Trial Outcomes & Findings for Hemodynamics Measurement in Radiofrequency Catheter Ablation (NCT NCT04227119)
NCT ID: NCT04227119
Last Updated: 2021-06-14
Results Overview
The systolic pressure values for atrium and ventricle heart chambers are presented for the irrigated ablation catheter derived measurements and the balloon tipped PA catheter measurements.
COMPLETED
NA
11 participants
Day 1 (during procedure)
2021-06-14
Participant Flow
Participants were recruited from the patient population at Emory University Hospital and the Emory Clinic, in Atlanta, Georgia who were scheduled to have a routine cardiac ablation using an irrigated ablation catheter for treatment of arrhythmia. Participants were enrolled between January 23, 2020 and May 18, 2020.
Participant milestones
| Measure |
Irrigated Ablation Catheter and Balloon Tipped PA Catheter
Participants undergoing cardiac ablation had cardiac pressures measured with an irrigated ablation catheter (standard protocol for this procedure) and a 5 French (5F) sized balloon tipped pulmonary artery (PA) catheter (for study purposes only).
The catheter will be positioned in either the left or right side of the heart.
|
|---|---|
|
Overall Study
STARTED
|
11
|
|
Overall Study
COMPLETED
|
11
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Race and Ethnicity were not collected from any participant.
Baseline characteristics by cohort
| Measure |
Irrigated Ablation Catheter and Balloon Tipped PA Catheter
n=11 Participants
Participants undergoing cardiac ablation had cardiac pressures measured with an irrigated ablation catheter (standard protocol for this procedure) and a 5F balloon tipped pulmonary artery (PA) catheter (for study purposes only).
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=11 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
6 Participants
n=11 Participants
|
|
Age, Categorical
>=65 years
|
5 Participants
n=11 Participants
|
|
Age, Continuous
|
66.2 years
STANDARD_DEVIATION 8.2 • n=11 Participants
|
|
Sex: Female, Male
Female
|
4 Participants
n=11 Participants
|
|
Sex: Female, Male
Male
|
7 Participants
n=11 Participants
|
|
Region of Enrollment
United States
|
11 Participants
n=11 Participants
|
|
Indication for Ablation
Atrial fibrillation
|
9 Participants
n=11 Participants
|
|
Indication for Ablation
Ventricular tachycardia
|
2 Participants
n=11 Participants
|
|
Ejection Fraction
|
47.3 percentage of blood pumped
STANDARD_DEVIATION 16.1 • n=11 Participants
|
|
Body Mass Index (BMI)
|
31.7 kg/m^2
STANDARD_DEVIATION 4.6 • n=11 Participants
|
|
Type of Atrial Fibrillation
Paroxysmal atrial fibrillation
|
4 Participants
n=11 Participants
|
|
Type of Atrial Fibrillation
Persistent atrial fibrillation
|
5 Participants
n=11 Participants
|
|
Type of Atrial Fibrillation
No atrial fibrillation
|
2 Participants
n=11 Participants
|
PRIMARY outcome
Timeframe: Day 1 (during procedure)Population: Participants had the ablation catheter positioned in only one side of the heart. Nine participants had measurements obtained from the left atrium and ventricle chambers, and two participants had measurements obtained from the right atrium and ventricle chambers.
The systolic pressure values for atrium and ventricle heart chambers are presented for the irrigated ablation catheter derived measurements and the balloon tipped PA catheter measurements.
Outcome measures
| Measure |
Irrigated Ablation Catheter
n=11 Participants
Participants undergoing cardiac ablation had cardiac pressures measured with an irrigated ablation catheter (standard protocol for this procedure).
|
Balloon Tipped PA Catheter
n=11 Participants
Participants undergoing cardiac ablation had cardiac pressures measured with a 5F balloon tipped pulmonary artery (PA) catheter (for study purposes only).
|
|---|---|---|
|
Systolic Pressure Value
Left ventricle
|
114.6 millimeters of mercury (mmHg)
Standard Deviation 33.4
|
105.9 millimeters of mercury (mmHg)
Standard Deviation 30.4
|
|
Systolic Pressure Value
Left atrium
|
26.5 millimeters of mercury (mmHg)
Standard Deviation 6.4
|
27.75 millimeters of mercury (mmHg)
Standard Deviation 7.4
|
|
Systolic Pressure Value
Right atrium
|
16.6 millimeters of mercury (mmHg)
Standard Deviation 5.1
|
13.6 millimeters of mercury (mmHg)
Standard Deviation 6.3
|
|
Systolic Pressure Value
Right ventricle
|
42.8 millimeters of mercury (mmHg)
Standard Deviation 3.6
|
54.1 millimeters of mercury (mmHg)
Standard Deviation 2.8
|
PRIMARY outcome
Timeframe: Day 1 (during procedure)Population: Participants had the ablation catheter positioned in only one side of the heart. Nine participants had measurements obtained from the left atrium and ventricle chambers, and two participants had measurements obtained from the right atrium and ventricle chambers.
The diastolic pressure values for ventricle heart chambers are presented for the irrigated ablation catheter derived measurements and the balloon tipped PA catheter measurements.
Outcome measures
| Measure |
Irrigated Ablation Catheter
n=11 Participants
Participants undergoing cardiac ablation had cardiac pressures measured with an irrigated ablation catheter (standard protocol for this procedure).
|
Balloon Tipped PA Catheter
n=11 Participants
Participants undergoing cardiac ablation had cardiac pressures measured with a 5F balloon tipped pulmonary artery (PA) catheter (for study purposes only).
|
|---|---|---|
|
Diastolic Pressure Value
Left ventricle
|
15.3 mmHg
Standard Deviation 4.0
|
17.3 mmHg
Standard Deviation 3.2
|
|
Diastolic Pressure Value
Right ventricle
|
20.2 mmHg
Standard Deviation 5.2
|
15.8 mmHg
Standard Deviation 4.6
|
POST_HOC outcome
Timeframe: Day 1 (during procedure)Population: This analysis includes the nine participants had measurements obtained from the left ventricle chamber.
The systolic pressure values per time, in milliseconds (ms), for the left ventricle chamber is presented for the irrigated ablation catheter derived measurements and the balloon tipped PA catheter measurements.
Outcome measures
| Measure |
Irrigated Ablation Catheter
n=9 Participants
Participants undergoing cardiac ablation had cardiac pressures measured with an irrigated ablation catheter (standard protocol for this procedure).
|
Balloon Tipped PA Catheter
n=9 Participants
Participants undergoing cardiac ablation had cardiac pressures measured with a 5F balloon tipped pulmonary artery (PA) catheter (for study purposes only).
|
|---|---|---|
|
Systolic Pressure Per Millisecond
|
0.49 mmHg/ms
Standard Deviation 0.15
|
0.53 mmHg/ms
Standard Deviation 0.11
|
Adverse Events
Irrigated Ablation Catheter
Balloon Tipped PA Catheter
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place