Trial Outcomes & Findings for TOKA: Custom Made Devices for High Tibial Osteotomy (HTO) - Clinical Investigation (NCT NCT04574570)
NCT ID: NCT04574570
Last Updated: 2025-08-03
Results Overview
The device morphology assessment is performed by verifying the matching between the planned correction and the post-operative imaging results, along with the investigation of the maintenance of the desired correction at the follow-up meetings. These results are measured through the correction angle, using image-based techniques.
COMPLETED
NA
25 participants
At 6 months after surgery
2025-08-03
Participant Flow
Participant milestones
| Measure |
Interventional Patient
Patients selected and operated for personalised High Tibial Ostotomy (HTO) using theTOKA®
|
|---|---|
|
Overall Study
STARTED
|
25
|
|
Overall Study
COMPLETED
|
25
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
TOKA: Custom Made Devices for High Tibial Osteotomy (HTO) - Clinical Investigation
Baseline characteristics by cohort
| Measure |
Interventional Patient
n=25 Participants
Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
|
|---|---|
|
Age, Continuous
|
54.4 years
STANDARD_DEVIATION 7.4 • n=5 Participants
|
|
Sex: Female, Male
Female
|
6 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
19 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Hispanic or Latino
|
0 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Not Hispanic or Latino
|
0 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Unknown or Not Reported
|
25 Participants
n=5 Participants
|
|
Body Mass Index
|
26.8 kg/m2
STANDARD_DEVIATION 4.0 • n=5 Participants
|
PRIMARY outcome
Timeframe: At 6 months after surgeryPopulation: Operated Patients
The device morphology assessment is performed by verifying the matching between the planned correction and the post-operative imaging results, along with the investigation of the maintenance of the desired correction at the follow-up meetings. These results are measured through the correction angle, using image-based techniques.
Outcome measures
| Measure |
Interventional Patient
n=25 Participants
Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
High Tibial Osteotomy: Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
|
|---|---|
|
Morphology Assessment Via Correction Angle Assessment
|
1.5 Degree
Standard Deviation 2.0
|
PRIMARY outcome
Timeframe: At 6 months after surgeryPopulation: Operated patients
The device morphology assessment is performed by verifying the matching between the planned correction and the post-operative imaging results, along with the investigation of the maintenance of the desired correction at the follow-up meetings. These results are measured through the hip-knee-ankle angle (HKA - mechanical axis) reported in degree, using image-based techniques.
Outcome measures
| Measure |
Interventional Patient
n=25 Participants
Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
High Tibial Osteotomy: Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
|
|---|---|
|
Morphology Assessment Via Hip-knee-ankle Angle Assessment
|
1.5 Degree
Standard Deviation 2.0
|
PRIMARY outcome
Timeframe: At 6 months after surgeryThe device morphology assessment is performed by verifying the matching between the planned correction and the post-operative imaging results, along with the investigation of the maintenance of the desired correction at the follow-up meetings. These results are measured through the Mikulicz point reported in % of the tibial width from the medial to the lateral region, using image-based techniques.
Outcome measures
| Measure |
Interventional Patient
n=25 Participants
Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
High Tibial Osteotomy: Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
|
|---|---|
|
Morphology Assessment Via Mikulicz Point Assessment
|
-9.7 % of medial-lateral tibial plateau width
Standard Deviation 8.4
|
PRIMARY outcome
Timeframe: At 6 months after surgeryPopulation: Operated patiemts
The device morphology assessment is performed by verifying the matching between the planned correction and the post-operative imaging results, along with the investigation of the maintenance of the desired correction at the follow-up meetings. These results are measured through the posterior slope reported in degree, using image-based techniques.
Outcome measures
| Measure |
Interventional Patient
n=25 Participants
Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
High Tibial Osteotomy: Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
|
|---|---|
|
Morphology Assessment Via Posterior Slope Assessment
|
0.1 Degree
Standard Deviation 0.3
|
SECONDARY outcome
Timeframe: Before surgeryPopulation: Operated Patients - Max Knee Flexion Rotation
The functional assessment is performed via three dimensional gait analysis of the patients before surgery. Relevant kinematic results are reported in terms of joint rotations (in degrees)..
Outcome measures
| Measure |
Interventional Patient
n=25 Participants
Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
High Tibial Osteotomy: Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
|
|---|---|
|
Functional Assessment Via Gait Analysis - Kinematics
|
70.5 Degree
Standard Deviation 5.0
|
SECONDARY outcome
Timeframe: Before surgeryPopulation: Operated Patients - Max Abduction Moment
The functional assessment is performed via three dimensional gait analysis of the patients before surgery. Relevant kinetic results are reported in terms of joint moments (in N\*mm, normalized to patient's body weight times height).
Outcome measures
| Measure |
Interventional Patient
n=25 Participants
Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
High Tibial Osteotomy: Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
|
|---|---|
|
Functional Assessment Via Gait Analysis - Kinetics
|
3.8 % of Body Weight * height
Standard Deviation 1.2
|
SECONDARY outcome
Timeframe: 6 months after surgery.Population: Operated patients - KOOS (total) at 6 month after surgery
The functional assessment is performed via clinical scoring using the Knee Osteoarthritis Outcome Score (KOOS). KOOS value range: \[min / max\]= 0 - 100, higher values mean better outcome.
Outcome measures
| Measure |
Interventional Patient
n=25 Participants
Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
High Tibial Osteotomy: Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
|
|---|---|
|
Functional Assessment Via Clinical Scoring - Knee Osteoarthritis Outcome Score
|
81.2 score on a scale
Standard Deviation 9.7
|
SECONDARY outcome
Timeframe: 6 months after surgery.Population: Operated patients - European Quality of life via 5-Dimensions questionnaire (total) at 6 month after surgery
The functional assessment is performed via clinical scoring using the European Quality of life via 5-Dimensions questionnaire (EQ-5D). In this applied scoring system, respondents rank their health status into 5 dimensions: mobility, self-care, usual activities, pain/discomfort, and anxiety/depression. Each dimension can be rated from 1 to 3 as follows: 1=no problem, 2=moderate problems or 3=severe problems. The final total score is the sum of the scores given to each dimension and ranges from 5 (best outcome) to 15 (worst outcome).
Outcome measures
| Measure |
Interventional Patient
n=25 Participants
Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
High Tibial Osteotomy: Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
|
|---|---|
|
Functional Assessment Via Clinical Scoring - European Quality of Life Via 5-Dimensions Questionnaire
|
7.5 units on scale
Standard Deviation 0.9
|
SECONDARY outcome
Timeframe: 6 month after surgery.Population: Pot operative patients at 6 month after surgery
The functional assessment is performed via clinical scoring using the Tegner score. Tegner scale value range: \[min / max\]= 0 - 10, higher values mean better outcome.
Outcome measures
| Measure |
Interventional Patient
n=25 Participants
Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
High Tibial Osteotomy: Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
|
|---|---|
|
Functional Assessment Via Clinical Scoring - Tegner Score
|
4.2 score on a scale
Standard Deviation 1.3
|
SECONDARY outcome
Timeframe: 6 months after surgery.Population: Operated patients - KSS (total) at 6 months after surgery
The functional assessment is performed via clinical scoring using the knee society system score (KSS). KSS value range: \[min / max\]= 0 - 100, higher values mean better outcome.
Outcome measures
| Measure |
Interventional Patient
n=25 Participants
Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
High Tibial Osteotomy: Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
|
|---|---|
|
Functional Assessment Via Clinical Scoring - Knee Society System Score
|
93.5 score on a scale
Standard Deviation 13.7
|
SECONDARY outcome
Timeframe: 6 months after surgery.Population: Operated patients - VAS at 6 month after surgery
The functional assessment is performed via clinical scoring using the Visual Analogue Scale (VAS). VAS is based on a 10-cm long straight line with one end meaning no pain and the other end meaning the worst pain. The score is from 0 to 10 (0= no pain, 10= worst pain ever)
Outcome measures
| Measure |
Interventional Patient
n=25 Participants
Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
High Tibial Osteotomy: Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
|
|---|---|
|
Functional Assessment Via Clinical Scoring - Visual Analogue Scale
|
1.3 score on a scale
Standard Deviation 1.0
|
SECONDARY outcome
Timeframe: Before surgery and at 1, 3, 6, 12 and 24 months after surgery.Population: Operated patients at 6 months after surgery
The functional assessment is performed via x-ray examinations by reporting knee joint alligments in degrees..
Outcome measures
| Measure |
Interventional Patient
n=25 Participants
Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
High Tibial Osteotomy: Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
|
|---|---|
|
The Functional Assessment Via X-ray Examinations
|
181.9 Degree
Standard Deviation 2.6
|
SECONDARY outcome
Timeframe: 6 months after surgeryPopulation: Operated Patients - Max Knee Flexion Rotation at 6 months after surgery
The functional assessment is performed via three dimensional gait analysis of the patients before and after surgery. Relevant kinematic results are reported in terms of joint rotations (in degrees)..
Outcome measures
| Measure |
Interventional Patient
n=25 Participants
Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
High Tibial Osteotomy: Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
|
|---|---|
|
Functional Assessment Via Gait Analysis - Kinematics
|
58 Degree
Standard Deviation 6.0
|
SECONDARY outcome
Timeframe: at 6 months after surgeryPopulation: Operated Patients - Max Abduction Moment
The functional assessment is performed via three dimensional gait analysis of the patients before and after surgery. Relevant kinematic results are reported in terms of joint moments (in N\*mm, normalized to patient's body weight times height).
Outcome measures
| Measure |
Interventional Patient
n=25 Participants
Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
High Tibial Osteotomy: Personalised High Tibial Osteotomy (HTO) using a patient-specific fixation plate (TOKA®)
|
|---|---|
|
Functional Assessment Via Gait Analysis - Kinetics
|
2.8 % of Body Weight * height
Standard Deviation 0.8
|
Adverse Events
Interventional Patient
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Dr Stefano Zaffagnini
IRCCS Istituto Ortopedico Rizzoli
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place