Trial Outcomes & Findings for Ankle Foot Orthosis Comparative Effect (NCT NCT03901053)
NCT ID: NCT03901053
Last Updated: 2026-01-29
Results Overview
The Patient Reported Outcome Information System (PROMIS) physical function Computer Adaptive Test (CAT) is a computerized assessment measuring physical function. It is scored using a T-score in which 50 is the mean of a relevant reference population and 10 is the standard deviation (SD) of that population. In a given PROMIS domain, a T-score above 50 represents more of the measured variable than the population average.
COMPLETED
NA
54 participants
3 months
2026-01-29
Participant Flow
Four (4) of the participants in this study failed the post-informed consent screening checklist, and were excluded form the study before being assigned to either group. The fifty (50) eligible participants were randomized to one of the two groups.
Participant milestones
| Measure |
Reaktiv, PhatBrace
Participants in this arm will receive the Reaktiv AFO made by FabTech Systems LLC first, then the PhatBrace AFO by Bio-Mechanical Composites Inc. second.
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
PhatBrace, Reaktiv
Participants in this arm will receive the PhatBrace AFO by Bio-Mechanical Composites Inc. first, then the Reaktiv AFO made by FabTech Systems LLC second.
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Overall Study
STARTED
|
25
|
25
|
|
Overall Study
COMPLETED
|
12
|
14
|
|
Overall Study
NOT COMPLETED
|
13
|
11
|
Reasons for withdrawal
| Measure |
Reaktiv, PhatBrace
Participants in this arm will receive the Reaktiv AFO made by FabTech Systems LLC first, then the PhatBrace AFO by Bio-Mechanical Composites Inc. second.
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
PhatBrace, Reaktiv
Participants in this arm will receive the PhatBrace AFO by Bio-Mechanical Composites Inc. first, then the Reaktiv AFO made by FabTech Systems LLC second.
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Overall Study
Lost to Follow-up
|
6
|
2
|
|
Overall Study
Withdrawal by Subject
|
4
|
3
|
|
Overall Study
Unsatisfactory Device
|
1
|
2
|
|
Overall Study
No longer meets study criteria
|
1
|
3
|
|
Overall Study
Subject's Medical Status
|
0
|
1
|
|
Overall Study
Screening Failure
|
1
|
0
|
Baseline Characteristics
Ankle Foot Orthosis Comparative Effect
Baseline characteristics by cohort
| Measure |
Reaktiv, PhatBrace
n=12 Participants
Participants in this arm will receive the Reaktiv AFO made by FabTech Systems LLC first, then the PhatBrace AFO by Bio-Mechanical Composites Inc. second.
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
PhatBrace, Reaktiv
n=14 Participants
Participants in this arm will receive the PhatBrace AFO by Bio-Mechanical Composites Inc. first, then the Reaktiv AFO made by FabTech Systems LLC second.
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
Total
n=26 Participants
Total of all reporting groups
|
|---|---|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=35 Participants
|
0 Participants
n=4328 Participants
|
0 Participants
n=8687 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
12 Participants
n=35 Participants
|
13 Participants
n=4328 Participants
|
25 Participants
n=8687 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=35 Participants
|
1 Participants
n=4328 Participants
|
1 Participants
n=8687 Participants
|
|
Age, Continuous
|
38.3 years
STANDARD_DEVIATION 10.4 • n=35 Participants
|
44.5 years
STANDARD_DEVIATION 10.8 • n=4328 Participants
|
41.7 years
STANDARD_DEVIATION 10.9 • n=8687 Participants
|
|
Sex: Female, Male
Female
|
4 Participants
n=35 Participants
|
4 Participants
n=4328 Participants
|
8 Participants
n=8687 Participants
|
|
Sex: Female, Male
Male
|
8 Participants
n=35 Participants
|
10 Participants
n=4328 Participants
|
18 Participants
n=8687 Participants
|
|
Ethnicity (NIH/OMB)
Hispanic or Latino
|
2 Participants
n=35 Participants
|
1 Participants
n=4328 Participants
|
3 Participants
n=8687 Participants
|
|
Ethnicity (NIH/OMB)
Not Hispanic or Latino
|
10 Participants
n=35 Participants
|
13 Participants
n=4328 Participants
|
23 Participants
n=8687 Participants
|
|
Ethnicity (NIH/OMB)
Unknown or Not Reported
|
0 Participants
n=35 Participants
|
0 Participants
n=4328 Participants
|
0 Participants
n=8687 Participants
|
|
Race (NIH/OMB)
American Indian or Alaska Native
|
1 Participants
n=35 Participants
|
0 Participants
n=4328 Participants
|
1 Participants
n=8687 Participants
|
|
Race (NIH/OMB)
Asian
|
0 Participants
n=35 Participants
|
0 Participants
n=4328 Participants
|
0 Participants
n=8687 Participants
|
|
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
|
0 Participants
n=35 Participants
|
0 Participants
n=4328 Participants
|
0 Participants
n=8687 Participants
|
|
Race (NIH/OMB)
Black or African American
|
0 Participants
n=35 Participants
|
0 Participants
n=4328 Participants
|
0 Participants
n=8687 Participants
|
|
Race (NIH/OMB)
White
|
10 Participants
n=35 Participants
|
13 Participants
n=4328 Participants
|
23 Participants
n=8687 Participants
|
|
Race (NIH/OMB)
More than one race
|
1 Participants
n=35 Participants
|
1 Participants
n=4328 Participants
|
2 Participants
n=8687 Participants
|
|
Race (NIH/OMB)
Unknown or Not Reported
|
0 Participants
n=35 Participants
|
0 Participants
n=4328 Participants
|
0 Participants
n=8687 Participants
|
PRIMARY outcome
Timeframe: 3 monthsThe Patient Reported Outcome Information System (PROMIS) physical function Computer Adaptive Test (CAT) is a computerized assessment measuring physical function. It is scored using a T-score in which 50 is the mean of a relevant reference population and 10 is the standard deviation (SD) of that population. In a given PROMIS domain, a T-score above 50 represents more of the measured variable than the population average.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=26 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
PROMIS Patient Reported Outcomes for Physical Function
|
44.78 score on a scale
Standard Deviation 6.01
|
45.02 score on a scale
Standard Deviation 8.35
|
PRIMARY outcome
Timeframe: 3 monthsThe Patient Reported Outcome Information System (PROMIS) pain interference Computer Adaptive Test (CAT) is a computerized assessment measuring pain interference. It is scored using a T-score in which 50 is the mean of a relevant reference population and 10 is the standard deviation (SD) of that population. In a given PROMIS domain, a T-score above 50 represents more of the measured variable than the population average.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=26 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
PROMIS Patient Reported Outcomes for Pain Interference
|
55.22 T-score
Standard Deviation 6.30
|
53.49 T-score
Standard Deviation 6.52
|
PRIMARY outcome
Timeframe: 3 monthsThe Activities-Specific Balance Confidence (ABC) scale provides a standardized evaluation of balance confidence across a range of common daily tasks. Higher scores indicate more confidence, with 0 = no confidence and 100 = full confidence.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=26 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Activities-Specific Balance Confidence (ABC)
|
76.37 units on a scale
Standard Deviation 19.66
|
76.92 units on a scale
Standard Deviation 18.92
|
PRIMARY outcome
Timeframe: 3 monthsSatisfaction with device will be assessed using the Orthotics Prosthetics Users' Survey Satisfaction With Device Score (11-55). Higher scores indicate a better outcome.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=26 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Satisfaction With Device (OPUS - CSD)
|
35.88 score on a scale
Standard Deviation 6.13
|
36.04 score on a scale
Standard Deviation 8.06
|
PRIMARY outcome
Timeframe: 3 monthsComfort scores range from 0 = most uncomfortable to 10 = most comfortable.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=26 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Modified Socket Comfort Score (Comfort)
|
7.12 score on a scale
Standard Deviation 2.10
|
7.12 score on a scale
Standard Deviation 1.93
|
PRIMARY outcome
Timeframe: 3 monthsSmoothness scores range from 0 = least smooth to 10 = most smooth.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=26 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Modified Socket Comfort Score (Comfort and Smoothness)
|
7.65 score on a scale
Standard Deviation 1.92
|
7.46 score on a scale
Standard Deviation 2.06
|
PRIMARY outcome
Timeframe: 3 monthsPopulation: One participant failed to complete in-person physical performance measures in the PhatBrace condition, and their reported numerical pain rating was unable to be collected as a result.
Pain will be assessed using a standard 11-point numerical pain rating scale, in which 0 = no pain and 10 = worst pain imaginable.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=25 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Numerical Pain Rating Scale
|
1.85 score on a scale
Standard Deviation 2.03
|
1.68 score on a scale
Standard Deviation 1.75
|
PRIMARY outcome
Timeframe: After 3 months of accomodation with the final device (Crossover study)The participant will rank order their preference for the Reaktiv and Phatbrace on a questionnaire.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=26 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Participant Device Preference
|
10 Participants
|
16 Participants
|
PRIMARY outcome
Timeframe: 3 monthsPopulation: One participant did not complete in-person physical performance measures in the PhatBrace condition, and their four-square step test time was unable to be collected as a result.
The 4SST is a standardized timed test of balance and agility. Measured in seconds with lower scores indicating better balance and agility.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=25 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Four-square Step Test (4SST- Timed)
|
7.39 Seconds
Standard Deviation 1.72
|
7.06 Seconds
Standard Deviation 1.78
|
PRIMARY outcome
Timeframe: 3 monthsPopulation: One participant did not complete in-person physical performance measures in the PhatBrace condition, and their self-selected walking velocity was unable to be collected as a result.
SSWV will be assessed using the timed 10 meter walk test. Measured in seconds.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=25 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Self-selected Walking Velocity (SSWV - Timed)
|
4.46 Seconds
Standard Deviation 0.58
|
4.49 Seconds
Standard Deviation 0.60
|
PRIMARY outcome
Timeframe: 3 monthsPopulation: One participant did not complete in-person physical performance measures in the PhatBrace condition, and their 10 meter shuttle run was unable to be collected as a result.
The 10 meter shuttle run is a well-established timed measure of speed. Measured in seconds.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=25 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
10 Meter Shuttle Run (10M Shuttle - Timed)
|
19.05 Seconds
Standard Deviation 5.81
|
19.05 Seconds
Standard Deviation 6.41
|
PRIMARY outcome
Timeframe: 3 monthsPopulation: One participant did not complete the in-person biomechanics collection in the PhatBrace condition, and their ankle joint power was unable to be collected as a result.
Peak ankle joint moment (W/kg) during gait.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=25 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Ankle Joint Power
|
1.43 W/kg
Standard Deviation 0.17
|
1.29 W/kg
Standard Deviation 0.27
|
PRIMARY outcome
Timeframe: 3 monthsPopulation: One participant did not complete the in-person biomechanics collection in the PhatBrace condition, and their center of pressure velocity was unable to be collected as a result.
Timing of peak center of pressure velocity (percent stance) during gait.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=25 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Center of Pressure Velocity Timing
|
25.81 percent of gait cycle
Standard Deviation 12.69
|
15.11 percent of gait cycle
Standard Deviation 8.83
|
PRIMARY outcome
Timeframe: 3 monthsPopulation: One participant did not complete the in-person biomechanics collection in the PhatBrace condition, and their center of pressure velocity was unable to be collected as a result.
Magnitude of peak center of pressure velocity (m/s) during gait.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=25 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Center of Pressure Velocity Magnitude
|
0.89 m/s
Standard Deviation 0.29
|
0.75 m/s
Standard Deviation 0.21
|
SECONDARY outcome
Timeframe: 3 monthsThe Patient Reported Outcome Information System (PROMIS) satisfaction with participation in social activities Computer Adaptive Test (CAT) is a computerized assessment measuring satisfaction with participation in social activities. It is scored using a T-score in which 50 is the mean of a relevant reference population and 10 is the standard deviation (SD) of that population. In a given PROMIS domain, a T-score above 50 represents more of the measured variable than the population average.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=26 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
PROMIS Patient Reported Outcomes for Satisfaction With Participation in Social Activities
|
49.39 T-score
Standard Deviation 7.58
|
48.42 T-score
Standard Deviation 7.22
|
SECONDARY outcome
Timeframe: BaselineParticipants will report their physical activity using the Paffenbarger Physical Activity Questionnaire. Average hours of vigorous activity completed during the week and during the weekend will be recorded separately. Participants completed this questionnaire at the beginning of the study prior and reported baseline physical activity without wearing either CDO.
Outcome measures
| Measure |
Reaktiv
n=12 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=14 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Paffenbarger Physical Activity Questionnaire
Average on weekday
|
2.17 Hours
Standard Deviation 2.25
|
1.66 Hours
Standard Deviation 1.81
|
|
Paffenbarger Physical Activity Questionnaire
Average on weekend
|
1.75 Hours
Standard Deviation 2.09
|
2.39 Hours
Standard Deviation 2.37
|
SECONDARY outcome
Timeframe: 3 monthsPopulation: One participant did not complete the in-person biomechanics collection in the PhatBrace condition, and their ankle joint moment was unable to be collected as a result.
Peak ankle joint moment (Nm/kg) during gait.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=25 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Ankle Joint Moment
|
1.43 Nm/kg
Standard Deviation 0.17
|
1.29 Nm/kg
Standard Deviation 0.27
|
OTHER_PRE_SPECIFIED outcome
Timeframe: 3 monthsThe Patient Reported Outcome Information System (PROMIS) pain behavior Computer Adaptive Test (CAT) is a computerized assessment measuring pain behavior. It is scored using a T-score in which 50 is the mean of a relevant reference population and 10 is the standard deviation (SD) of that population. In a given PROMIS domain, a T-score above 50 represents more of the measured variable than the population average.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=26 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
PROMIS Patient Reported Outcomes for Pain Behavior
|
55.22 T-score
Standard Deviation 5.60
|
54.45 T-score
Standard Deviation 7.79
|
OTHER_PRE_SPECIFIED outcome
Timeframe: 3 monthsThe Patient Reported Outcome Information System (PROMIS) depression Computer Adaptive Test (CAT) is a computerized assessment measuring depression. It is scored using a T-score in which 50 is the mean of a relevant reference population and 10 is the standard deviation (SD) of that population. In a given PROMIS domain, a T-score above 50 represents more of the measured variable than the population average.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=26 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
PROMIS Patient Reported Outcomes for Depression
|
48.24 T-score
Standard Deviation 9.72
|
48.35 T-score
Standard Deviation 8.86
|
OTHER_PRE_SPECIFIED outcome
Timeframe: 3 monthsThe Patient Reported Outcome Information System (PROMIS) satisfaction with participation in social roles Computer Adaptive Test (CAT) is a computerized assessment measuring satisfaction with participation in social roles. It is scored using a T-score in which 50 is the mean of a relevant reference population and 10 is the standard deviation (SD) of that population. In a given PROMIS domain, a T-score above 50 represents more of the measured variable than the population average.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=26 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
PROMIS Patient Reported Outcomes for Satisfaction With Participation in Social Roles
|
49.37 T-score
Standard Deviation 8.56
|
48.70 T-score
Standard Deviation 7.48
|
OTHER_PRE_SPECIFIED outcome
Timeframe: 3 monthsSatisfaction with services will be assessed using the Orthotics Prosthetics Users' Survey Satisfaction With Services Score (10-50). Higher scores indicate a better outcome.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=26 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Satisfaction With Services (OPUS - CSS)
|
40.04 score on a scale
Standard Deviation 8.47
|
38.65 score on a scale
Standard Deviation 8.75
|
OTHER_PRE_SPECIFIED outcome
Timeframe: 3 monthsPopulation: One participant did not complete in-person physical performance measures in the PhatBrace condition, and their sit-to-stand time was unable to be collected as a result.
STS5 is a well-established timed measure of lower limb muscle strength and power. Participants are instructed to stand up and sit down 5 times as fast as possible.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=25 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Sit to Stand 5 Times (STS5 - Timed)
|
8.94 Seconds
Standard Deviation 2.20
|
8.82 Seconds
Standard Deviation 2.55
|
OTHER_PRE_SPECIFIED outcome
Timeframe: 3 monthsPopulation: One participant did not complete in-person physical performance measures in the PhatBrace condition, and timed stair ascent was unable to be collected as a result.
The timed stair ascent test is a measure of mobility, agility, and lower limb muscle strength. Participants are instructed to ascend a flight of 12 steps as quickly and safely as possible while contacting every step.
Outcome measures
| Measure |
Reaktiv
n=26 Participants
Reaktiv AFO: This device is a carbon fiber ankle foot orthosis (brace) that contacts the foot and lower leg. It consists of a foot plate that is designed to protect the foot, a carbon fiber strut on the back that stores and returns energy, and a cuff that goes below the knee. It has a custom geometry and alignment that is specific to the individual.
|
PhatBrace
n=25 Participants
PhatBrace AFO: Although lighter and thinner than the Reaktiv it has a similar overall design. The posterior strut is typically longer and the footplate is thinner, more compliant and less customized. Further, the device is formed into a single piece.
|
|---|---|---|
|
Timed Stair Ascent (TSA - Timed)
|
5.82 Seconds
Standard Deviation 2.97
|
5.43 Seconds
Standard Deviation 2.60
|
Adverse Events
Reaktiv
PhatBrace
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place