Lifelight Non-Invasive Blood Pressure and Heart Rate Validation Study

NCT ID: NCT06535152

Last Updated: 2024-08-02

Study Results

Results pending

The study team has not published outcome measurements, participant flow, or safety data for this trial yet. Check back later for updates.

Basic Information

Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.

Recruitment Status

COMPLETED

Clinical Phase

NA

Total Enrollment

85 participants

Study Classification

INTERVENTIONAL

Study Start Date

2023-05-18

Study Completion Date

2023-08-03

Brief Summary

Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.

Lifelight is a novel, calibration-free, software-only medical device that contactlessly measures pulse rate and blood pressure by measuring remote photoplethysmography (rPPG) signals of a patient's face using a standard smartphone or tablet camera.

By conducting this validation study we aimed to demonstrate the accuracy of Lifelight for classifying the full range of clinically relevant blood pressures as hypertensive or non-hypertensive and demonstrate its accuracy for measuring pulse rate and the blood pressure of people with blood pressures relevant to Stage 1 hypertension.

Detailed Description

Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.

Lifelight was investigated in a study that followed the data collection and data analysis methodology outlined in ISO 81060-2:2018/AMD 1:2020 "Non-invasive Sphygmomanometers - Part 2: Clinical investigation of automated measurement type". This validation study was conducted by independent laboratory Element Materials Technology Boulder (formerly Clinimark), a global leader for clinical testing of vital sign data for medical devices and consumer wearable products. The study generated data from 85 people aged 18-85 with the wide-ranging distribution of blood pressures specified in ISO 81060-2:2018/AMD 1:2020.The accuracy of Lifelight's blood pressure measurements was assessed by comparing Lifelight's measurements with measurements made by dual-observer manual auscultation using the same-arm sequential method specified in ISO 81060-2:2018/AMD 1:2020. The accuracy of Lifelight's pulse rate measurements was assessed by comparing Lifelight's measurements with concurrent ECG-derived heart rate values. Data was collected in an Independent Review Board-approved study conducted by independent laboratory Clinimark (now Element) in the USA and repurposed for the testing of the new improved model in the latest version of the Lifelight® EA device. The data collected consists of 85 participants, with three readings each. The participants were resting before their measurements were taken to ensure stable blood pressure. For each Lifelight measurement, two sets of blood pressure measurements were taken using dual-observer manual auscultation. The first measurement was taken before the Lifelight measurement and the second just after. If there was too much variation between the before and after measurement, or between the two independent observers, the measurement was discarded and repeated. For this validation procedure, the measurements taken during the original study were re-processed in the new updated algorithm.

Conditions

See the medical conditions and disease areas that this research is targeting or investigating.

Hypertension

Study Design

Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.

Allocation Method

NA

Intervention Model

SINGLE_GROUP

Up to eight pairs of reference and Lifelight recordings (starting and ending with reference recordings) were taken sequentially to obtain a minimum of three valid paired reference and Lifelight BP measurements. At least 60 seconds elapsed between each BP determination.
Primary Study Purpose

OTHER

Blinding Strategy

NONE

Study Groups

Review each arm or cohort in the study, along with the interventions and objectives associated with them.

Lifelight and BP measurement

The accuracy of Lifelight's blood pressure measurements was assessed by comparing Lifelight's measurements with measurements made by dual-observer manual auscultation using the same-arm sequential method specified in ISO 81060-2:2018/AMD 1:2020. The accuracy of Lifelight's pulse rate measurements was assessed by comparing Lifelight's measurements with concurrent ECG-derived heart rate values. The participants were resting before their measurements were taken to ensure stable blood pressure. For each Lifelight measurement, two sets of blood pressure measurements were taken using dual-observer manual auscultation. The first measurement was taken before the Lifelight measurement and the second just after. If there was too much variation between the before and after measurement, or between the 2 independent observers, the measurement was discarded and repeated. For this validation procedure, the measurements taken during the original study were re-processed in the new updated algorithm.

Group Type EXPERIMENTAL

Lifelight

Intervention Type DEVICE

Lifelight is a artificial-intelligence software application that enables completely contactless spot measurements (i.e., one-off measurements without prior calibration) of PR and BP. Based on the science of remote photoplethysmography (rPPG), Lifelight works by detecting tiny changes in the colour of facial skin that occur every time the heart beats. All that is required to generate PR and BP measurements is that the software application is downloaded onto a computer device (e.g., smartphone or tablet) with a standard camera, and that the patient's face stays in the line of sight of the camera for up to 60 seconds.

Manual sphygmomanometer

Intervention Type DEVICE

Blood Pressure Measurement

Interventions

Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.

Lifelight

Lifelight is a artificial-intelligence software application that enables completely contactless spot measurements (i.e., one-off measurements without prior calibration) of PR and BP. Based on the science of remote photoplethysmography (rPPG), Lifelight works by detecting tiny changes in the colour of facial skin that occur every time the heart beats. All that is required to generate PR and BP measurements is that the software application is downloaded onto a computer device (e.g., smartphone or tablet) with a standard camera, and that the patient's face stays in the line of sight of the camera for up to 60 seconds.

Intervention Type DEVICE

Manual sphygmomanometer

Blood Pressure Measurement

Intervention Type DEVICE

Eligibility Criteria

Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.

Inclusion Criteria

* Healthy volunteers aged 18 to 85 years old

Exclusion Criteria

* Participants were excluded from the study if they were medically unsuitable for participation at time of visit, any heart dysrhythmias (except respiratory sinus arrhythmia) as confirmed with a 3 lead ECG, compromised circulation or peripheral vascular disease, clotting disorders, female participants who were pregnant or trying to get pregnant, excessive facial hair, and conditions that affect the skin, such as anaemia, jaundice, rosacea, psoriasis, acute acne, and erythropoietic protoporphyria.

Participants could choose to withdraw themselves from the study without prejudice or they could be withdrawn by study investigators for predetermined reasons. Data excluded from the analysis was documented with justifications.
Minimum Eligible Age

18 Years

Maximum Eligible Age

85 Years

Eligible Sex

ALL

Accepts Healthy Volunteers

Yes

Sponsors

Meet the organizations funding or collaborating on the study and learn about their roles.

Element Materials Technology

INDUSTRY

Sponsor Role collaborator

Mind Over Matter Medtech Ltd

INDUSTRY

Sponsor Role collaborator

Xim Limited

INDUSTRY

Sponsor Role lead

Responsible Party

Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.

Responsibility Role SPONSOR

Principal Investigators

Learn about the lead researchers overseeing the trial and their institutional affiliations.

Arthur Cabrera, MD

Role: PRINCIPAL_INVESTIGATOR

Investigator has left Element Materials Technology

Locations

Explore where the study is taking place and check the recruitment status at each participating site.

Element Materials Technology

Louisville, Colorado, United States

Site Status

Countries

Review the countries where the study has at least one active or historical site.

United States

Related Links

Access external resources that provide additional context or updates about the study.

https://preprints.jmir.org/preprint/57241

Contactless and Calibration-free Blood Pressure and Pulse Rate Monitor for Screening and Monitoring of Hypertension: Cross-sectional Validation Study

Other Identifiers

Review additional registry numbers or institutional identifiers associated with this trial.

PR 2021-443

Identifier Type: -

Identifier Source: org_study_id