Hybrid Closed Loop Insulin Delivery System in Hypoglycemia
NCT ID: NCT03215914
Last Updated: 2024-10-24
Study Results
Outcome measurements, participant flow, baseline characteristics, and adverse events have been published for this study.
View full resultsBasic Information
Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.
COMPLETED
NA
10 participants
INTERVENTIONAL
2017-08-01
2024-10-01
Brief Summary
Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.
Related Clinical Trials
Explore similar clinical trials based on study characteristics and research focus.
Hybrid Closed Loop Pivotal Trial in Type 1 Diabetes
NCT02463097
Safety and Effectiveness Study of a Closed Loop System Maintaining Patients' Glucose Levels During an Overnight Period
NCT01712594
Closed-Loop Glucagon Administration For Hypoglycemia Treatment
NCT02181127
24/7 Closed-loop in Older Subjects With Type 1 Diabetes
NCT04025762
Four Way Crossover Closed Loop With Exercise Detection
NCT02862730
Detailed Description
Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.
Screening Phase At the screening visit, the study details and procedures will be discussed with a research coordinator and at least one of the PI or the research nurse practitioner. The potential participant is given adequate time to ask questions and review the informed consent document. Once satisfied that all questions have been answered, the potential participant will either decline to participate or sign the informed consent document. This may occur at a subsequent visit if the potential participant desires, in order to think further about what participation means and/or to consult with family, friends and/or a personal physician. The consent form is signed in the presence of a witness (research coordinator +/- family member). All participants must read, sign, and date a consent form before entering the study, undergoing physical examination or undergoing any testing. The informed consent form will be revised whenever important new safety information is available, whenever the protocol is amended, and/or whenever any new information becomes available that may affect participation in the studies.Eligibility will be confirmed through the performance of a history and physical examination by the PI or the research nurse practitioner, EKG, urine pregnancy test (if applicable), serum chemistries, TSH, cell counts, HbA1c and C-peptide, completion of glycemic lability and hypoglycemia awareness and hypoglycemia severity questionnaires, placement of a 7 day blinded CGM (iPro 2, Medtronic Diabetes, Northridge, CA) unless on CGM available for downloading, and 7 day accelerometry (WGT3X-BT, Actigraph LLC,) to define the nocturnal period.
Only after all eligibility criteria (inclusion and exclusion) are met, will a potential subject be enrolled. Repeated clinical testing throughout the study will ensure the continued safety and minimization of risk for the enrolled participants.
Study Intervention Phase Eligible subjects will complete a baseline assessment of glucose counterregulation by stepped- hyperinsulinemic hypoglycemic clamp prior to starting intervention with the hybrid closed-loop system (MiniMed 670G system, Medtronic Diabetes, Northridge, CA). Unless this system becomes available to the subject via their insurance carrier, one will be provided for them. Subjects will also receive a study glucometer (Contour Next Link 2.4, Bayer, Indianapolis, IN) that communicates with the MiniMed 670G insulin pump for bolus dosing calculation and glucose sensor calibration. Subjects who cannot maintain \> 80% (or 6/7 day) compliance with the sensor component as assessed at each study visit may be dropped since less compliance has not been associated with any benefit of CGM to glycemic control and limits the potential for benefit from LGS on hypoglycemia avoidance. Accuracy of the sensor will be assessed at each visit through devise download and interpretation. Study visits will occur weekly for the first month, then monthly until month 6, and then every 3 months until month 18. This schedule will allow for determination of possible benefit from hybrid closed- loop insulin delivery on glucose counterregulation after 6 months of intensive provider support, and then for assessment of the durability or potential further gains in beneficial effects after another 12 months of more typical provider interaction occurring every 3 months. Weekly visits may be performed via telephone with uploading devise data to Care Link for review and interpretation. Uploaded or downloaded insulin delivery, blood and sensor glucose monitoring, insulin dose settings and CGM calibration accuracy, alert settings, time spent in auto and manual modes, and LGS threshold and activity will be assessed at each visit, targeting \> 80% CGM and LGS compliance, adjusting basal and bolus insulin dosing in order to minimize glycemic excursions while maximizing hypoglycemia (\< 60 mg/dl) avoidance, with adjustment of alarms set to alert the subject to rapidly increasing or decreasing glucose and predict the occurrence of elevated or low blood glucose. During manual mode, target glucose ranges will be 90 - 140 mg/dl before meals, \< 180 mg/dl after meals, and 120 - 160 mg/dl at bedtime, with correction dosing to no lower than 100 mg/dl during the day, and 120 mg/dl overnight. Alarm settings may be individualized to target these ranges, but the hypoglycemia alarm for LGS will not be set lower than 70 mg/dl. During auto mode, the automated interprandial basal insulin delivery will adjust according to the closed-loop algorithm to target a sensor glucose of 120 mg/dl, which may be temporality increased to 150 mg/dl if needed to further minimize exposure to hypoglycemia during exercise or overnight. Prior to each 3 monthly visit, subjects will wear an actigraph monitor ( WGT3X-BT, Actigraph LLC,) for three weeks in order to define the nocturnal period. Every 6 months measures of hypoglycemia awareness (Clarke score) and severity (HYPO score), and the glycemic lability index (LI) will be calculated from questionnaires, event diaries, and device downloads, respectively. At 6 months and at 18 months, subjects will again undergo assessment of glucose counterregulation by stepped-hyperinsulinemic hypoglycemic clamp testing.
Conditions
See the medical conditions and disease areas that this research is targeting or investigating.
Study Design
Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.
NA
SINGLE_GROUP
TREATMENT
NONE
Study Groups
Review each arm or cohort in the study, along with the interventions and objectives associated with them.
Subjects with type 1 diabetes using MiniMed 670G system
Eligible subjects with type 1 diabetes will initiate hybrid closed-loop insulin delivery based on interstitial glucose monitoring via the MiniMed 670G system according to Medtronic's labeling. This system combines subject-delivered pre-meal boluses with automatic interprandial insulin delivery that includes automated functions for both predictive and threshold suspension of insulin delivery intended to minimize exposure to glucose levels \< 70 mg/dl.
MiniMed 670G system
Eligible subjects with type 1 diabetes will initiate hybrid closed-loop insulin delivery based on interstitial glucose monitoring via the MiniMed 670G system according to Medtronic's labeling. This system combines subject-delivered pre-meal boluses with automatic interprandial insulin delivery that includes automated functions for both predictive and threshold suspension of insulin delivery intended to minimize exposure to glucose levels \< 70 mg/dl.
Interventions
Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.
MiniMed 670G system
Eligible subjects with type 1 diabetes will initiate hybrid closed-loop insulin delivery based on interstitial glucose monitoring via the MiniMed 670G system according to Medtronic's labeling. This system combines subject-delivered pre-meal boluses with automatic interprandial insulin delivery that includes automated functions for both predictive and threshold suspension of insulin delivery intended to minimize exposure to glucose levels \< 70 mg/dl.
Eligibility Criteria
Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.
Inclusion Criteria
* Subjects who are able to provide written informed consent and to comply with the procedures of the study protocol.
* Clinical history compatible with type 1 diabetes with disease onset \< 40 years of age and insulin dependent for \> 10 years.
* Absent C-peptide (\< 0.3 ng/ml).
* Involvement in intensive diabetes management defined as the use of basal-bolus insulin analog delivery by multi-dose injection (MDI) or continuous subcutaneous insulin infusion (CSII) together with self-monitoring of blood glucose values more than 3 times daily with or without continuous glucose monitoring (CGM) under the direction of an endocrinologist, diabetologist, or diabetes nurse practitioner with at least 3 clinical evaluations during the previous 12 months.
* Hypoglycemia unawareness manifested by a Clarke score of 4 or more AND at least 1 of the following: HYPO score greater than or equal to the 90th percentile (1047); OR marked glycemic lability defined by a glycemic lability index (LI) score greater than or equal to the 90th percentile (433 mmol/l2/h•wk-1); OR a composite of a HYPO score greater than or equal to the 75th percentile (423) and a LI greater than or equal to the 75th percentile (329) (Senior et al., 2015).
* Documented \> 5% time spent in the hypoglycemic range (glucose \< 60 mg/dl) by 7 day real- time or blinded CGM; at least one episode of hypoglycemic during the 7 days must occur overnight.
Exclusion Criteria
* Insulin requirement of ≥ 1.0 units/kg•day.
* HbA1c ≥ 10%.
* Untreated proliferative diabetic retinopathy.
* Uncontrolled hypertension: systolic blood pressure \> 160 mmHg or diastolic blood pressure \> 100 mmHg.
* Active cardiovascular disease
* Abnormal kidney function: eGFR \< 60 ml/min/1.73 m2.
* Abnormal liver function: persistent elevation of liver function tests \> 1.5 times the upper limit of normal.
* Untreated hypothyroidism, Addison's disease, or Celiac disease.
* Anemia: baseline hemoglobin concentration \< 11 g/dl in women and \< 12 g/dl in men.
* Presence of a seizure disorder not related to prior severe hypoglycemia.
* Use of glucocorticoids greater than 5 mg of prednisone daily, or an equivalent physiologic dose of hydrocortisone.
* For female participants of child-bearing potential: Positive pregnancy test, presently breast-feeding, or unwillingness to use effective contraceptive measures for the duration of study participation. Oral contraceptives, intra-uterine devices, Norplant®, Depo-Provera®, and barrier devices with spermicide are acceptable contraceptive methods; condoms used alone are not acceptable.
* Treatment with any anti-diabetic medication other than insulin within 4 weeks of enrollment.
* Use of any investigational agents within 4 weeks of enrollment.
* Any medical condition that, in the opinion of the PI, will interfere with the safe completion of the study
25 Years
70 Years
ALL
No
Sponsors
Meet the organizations funding or collaborating on the study and learn about their roles.
National Institute of Diabetes and Digestive and Kidney Diseases (NIDDK)
NIH
University of Pennsylvania
OTHER
Responsible Party
Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.
Michael R. Rickels, MD, MS
Professor of Medicine
Principal Investigators
Learn about the lead researchers overseeing the trial and their institutional affiliations.
Michael R Rickels, MD
Role: PRINCIPAL_INVESTIGATOR
University of Pennsylvania
Locations
Explore where the study is taking place and check the recruitment status at each participating site.
Clinical and Translational Research Center, Hospital of University of Pennsylvania
Philadelphia, Pennsylvania, United States
Rodebaugh Diabetes Center, University of Pennsylvania
Philadelphia, Pennsylvania, United States
University of Pennsylvania - Institute for Diabetes, Obesity and Metabolism
Philadelphia, Pennsylvania, United States
Countries
Review the countries where the study has at least one active or historical site.
References
Explore related publications, articles, or registry entries linked to this study.
Malone SK, Matus AM, Flatt AJ, Peleckis AJ, Grunin L, Yu G, Jang S, Weimer J, Lee I, Rickels MR, Goel N. Prolonged Use of an Automated Insulin Delivery System Improves Sleep in Long-Standing Type 1 Diabetes Complicated by Impaired Awareness of Hypoglycemia. J Diabetes Sci Technol. 2024 Nov;18(6):1416-1423. doi: 10.1177/19322968231182406. Epub 2023 Jul 14.
Malone SK, Peleckis AJ, Grunin L, Yu G, Jang S, Weimer J, Lee I, Rickels MR, Goel N. Characterizing Glycemic Control and Sleep in Adults with Long-Standing Type 1 Diabetes and Hypoglycemia Unawareness Initiating Hybrid Closed Loop Insulin Delivery. J Diabetes Res. 2021 Feb 12;2021:6611064. doi: 10.1155/2021/6611064. eCollection 2021.
Provided Documents
Download supplemental materials such as informed consent forms, study protocols, or participant manuals.
Document Type: Study Protocol and Statistical Analysis Plan
Document Type: Informed Consent Form
Related Links
Access external resources that provide additional context or updates about the study.
Institute for Diabetes, Obesity and Metabolism Research Studies
Research Studies at the University of Pennsylvania
Perelman School of Medicine / Faculty Search / Michael Rickels, M.D., M.S
Other Identifiers
Review additional registry numbers or institutional identifiers associated with this trial.
827557
Identifier Type: -
Identifier Source: org_study_id
More Related Trials