Safety and Performance of NivolisMonitor and NivolisAnton for Monitoring Patients on Ventilatory Assistance at Home
NCT ID: NCT05698602
Last Updated: 2024-04-29
Study Results
The study team has not published outcome measurements, participant flow, or safety data for this trial yet. Check back later for updates.
Basic Information
Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.
ACTIVE_NOT_RECRUITING
NA
10 participants
INTERVENTIONAL
2023-04-24
2025-12-31
Brief Summary
Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.
The main objective of this study is to show that the use of NivolisMonitor and NivolisAnton is done in complete safety and that the clinical performances of these two devices are achieved. For this, the data measured by NivolisMonitor will be compared with the data provided by the treatment devices used by patients at home and the data provided by NivolisAnton will be compared with the data reported by the transcutaneous capnia monitor.
The secondary objectives concern the evaluation of the acceptability/usability of the devices by patients and healthcare professionals.
This evaluation will take the form of an observational clinical study on 10 patients, including :
* a selection of patients already treated with NIV or HDN according to the inclusion/non-inclusion criteria and submission of the information note
* an inclusion visit taking place during a routine pulmonological follow-up visit for these patients with collection of the free and informed consent of each patient. The blood gas data collected during this routine consultation will be compiled in the study observation book.
* a 3-day monitoring period:
* use over 3 days of the NivolisMonitor device at home, with on the last day the completion of the patient acceptability/usability questionnaire
* overnight use of NivolisAnton
* a teleconsultation to close the study for each patient
* completion of the healthcare professional acceptability/usability questionnaire at the end of the study
Related Clinical Trials
Explore similar clinical trials based on study characteristics and research focus.
Home Monitoring During Chemoradiation
NCT06597240
Novel Non-Invasive Monitoring Parameter in a Hospital Setting
NCT01306201
Telemonitoring in Home Mechanical Ventilation
NCT06143384
ICU-Recover Box 2.0, Smart Technology for Home Monitoring of ICU Patients
NCT07162948
Telematics Device for Daily Monitoring of Vital Signs in Cancer
NCT05650814
Detailed Description
Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.
Since the early 2000s, there has been growing interest in remote monitoring and telemonitoring of patients with chronic respiratory pathologies, in particular patients treated with Non-Invasive Ventilation (NIV). Thus, many studies have focused on the benefits of remote monitoring or telemonitoring. Although the concept of remote home ventilation monitoring seems intuitively useful, analysis of the literature indicates a need for trials with solid evidence. In the same way, nocturnal recording of transcutaneous capnia has shown its reliability and its interest for the follow-up of patients under respiratory assistance at home.
NivolisMonitor and NivolisAnton are new devices developed by Vivardis.
Study Design :
10 patients must be included in this one arm interventional study.
Nivolis study consists in :
* a pre-inclusion visit : V0
* an inclusion visit : V1
* the 3-days study period
* a end of study visit : V2
Pre-inclusion visit V0 :
The patients included in this study are patients already treated with a Home Ventilatory Assistance (NIV) or Home Nasal High Flow device (HDN). Their care is provided by a Home Health Provider (HHP).
Pre-inclusion of patients is carried out by the investigating pulmonologist and the HHP, it aims to:
1. Check the inclusion and non-inclusion criteria
2. And if the patient is eligible, inform the patient of the study, provide him with the information note and the form for obtaining his free and informed consent.
After having been informed, the patients have a period of reflection (1 week) before the inclusion visit during which the pulmonologist will collect the free, informed and written consent of the participant.
Inclusion Visit V1 :
This visit is a classic pulmonological consultation for the follow-up of a patient with Chronic Respiratory Failure by a pulmonologist.
This visit will take place as follows:
1. Collection of the patient's free and informed consent. A copy of the information note and the signed consent is given to the participant. The original copy is kept by the investigating doctor, a third copy is kept by the sponsor.
2. The patient will benefit from his usual consultation by the pulmonologist
3. NIV or HDN device settings: NIV or HDN settings
Data collected:
* Arterial gasometry
* NIV Settings
* HDN settings
* Oxygen therapy
3-days study period : For each patient enrolled in the study, at home, a NivolisMonitor recording for 3 days and a NivolisAnton recording for one night are performed.
The installation of NivolisMonitor and NivolisAnton is carried out by the Home Health Provider who supports the patient as part of the provision of his Ventilatory Assistance or Nasal High Flow device.
End of Study Visit V2 :
This visit is a pneumological teleconsultation for the follow-up of a patient with Respiratory sufficiency Chronic by a pulmonologist.
Data collected during this visit :
* NIV settings and usage
* HDN settings
* Sentec SDM Monitor data
* NivolisMonitor data
* NivolisAnton data
* Usability/Acceptability:
Data Management :
All data concerning visits V0, V 1, V2 will be entered into a CRF. Neither the CRF nor the database contains the identity of the patient (surname, first name). The patient is identified in these 2 documents by a number. The match between names and identification numbers is kept in another file, kept separate from medical information; He ensures the follow-up of the patients participating in the study.
The database is centralized, backed up daily. All these operations are performed continuously, that is, as the data is collected.
Statistics :
Statistical analysis of the data will be performed using IBM SPSS statistical analysis software.
All data are expressed as a mean ± standard deviation, median, frequency or percentage, if applicable.
The materiality level (level α ) is set at 0.05.
Statistical analysis includes:
* A description of patients includes:
1. Number of patients
2. Age, sex, weight, height, BMI
3. Pathologies
* A description of deviations from the protocol: number and % of patients incorrectly included, patients lost at follow-up, patients who did not use the telemedicine system
* A description of all data collected during V1 and V2, as well as during the 3-day collection period.
The frequency distributions will be analyzed with the Chi² test.
For NivolisMonitor:
The Anova analysis will be used to analyze all quantitative data collected during V1 and V2 and during the 3-day collection period.
Data collected from VNI software and NivolisMonitor will be analyzed with a single-factor Anova analysis. A linear regression analysis will be conducted, in addition, a visual analysis by the Bland and Altman method will be performed.
For NivolisAnton:
The comparison between the data collected by NivolisAnton and the Sentec SDM Transcutaneous Capnia Monitor will be performed using Anova Analysis (repeated measurements), one-factor regression analysis, and Bland and Altman visual evaluation.
For the evaluation of Usability/Acceptability, the data collected and the responses to the questionnaires will be expressed in frequency (percentage).
Conditions
See the medical conditions and disease areas that this research is targeting or investigating.
Study Design
Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.
NA
SINGLE_GROUP
DIAGNOSTIC
NONE
Study Groups
Review each arm or cohort in the study, along with the interventions and objectives associated with them.
NivolisMonitor and NivolisAnton Arm
For each patient enrolled in the study, a NivolisMonitor device is added to the usual ventilatory device for 3 days. NivolisMonitor is connected to the patient circuit, between the ventilatory device and the patient interface, ideally at the patient end of the circuit or possibly at the end of the breathing support device of the circuit (NIV or HDN).
NivolisMonitor continuously measures pressure (P in cm H20) and flow (F in L.min-1) as well as temperature (T in degree Celsisus), relative humidity (RH in percent RH) and, optionally, the Fi02 (in percent).
In addition, an overnight transcutaneous capnia recording will be performed using a Sentec SDM monitor to which a NivolisAnton device will be connected and will be able to collect data and transmit it via remote monitoring.
NivolisMonitor
NivolisMonitor comprises a sensing element and controller to measure respiratory values, such as pressure and flow optionnally temperature, relative humidity and oxygen concentration, of a patient being treated with a respiratory therapy devices NIV, CPAP or HDN.
NivolisMonitor is inserted between the respiratory assistance device and the mask or cannula inline with the standard patient circuit used routinely with the respiratory assistance device.
NivolisMonitor functions are:
* measure the respiratory values from the builtin sensor (Pressure, Flow, Temperature, Humidity, Oxygen fraction),
* store the measurements in the embedded memory
* extract from the measurement basic respiratory markers:
* detect if and when a patient is using the respiratory assist device,
* calculate leakage
* calculate respiratory frequency
* transmit the measured values via the NivolisBox component to NivolisPortal
NivolisAnton
NivolisAnton device enables wireless and remote communication from tcPCO2 monitors which measure transcuteanously CO2 and O2 partial pressures in the blood (tcPCO2 \& tcPO2) .
NivolisAnton device connects to the serial port of tcPCO2 monitor, detects if the monitor is actively measuring blood gas values, collects the measurements from the monitor by following manufacturer's communication protocol and stores the collected values in its internal memory. When applicable per the manufacturer recommendations, NivolisAnton collects the values to apply the drift correction as per the manufacturer recommendations.
NivolisAnton transmits the collected data to external receiving devices, such as Medical Devices (NivolisMonitor or another Medical Device that is granted approved data access by Vivardis) via Bluetooth or data servers (NivolisBox as an example) via WIFI. This data is thereafter transmitted by previously mentioned devices to the NivolisPortal server for archiving.
Interventions
Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.
NivolisMonitor
NivolisMonitor comprises a sensing element and controller to measure respiratory values, such as pressure and flow optionnally temperature, relative humidity and oxygen concentration, of a patient being treated with a respiratory therapy devices NIV, CPAP or HDN.
NivolisMonitor is inserted between the respiratory assistance device and the mask or cannula inline with the standard patient circuit used routinely with the respiratory assistance device.
NivolisMonitor functions are:
* measure the respiratory values from the builtin sensor (Pressure, Flow, Temperature, Humidity, Oxygen fraction),
* store the measurements in the embedded memory
* extract from the measurement basic respiratory markers:
* detect if and when a patient is using the respiratory assist device,
* calculate leakage
* calculate respiratory frequency
* transmit the measured values via the NivolisBox component to NivolisPortal
NivolisAnton
NivolisAnton device enables wireless and remote communication from tcPCO2 monitors which measure transcuteanously CO2 and O2 partial pressures in the blood (tcPCO2 \& tcPO2) .
NivolisAnton device connects to the serial port of tcPCO2 monitor, detects if the monitor is actively measuring blood gas values, collects the measurements from the monitor by following manufacturer's communication protocol and stores the collected values in its internal memory. When applicable per the manufacturer recommendations, NivolisAnton collects the values to apply the drift correction as per the manufacturer recommendations.
NivolisAnton transmits the collected data to external receiving devices, such as Medical Devices (NivolisMonitor or another Medical Device that is granted approved data access by Vivardis) via Bluetooth or data servers (NivolisBox as an example) via WIFI. This data is thereafter transmitted by previously mentioned devices to the NivolisPortal server for archiving.
Eligibility Criteria
Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.
Inclusion Criteria
* COPD patient treated with NIV or HDN according to current recommendations
* Patient with Obesity Hypoventilation Syndrome treated with NIV or HDN according to current recommendations
Exclusion Criteria
* Patients treated with NIV other than Philipps Dreamstation or Resmed Lumis
* Hospitalization leading to NIV discontinuation for at least 1 week during the screening phase
* Person deprived of liberty
* Adult protected by law
* Pregnant and breastfeeding women
18 Years
ALL
No
Sponsors
Meet the organizations funding or collaborating on the study and learn about their roles.
VIVARDIS
INDUSTRY
Responsible Party
Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.
Principal Investigators
Learn about the lead researchers overseeing the trial and their institutional affiliations.
Jean-Louis Pépin, Professor
Role: PRINCIPAL_INVESTIGATOR
University Hospital, Grenoble
Locations
Explore where the study is taking place and check the recruitment status at each participating site.
CHU Grenoble Alpes
Grenoble, , France
Countries
Review the countries where the study has at least one active or historical site.
Other Identifiers
Review additional registry numbers or institutional identifiers associated with this trial.
Nivolis
Identifier Type: -
Identifier Source: org_study_id
More Related Trials
Additional clinical trials that may be relevant based on similarity analysis.