Safety of Body Composition Analysis in Heart Failure Patients With Implantable Cardioverter Defibrillators (ICDs)

NCT ID: NCT01568606

Last Updated: 2020-03-10

Study Results

Results available

Outcome measurements, participant flow, baseline characteristics, and adverse events have been published for this study.

View full results

Basic Information

Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.

Recruitment Status

COMPLETED

Clinical Phase

NA

Total Enrollment

20 participants

Study Classification

INTERVENTIONAL

Study Start Date

2011-01-31

Study Completion Date

2011-03-31

Brief Summary

Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.

The purpose of this study is to define the safety of using bioimpedance for analysis of body composition in heart failure patients with ICD devices. The Biospace America InBody 520 scale, using direct segmental multifrequency bioimpedance, will be utilized to assess patients' fat mass, lean muscle mass, and edema status.

Although the Biospace America InBody 520 scale is routinely used to analyze body composition in various settings including the Ahmanson-UCLA Cardiomyopathy clinic, due to theoretical concerns of safety, bioimpedance has not been routinely used in patients with ICDs. The investigators hope that this study will allow us to routinely analyze body composition in heart failure patients with ICDs, information which can be used to help guide dietary, exercise, and medical prescriptions for the investigators heart failure patients.

Detailed Description

Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.

Up to one half of heart failure (HF) patients are overweight and obese, as measured by body mass index (BMI). Comprehensive body composition assessment utilizing bioimpedance scales wuch as Biospace America InBody 520 scale, can give accurate measurements of fat mass, lean muscle mass, and edematous water weight. Body composition analysis is routinely used in many clinical settings, including the Ahmanson-UCLA Cardiomyopathy Center. However, bioimpedance scales are generally not used in patients with implantable cardioverter-defibrillators (ICDs), a group that represents a majority of our HF patient population. The safety of InBody 520 technology - direct segmental multifrequency (5, 100, 500 KHz) bioimpedance analysis - in patients with ICDs has not been formally evaluated. There is a theoretic concern that the ICD may misinterpret the applied currents for a cardiac arrhythmia and possibly cause a discharge from the ICD. However, there has been no documentation of any adverse events in patients with ICDs using the InBody 520 or other bioimpedance scales. Furthermore, thoracic bioimpedance analysis (BioZ ICG, 70 Khz) has been documented to be successfully utilized without adverse effects in patients with HF and ICDs. This study will define the safety of using the InBody 520 scale for analysis of body composition in patients with ICD devices. Subjects with HF and and ICD will have their ICD interrogated and continually monitored by an electrophysiologist before, during, and after the 30-50 seconds of bioimpedance analysis, which involves standing on the InBody 520 scale. If there is any suggestion of the ICD interpreting applied frequencies as a cardiac arrhythmia, the patient will be removed from the scale and the electrophysiologist will simultaneously temporarily deactivate their ICD device to prevent discharge. Twenty subjects (from the following four ICD companies: Medtronic, St. Jude, Guidant, and Biotronik) will participate in our study.

Conditions

See the medical conditions and disease areas that this research is targeting or investigating.

Heart Failure Obesity

Study Design

Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.

Allocation Method

NA

Intervention Model

SINGLE_GROUP

Primary Study Purpose

SCREENING

Blinding Strategy

NONE

Study Groups

Review each arm or cohort in the study, along with the interventions and objectives associated with them.

Body Composition Analysis InBody Scale

Group Type EXPERIMENTAL

Body Composition Analysis InBody Scale

Intervention Type DEVICE

Patients will have their ICD interrogated using a pacemaker Programmer specifically designed from one of the 4 pacemaker manufacturers: St. Jude, Medtronic, Guidant, and Biotronik. Depending on the type of pacemaker, interrogation can be done either wirelessly or by placing a wand over the pacemaker. The ICD will be turned to a "monitor-only" mode to disable risk of ICD shock by the supervising electrophysiologist. Patient will be asked to step on the InBody 520 scale to have their body composition analyzed. Patient will continue to have their ICD actively monitored. Analysis will take approximately 30 - 50 seconds. At the end of analysis, ICD device will be returned to default settings by supervising electrophysiologist.

Interventions

Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.

Body Composition Analysis InBody Scale

Patients will have their ICD interrogated using a pacemaker Programmer specifically designed from one of the 4 pacemaker manufacturers: St. Jude, Medtronic, Guidant, and Biotronik. Depending on the type of pacemaker, interrogation can be done either wirelessly or by placing a wand over the pacemaker. The ICD will be turned to a "monitor-only" mode to disable risk of ICD shock by the supervising electrophysiologist. Patient will be asked to step on the InBody 520 scale to have their body composition analyzed. Patient will continue to have their ICD actively monitored. Analysis will take approximately 30 - 50 seconds. At the end of analysis, ICD device will be returned to default settings by supervising electrophysiologist.

Intervention Type DEVICE

Other Intervention Names

Discover alternative or legacy names that may be used to describe the listed interventions across different sources.

Body Composition Analysis InBody 520 Scale

Eligibility Criteria

Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.

Inclusion Criteria

* HF patients of any etiology who currently have an implanted ICD

Exclusion Criteria

* patients who are non-ambulatory
* have a physical disability making them unable to stand on the InBody
* those who are above the height and weight maximums for the device \[Height \> 220cm (7'2.6''), Weight \> 250kg (551lb)\]
* those who are pregnant
Minimum Eligible Age

18 Years

Eligible Sex

ALL

Accepts Healthy Volunteers

No

Sponsors

Meet the organizations funding or collaborating on the study and learn about their roles.

University of California, Los Angeles

OTHER

Sponsor Role lead

Responsible Party

Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.

Tamara Horwich

Dr. Tamara Horwich, M.D., M.S.

Responsibility Role PRINCIPAL_INVESTIGATOR

Principal Investigators

Learn about the lead researchers overseeing the trial and their institutional affiliations.

Tamara Horwich, MD, MS

Role: PRINCIPAL_INVESTIGATOR

University of California, Los Angeles

Locations

Explore where the study is taking place and check the recruitment status at each participating site.

Ahmanson-UCLA Cardiomyopathy Center

Los Angeles, California, United States

Site Status

Countries

Review the countries where the study has at least one active or historical site.

United States

References

Explore related publications, articles, or registry entries linked to this study.

Buch E, Bradfield J, Larson T, Horwich T. Effect of bioimpedance body composition analysis on function of implanted cardiac devices. Pacing Clin Electrophysiol. 2012 Jun;35(6):681-4. doi: 10.1111/j.1540-8159.2012.03377.x. Epub 2012 Mar 27.

Reference Type RESULT
PMID: 22452409 (View on PubMed)

Other Identifiers

Review additional registry numbers or institutional identifiers associated with this trial.

UCLA M-IRB1 # 10-001055

Identifier Type: -

Identifier Source: org_study_id