Safety of Body Composition Analysis in Heart Failure Patients With Implantable Cardioverter Defibrillators (ICDs)
NCT ID: NCT01568606
Last Updated: 2020-03-10
Study Results
Outcome measurements, participant flow, baseline characteristics, and adverse events have been published for this study.
View full resultsBasic Information
Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.
COMPLETED
NA
20 participants
INTERVENTIONAL
2011-01-31
2011-03-31
Brief Summary
Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.
Although the Biospace America InBody 520 scale is routinely used to analyze body composition in various settings including the Ahmanson-UCLA Cardiomyopathy clinic, due to theoretical concerns of safety, bioimpedance has not been routinely used in patients with ICDs. The investigators hope that this study will allow us to routinely analyze body composition in heart failure patients with ICDs, information which can be used to help guide dietary, exercise, and medical prescriptions for the investigators heart failure patients.
Related Clinical Trials
Explore similar clinical trials based on study characteristics and research focus.
Bioimpedance Analysis in Chronic Heart Failure
NCT02662439
Biomonitoring and Cardiorenal Syndrome in Heart Failure(BIONICS-HF) Trial
NCT01570153
Cardiac Resynchronization Therapy Defibrillator Based Impedance Monitoring Study
NCT00500838
MIBG Scintigraphy as a Tool for Selecting Patients Requiring Implantable Cardioverter Defibrillator (ICD)
NCT01185756
BIA and Cardiac Implantable Electronic Devices
NCT03045822
Detailed Description
Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.
Conditions
See the medical conditions and disease areas that this research is targeting or investigating.
Study Design
Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.
NA
SINGLE_GROUP
SCREENING
NONE
Study Groups
Review each arm or cohort in the study, along with the interventions and objectives associated with them.
Body Composition Analysis InBody Scale
Body Composition Analysis InBody Scale
Patients will have their ICD interrogated using a pacemaker Programmer specifically designed from one of the 4 pacemaker manufacturers: St. Jude, Medtronic, Guidant, and Biotronik. Depending on the type of pacemaker, interrogation can be done either wirelessly or by placing a wand over the pacemaker. The ICD will be turned to a "monitor-only" mode to disable risk of ICD shock by the supervising electrophysiologist. Patient will be asked to step on the InBody 520 scale to have their body composition analyzed. Patient will continue to have their ICD actively monitored. Analysis will take approximately 30 - 50 seconds. At the end of analysis, ICD device will be returned to default settings by supervising electrophysiologist.
Interventions
Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.
Body Composition Analysis InBody Scale
Patients will have their ICD interrogated using a pacemaker Programmer specifically designed from one of the 4 pacemaker manufacturers: St. Jude, Medtronic, Guidant, and Biotronik. Depending on the type of pacemaker, interrogation can be done either wirelessly or by placing a wand over the pacemaker. The ICD will be turned to a "monitor-only" mode to disable risk of ICD shock by the supervising electrophysiologist. Patient will be asked to step on the InBody 520 scale to have their body composition analyzed. Patient will continue to have their ICD actively monitored. Analysis will take approximately 30 - 50 seconds. At the end of analysis, ICD device will be returned to default settings by supervising electrophysiologist.
Other Intervention Names
Discover alternative or legacy names that may be used to describe the listed interventions across different sources.
Eligibility Criteria
Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.
Inclusion Criteria
Exclusion Criteria
* have a physical disability making them unable to stand on the InBody
* those who are above the height and weight maximums for the device \[Height \> 220cm (7'2.6''), Weight \> 250kg (551lb)\]
* those who are pregnant
18 Years
ALL
No
Sponsors
Meet the organizations funding or collaborating on the study and learn about their roles.
University of California, Los Angeles
OTHER
Responsible Party
Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.
Tamara Horwich
Dr. Tamara Horwich, M.D., M.S.
Principal Investigators
Learn about the lead researchers overseeing the trial and their institutional affiliations.
Tamara Horwich, MD, MS
Role: PRINCIPAL_INVESTIGATOR
University of California, Los Angeles
Locations
Explore where the study is taking place and check the recruitment status at each participating site.
Ahmanson-UCLA Cardiomyopathy Center
Los Angeles, California, United States
Countries
Review the countries where the study has at least one active or historical site.
References
Explore related publications, articles, or registry entries linked to this study.
Buch E, Bradfield J, Larson T, Horwich T. Effect of bioimpedance body composition analysis on function of implanted cardiac devices. Pacing Clin Electrophysiol. 2012 Jun;35(6):681-4. doi: 10.1111/j.1540-8159.2012.03377.x. Epub 2012 Mar 27.
Other Identifiers
Review additional registry numbers or institutional identifiers associated with this trial.
UCLA M-IRB1 # 10-001055
Identifier Type: -
Identifier Source: org_study_id
More Related Trials