SpO2 Accuracy Comparison of Medline ReNewal Sensors to Arterial Blood CO-Oximetry

NCT ID: NCT03565679

Last Updated: 2019-06-18

Study Results

Results available

Outcome measurements, participant flow, baseline characteristics, and adverse events have been published for this study.

View full results

Basic Information

Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.

Recruitment Status

COMPLETED

Clinical Phase

NA

Total Enrollment

10 participants

Study Classification

INTERVENTIONAL

Study Start Date

2018-05-09

Study Completion Date

2018-05-10

Brief Summary

Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.

The purpose of this clinical study is to validate the SpO2 accuracy of the Medline ReNewal pulse oximetry sensors during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. The end goal is to provide supporting documentation for the SpO2 accuracy validation of the ReNewal sensors.

It is required that the Accuracy Root Mean Square (ARMS) performance of the ReNewal pulse oximetry sensors will meet a specification of 3 or better in non-motion conditions for the range of 70 - 100% SaO2 thereby demonstrating an acceptable SpO2 accuracy performance specification.

Detailed Description

Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. Pulse oximetry monitoring is considered a standard physiological measurement and is used by clinicians in everyday situations to estimate arterial oxygen saturation. Because an arterial sample of blood is not required to make the measurement, the pulse oximeter can provide non-invasive real time information. The clinical use of pulse oximeters has reduced the frequency and necessity of invasive arterial blood sampling and has helped to improve patient safety by providing continuous information to clinicians about patients' oxygenation status.

The Medline ReNewal Reprocessed pulse oximeter sensors are indicated for single patient use when continuous noninvasive arterial oxygen saturation and pulse rate monitoring are required. Following clinical use, each sensor is returned to Medline ReNewal where it is cleaned, disinfected, and packaged for an additional clinical use. Each sensor is tracked through the reprocessing cycle to monitor the number of times the sensor has been reprocessed. The purpose of this study will be to test the accuracy of the SpO2 component of the devices.

The SpO2 accuracy performance of ReNewal pulse oximetry sensors will be evaluated during non-motion conditions over the range of 70-100% SaO2 and compared to arterial blood samples assessed by CO-Oximetry. A minimum of 10 healthy adult subjects, ranging in pigmentation from light to dark, will be enrolled in the study to meet the study design requirements defined by International Organization for Standardization (ISO) 80601-2-61:2011 and by the FDA's Guidance for Pulse Oximeters (March 4, 2013). The subjects will have an arterial catheter placed in the radial artery to allow for simultaneous blood samples during stable plateaus of induced hypoxic levels. The investigational devices will be placed on the fingers of both hands for the test sites per the instructions for use. Simultaneous data collection will be set up for each of the systems under test. For the data analysis, the control oximeter will be used to assess the stability of each data point. Data that is found to be unstable will be removed prior to the comparative analysis. Next the CO-Oximeter data will be reviewed to make sure it does not contain any anomalous values such as elevated, low or inconsistent carboxyhemoglobin (COHb), methemoglobin (MetHb), or total hemoglobin (tHb) data. Anomalous values will be removed from the analysis prior to pairing of the SpO2 and SaO2 data. The Accuracy Root Mean Square (ARMS) calculation is used to determine the SpO2 accuracy performance. Success will be achieved with an ARMS of 3 or better showing equivalence to the Gold Standard Reference CO-Oximetry providing documentation to support SpO2 accuracy claims for the investigational device(s).

The study population will include 10-15 healthy non-smoking (or has refrained from smoking for 2 days) competent adults 18-50 years of age. The subject selection will be an equitable distribution of males and females of any race with varying skin tones including at least 2 darkly pigmented subjects or 15% of the subject pool, whichever is larger. Data collection will occur over a 2-5 day period for this study population. Follow-up with each subject will be conducted within 5 days following participation in the study and will be conducted via telephone, text or email.

Conditions

See the medical conditions and disease areas that this research is targeting or investigating.

Arterial Oxygen Saturation Healthy

Study Design

Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.

Allocation Method

NON_RANDOMIZED

Intervention Model

PARALLEL

Subjects will recline for the study. Reference sensors will be placed on each subject to evaluate the SpO2 accuracy and performance. Shield material may be used between any adjacent finger sensors to prevent optical crosstalk. Simultaneous data collection will be set up for devices under test. The data from the test devices will be collected by the sponsor or trained Clinimark Study staff. Data will be collected for 1 second intervals data analysis. The SpO2 accuracy of the test devices will be evaluated over the oxygen saturation range between 70-100%.
Primary Study Purpose

DEVICE_FEASIBILITY

Blinding Strategy

NONE

Study Groups

Review each arm or cohort in the study, along with the interventions and objectives associated with them.

Masimo SpO2 Adhesive Sensors, Adtx (1859) & (2329)

Reference sensors from the reprocessed oximeter device will be placed on each subject to evaluate the SpO2 accuracy and performance.

Group Type EXPERIMENTAL

MEDLINE RENEWAL PULSE OXIMETRY SENSORS

Intervention Type DEVICE

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. Pulse oximetry monitoring is considered a standard physiological measurement and is used by clinicians in everyday situations to estimate arterial oxygen saturation. Because an arterial sample of blood is not required to make the measurement, the pulse oximeter can provide non-invasive real time information.

Covidien Nellcor SpO2 Sensor, MAX-A and MAX-N

A whole blood analyzer (CO-Oximeter) is used as the reference standard device for obtaining the functional SaO2 value from arterial blood samples obtained during the study.

Group Type SHAM_COMPARATOR

CO-OXIMETRY SENSORS

Intervention Type DEVICE

A whole blood analyzer (CO-Oximeter) is used as the reference standard device for obtaining the functional SaO2 value from arterial blood samples obtained during the study.

Interventions

Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.

MEDLINE RENEWAL PULSE OXIMETRY SENSORS

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. Pulse oximetry monitoring is considered a standard physiological measurement and is used by clinicians in everyday situations to estimate arterial oxygen saturation. Because an arterial sample of blood is not required to make the measurement, the pulse oximeter can provide non-invasive real time information.

Intervention Type DEVICE

CO-OXIMETRY SENSORS

A whole blood analyzer (CO-Oximeter) is used as the reference standard device for obtaining the functional SaO2 value from arterial blood samples obtained during the study.

Intervention Type DEVICE

Eligibility Criteria

Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.

Inclusion Criteria

* 10-15 Adults with a minimum of 3 males and a minimum of 3 females, with the balance made up of either
* Subject must have the ability to understand and provide written informed consent
* Subject is 18 to 50 years of age
* Subject must be willing and able to comply with study procedures and duration
* Subject is a non-smoker or who has not smoked within 2 days prior to the study

Exclusion Criteria

* Subject is considered as being morbidly obese (defined as BMI \>39.5)
* Compromised circulation, injury, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)
* Females who are pregnant, who are trying to get pregnant, or have a urine test positive for pregnancy on the day of the study
* Smoker Subjects who have refrained will be screened for COHb levels \>3% as assessed with a Masimo Radical 7 (Rainbow)
* Subjects with known respiratory conditions such as: (self-reported)

* uncontrolled / severe asthma,
* flu,
* pneumonia / bronchitis,
* shortness of breath / respiratory distress,
* respiratory or lung surgery,
* emphysema, chronic obstructive pulmonary disease (COPD), lung disease
* Subjects with known heart or cardiovascular conditions such as: (self-reported, except for blood pressure and ECG review)

* hypertension: systolic \>140mmHg, Diastolic \>90mmHg on 3 consecutive readings (reviewed during health screen).
* have had cardiovascular surgery
* Chest pain (angina)
* heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia (reviewed during health screen)
* previous heart attack
* blocked artery
* unexplained shortness of breath
* congestive heart failure (CHF)
* history of stroke
* transient ischemic attack
* carotid artery disease
* myocardial ischemia
* myocardial infarction
* cardiomyopathy
* Self-reported health conditions as identified in the Health Assessment Form (self-reported)

* diabetes,
* uncontrolled thyroid disease,
* kidney disease / chronic renal impairment,
* history of seizures (except childhood febrile seizures),
* epilepsy,
* history of unexplained syncope,
* recent history of frequent migraine headaches,
* recent head injury
* cancer / chemotherapy
* Subjects with known clotting disorders (self-reported)

* history of bleeding disorders or personal history of prolonged bleeding from injury
* history of blood clots
* hemophilia
* current use of blood thinner: prescription or daily use of aspirin
* Subjects with severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors (self-reported)
* Subjects with prior or known allergies to iodine or lidocaine (or similar pharmacological agents, e.g. Novocaine)
* Failure of the Perfusion Index Ulnar/Ulnar+Radial Ratio test
* Unwillingness or inability to remove colored nail polish from test digits.
* Other known health condition, should be considered upon disclosure in health assessment form
Minimum Eligible Age

18 Years

Maximum Eligible Age

50 Years

Eligible Sex

ALL

Accepts Healthy Volunteers

Yes

Sponsors

Meet the organizations funding or collaborating on the study and learn about their roles.

Clinimark, LLC

OTHER

Sponsor Role collaborator

Medline Industries

INDUSTRY

Sponsor Role lead

Responsible Party

Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.

Responsibility Role SPONSOR

Principal Investigators

Learn about the lead researchers overseeing the trial and their institutional affiliations.

Arthur Cabrera, MD

Role: PRINCIPAL_INVESTIGATOR

Avista Adventist Hospital, Staff Anesthesiologist

Locations

Explore where the study is taking place and check the recruitment status at each participating site.

Clinimark Desaturation Laboratory, Site ID# 001

Louisville, Colorado, United States

Site Status

Countries

Review the countries where the study has at least one active or historical site.

United States

Provided Documents

Download supplemental materials such as informed consent forms, study protocols, or participant manuals.

Document Type: Study Protocol and Statistical Analysis Plan

View Document

Other Identifiers

Review additional registry numbers or institutional identifiers associated with this trial.

PR 2018-268

Identifier Type: -

Identifier Source: org_study_id