RCT of Mesh Versus Jet Nebulizers on Clinical Outcomes During Mechanical Ventilation in the Intensive Care Unit

NCT ID: NCT02056119

Last Updated: 2023-03-21

Study Results

Results available

Outcome measurements, participant flow, baseline characteristics, and adverse events have been published for this study.

View full results

Basic Information

Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.

Recruitment Status

COMPLETED

Clinical Phase

NA

Total Enrollment

120 participants

Study Classification

INTERVENTIONAL

Study Start Date

2013-09-30

Study Completion Date

2014-10-31

Brief Summary

Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.

Aerosol delivery during mechanical ventilation has long been a long debated topic. As evidence-based knowledge about the delivery of aerosol to the lungs of mechanically ventilated patients increases, one piece of the puzzle has remained unexplored; measurement of clinically relevant outcomes. The primary aim of this research is to compare clinical outcomes (ventilator-associated events (VAEs), length of stay (LOS) in intensive care unit (ICU), and total days on mechanical ventilation) when using a traditional jet nebulizer versus a newer generation vibrating mesh nebulizer during mechanical ventilation. The secondary aim of this research is to identify source of bacteria by obtaining cultures of each nebulizer and ventilator circuit and plating them for colony growth and identification.

Detailed Description

Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.

This pilot study was designed as a randomized, controlled trial comparing clinical outcomes and contamination rates in jet nebulizers versus vibrating mesh nebulizers. The sample size will be one of consecutive convenience over an entire year (September 2013-September 2014) to include all seasons. The study will be performed at Rush University Medical Center (RUMC) in Chicago, IL.

Each mechanically ventilated patient within inclusion criteria, with a physician order for aerosol treatment, will be randomized to either the jet nebulizer or mesh nebulizer group using SPSS computer software. The respiratory care staff will place the nebulization device in the ventilator circuit per hospital protocol. Both devices, jet and mesh nebulizers, are currently standard practice; therefore hospital protocol will be followed regarding placement and administration of ordered aerosol treatment.

The study subject will remain on the device to which they were randomized for the duration of their hospital stay to obtain clinical outcome data. Study staff to be notified of patient extubation from mechanical ventilation, discharge from intensive care unit, or expiration; this will conclude participant involvement in the study. Retrospective clinical outcome data will be obtained from the subject's electronic medical chart at conclusion of study.

Retrospective data to include:

1. LOS in ICU (days)
2. LOS on mechanical ventilator (days)
3. Drug administration information

1. Drug and dose ordered
2. Number of treatments delivered
3. Treatment type

* Continuous
* Intermittent
4. Ventilator-Associated Tracheobronchitis (# VATs)16

a. Fever, increased volume and purulence of secretions, a positive culture (quantitative or semi-quantitative) of a respiratory sample (tracheal aspirates or bronchoscopic specimens), and the absence of a new or an evolving pulmonary infiltrate in the chest x-ray in a patient on mechanical ventilation \>48 hours
5. Ventilator-Associated Events in Adults15

1. Ventilator-Associated Condition (# VACs)

* After a period of stability or improvement on the ventilator, the patient has at least one of the following indications of worsening oxygenation:
* Minimum daily FiO2 values increase ≥ 0.20 over baseline and remain at or above that increased level for ≥ 2 calendar days. -
* Minimum daily PEEP values increase ≥ 3 cm H2O over baseline and remain at or above that increased level for ≥ 2 calendar days.
2. Infection-related Ventilator-Associated Complication (# IVACs)

* On or after calendar day 3 of mechanical ventilation and within 2 calendar days before or after the onset of worsening oxygenation, the patient meets both of the following criteria:
* Temperature \> 38˚ C or \< 36˚ C, or white blood cell count ≥ 12,000 cells/mm3 or ≤ 4,000 cells/mm3. AND
* A new antimicrobial agent(s) is started, and is continued for ≥ 4 calendar days.
3. Possible Ventilator-Associated Pneumonia (# Possible VAPs)

* On or after calendar day 3 of mechanical ventilation and within 2 calendar days before or after the onset of worsening oxygenation, ONE of the following criteria is met:
* Purulent respiratory secretions
* Positive culture
4. Probably Ventilator-Associated Pneumonia (# Probable VAPs)

* On or after calendar day 3 of mechanical ventilation and within 2 calendar days before or after the onset of worsening oxygenation, ONE of the following criteria is met:
* Purulent respiratory secretions (from one or more specimen collections- and defined as for possible VAP) and one of the following (positive culture of endotracheal aspirate, positive culture of bronchoalveolar lavage, positive culture of lung tissue, positive culture of protected specimen brush
* One of the following without requirement for purulent respiratory secretions (positive pleural fluid culture, positive lung histopathology, positive diagnostic test for Legionella, positive diagnostic test on respiratory secretions for influenza virus, respiratory syncytial virus, adenovirus, parainfluenza virus)

The Jet Nebulizer Protocol:

* Physician order received for subject; randomization to jet nebulizer occurred.
* Jet nebulizer, Misty Max 10™ (Carefusion, CA), placed into spring-loaded t-piece between the humidifier and the ventilator (approximately 15 cm from the gas outlet).
* Aerosol treatment delivered per physician order at flow rates of 8-10 L/min.
* Jet nebulizer to be replaced every 3 days per hospital protocol.

1. Prior to every nebulizer change or every 3 days, study staff to be notified in order to obtain cultures.
2. Cultures

* Jet Nebulizer Culture (3 plates): Sample to be collected from jet nebulizer by running oxygen at a flow of 8-10 L/min through nebulizer (3 mL normal saline in the reservoir) while directing the nebulizer towards plate for 1 minute. Process to be repeated for 3 different plates. Plates to be labeled, have edges sealed with tape, be placed in biohazard bag and sent to microbiology lab for analysis.

1. Plate 1: MacConkey agar (id gram (-) organisms)
2. Plate 2: Chocolate agar (id fastidious organisms)
3. Plate 3: Blood agar (id gram(+) and gram(-) organisms)
* Ventilator Circuit Culture: Material to be collected from rim of inspiratory ventilator circuit at wye by rubbing a sterile swab along circuit 3x in a circular motion. Swab to be sealed in test tube with lid and sent to lab for analysis.
* Sputum samples: Data obtained per electronic medical chart. Sputum sample data to be analyzed in study when ordered by physician.
* After obtaining cultures, the jet nebulizer will be replaced with new one and labeled per hospital protocol.
* Further cultures are not necessary unless patient is identified by study staff to have VAE. (Appendix A)
* Study staff to be notified of patient extubation from mechanical ventilation, discharge from intensive care unit or expiration.

The Vibrating Mesh Nebulizer protocol:

* Physician order received for subject; randomization to vibrating mesh nebulizer occurred.
* Aeroneb® Solo (Aerogen, Galway, Ireland) vibrating mesh nebulizer to be placed before the heater on the "dry side" of the heater water chamber on the inspiratory ventilator circuit.
* Aerosol treatment delivered per physician order.
* Mesh nebulizer to be replaced every 30 days per hospital protocol.

1. Prior to every nebulizer change or every 3 days, study staff to be notified in order to obtain cultures.
2. Cultures

* Vibrating Mesh Nebulizer Culture (3 plates): Sample to be collected from vibrating mesh nebulizer while directing the nebulizer(3 mL normal saline in the reservoir) towards plate for 1 minute. Process to be repeated for 3 different plates. Plates to be labeled, have edges sealed with tape, be placed in biohazard bag and sent to microbiology lab for analysis.

1. Plate 1: MacConkey agar (id gram (-) organisms)
2. Plate 2: Chocolate agar (id fastidious organisms)
3. Plate 3: Blood agar (id gram(+) and gram(-) organisms)
* Ventilator Circuit Culture: Material to be collected from rim of inspiratory ventilator circuit at wye by rubbing a sterile swab along circuit 3x in a circular motion. Swab to be sealed in test tube with lid and sent to lab for analysis.
* Sputum samples: Data obtained per electronic medical chart. Sputum sample data to be analyzed in study when ordered by physician.
* Vibrating mesh nebulizer replaced with new one and labeled per hospital protocol.
* Further cultures not necessary unless patient identified by study staff to have VAE. (appendix A)
* Study staff to be notified immediately of patient extubation from mechanical ventilation, discharge from intensive care unit or expiration.

Conditions

See the medical conditions and disease areas that this research is targeting or investigating.

Respiratory Insufficiency

Study Design

Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.

Allocation Method

RANDOMIZED

Intervention Model

PARALLEL

Primary Study Purpose

BASIC_SCIENCE

Blinding Strategy

NONE

Study Groups

Review each arm or cohort in the study, along with the interventions and objectives associated with them.

Jet Nebulizer Arm

The Jet Nebulizer Protocol:

* Physician order received for subject, randomization to jet nebulizer occurred.
* Jet nebulizer, Misty Max 10™ (Carefusion, CA), placed into spring loaded t-piece between the humidifier and the ventilator (approximately 15 cm from the gas outlet).
* Aerosol treatment delivered per physician order at flow rates of 8-10 L/min.
* Jet nebulizer to be replaced every 3 days per hospital protocol. a. Prior to every nebulizer change or every 3 days, study staff to be notified in order to obtain cultures.

Group Type ACTIVE_COMPARATOR

Jet Nebulizer Arm

Intervention Type DEVICE

* Physician order received for subject, randomization to jet nebulizer occurred.
* Jet nebulizer, Misty Max 10™ (Carefusion, CA), placed into spring loaded t-piece between the humidifier and the ventilator (approximately 15 cm from the gas outlet).
* Aerosol treatment delivered per physician order at flow rates of 8-10 L/min.
* Jet nebulizer to be replaced every 3 days per hospital protocol. a. Prior to every nebulizer change or every 3 days, study staff to be notified in order to obtain cultures.

Vibrating Mesh Nebulizer Arm

* Physician order received for subject, randomization to vibrating mesh nebulizer occurred.
* Aeroneb® Solo (Aerogen, Galway, Ireland) vibrating mesh nebulizer to be placed before the heater on the "dry side" of the heater water chamber on the inspiratory ventilator circuit.
* Aerosol treatment delivered per physician order.
* Mesh nebulizer to be replaced every 30 days per hospital protocol. a. Prior to every nebulizer change or every 3 days, study staff to be notified in order to obtain cultures.

Group Type EXPERIMENTAL

Vibrating Mesh Nebulizer Arm

Intervention Type DEVICE

* Physician order received for subject, randomization to vibrating mesh nebulizer occurred.
* Aeroneb® Solo (Aerogen, Galway, Ireland) vibrating mesh nebulizer to be placed before the heater on the "dry side" of the heater water chamber on the inspiratory ventilator circuit.
* Aerosol treatment delivered per physician order.
* Mesh nebulizer to be replaced every 30 days per hospital protocol. a. Prior to every nebulizer change or every 3 days, study staff to be notified in order to obtain cultures.

Interventions

Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.

Vibrating Mesh Nebulizer Arm

* Physician order received for subject, randomization to vibrating mesh nebulizer occurred.
* Aeroneb® Solo (Aerogen, Galway, Ireland) vibrating mesh nebulizer to be placed before the heater on the "dry side" of the heater water chamber on the inspiratory ventilator circuit.
* Aerosol treatment delivered per physician order.
* Mesh nebulizer to be replaced every 30 days per hospital protocol. a. Prior to every nebulizer change or every 3 days, study staff to be notified in order to obtain cultures.

Intervention Type DEVICE

Jet Nebulizer Arm

* Physician order received for subject, randomization to jet nebulizer occurred.
* Jet nebulizer, Misty Max 10™ (Carefusion, CA), placed into spring loaded t-piece between the humidifier and the ventilator (approximately 15 cm from the gas outlet).
* Aerosol treatment delivered per physician order at flow rates of 8-10 L/min.
* Jet nebulizer to be replaced every 3 days per hospital protocol. a. Prior to every nebulizer change or every 3 days, study staff to be notified in order to obtain cultures.

Intervention Type DEVICE

Other Intervention Names

Discover alternative or legacy names that may be used to describe the listed interventions across different sources.

Aeroneb Solo (Aerogen, Galway, Ireland) Misty Max 10 nebulizer (Carefusion, California)

Eligibility Criteria

Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.

Inclusion Criteria

* Age \> 1 y/o
* Intubated with endotracheal tube (ETT) and on mechanical ventilation
* Patient in all intensive care units at Rush University Medical Center
* Physician order for aerosolized breathing treatment in medical record

Exclusion Criteria

* Age \< 1 y/o
* Diaphragmatic paralysis
* Paraplegic or high spinal cord injury
* Brain death
Minimum Eligible Age

1 Year

Eligible Sex

ALL

Accepts Healthy Volunteers

No

Sponsors

Meet the organizations funding or collaborating on the study and learn about their roles.

Rush University Medical Center

OTHER

Sponsor Role lead

Responsible Party

Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.

Responsibility Role SPONSOR

Principal Investigators

Learn about the lead researchers overseeing the trial and their institutional affiliations.

Meagan N Dubosky, MS

Role: PRINCIPAL_INVESTIGATOR

Rush University, Respiratory Care Research Coordinator

Locations

Explore where the study is taking place and check the recruitment status at each participating site.

Rush University Medical Center

Chicago, Illinois, United States

Site Status

Countries

Review the countries where the study has at least one active or historical site.

United States

Other Identifiers

Review additional registry numbers or institutional identifiers associated with this trial.

13070203

Identifier Type: -

Identifier Source: org_study_id