Clinical Study of Wrist Blood Pressure Monitor in Conformance With the ANSI/AAMI SP10 Standard

NCT ID: NCT01662947

Last Updated: 2012-08-13

Study Results

Results pending

The study team has not published outcome measurements, participant flow, or safety data for this trial yet. Check back later for updates.

Basic Information

Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.

Recruitment Status

COMPLETED

Clinical Phase

NA

Total Enrollment

85 participants

Study Classification

INTERVENTIONAL

Study Start Date

2012-05-31

Study Completion Date

2012-06-30

Brief Summary

Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.

The clinical protocol of the clinical testing of this device:

1. Objective of the test: To verify the efficiency of device.
2. Test methods and procedures: Clinical test in two positions: Seated and Supine.
3. DUT: Transtek Wrist Blood Pressure Monitor, Model: TMB-1117. Cuff size: 13.5 - 21.5 cm.
4. Comparison device: Yuyue Medical Blood Pressure Meter, YYBP-212, accuracy: ±1 mmHg and range: 0-300 mmHg.
5. Study endpoints: Comply with ANSI/AAMI SP10-2002 standard.
6. Statistical methodology used: Description of statistical methods.
7. Result: Meet the requirements of SP10.

Detailed Description

Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.

Hospital Information

Hospital Information The data was collected by Zhongshan City People's Hospital Clinical Investigator Team at Zhongshan City People's Hospital, No.2 Sunwen Dong Road, Zhongshan, Guangdong 528403, P. R. China.

Investigator: Dr. Cao Yuedong, Leader; Li Qidong, Nurse A; Chen Wenqian, Nurse B.

Contact Dr. Cao Yuedong Tel: +86 760 88823818

Used Equipments DUT (Device Under Test): Transtek Wrist Blood Pressure Monitor, TMB-1117 Reference Device: Yuyue Medical Blood Pressure Meter, YYBP-212, accuracy: ±1 mmHg and range: 0-300 mmHg.

Test Protocol

1. Test Purpose:

The aim of clinical test is to determine, for the purpose of design qualification, the overall performance of the system meeting the following requirement: For systolic and diastolic pressures, treated separately, the mean difference of the paired measurement of the test system and the comparison system shall be ± 5 mmHg or less, with standard deviation of 8 mmHg or less.
2. Target Subject:

1\) The subject database shall contain at least 85 subjects. 2) The distribution of blood pressure of the subjects required to be: At least 10 percent below 100 mmHg systolic; at least 10 percent above 160 mmHg systolic; at least 10 percent below 60 mmHg diastolic; at least 10 percent above 100 mmHg diastolic; with the remainder distributed between these outer limits.

3\. Identification and precision of equipment used According to ANSI/AAMI SP10, the maximum measurement error of the non-automatic sphygmomanometers used in the comparing test shall not exceed 1.0mmHg at the temperature of the test. The manometer was used to verify its accuracy before test.

The clinical study was conducted according to the protocol described in the ANSI/AAMI SP10 with 85 subjects.

4\. Test procedures (auscultatory method, same-limb sequential measurement was chosen)

1. Record the age, gender, and circumference of the left wrist of the subject.
2. Let subject seated or supine calmly for five minutes before test.
3. Measure heart rate in one minute by palpation on radial artery, record it.
4. Two observers shall make simultaneous, blinded, blood pressure determination on the subject's left arm. Each nurse will write down the reading on a small piece of paper, make a tick if there is an auscultatory gap and then submit to the recorder.
5. The two readings separately from two observers will be recorded, and if two observers agree that they find an auscultatory gap this time, it will also be recorded. The wrist cuff will be take off, after 90 seconds rest, TMB-1117 cuff will be taken on and tested, and the readings (systolic, diastolic, mean BP and Heart rate) be recorded.
6. For each subject, repeat 3) and 5) to get 3 measurements.
7. Repeat step 1) to 6) for every subject.

5\. Note:

1. No motion and speaking are allowed during the measurement.
2. The reference device's cuff release at the rate about 3 mmHg/s so to ensure an accurate reading of observer.
3. The arm, wrist and the haear of subject to keep on the same height when measurement.
4. Test environment: Temperature: 20±1 ℃; Relative humidity: 40\~50%.

Conditions

See the medical conditions and disease areas that this research is targeting or investigating.

Efficacy of Wrist BPM Measurement Hypertension

Keywords

Explore important study keywords that can help with search, categorization, and topic discovery.

blood pressure

Study Design

Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.

Allocation Method

NON_RANDOMIZED

Intervention Model

FACTORIAL

Primary Study Purpose

DIAGNOSTIC

Blinding Strategy

NONE

Study Groups

Review each arm or cohort in the study, along with the interventions and objectives associated with them.

DUT Arm

DUT: Transtek Wrist Blood Pressure Monitor TMB-1117

Measurement: Blood Pressure

Groups/Cohorts: DUT

Group Type EXPERIMENTAL

Blood pressure Transtek Wrist Blood Pressure Monitor, TMB-1117

Intervention Type DEVICE

Systolic, Diastolic pressure.

Reference Arm

Reference Device: Yuyue Medical Blood Pressure Meter, YYBP-212, accuracy: ±1mmHg and range: 0-300mmHg.

Measurement: Blood Pressure

Groups/Cohorts: Reference

Group Type EXPERIMENTAL

Blood pressure Reference Device: Yuyue Medical Blood Pressure Meter, YYBP-2

Intervention Type DEVICE

Systolic, Diastolic pressure

Interventions

Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.

Blood pressure Transtek Wrist Blood Pressure Monitor, TMB-1117

Systolic, Diastolic pressure.

Intervention Type DEVICE

Blood pressure Reference Device: Yuyue Medical Blood Pressure Meter, YYBP-2

Systolic, Diastolic pressure

Intervention Type DEVICE

Other Intervention Names

Discover alternative or legacy names that may be used to describe the listed interventions across different sources.

DUT TMB-1117 measurement Reference mercury sphygmomanometer measurement

Eligibility Criteria

Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.

Inclusion Criteria

* male,female,

Exclusion Criteria

* below 18 years old
Minimum Eligible Age

18 Years

Eligible Sex

ALL

Accepts Healthy Volunteers

Yes

Sponsors

Meet the organizations funding or collaborating on the study and learn about their roles.

BTS International

OTHER

Sponsor Role collaborator

Leo Wang

OTHER

Sponsor Role lead

Responsible Party

Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.

Leo Wang

Team leader

Responsibility Role SPONSOR_INVESTIGATOR

Principal Investigators

Learn about the lead researchers overseeing the trial and their institutional affiliations.

Leo Wang, Leader

Role: STUDY_CHAIR

BTS International

Locations

Explore where the study is taking place and check the recruitment status at each participating site.

Zhongshan City People's Hospital

Zhongshan, Guangdong, China

Site Status

Countries

Review the countries where the study has at least one active or historical site.

China

References

Explore related publications, articles, or registry entries linked to this study.

Alpert BS. Validation of the Tiba Medical Ambulo 2400 ambulatory blood pressure monitor to the ISO Standard and BHS protocol. Blood Press Monit. 2010 Oct;15(5):275-7. doi: 10.1097/MBP.0b013e32833c8b39.

Reference Type BACKGROUND
PMID: 20559140 (View on PubMed)

Related Links

Access external resources that provide additional context or updates about the study.

Other Identifiers

Review additional registry numbers or institutional identifiers associated with this trial.

BTS-TRANS12001

Identifier Type: OTHER

Identifier Source: secondary_id

Wrist BPM TMB-1117

Identifier Type: -

Identifier Source: org_study_id