Rainbow R2-25 vs. R1-25 SpHb Sub-Range Performance Equivalence

NCT ID: NCT03134326

Last Updated: 2017-07-27

Study Results

Results available

Outcome measurements, participant flow, baseline characteristics, and adverse events have been published for this study.

View full results

Basic Information

Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.

Recruitment Status

COMPLETED

Clinical Phase

NA

Total Enrollment

21 participants

Study Classification

INTERVENTIONAL

Study Start Date

2009-06-18

Study Completion Date

2009-07-02

Brief Summary

Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.

In this study, the concentration of hemoglobin was measured using noninvasive pulse oximeter sensors and compared against the hemoglobin measurement from the subject's blood sample under controlled conditions. Data collected in the study was in the sub-range of 12-17 g/dL in the device's 8-17 g/dL specification range.

The purpose of the study was to evaluate a change in the sensor component and to assess performance equivalence in the sub-range. The performance of non-invasive hemoglobin is not uniform across the accuracy range.

Detailed Description

Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.

Conditions

See the medical conditions and disease areas that this research is targeting or investigating.

Healthy

Study Design

Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.

Allocation Method

NA

Intervention Model

SINGLE_GROUP

Primary Study Purpose

OTHER

Study Groups

Review each arm or cohort in the study, along with the interventions and objectives associated with them.

Test group

All subjects will be enrolled in the test group and will receive both R1-25 and R2-25 Pulse Oximeter Sensors

Group Type EXPERIMENTAL

R1-25 and R2-25 Pulse Oximeter Sensor

Intervention Type DEVICE

Interventions

Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.

R1-25 and R2-25 Pulse Oximeter Sensor

Intervention Type DEVICE

Eligibility Criteria

Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.

Inclusion Criteria

* Greater than 18 years old
* Healthy

Exclusion Criteria

* Do not understand the study and the risks involved
* Deemed ineligible by study test personnel
Minimum Eligible Age

18 Years

Eligible Sex

ALL

Accepts Healthy Volunteers

Yes

Sponsors

Meet the organizations funding or collaborating on the study and learn about their roles.

Masimo Corporation

INDUSTRY

Sponsor Role lead

Responsible Party

Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.

Responsibility Role SPONSOR

Locations

Explore where the study is taking place and check the recruitment status at each participating site.

Masimo Clinical Lab

Irvine, California, United States

Site Status

Countries

Review the countries where the study has at least one active or historical site.

United States

Other Identifiers

Review additional registry numbers or institutional identifiers associated with this trial.

TR18735-1029

Identifier Type: -

Identifier Source: org_study_id

More Related Trials