Is the RAPDx Pupillograph Able to Distinguish Between Glaucoma Subjects and Healthy Subjects?

NCT ID: NCT02526693

Last Updated: 2019-10-23

Study Results

Results available

Outcome measurements, participant flow, baseline characteristics, and adverse events have been published for this study.

View full results

Basic Information

Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.

Recruitment Status

COMPLETED

Clinical Phase

NA

Total Enrollment

104 participants

Study Classification

INTERVENTIONAL

Study Start Date

2014-06-30

Study Completion Date

2015-08-31

Brief Summary

Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.

The Konan RAPDx (Konan Medical USA, Irvine, CA) is a newly patented pupillography device.The aims of this study are to assess the ability of the RAPDx to distinguish between healthy subjects and patients with confirmed glaucoma using standard testing sequences developed for use at the Wills Eye Hospital Glaucoma Research Center and to determine the combination of demographic, clinical, and RAPDx testing parameters which allow for maximum sensitivity and specificity.

Detailed Description

Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.

The RAPDx utilizes noninvasive digital, high-definition, infrared machine-vision with eye-tracking and automated blink detection technology to analyze and quantify the pupillary response to light.

During the scheduled appointment, all patients will receive an undilated fundus examination by the attending ophthalmologist. The following data will be collected; Demographic information, Visual acuity, Intraocular pressure (IOP) measured by Goldmann applanation tonometry, Disc damage likelihood scale (DDLS), Vertical cup/disc ratio, Gonioscopy (if not documented in the chart within the past 2 years) and Humphrey visual field examination.

Each participant will undergo RAPDx testing with two different testing sequences. They are separated by a 10-second resting period during which the patient is instructed to close his or her eyes. The exam may be paused at any time and re-alignment may be performed during any pause. The two testing sequences are:

1. Standard Factory Setting: 0.1-second stimuli with 2-second inter-stimuli pauses;
2. Custom Setting: 3-second stimuli with 1-second inter-stimuli pauses

Conditions

See the medical conditions and disease areas that this research is targeting or investigating.

Glaucoma

Study Design

Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.

Allocation Method

NON_RANDOMIZED

Intervention Model

PARALLEL

Primary Study Purpose

DIAGNOSTIC

Blinding Strategy

NONE

Study Groups

Review each arm or cohort in the study, along with the interventions and objectives associated with them.

Glaucoma Patients

Glaucoma patients recruited from Wills Eye Hospital Glaucoma Service will be tested with the relative afferent pupillary defect test (RAPDx) Pupillometer. The noninvasive RAPDx measures the pupils response during light stimulation.

Group Type OTHER

Pupillometer

Intervention Type DIAGNOSTIC_TEST

The Konan RAPDx (relative afferent pupillary defect) (Konan Medical USA, Irvine, CA) pupillometer utilizes digital, high-definition, infrared machine-vision with eye-tracking and automated blink detection technology to analyze and quantify pupillary response to light.

Healthy Controls

Healthy subjects with no eye diseases recruited from Wills Eye Hospital Glaucoma Service staff, family and friends will be tested with the relative afferent pupillary defect test (RAPDx) Pupillometer. The noninvasive RAPDx measures the pupils response during light stimulation.

Group Type OTHER

Pupillometer

Intervention Type DIAGNOSTIC_TEST

The Konan RAPDx (relative afferent pupillary defect) (Konan Medical USA, Irvine, CA) pupillometer utilizes digital, high-definition, infrared machine-vision with eye-tracking and automated blink detection technology to analyze and quantify pupillary response to light.

Interventions

Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.

Pupillometer

The Konan RAPDx (relative afferent pupillary defect) (Konan Medical USA, Irvine, CA) pupillometer utilizes digital, high-definition, infrared machine-vision with eye-tracking and automated blink detection technology to analyze and quantify pupillary response to light.

Intervention Type DIAGNOSTIC_TEST

Other Intervention Names

Discover alternative or legacy names that may be used to describe the listed interventions across different sources.

Relative afferent pupillary defect pupillometer

Eligibility Criteria

Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.

Inclusion Criteria

Glaucoma patients:

* optic nerve damage (neuroretinal rim notch, asymmetric inter-eye cup to disc (c/d\_ ratio \>0.2 or disc damage likelihood scale (DDLS) \>2, or absence of neuroretinal rim not due to other cause)
* glaucomatous visual field (VF) deficits (cluster of 3 or more points on pattern deviation plot depressed below 5% level, at least 1 depressed below 1% level; OR corrected pattern standard deviation/pattern standard deviation significant at P \<0.05; or glaucoma hemifield test "outside normal limits") with good reliability indices (fixation losses, false-positive rate, false-negative rate each \< 33%).

Healthy subjects:

* normal optic nerve exam
* normal reliable VF (Humphrey mean deviation (MD) \>-2 or Octopus MD ≤0.8; fixation losses, false-positive rate, and false-negative rate each \< 33%)
* open angles gonioscopy.

Exclusion Criteria

* Abnormal ocular motility preventing binocular fixation (e.g. strabismus, nystagmus).
* Any condition preventing adequate visualization and examination of pupil or optic nerve (e.g. dense corneal opacities or lens opacities).
* Active infection of anterior or posterior segments of the eye.
* Any intraocular surgical or laser procedure within previous 4 weeks.
* Any non-glaucomatous condition causing RAPD, anisocoria or corectopia (ex. optic neuropathy, Horner's syndrome, previous iris injury due to trauma or surgery, etc.).
* Subjects under age 18 or subjects presently housed in correctional facility.
Minimum Eligible Age

18 Years

Eligible Sex

ALL

Accepts Healthy Volunteers

Yes

Sponsors

Meet the organizations funding or collaborating on the study and learn about their roles.

Wills Eye

OTHER

Sponsor Role lead

Responsible Party

Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.

George L. Spaeth MD

Principal Investigator

Responsibility Role PRINCIPAL_INVESTIGATOR

Principal Investigators

Learn about the lead researchers overseeing the trial and their institutional affiliations.

George Spaeth, MD

Role: PRINCIPAL_INVESTIGATOR

Wills Eye Hospital

Other Identifiers

Review additional registry numbers or institutional identifiers associated with this trial.

14-400E

Identifier Type: -

Identifier Source: org_study_id

More Related Trials

Additional clinical trials that may be relevant based on similarity analysis.

Validation of i-ROP DL to Detect More Than Mild ROP
NCT06109285 ACTIVE_NOT_RECRUITING NA
Efficacy and Safety of XEN63 Gel Implant
NCT06844240 ENROLLING_BY_INVITATION