Trial Outcomes & Findings for Electromyographic and Acceleromyographic Monitoring in Restricted Arm Movement Surgical Setting (NCT NCT04352140)

NCT ID: NCT04352140

Last Updated: 2023-02-17

Results Overview

The number of subject's who's residual neuromuscular blockage units are the same on both the TetraGraph and the ToFscan

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

60 participants

Primary outcome timeframe

Up to 1 hour postoperatively

Results posted on

2023-02-17

Participant Flow

Participant milestones

Participant milestones
Measure
Dominant Hand
Subject who received an elective surgical procedure had TetraGraph device lead placement on the dominant hand, ToFscan placed on non-dominant hand Tetragraph: FDA approved neuromuscular transmission monitor capable of measuring the depth of neuromuscular block in anesthetized patients who received neuromuscular blocking agents. TetraGraph uses EMG to measure the muscle action potentials that are generated in response to electrical neurostimulation via skin electrodes. ToFscan: ToFscan is a nerve stimulator module used for the measurement of neuromuscular transmission via accelerometry. ToFscan was developed by Drager Technologies, Canada, and it uses a three-dimensional piezoelectric sensor that attaches to the thumb via a hand adapter to measure acceleration in multiple planes (Murphy et al, 2018). Following simultaneous ulnar nerve stimulation on each arm, the response of the adductor pollicis will be measured
Non-dominant Hand
Subject who received an elective surgical procedure had TetraGraph device lead placement on the non-dominant hand, ToFscan placed on dominant hand Tetragraph: FDA approved neuromuscular transmission monitor capable of measuring the depth of neuromuscular block in anesthetized patients who received neuromuscular blocking agents. TetraGraph uses EMG to measure the muscle action potentials that are generated in response to electrical neurostimulation via skin electrodes. ToFscan: ToFscan is a nerve stimulator module used for the measurement of neuromuscular transmission via accelerometry. ToFscan was developed by Drager Technologies, Canada, and it uses a three-dimensional piezoelectric sensor that attaches to the thumb via a hand adapter to measure acceleration in multiple planes (Murphy et al, 2018). Following simultaneous ulnar nerve stimulation on each arm, the response of the adductor pollicis will be measured
Overall Study
STARTED
30
30
Overall Study
COMPLETED
30
30
Overall Study
NOT COMPLETED
0
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Electromyographic and Acceleromyographic Monitoring in Restricted Arm Movement Surgical Setting

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Dominant Hand
n=30 Participants
Subject who received an elective surgical procedure had TetraGraph device lead placement on the dominant hand, ToFscan placed on non-dominant hand Tetragraph: FDA approved neuromuscular transmission monitor capable of measuring the depth of neuromuscular block in anesthetized patients who received neuromuscular blocking agents. TetraGraph uses EMG to measure the muscle action potentials that are generated in response to electrical neurostimulation via skin electrodes. ToFscan: ToFscan is a nerve stimulator module used for the measurement of neuromuscular transmission via accelerometry. ToFscan was developed by Drager Technologies, Canada, and it uses a three-dimensional piezoelectric sensor that attaches to the thumb via a hand adapter to measure acceleration in multiple planes (Murphy et al, 2018). Following simultaneous ulnar nerve stimulation on each arm, the response of the adductor pollicis will be measured
Non-dominant Hand
n=30 Participants
Subject who received an elective surgical procedure had TetraGraph device lead placement on the non-dominant hand, ToFscan placed on dominant hand Tetragraph: FDA approved neuromuscular transmission monitor capable of measuring the depth of neuromuscular block in anesthetized patients who received neuromuscular blocking agents. TetraGraph uses EMG to measure the muscle action potentials that are generated in response to electrical neurostimulation via skin electrodes. ToFscan: ToFscan is a nerve stimulator module used for the measurement of neuromuscular transmission via accelerometry. ToFscan was developed by Drager Technologies, Canada, and it uses a three-dimensional piezoelectric sensor that attaches to the thumb via a hand adapter to measure acceleration in multiple planes (Murphy et al, 2018). Following simultaneous ulnar nerve stimulation on each arm, the response of the adductor pollicis will be measured
Total
n=60 Participants
Total of all reporting groups
Age, Continuous
60.2 years
STANDARD_DEVIATION 12.49 • n=5 Participants
60.6 years
STANDARD_DEVIATION 14.04 • n=7 Participants
60.37 years
STANDARD_DEVIATION 13.18 • n=5 Participants
Sex: Female, Male
Female
13 Participants
n=5 Participants
12 Participants
n=7 Participants
25 Participants
n=5 Participants
Sex: Female, Male
Male
17 Participants
n=5 Participants
18 Participants
n=7 Participants
35 Participants
n=5 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
Race (NIH/OMB)
Asian
2 Participants
n=5 Participants
0 Participants
n=7 Participants
2 Participants
n=5 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
Race (NIH/OMB)
Black or African American
0 Participants
n=5 Participants
1 Participants
n=7 Participants
1 Participants
n=5 Participants
Race (NIH/OMB)
White
28 Participants
n=5 Participants
29 Participants
n=7 Participants
57 Participants
n=5 Participants
Race (NIH/OMB)
More than one race
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
Region of Enrollment
Hungary
15 participants
n=5 Participants
15 participants
n=7 Participants
30 participants
n=5 Participants
Region of Enrollment
United States
15 participants
n=5 Participants
15 participants
n=7 Participants
30 participants
n=5 Participants

PRIMARY outcome

Timeframe: Up to 1 hour postoperatively

The number of subject's who's residual neuromuscular blockage units are the same on both the TetraGraph and the ToFscan

Outcome measures

Outcome measures
Measure
Dominant Hand
n=30 Participants
Subject who received an elective surgical procedure had TetraGraph device lead placement on the dominant hand, ToFscan placed on non-dominant hand Tetragraph: FDA approved neuromuscular transmission monitor capable of measuring the depth of neuromuscular block in anesthetized patients who received neuromuscular blocking agents. TetraGraph uses EMG to measure the muscle action potentials that are generated in response to electrical neurostimulation via skin electrodes. ToFscan: ToFscan is a nerve stimulator module used for the measurement of neuromuscular transmission via accelerometry. ToFscan was developed by Drager Technologies, Canada, and it uses a three-dimensional piezoelectric sensor that attaches to the thumb via a hand adapter to measure acceleration in multiple planes (Murphy et al, 2018). Following simultaneous ulnar nerve stimulation on each arm, the response of the adductor pollicis will be measured
Non-dominant Hand
n=30 Participants
Subject who received an elective surgical procedure had TetraGraph device lead placement on the non-dominant hand, ToFscan placed on dominant hand Tetragraph: FDA approved neuromuscular transmission monitor capable of measuring the depth of neuromuscular block in anesthetized patients who received neuromuscular blocking agents. TetraGraph uses EMG to measure the muscle action potentials that are generated in response to electrical neurostimulation via skin electrodes. ToFscan: ToFscan is a nerve stimulator module used for the measurement of neuromuscular transmission via accelerometry. ToFscan was developed by Drager Technologies, Canada, and it uses a three-dimensional piezoelectric sensor that attaches to the thumb via a hand adapter to measure acceleration in multiple planes (Murphy et al, 2018). Following simultaneous ulnar nerve stimulation on each arm, the response of the adductor pollicis will be measured
Accuracy of Values Between Tetragraph and ToFscan
0 Participants
2 Participants

SECONDARY outcome

Timeframe: Up to 1 hour postoperatively

Number of patients with train of four ratio \< 0.9 after administration of reversal agent.

Outcome measures

Outcome measures
Measure
Dominant Hand
n=30 Participants
Subject who received an elective surgical procedure had TetraGraph device lead placement on the dominant hand, ToFscan placed on non-dominant hand Tetragraph: FDA approved neuromuscular transmission monitor capable of measuring the depth of neuromuscular block in anesthetized patients who received neuromuscular blocking agents. TetraGraph uses EMG to measure the muscle action potentials that are generated in response to electrical neurostimulation via skin electrodes. ToFscan: ToFscan is a nerve stimulator module used for the measurement of neuromuscular transmission via accelerometry. ToFscan was developed by Drager Technologies, Canada, and it uses a three-dimensional piezoelectric sensor that attaches to the thumb via a hand adapter to measure acceleration in multiple planes (Murphy et al, 2018). Following simultaneous ulnar nerve stimulation on each arm, the response of the adductor pollicis will be measured
Non-dominant Hand
n=30 Participants
Subject who received an elective surgical procedure had TetraGraph device lead placement on the non-dominant hand, ToFscan placed on dominant hand Tetragraph: FDA approved neuromuscular transmission monitor capable of measuring the depth of neuromuscular block in anesthetized patients who received neuromuscular blocking agents. TetraGraph uses EMG to measure the muscle action potentials that are generated in response to electrical neurostimulation via skin electrodes. ToFscan: ToFscan is a nerve stimulator module used for the measurement of neuromuscular transmission via accelerometry. ToFscan was developed by Drager Technologies, Canada, and it uses a three-dimensional piezoelectric sensor that attaches to the thumb via a hand adapter to measure acceleration in multiple planes (Murphy et al, 2018). Following simultaneous ulnar nerve stimulation on each arm, the response of the adductor pollicis will be measured
Incidence of Residual Neuromuscular Blockade
3 Participants
2 Participants

Adverse Events

Dominant Hand

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Non-dominant Hand

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

J. Ross Renew, M.D.

Mayo Clinic

Phone: 904-956-3325

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place