Trial Outcomes & Findings for Performance Evaluation of Blood Glucose Monitoring System Using ISO Study Design (NCT NCT01575080)

NCT ID: NCT01575080

Last Updated: 2016-02-29

Results Overview

Untrained subjects with diabetes self-test fingerstick blood using the Blood Glucose Monitoring System (BGMS). BGMS results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results are used to calculate the number of BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) or +/- 15% (\>=100mg/dL YSI capillary plasma).

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

102 participants

Primary outcome timeframe

1 hour

Results posted on

2016-02-29

Participant Flow

Participant milestones

Participant milestones
Measure
Intended Users of the Monitoring System
Untrained subjects with diabetes use Contour TS Blood Glucose Monitoring System.
Overall Study
STARTED
102
Overall Study
COMPLETED
102
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Performance Evaluation of Blood Glucose Monitoring System Using ISO Study Design

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Intended Users of the Monitoring System
n=102 Participants
Untrained subjects with diabetes use Contour TS Blood Glucose Monitoring System.
Age, Continuous
55 years
n=5 Participants
Sex: Female, Male
Female
53 Participants
n=5 Participants
Sex: Female, Male
Male
49 Participants
n=5 Participants
Region of Enrollment
United States
102 participants
n=5 Participants

PRIMARY outcome

Timeframe: 1 hour

Population: 102 untrained subjects tested one test strip lot on the Contour TS Blood Glucose Monitoring System.

Untrained subjects with diabetes self-test fingerstick blood using the Blood Glucose Monitoring System (BGMS). BGMS results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results are used to calculate the number of BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) or +/- 15% (\>=100mg/dL YSI capillary plasma).

Outcome measures

Outcome measures
Measure
Intended Users of the Monitoring System
n=102 BG test results
Untrained subjects with diabetes use Contour TS Blood Glucose Monitoring System.
Number of Self-Test Fingerstick Blood Glucose Results Within +/- 15mg/dL (<100mg/dL) or Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method
84 Number of BG Test Results

SECONDARY outcome

Timeframe: 1 hour

Population: 102 untrained subjects tested one test strip lot on the Contour TS Blood Glucose Monitoring System

Untrained subjects with diabetes self-test Alternative Site (AST) Palm blood using the Blood Glucose Monitoring System (BGMS). BGMS AST palm results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma BG results are used to calculate the number of AST palm BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) or +/- 15% (\>=100mg/dL YSI capillary plasma).

Outcome measures

Outcome measures
Measure
Intended Users of the Monitoring System
n=102 BG test results
Untrained subjects with diabetes use Contour TS Blood Glucose Monitoring System.
Number of Self-Test Alternative Site (Palm) Blood Glucose Results Within +/- 15mg/dL (<100mg/dL) or Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method
78 Number of BG Test Results

SECONDARY outcome

Timeframe: 1 hour

Population: 102 untrained subjects tested one test strip lot using the Contour TS BG Monitoring System. One subject was unable to obtain sufficient sample volume to test. Another subject's meter displayed

Untrained subjects with diabetes self-test Alternative Site (AST) Forearm blood using the Blood Glucose Monitoring System (BGMS). BGMS AST forearm results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma BG results are used to calculate the number of AST forearm BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) or +/- 15% (\>=100mg/dL YSI capillary plasma).

Outcome measures

Outcome measures
Measure
Intended Users of the Monitoring System
n=100 BG test results
Untrained subjects with diabetes use Contour TS Blood Glucose Monitoring System.
Number of Self-Test Alternative Site (Forearm) Blood Glucose Results Within +/- 15mg/dL (<100mg/dL) or Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method
83 Number of BG Test Results

SECONDARY outcome

Timeframe: 1 hour

Subjects will complete short questionnaires to provide feedback on the labeling materials and system ease of use. Subjects may respond 'Strongly Agree' 'Agree' 'Neutral' 'Disagree' 'Strongly Disagree'.

Outcome measures

Outcome measures
Measure
Intended Users of the Monitoring System
n=102 Participants
Untrained subjects with diabetes use Contour TS Blood Glucose Monitoring System.
Number of Subject Responses That 'Agree' or 'Strongly Agree' With Questionnaire Statements
I found it easy to do a blood test w this meter
98 participants
Number of Subject Responses That 'Agree' or 'Strongly Agree' With Questionnaire Statements
Meter display (test result) is easy to read
102 participants
Number of Subject Responses That 'Agree' or 'Strongly Agree' With Questionnaire Statements
Easy to see and understand the test results
101 participants
Number of Subject Responses That 'Agree' or 'Strongly Agree' With Questionnaire Statements
I like the overall meter design
80 participants
Number of Subject Responses That 'Agree' or 'Strongly Agree' With Questionnaire Statements
Meter is easy to handle
90 participants
Number of Subject Responses That 'Agree' or 'Strongly Agree' With Questionnaire Statements
Easy to handle the individual test strips
64 participants
Number of Subject Responses That 'Agree' or 'Strongly Agree' With Questionnaire Statements
Instructions (User Guide) easy to understand
97 participants
Number of Subject Responses That 'Agree' or 'Strongly Agree' With Questionnaire Statements
Instructions clearly explain how to run a test
101 participants
Number of Subject Responses That 'Agree' or 'Strongly Agree' With Questionnaire Statements
Instructions clearly explain meter error message
98 participants
Number of Subject Responses That 'Agree' or 'Strongly Agree' With Questionnaire Statements
I found this meter easy to use
100 participants
Number of Subject Responses That 'Agree' or 'Strongly Agree' With Questionnaire Statements
I found this meter intuitive
96 participants
Number of Subject Responses That 'Agree' or 'Strongly Agree' With Questionnaire Statements
I found this meter practical
97 participants

Adverse Events

Intended Users of the Monitoring System

Serious events: 0 serious events
Other events: 2 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Other adverse events
Measure
Intended Users of the Monitoring System
n=102 participants at risk
Untrained subjects with diabetes use Contour TS Blood Glucose Monitoring System.
Endocrine disorders
hypoglycemia
2.0%
2/102 • Number of events 2

Additional Information

Carmine Greene

Ascensia Diabetes Care

Phone: 574-257-3040

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place