Accuracy of Zero-flux and Ingestible Thermometers

NCT ID: NCT02121574

Last Updated: 2016-01-12

Study Results

Results pending

The study team has not published outcome measurements, participant flow, or safety data for this trial yet. Check back later for updates.

Basic Information

Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.

Recruitment Status

COMPLETED

Total Enrollment

30 participants

Study Classification

OBSERVATIONAL

Study Start Date

2014-09-30

Study Completion Date

2015-06-30

Brief Summary

Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.

Background When a patient is anesthetised and undergoes an operation, their body temperature falls. Managing a body temperature properly by keeping it within normal limits has been shown to improve recovery from surgery.

A vital part of managing temperature is being able to measure it properly. There are several ways of doing so, which consist of invasive methods such as temperature probes in the oesophagus and bladder, and non invasive methods such as infrared scanners and ear canal probes. During an operation, temperature changes as small as 0.5 °C can affect outcomes but, currently, only invasive methods are reliable enough to detect this. Oesophageal probes can't be used in patients who are not anaesthetised during their operations, and it is inappropriate to insert a urinary catheter should it not be otherwise needed.

The purpose of this study is to look at two new methods of measuring a patient's temperature which would help us in these situations. The first is called zero-flux thermometry and involves placing an adhesive pad to the forehead, which gives constant measurement of a patient's core temperature. The second involves swallowing an ingestible capsule that is no bigger than a normal pill. This technique has been very useful in research tracking a patient's core temperature as they are transferred through different areas over a longer period of time. Whilst these two methods are potentially very useful, they have never been tested on hospital patients.

Aims:

The investigators want to determine whether readings from these thermometers are accurate enough to be used in patients before, during and after surgery. Should they prove to be sufficiently accurate it will allow us to improve patient care through both clinical practice and research.

Plan:

During operations, temperature is routinely measured in line with NICE guidelines with a thermometer placed in the nose or oesophagous. The investigators will not be changing routine practice in any way, but checking measurements from the new devices against the proven, routine ones.

The investigators plan to record measurements from each device every minute throughout the operation in 20 patients. The investigators will then perform statistical tests to check their accuracy.

Study Hypothesis: Zero flux and ingestible thermometers are not significantly different at detecting patients body temperature than oesophageal temperature probes

Detailed Description

Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.

Summary:

The study has been designed to provide information on the suitability and accuracy of two methods of temperature measurement in the perioperative period.

Background:

Good perioperative temperature management is an important means of improving patient outcome \[1\].

An essential part of optimal management is accurate temperature measurement. The gold standard is pulmonary artery catheter (PAC) measurement. However, these are used infrequently in British practice due to the high rates of associated complications. Urinary bladder and oesophageal temperatures are sufficiently accurate but are unsuitable for many patients \[2\]. For example it is not possible to use oesophageal probes in awake patients and many patients are not catheterised.

Although non-invasive and relatively cheap means of measuring temperature (such as aural canal probes and infrared scanners) exist, they are neither accurate nor consistent enough to distinguish the 0.5˚C differences which have a significant effect on patient outcomes \[2\].

In this study, the investigators plan to look at two methods of temperature measurement which potentially offer significant improvements to surgical care.

The first is a new method known as zero-flux thermometry. This is completely non-invasive, involving no more than placing an adhesive pad to the forehead. Connected up to a monitor, this will equilibrate with the patient's core temperature and give a constant read-out.

The second is an ingestible capsule. These have become an important tool for measuring both hypothermia and hyperthermia in other settings. Their accuracy has not, however, been tested in hospital patients. This technology, although too expensive for routine use, offers significant advantages for research as they can provide a constant record of temperature throughout the patient's stay in hospital. For example, the investigators' audit data show that if patients admitted with fractured neck of femur become hypothermic at any time, their mortality increases. Subsequent to this we showed that patient temperatures drop significantly between ambulance, A\&E and admission to the ward. These parts of the patient journey are possible targets for the prevention of hypothermia and the continuous monitoring offered by these devices will allow us to accurately test the efficacy of the interventions.

Aims and objectives:

The aim of this study is to determine the accuracy in the perioperative period of zero-flux and ingestible temperature readings.

Previous studies have shown that oesophageal and nasopharyngeal measurements can be taken as an accurate reflection of core temperature \[2\]. The consistency and accuracy of the other two will be compared to this.

Plan of investigation:

This study will be purely observational. There will be no interference with current practice. Nasopharyngeal/oesophageal probes are routinely placed, according to NICE guidelines, in these patients. The zero-flux thermometer is non-invasive and the ingestible capsule no bigger than a pill, so neither provide any additional risk to the patient.

Temperature readings will be downloaded from each of the three devices with a maximum interval of 1 minute. The investigators have calculated that readings from 15 subjects will provide sufficient power. The investigators plan to recruit 20 patients. Analysis of agreement between the methods will be performed as suggested by Bland and Altman\[3\].

Conditions

See the medical conditions and disease areas that this research is targeting or investigating.

Surgery Hypothermia Anesthesia

Study Design

Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.

Study Time Perspective

CROSS_SECTIONAL

Study Groups

Review each arm or cohort in the study, along with the interventions and objectives associated with them.

Zero-flux and ingestible thermometer

Ingestion of a capsule thermometer pre-operatively Attachment of a zero flux temperature electrode intraoperatively Standard oesophageal thermometer

Zero flux and ingestible thermometers

Intervention Type DEVICE

Interventions

Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.

Zero flux and ingestible thermometers

Intervention Type DEVICE

Eligibility Criteria

Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.

Inclusion Criteria

Participants will be patients undergoing elective gynaecological, vascular or urological surgery at the Royal Sussex County Hospital, and as such will be seen at a pre-operative assessment clinic as part of their normal work up for theatre.

Exclusion Criteria

* Lacking mental capacity
* Not speaking English
* Malignancy in medical history
* Undergoing open abdominal surgery
Eligible Sex

ALL

Accepts Healthy Volunteers

No

Sponsors

Meet the organizations funding or collaborating on the study and learn about their roles.

Brighton and Sussex University Hospitals NHS Trust

OTHER

Sponsor Role lead

Responsible Party

Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.

Dr C. Mark Harper

Dr C. Mark Harper Consultant Anaesthetist

Responsibility Role PRINCIPAL_INVESTIGATOR

Principal Investigators

Learn about the lead researchers overseeing the trial and their institutional affiliations.

James M Jack, MBBS BSc

Role: PRINCIPAL_INVESTIGATOR

BSUH NHS Trust

Locations

Explore where the study is taking place and check the recruitment status at each participating site.

Royal Sussex County Hospital

Brighton, East Sussex, United Kingdom

Site Status

Royal Sussex County Hospital

Brighton, East Sussex, United Kingdom

Site Status

Countries

Review the countries where the study has at least one active or historical site.

United Kingdom

References

Explore related publications, articles, or registry entries linked to this study.

Jack JM, Ellicott H, Jones CI, Bremner SA, Densham I, Harper CM. Determining the accuracy of zero-flux and ingestible thermometers in the peri-operative setting. J Clin Monit Comput. 2019 Dec;33(6):1113-1118. doi: 10.1007/s10877-019-00252-9. Epub 2019 Jan 8.

Reference Type DERIVED
PMID: 30623279 (View on PubMed)

Other Identifiers

Review additional registry numbers or institutional identifiers associated with this trial.

WKR0-2013-0051

Identifier Type: -

Identifier Source: org_study_id

More Related Trials

Additional clinical trials that may be relevant based on similarity analysis.