< 2%, CtHb \> 10mg/dL)
* Healthy adult capable of undergoing controlled hypoxemia to the level called for in the protocol with minimal medical risk
Exclusion Criteria
* Individual subject to conditions that result in elevated levels of methemoglobin
* Subject who would be placed at undue medical risk associated with any procedure called for in the protocol
* Subject with open wounds
* Pregnant woman (negative pregnancy test needed)
* Subject allergic to silicon, polyamide and silver yarn
* Subject with cardiorespiratory disorders likely to worsen with thoracic pressure applied by the textile band
* Subject with mental or motor impairment preventing him from expressing pain
* Subject with blood or skin disorder that may impacting results
* Subject with behavioral disorders, too agitated or too aggressive
* Subject with sensorial disorders, insensible to skin pain
* Subject susceptible to tension/pressure based headaches
18 Years
ALL
Yes
Sponsors
Meet the organizations funding or collaborating on the study and learn about their roles.
NAMSA
OTHER
Université Catholique de Louvain
OTHER
BioSerenity
INDUSTRY
Responsible Party
Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.
Principal Investigators
Learn about the lead researchers overseeing the trial and their institutional affiliations.
marc FRANCAUX
Role: PRINCIPAL_INVESTIGATOR
UNIVERSITE CATHOLIQUE DE LOUVAIN
Locations
Explore where the study is taking place and check the recruitment status at each participating site.
Université catholique louvain la neuve, centre sportif
Louvain-la-Neuve, , Belgium
Countries
Review the countries where the study has at least one active or historical site.
Provided Documents
Download supplemental materials such as informed consent forms, study protocols, or participant manuals.
Document Type: Study Protocol and Statistical Analysis Plan
Other Identifiers
Review additional registry numbers or institutional identifiers associated with this trial.
BIOSERENITY_80601-2-61
Identifier Type: -
Identifier Source: org_study_id
More Related Trials
Additional clinical trials that may be relevant based on similarity analysis.