Nasotracheal Suction With Tiemann Catheter Compared to the Classic Technique With the Suction Catheter

NCT ID: NCT02261428

Last Updated: 2015-03-17

Study Results

Results available

Outcome measurements, participant flow, baseline characteristics, and adverse events have been published for this study.

View full results

Basic Information

Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.

Recruitment Status

COMPLETED

Clinical Phase

NA

Total Enrollment

20 participants

Study Classification

INTERVENTIONAL

Study Start Date

2014-05-31

Study Completion Date

2014-09-30

Brief Summary

Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.

The study compares the effectiveness of Tiemann and Suction catheters, with regard to attempts to enter the trachea and the time required for the procedure.

Detailed Description

Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.

The population of the study was twenty (20) patients, hospitalized in General and Cardiothoracic ICU, mostly with respiratory failure of various etiologies, the majority of which has complicated with Unit Myoneuropathy. This state confirmed by medical research council (MRC) Sum score which gave a result less than 48/6011 which corresponds to thirteen (13) patients in the total population of the study.

The randomization of the study achieved using the sealed envelopes method and associated with the catheter to be first used in the suction9,10 (Tiemann or suction catheter).

Twenty (20) sealed envelopes were prepared from a team not related to the study. The sealed envelopes had an inside paper with the sign "T" for the first ten (10) envelopes and "N" for the rest ten (10) of them.

When a new patient met the criteria for this study, one sealed envelope was randomly selected to determine the technique would come first: "T" for Tiemann technique or "N" for Suction catheter technique.

Furthermore the following parameters tabulated for each patient:

* Sex
* Unit
* Cause input
* First catheter used
* Second catheter used
* Hospitalization day
* Age
* Possible causes weakness
* Percentage oxygen mixture fraction of inspired oxygen (FiO2) and
* Oxygen partial pressure (PaO2) before any action

In addition to the above, all the parameters needed for calculating the Sequential Organ Failure Assessment score (SOFA) were tabulated, such as:

* Bilirubin
* The Glasgow Coma Score
* Platelets
* Mean arterial pressure
* Creatinine and urine 24 hours

Finally, it was recorded for each patient before and after the two catheterizations as indicators of disturbance and physiological burden attached to the intervention:

* The systolic and diastolic blood pressure
* The respiratory rate
* The heart rate
* The saturation of hemoglobin
* The blood gases, which recorded the partial pressure of oxygen (PaO2), carbon dioxide carbon dioxide partial pressure (PaCO2), the Ph and bicarbonates (HCO3).

A first check took place to identify each patient who met the admission criteria for the study.

The procedure required two people skilled in the nasotracheal suction (NTS), the first one to perform the NTS and the other one to record the time and the attempts needed, and it was beginning by correcting the posture of each patient on the bed, when the Great Trochanter was more than 20 centimeters away from the hinge of the bed.

The pillow beneath the patient's head was removed to have an extended position, the first blood gas analysis was done and the parameters mentioned above were recorded in the relevant tables.

To determine the access of the trachea, it was used a stethoscope to hear turbulent flow on the anterior surface of the chest.

The whole procedure comprises two "steps":

a) Achieving of the entrance into the nasopharynx b) Achieving of the entrance into the trachea

1. To achieve the entrance into the nasopharynx:

* It was selected the opposite nostril that Levine was placed.
* The insertion of the catheter was always opposite to the medial canthus of the eye.
* If it was found resistance and the catheter was not inserted to the nasopharynx, it was pulled slightly outwardly and then inwardly.

Each kind of micromanipulation recorded as attempt. The maximum number of attempts were agreed in Protocol Design (beyond which the attempts would stop) was five (5). In fact, in no cases was exceeded the limit of five attempts.

The time recorded corresponds to the time needed for the successful advancement of the catheter in the nasopharynx through the nostril (T1).
2. Achieving of the entrance into the trachea:

The successful insertion of the catheter into the trachea was confirmed by another member of the team using stethoscope and was determined by:

i. Hearing of turbulent flow when closing the "fingertip" ii. Insertion without resistance iii. Tickly cough during the insertion.

The insertion to trachea was considered to be unsuccessful as the catheter insert into the esophagus and thus:

i. There was no production of turbulent airflow when the hole of "fingertip"was closed ii. It was found resistance during the insertion of the catheter iii. There was a grimace of disgust on the patient's face. As the insertion was considered to be unsuccessful according to the above conditions, catheter was drawn slightly outward (less than one centimeter) and redirected to the trachea.

Each kind of micromanipulation recorded as attempt. The maximum number of attempts were agreed in Protocol Design was ten (10). Beyond the above limit the insertion was considered to be unsuccessful and repeated after half an hour15 with the other catheter (half an hour is the sufficient time for the patient to return to his initial condition).

The time of the second "step" recorded, corresponds to the time needed for the successful promotion of the catheter from the nasopharynx to the trachea (T2).

Finally it was recorded the time needed for each suction which was tabulated below as (T3).

The summation of (T1), (T2) and (T3) was reflected the total time needed for the whole procedure which was followed by a new blood gas analysis.

All parameters were recorded by the team involved to the study. The catheters used for this study were 14 French dia. of the same industry and from the same material (medical PVC).

As it was mentioned above, the whole idea based on the J-shape of the Tiemann catheter which is useful for the orientation of the catheter during the insertion to the nasopharynx first and then into the trachea and placed always downwards.

The oxygen face mask was maintained during the insertion of the catheter in order to continue the patient's oxygenation and avoid hypoxemia. This was achieved by leaving outside the oxygen face mask only the opposite nostril of the nasogastric tube.

No extra oxygen was given to patients before the procedure as pre-oxygenation.

The size of the sample was calculated based on a pilot study and was determined to twenty (20) patients were not included in the main study. Thus ten (10) patients underwent an equal number of nasotracheal suctioning with Tiemann catheter and ten (10) with Suction catheter.

For standard deviation σ1=3.9, and an error α=0,05 and power 80, the sample size must be bigger than twelve (12).

From all the above the investigators summarize that the sample size of twenty (20) patients was adequate.

For the statistical analysis was used the SPSS 20 Inc., Chicago, IL, and for paired values the Paired T-Test and Wilcoxon Signed Rank Test.

For paired nominal parameters was used the McNemar Χ2 test. To determine normal distributions, was used the Kolmogorov-Smirnov test. In the majority of cases data are presented as Median (Range) in abnormal distributions and Mean (SD) in normal.

Odds ratio was used to quantify how strongly Tiemann associated with Suction catheter in total attempts.

Significance was taken as P \< 0.05. To determine the difference in time needed by using the two catheters was used Kaplan-Mayer analysis and long-rank test.

Conditions

See the medical conditions and disease areas that this research is targeting or investigating.

Muscle Weakness

Study Design

Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.

Allocation Method

RANDOMIZED

Intervention Model

CROSSOVER

Primary Study Purpose

SUPPORTIVE_CARE

Blinding Strategy

SINGLE

Outcome Assessors

Study Groups

Review each arm or cohort in the study, along with the interventions and objectives associated with them.

Catheter Tiemann

We tested the ability of Tiemman catheter to access trachea and aspirate bronchial secretions, measuring number of attempts and time required for the intervention

Group Type EXPERIMENTAL

Catheter Tiemann

Intervention Type DEVICE

We tested the ability of Tiemann catheter to access trachea in order to aspirate bronchial secretions in ICU patients

Suction catheter

We tested the ability of suction catheter to access trachea and aspirate bronchial secretions, measuring number of attempts and time required for the intervention

Group Type ACTIVE_COMPARATOR

Suction catheter

Intervention Type DEVICE

We tested the ability of Suction catheter to access trachea in order to aspirate bronchial secretions in ICU patients

Interventions

Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.

Catheter Tiemann

We tested the ability of Tiemann catheter to access trachea in order to aspirate bronchial secretions in ICU patients

Intervention Type DEVICE

Suction catheter

We tested the ability of Suction catheter to access trachea in order to aspirate bronchial secretions in ICU patients

Intervention Type DEVICE

Eligibility Criteria

Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.

Inclusion Criteria

* ICU patients with nasogastric tube and spontaneous breathing without artificial airway
* Inability to mobilize and handle secretions
* Glasgow Coma Scale (GCS) \>12/15
* PaO2/FiO2 ratio \>100

Exclusion Criteria

* Active nasal bleeding
* international normalized ratio (INR)\>3
* Glasgow Coma Scale (GCS) \<11/15
* Facial or basal cranial fractures
* Tumor in trachea
* Pregnancy
* ICU patients with endotracheal tube or tracheostomy tube
Minimum Eligible Age

45 Years

Maximum Eligible Age

85 Years

Eligible Sex

ALL

Accepts Healthy Volunteers

No

Sponsors

Meet the organizations funding or collaborating on the study and learn about their roles.

Grigoriadis Konstantinos

OTHER

Sponsor Role lead

Responsible Party

Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.

Grigoriadis Konstantinos

Physical Therapist

Responsibility Role SPONSOR_INVESTIGATOR

Principal Investigators

Learn about the lead researchers overseeing the trial and their institutional affiliations.

Dimitrios Angouras, MD

Role: STUDY_DIRECTOR

Athens University School of Medicine, Department of Cardiothoracic Surgery

Theodoros Xanthos, MD

Role: STUDY_DIRECTOR

Athens University School of Medicine, Postgraduate program in CPR (moderator)

Konstantinos Grigoriadis, PT

Role: PRINCIPAL_INVESTIGATOR

Attikon Hospital

Katerina Flevari, MD

Role: STUDY_CHAIR

Attikon Hospital

References

Explore related publications, articles, or registry entries linked to this study.

Roppolo LP, Vilke GM, Chan TC, Krishel S, Hayden SR, Rosen P, Trione M. Nasotracheal intubation in the emergency department, revisited. J Emerg Med. 1999 Sep-Oct;17(5):791-9. doi: 10.1016/s0736-4679(99)00085-2.

Reference Type BACKGROUND
PMID: 10499691 (View on PubMed)

Saissy JM, Boussignac G, Cheptel E, Rouvin B, Fontaine D, Bargues L, Levecque JP, Michel A, Brochard L. Efficacy of continuous insufflation of oxygen combined with active cardiac compression-decompression during out-of-hospital cardiorespiratory arrest. Anesthesiology. 2000 Jun;92(6):1523-30. doi: 10.1097/00000542-200006000-00007.

Reference Type BACKGROUND
PMID: 10839900 (View on PubMed)

Grigoriadis KEpsilon, Angouras DC, Flevari A, Xathos T. Comparison of the Feasibility and Safety of Nasotracheal Suctioning With Curved Edge Catheter Versus Conventional Suction Catheter in Critically Ill Subjects: A Prospective Randomized Crossover Trial. Respir Care. 2015 Dec;60(12):1826-33. doi: 10.4187/respcare.03875. Epub 2015 Oct 20.

Reference Type DERIVED
PMID: 26487746 (View on PubMed)

Pode D, Manny J. A simplified method for repeated nasotracheal suction. Surg Gynecol Obstet. 1984 Aug;159(2):173-4.

Reference Type BACKGROUND
PMID: 6463830 (View on PubMed)

Brucia J, Rudy E. The effect of suction catheter insertion and tracheal stimulation in adults with severe brain injury. Heart Lung. 1996 Jul-Aug;25(4):295-303. doi: 10.1016/s0147-9563(96)80065-3.

Reference Type BACKGROUND
PMID: 8836745 (View on PubMed)

Gemma M, Tommasino C, Cerri M, Giannotti A, Piazzi B, Borghi T. Intracranial effects of endotracheal suctioning in the acute phase of head injury. J Neurosurg Anesthesiol. 2002 Jan;14(1):50-4. doi: 10.1097/00008506-200201000-00010.

Reference Type BACKGROUND
PMID: 11773824 (View on PubMed)

Marlow TJ, Goltra DD Jr, Schabel SI. Intracranial placement of a nasotracheal tube after facial fracture: a rare complication. J Emerg Med. 1997 Mar-Apr;15(2):187-91. doi: 10.1016/s0736-4679(96)00356-3.

Reference Type BACKGROUND
PMID: 9144060 (View on PubMed)

Sloan EP, VanRooyen MJ. Suction catheter-assisted nasotracheal intubation. Acad Emerg Med. 1994 Jul-Aug;1(4):388-90. doi: 10.1111/j.1553-2712.1994.tb02651.x.

Reference Type BACKGROUND
PMID: 7614287 (View on PubMed)

Other Identifiers

Review additional registry numbers or institutional identifiers associated with this trial.

AttikonH

Identifier Type: -

Identifier Source: org_study_id

More Related Trials

Additional clinical trials that may be relevant based on similarity analysis.

Use of TrachPhone in Tracheostomized Patients
NCT06487104 ENROLLING_BY_INVITATION NA
Neck Drains for SLNDs
NCT03113526 COMPLETED NA