Trial Outcomes & Findings for St.Jude Medical Cardiac Lead Assessment Study (NCT NCT01507987)

NCT ID: NCT01507987

Last Updated: 2021-08-23

Results Overview

The prevalence of electrical dysfunction was calculated as the number of leads that have electrical dysfunction through three years of follow-up for the 4 lead groups divided by total number of leads in enrolled patients in each of the 4 lead groups.

Recruitment status

COMPLETED

Target enrollment

2216 participants

Primary outcome timeframe

This outcome is assessed at enrollment and over the course of 3 years of follow-up

Results posted on

2021-08-23

Participant Flow

The study enrolled a total of 2,216 subjects at 45 sites to achieve an evaluable sample size of at least 1,401 subjects. The first subject enrolled in this study on March 20, 2013 and enrollment was completed in June, 2016 with the last follow-up completed in June 13 2019.

A subject that has at least one market released lead implanted was eligible for enrollment in the study. Subjects could have more than one study lead implanted. Each lead group was considered separately for independent analyses.

Unit of analysis: implants

Participant milestones

Participant milestones
Measure
Durata Leads
Durata leads implanted between 2008-2010
Riata Leads
Riata leads implanted between 2002-2009
Riata ST
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
QuickSite/QuickFlex implanted 2006-2010
Overall Study
STARTED
982 982
670 670
458 458
737 737
Overall Study
COMPLETED
618 618
335 335
252 252
480 480
Overall Study
NOT COMPLETED
364 364
335 335
206 206
257 257

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

St.Jude Medical Cardiac Lead Assessment Study

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
All Subjects Enrolled in the Study
n=2216 Participants
Demographics for all enrolled patients (N=2216)
Age, Continuous
67 years
STANDARD_DEVIATION 12 • n=5 Participants
Sex: Female, Male
Female
600 Participants
n=5 Participants
Sex: Female, Male
Male
1616 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
56 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
2158 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
2 Participants
n=5 Participants
Race (NIH/OMB)
American Indian or Alaska Native
8 Participants
n=5 Participants
Race (NIH/OMB)
Asian
91 Participants
n=5 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
3 Participants
n=5 Participants
Race (NIH/OMB)
Black or African American
335 Participants
n=5 Participants
Race (NIH/OMB)
White
1770 Participants
n=5 Participants
Race (NIH/OMB)
More than one race
0 Participants
n=5 Participants
Race (NIH/OMB)
Unknown or Not Reported
9 Participants
n=5 Participants
Region of Enrollment
Canada
48 participants
n=5 Participants
Region of Enrollment
United States
2085 participants
n=5 Participants
Region of Enrollment
Japan
83 participants
n=5 Participants

PRIMARY outcome

Timeframe: This outcome is assessed at enrollment and over the course of 3 years of follow-up

Population: All SJM leads enrolled in the study with Durata, Riata, Riata ST, or QuickSite/QuickFlex leads

The prevalence of electrical dysfunction was calculated as the number of leads that have electrical dysfunction through three years of follow-up for the 4 lead groups divided by total number of leads in enrolled patients in each of the 4 lead groups.

Outcome measures

Outcome measures
Measure
Durata Leads
n=982 implanted leads
Durata leads implanted between 2008-2010
Riata Leads
n=670 implanted leads
Riata leads implanted between 2002-2009
Riata ST
n=458 implanted leads
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
n=737 implanted leads
QuickSite/QuickFlex implanted 2006-2010
Riata Leads Lead Age: 6-less Than 7 Years
lead age: 6-less than 7 years
Riata Leads Lead Age: 7 Years or Greater
lead age: 7 years or greater
Riata ST Lead Age: Less Than 5.5 Years
lead age: less than 5.5 years
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years
lead age: greater than or equal to 5.5 years
QuickSite/QuickFlex Lead Age: Less Than 5 Years
lead age: less than 5 years
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years
lead age: 5 to less than 6 years
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years
lead age: greater than or equal to 6 years
Number of Leads With Electrical Dysfunction
24 implanted leads
27 implanted leads
15 implanted leads
2 implanted leads

PRIMARY outcome

Timeframe: This outcome is assessed at enrollment

Population: All SJM leads enrolled in the study with Durata, Riata, Riata ST, or QuickSite/QuickFlex leads

The prevalence of externalized conductors was calculated as the number of leads that have been shown to have externalized conductors by imaging at the time of enrollment in each of the 4 lead groups divided by total number of leads in enrolled patients in each of the 4 lead groups

Outcome measures

Outcome measures
Measure
Durata Leads
n=973 implanted leads
Durata leads implanted between 2008-2010
Riata Leads
n=660 implanted leads
Riata leads implanted between 2002-2009
Riata ST
n=452 implanted leads
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
n=723 implanted leads
QuickSite/QuickFlex implanted 2006-2010
Riata Leads Lead Age: 6-less Than 7 Years
lead age: 6-less than 7 years
Riata Leads Lead Age: 7 Years or Greater
lead age: 7 years or greater
Riata ST Lead Age: Less Than 5.5 Years
lead age: less than 5.5 years
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years
lead age: greater than or equal to 5.5 years
QuickSite/QuickFlex Lead Age: Less Than 5 Years
lead age: less than 5 years
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years
lead age: 5 to less than 6 years
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years
lead age: greater than or equal to 6 years
Number of Leads With Externalized Conductors
0 implanted leads
144 implanted leads
44 implanted leads
10 implanted leads

PRIMARY outcome

Timeframe: This outcome is assessed at enrollment

The prevalence of other visual lead anomalies (by each subcategory) was calculated as the number of leads that have been shown to have other visual anomaly (by each subcategory) at the time of enrollment in each of the 4 lead groups divided by total number of leads in enrolled patients in each of the 4 lead groups.

Outcome measures

Outcome measures
Measure
Durata Leads
n=974 implanted leads
Durata leads implanted between 2008-2010
Riata Leads
n=646 implanted leads
Riata leads implanted between 2002-2009
Riata ST
n=454 implanted leads
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
n=724 implanted leads
QuickSite/QuickFlex implanted 2006-2010
Riata Leads Lead Age: 6-less Than 7 Years
lead age: 6-less than 7 years
Riata Leads Lead Age: 7 Years or Greater
lead age: 7 years or greater
Riata ST Lead Age: Less Than 5.5 Years
lead age: less than 5.5 years
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years
lead age: greater than or equal to 5.5 years
QuickSite/QuickFlex Lead Age: Less Than 5 Years
lead age: less than 5 years
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years
lead age: 5 to less than 6 years
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years
lead age: greater than or equal to 6 years
Number of Other Visual Lead Anomalies by Each Subcategory (Fracture, Kink, Subclavian Crush, Other Irregularities)
7 implanted leads
19 implanted leads
7 implanted leads
4 implanted leads

PRIMARY outcome

Timeframe: This outcome is assessed from enrollment through the course of 3 years of follow-up

Population: The annual hazard rate was assessed separately in each of the 4 lead groups.

Annual hazard rate is the proportion of number of leads with electrical dysfunction to the total number of lead follow-up years.

Outcome measures

Outcome measures
Measure
Durata Leads
n=982 implanted leads
Durata leads implanted between 2008-2010
Riata Leads
n=670 implanted leads
Riata leads implanted between 2002-2009
Riata ST
n=458 implanted leads
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
n=737 implanted leads
QuickSite/QuickFlex implanted 2006-2010
Riata Leads Lead Age: 6-less Than 7 Years
lead age: 6-less than 7 years
Riata Leads Lead Age: 7 Years or Greater
lead age: 7 years or greater
Riata ST Lead Age: Less Than 5.5 Years
lead age: less than 5.5 years
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years
lead age: greater than or equal to 5.5 years
QuickSite/QuickFlex Lead Age: Less Than 5 Years
lead age: less than 5 years
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years
lead age: 5 to less than 6 years
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years
lead age: greater than or equal to 6 years
The Annual Hazard Rate of Lead Electrical Dysfunction in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.
0.010 leads with electrical dysfunction/years
0.019 leads with electrical dysfunction/years
0.014 leads with electrical dysfunction/years
0.001 leads with electrical dysfunction/years

PRIMARY outcome

Timeframe: year 1 follow-up

Population: The annual hazard rate was assessed separately in each of the 4 lead groups.

Annual hazard rate calculated as ratio of all leads that are adjudicated by the committee and a have definite outcome of presence of externalized conductors to total follow-up years from enrollment to year 1

Outcome measures

Outcome measures
Measure
Durata Leads
n=808 implanted leads
Durata leads implanted between 2008-2010
Riata Leads
n=375 implanted leads
Riata leads implanted between 2002-2009
Riata ST
n=316 implanted leads
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
n=556 implanted leads
QuickSite/QuickFlex implanted 2006-2010
Riata Leads Lead Age: 6-less Than 7 Years
lead age: 6-less than 7 years
Riata Leads Lead Age: 7 Years or Greater
lead age: 7 years or greater
Riata ST Lead Age: Less Than 5.5 Years
lead age: less than 5.5 years
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years
lead age: greater than or equal to 5.5 years
QuickSite/QuickFlex Lead Age: Less Than 5 Years
lead age: less than 5 years
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years
lead age: 5 to less than 6 years
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years
lead age: greater than or equal to 6 years
The Annual Hazard Rate of New Cases of Externalized Conductors (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.
0.0025 leads with externalized conductors/years
Interval 0.0003 to 0.0089
0.0480 leads with externalized conductors/years
Interval 0.0287 to 0.0748
0.0158 leads with externalized conductors/years
Interval 0.0052 to 0.0365
0.0288 leads with externalized conductors/years
Interval 0.0165 to 0.0463

PRIMARY outcome

Timeframe: year 1 to year 2 follow-up

Population: The annual hazard rate was assessed separately in each of the 4 lead groups.

Annual hazard rate calculated as ratio of all leads that are adjudicated by the committee and a have definite outcome of presence of externalized conductors to total follow-up years from year 1 to year 2

Outcome measures

Outcome measures
Measure
Durata Leads
n=687 implanted leads
Durata leads implanted between 2008-2010
Riata Leads
n=301 implanted leads
Riata leads implanted between 2002-2009
Riata ST
n=256 implanted leads
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
n=462 implanted leads
QuickSite/QuickFlex implanted 2006-2010
Riata Leads Lead Age: 6-less Than 7 Years
lead age: 6-less than 7 years
Riata Leads Lead Age: 7 Years or Greater
lead age: 7 years or greater
Riata ST Lead Age: Less Than 5.5 Years
lead age: less than 5.5 years
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years
lead age: greater than or equal to 5.5 years
QuickSite/QuickFlex Lead Age: Less Than 5 Years
lead age: less than 5 years
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years
lead age: 5 to less than 6 years
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years
lead age: greater than or equal to 6 years
The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.
0.0015 leads with externalized conductors/years
Interval 0.0 to 0.0081
0.0897 leads with externalized conductors/years
Interval 0.0599 to 0.1278
0.0234 leads with externalized conductors/years
Interval 0.0086 to 0.0503
0.0087 leads with externalized conductors/years
Interval 0.0024 to 0.022

PRIMARY outcome

Timeframe: year 2 to year 3 follow-up

Population: The annual hazard rate was assessed separately in each of the 4 lead groups.

Annual hazard rate calculated as ratio of all leads that are adjudicated by the committee and a have definite outcome of presence of externalized conductors to total follow-up years from year 2 to year 3

Outcome measures

Outcome measures
Measure
Durata Leads
n=570 implanted leads
Durata leads implanted between 2008-2010
Riata Leads
n=228 implanted leads
Riata leads implanted between 2002-2009
Riata ST
n=206 implanted leads
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
n=362 implanted leads
QuickSite/QuickFlex implanted 2006-2010
Riata Leads Lead Age: 6-less Than 7 Years
lead age: 6-less than 7 years
Riata Leads Lead Age: 7 Years or Greater
lead age: 7 years or greater
Riata ST Lead Age: Less Than 5.5 Years
lead age: less than 5.5 years
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years
lead age: greater than or equal to 5.5 years
QuickSite/QuickFlex Lead Age: Less Than 5 Years
lead age: less than 5 years
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years
lead age: 5 to less than 6 years
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years
lead age: greater than or equal to 6 years
The Annual Hazard Rate of New Cases of Externalized Conductors (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads.
0.0035 leads with externalized conductors/years
Interval 0.0004 to 0.0126
0.0658 leads with externalized conductors/years
Interval 0.0373 to 0.1062
0.0097 leads with externalized conductors/years
Interval 0.0012 to 0.0346
0.0110 leads with externalized conductors/years
Interval 0.003 to 0.028

PRIMARY outcome

Timeframe: This outcome is assessed from enrollment to year 1

Population: The annual hazard rate was assessed separately in each of the 4 lead groups.

Annual hazard rate calculated as proportion of leads adjudicated by the committee and a have definite outcome of presence of other visual lead anomaly to total follow up years from enrollment to year 1

Outcome measures

Outcome measures
Measure
Durata Leads
n=803 implanted leads
Durata leads implanted between 2008-2010
Riata Leads
n=461 implanted leads
Riata leads implanted between 2002-2009
Riata ST
n=343 implanted leads
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
n=571 implanted leads
QuickSite/QuickFlex implanted 2006-2010
Riata Leads Lead Age: 6-less Than 7 Years
lead age: 6-less than 7 years
Riata Leads Lead Age: 7 Years or Greater
lead age: 7 years or greater
Riata ST Lead Age: Less Than 5.5 Years
lead age: less than 5.5 years
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years
lead age: greater than or equal to 5.5 years
QuickSite/QuickFlex Lead Age: Less Than 5 Years
lead age: less than 5 years
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years
lead age: 5 to less than 6 years
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years
lead age: greater than or equal to 6 years
The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Enrollment to Year 1) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads
0.0025 leads w other visual lead anomaly/years
Interval 0.0003 to 0.009
0.0108 leads w other visual lead anomaly/years
Interval 0.0035 to 0.0251
0.0117 leads w other visual lead anomaly/years
Interval 0.0032 to 0.0296
0.0053 leads w other visual lead anomaly/years
Interval 0.0011 to 0.0153

PRIMARY outcome

Timeframe: This outcome is assessed from year 1 to year 2 of follow-up

Population: The annual hazard rate was assessed separately in each of the 4 lead groups.

Annual hazard rate calculated as proportion of leads adjudicated by the committee and a have definite outcome of presence of other visual lead anomaly to total follow up years from year 1 to year 2

Outcome measures

Outcome measures
Measure
Durata Leads
n=254 implanted leads
Durata leads implanted between 2008-2010
Riata Leads
n=374 implanted leads
Riata leads implanted between 2002-2009
Riata ST
n=286 implanted leads
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
n=482 implanted leads
QuickSite/QuickFlex implanted 2006-2010
Riata Leads Lead Age: 6-less Than 7 Years
lead age: 6-less than 7 years
Riata Leads Lead Age: 7 Years or Greater
lead age: 7 years or greater
Riata ST Lead Age: Less Than 5.5 Years
lead age: less than 5.5 years
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years
lead age: greater than or equal to 5.5 years
QuickSite/QuickFlex Lead Age: Less Than 5 Years
lead age: less than 5 years
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years
lead age: 5 to less than 6 years
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years
lead age: greater than or equal to 6 years
The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 1 to Year 2) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads
0 leads w other visual lead anomaly/years
Interval 0.0 to 0.0054
0.0027 leads w other visual lead anomaly/years
Interval 0.0001 to 0.0148
0.0070 leads w other visual lead anomaly/years
Interval 0.0008 to 0.025
0.0021 leads w other visual lead anomaly/years
Interval 0.0001 to 0.0115

PRIMARY outcome

Timeframe: This outcome is assessed from year 2 to year 3 follow-up

Population: The annual hazard rate was assessed separately in each of the 4 lead groups.

Annual hazard rate calculated as proportion of leads adjudicated by the committee and a have definite outcome of presence of other visual lead anomaly to total follow up years from year 2 to year 3

Outcome measures

Outcome measures
Measure
Durata Leads
n=565 implanted leads
Durata leads implanted between 2008-2010
Riata Leads
n=305 implanted leads
Riata leads implanted between 2002-2009
Riata ST
n=229 implanted leads
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
n=387 implanted leads
QuickSite/QuickFlex implanted 2006-2010
Riata Leads Lead Age: 6-less Than 7 Years
lead age: 6-less than 7 years
Riata Leads Lead Age: 7 Years or Greater
lead age: 7 years or greater
Riata ST Lead Age: Less Than 5.5 Years
lead age: less than 5.5 years
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years
lead age: greater than or equal to 5.5 years
QuickSite/QuickFlex Lead Age: Less Than 5 Years
lead age: less than 5 years
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years
lead age: 5 to less than 6 years
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years
lead age: greater than or equal to 6 years
The Annual Hazard Rate of New Cases of Other Visual Lead Anomalies by Each Subcategory (From Year 2 to Year 3) in the Riata, Riata ST, QuickSite/QuickFlex (1056T, 1058T, 1156T, 1158T) and Durata Leads
0 leads w other visual lead anomaly/years
Interval 0.0 to 0.0065
0.0066 leads w other visual lead anomaly/years
Interval 0.0008 to 0.0235
0 leads w other visual lead anomaly/years
Interval 0.0 to 0.016
0 leads w other visual lead anomaly/years
Interval 0.0 to 0.0095

SECONDARY outcome

Timeframe: This outcome is assessed from time of detecting EC till the course of 3 years of follow-up

Population: N at risk from detecting externalized conductors at each follow-up time point

Kaplan-Meier analysis of time from externalized conductors (EC) to electrical dysfunction, by year of initial implant. Count of cumulative lead anomaly events from time of detecting EC

Outcome measures

Outcome measures
Measure
Durata Leads
n=1 Implant Leads
Durata leads implanted between 2008-2010
Riata Leads
n=2 Implant Leads
Riata leads implanted between 2002-2009
Riata ST
n=2 Implant Leads
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
n=42 Implant Leads
QuickSite/QuickFlex implanted 2006-2010
Riata Leads Lead Age: 6-less Than 7 Years
n=67 Implant Leads
lead age: 6-less than 7 years
Riata Leads Lead Age: 7 Years or Greater
n=93 Implant Leads
lead age: 7 years or greater
Riata ST Lead Age: Less Than 5.5 Years
n=23 Implant Leads
lead age: less than 5.5 years
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years
n=34 Implant Leads
lead age: greater than or equal to 5.5 years
QuickSite/QuickFlex Lead Age: Less Than 5 Years
n=12 Implant Leads
lead age: less than 5 years
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years
n=12 Implant Leads
lead age: 5 to less than 6 years
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years
n=15 Implant Leads
lead age: greater than or equal to 6 years
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant
year 3 from detecting externalized conductor
NA lead anomaly events
no data available
NA lead anomaly events
no data available
NA lead anomaly events
no data available
2 lead anomaly events
3 lead anomaly events
6 lead anomaly events
1 lead anomaly events
4 lead anomaly events
NA lead anomaly events
no data available
0 lead anomaly events
0 lead anomaly events
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant
year 0 from detecting externalized conductor
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
1 lead anomaly events
1 lead anomaly events
3 lead anomaly events
0 lead anomaly events
1 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant
year 1 from detecting externalized conductor
0 lead anomaly events
NA lead anomaly events
no data available
NA lead anomaly events
no data available
2 lead anomaly events
1 lead anomaly events
5 lead anomaly events
0 lead anomaly events
2 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Electrical Dysfunction, by Year of Initial Implant
year 2 from detecting externalized conductor
NA lead anomaly events
no data available
NA lead anomaly events
no data available
NA lead anomaly events
no data available
2 lead anomaly events
2 lead anomaly events
6 lead anomaly events
1 lead anomaly events
2 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events

SECONDARY outcome

Timeframe: This outcome is assessed from time of detecting other visual lead anomaly till the course of 3 years of follow-up

Population: N at risk from detecting other visual lead anomaly

Kaplan-Meier analysis of time from other visual lead anomaly to electrical dysfunction, by year of initial implant. Count of cumulative lead anomaly events from time of detecting other visual lead anomaly

Outcome measures

Outcome measures
Measure
Durata Leads
n=5 Implant Leads
Durata leads implanted between 2008-2010
Riata Leads
n=2 Implant Leads
Riata leads implanted between 2002-2009
Riata ST
n=2 Implant Leads
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
n=7 Implant Leads
QuickSite/QuickFlex implanted 2006-2010
Riata Leads Lead Age: 6-less Than 7 Years
n=10 Implant Leads
lead age: 6-less than 7 years
Riata Leads Lead Age: 7 Years or Greater
n=10 Implant Leads
lead age: 7 years or greater
Riata ST Lead Age: Less Than 5.5 Years
n=6 Implant Leads
lead age: less than 5.5 years
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years
n=7 Implant Leads
lead age: greater than or equal to 5.5 years
QuickSite/QuickFlex Lead Age: Less Than 5 Years
n=2 Implant Leads
lead age: less than 5 years
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years
n=3 Implant Leads
lead age: 5 to less than 6 years
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years
n=3 Implant Leads
lead age: greater than or equal to 6 years
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant
1 years follow up from detecting other visual lead anomaly
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
1 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant
2 years follow up from detecting other visual lead anomaly
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
1 lead anomaly events
0 lead anomaly events
1 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant
0 years follow up from detecting other visual lead anomaly
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
1 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Other Visual Lead Anomaly to Electrical Dysfunction, by Year of Initial Implant
3 years follow up from detecting other visual lead anomaly
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
1 lead anomaly events
0 lead anomaly events
1 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events

SECONDARY outcome

Timeframe: within 30 days associated with study lead related interventions

Population: Subject has at least one market released Riata, Riata ST, QuickSite/QuickFlex, or Durata lead implanted

Number of participants with adverse events within 30 days associated with study lead related interventions for each lead group

Outcome measures

Outcome measures
Measure
Durata Leads
n=982 Participants
Durata leads implanted between 2008-2010
Riata Leads
n=670 Participants
Riata leads implanted between 2002-2009
Riata ST
n=458 Participants
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
n=737 Participants
QuickSite/QuickFlex implanted 2006-2010
Riata Leads Lead Age: 6-less Than 7 Years
lead age: 6-less than 7 years
Riata Leads Lead Age: 7 Years or Greater
lead age: 7 years or greater
Riata ST Lead Age: Less Than 5.5 Years
lead age: less than 5.5 years
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years
lead age: greater than or equal to 5.5 years
QuickSite/QuickFlex Lead Age: Less Than 5 Years
lead age: less than 5 years
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years
lead age: 5 to less than 6 years
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years
lead age: greater than or equal to 6 years
Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).
Lead dislodgement or migration
1 Participants
0 Participants
0 Participants
0 Participants
Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).
lead insulation damage
0 Participants
0 Participants
0 Participants
1 Participants
Number of Participants With an Adverse Event Through 30 Days Post-intervention for Lead (e.g. Extraction, Abandonment, Revision, Other).
phrenic nerve/diaphragmatic stimulation
0 Participants
0 Participants
0 Participants
1 Participants

SECONDARY outcome

Timeframe: This outcome is assessed from time of detection of externalized conductor to 3 years follow-up

Population: N at risk from time of externalized conductors at each follow-up period

time from externalized conductors to clinical intervention for all lead subgroups.

Outcome measures

Outcome measures
Measure
Durata Leads
n=1 Implant Leads
Durata leads implanted between 2008-2010
Riata Leads
n=2 Implant Leads
Riata leads implanted between 2002-2009
Riata ST
n=2 Implant Leads
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
n=42 Implant Leads
QuickSite/QuickFlex implanted 2006-2010
Riata Leads Lead Age: 6-less Than 7 Years
n=67 Implant Leads
lead age: 6-less than 7 years
Riata Leads Lead Age: 7 Years or Greater
n=93 Implant Leads
lead age: 7 years or greater
Riata ST Lead Age: Less Than 5.5 Years
n=23 Implant Leads
lead age: less than 5.5 years
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years
n=34 Implant Leads
lead age: greater than or equal to 5.5 years
QuickSite/QuickFlex Lead Age: Less Than 5 Years
n=12 Implant Leads
lead age: less than 5 years
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years
n=12 Implant Leads
lead age: 5 to less than 6 years
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years
n=15 Implant Leads
lead age: greater than or equal to 6 years
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant
3 years from detecting EC
NA lead anomaly events
no data available
1 lead anomaly events
1 lead anomaly events
4 lead anomaly events
8 lead anomaly events
18 lead anomaly events
6 lead anomaly events
6 lead anomaly events
1 lead anomaly events
0 lead anomaly events
0 lead anomaly events
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant
0 years from detecting EC
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
1 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant
1 year from detecting EC
0 lead anomaly events
1 lead anomaly events
1 lead anomaly events
2 lead anomaly events
5 lead anomaly events
11 lead anomaly events
3 lead anomaly events
4 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
Estimation of Cumulative Lead Anomaly Events From Time of Detecting Externalized Conductors to Clinical Intervention, by Year of Initial Implant
2 years from detecting EC
NA lead anomaly events
no data available
1 lead anomaly events
1 lead anomaly events
3 lead anomaly events
7 lead anomaly events
15 lead anomaly events
6 lead anomaly events
4 lead anomaly events
1 lead anomaly events
0 lead anomaly events
0 lead anomaly events

SECONDARY outcome

Timeframe: This outcome is assessed from time of detection of other visual lead anomaly until 3 years of follow-up

Population: N at risk from time of detecting visual lead anomaly at each follow-up period

Kaplan-Meier analysis of time from visual lead anomalies for all lead and lead age subcategories to clinical intervention

Outcome measures

Outcome measures
Measure
Durata Leads
n=5 Implant Leads
Durata leads implanted between 2008-2010
Riata Leads
n=2 Implant Leads
Riata leads implanted between 2002-2009
Riata ST
n=2 Implant Leads
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
n=7 Implant Leads
QuickSite/QuickFlex implanted 2006-2010
Riata Leads Lead Age: 6-less Than 7 Years
n=10 Implant Leads
lead age: 6-less than 7 years
Riata Leads Lead Age: 7 Years or Greater
n=10 Implant Leads
lead age: 7 years or greater
Riata ST Lead Age: Less Than 5.5 Years
n=6 Implant Leads
lead age: less than 5.5 years
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years
n=7 Implant Leads
lead age: greater than or equal to 5.5 years
QuickSite/QuickFlex Lead Age: Less Than 5 Years
n=2 Implant Leads
lead age: less than 5 years
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years
n=3 Implant Leads
lead age: 5 to less than 6 years
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years
n=3 Implant Leads
lead age: greater than or equal to 6 years
Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant
3 years from detection other visual lead anomaly
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
1 lead anomaly events
0 lead anomaly events
2 lead anomaly events
0 lead anomaly events
2 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant
0 years from detection other visual lead anomaly
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant
1 years from detection other visual lead anomaly
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
1 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
1 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
Estimation of Lead Anomaly Events From Time of Other Visual Lead Anomalies by Each Subcategory to Clinical Intervention, by Year of Initial Implant
2 years from detection other visual lead anomaly
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events
1 lead anomaly events
0 lead anomaly events
2 lead anomaly events
0 lead anomaly events
2 lead anomaly events
0 lead anomaly events
0 lead anomaly events
0 lead anomaly events

SECONDARY outcome

Timeframe: This outcome is assessed from enrollment till the course of 3 years of follow-up

Population: Patients with Riata or Riata ST leads.

Comparison of patients with EC to those without EC: lead size effect on number of patients with or without EC (7F and 8F are lead sizes, F = French). Riata leads have 8F diameter and Riata ST leads have 7F diameter. The Riata/Riata ST families of leads were used for this comparison only to determine whether the difference in lead size (with the silicon material and other lead design factors controlled for) had an effect on externalization of conductors.

Outcome measures

Outcome measures
Measure
Durata Leads
n=261 Participants
Durata leads implanted between 2008-2010
Riata Leads
n=851 Participants
Riata leads implanted between 2002-2009
Riata ST
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
QuickSite/QuickFlex implanted 2006-2010
Riata Leads Lead Age: 6-less Than 7 Years
lead age: 6-less than 7 years
Riata Leads Lead Age: 7 Years or Greater
lead age: 7 years or greater
Riata ST Lead Age: Less Than 5.5 Years
lead age: less than 5.5 years
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years
lead age: greater than or equal to 5.5 years
QuickSite/QuickFlex Lead Age: Less Than 5 Years
lead age: less than 5 years
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years
lead age: 5 to less than 6 years
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years
lead age: greater than or equal to 6 years
Comparison of Number of Patients With Externalized Conductors to Number of Patients Without Externalized Conductors (EC)
7F lead size (Riata ST)
57 Participants
395 Participants
Comparison of Number of Patients With Externalized Conductors to Number of Patients Without Externalized Conductors (EC)
8F lead size (Riata)
204 Participants
456 Participants

SECONDARY outcome

Timeframe: This outcome is assessed from enrollment till the course of 3 years of follow-up

Population: Patients implanted with SJM Riata, Riata ST, Durata, or QuickSite/QuickFlex leads. Leads were combined for this analysis to compare patient groups with and without electrical dysfunction based on age (irrespective of lead implant type)

Comparison of patients with ED to those without ED: subject age effect on ED

Outcome measures

Outcome measures
Measure
Durata Leads
n=68 Participants
Durata leads implanted between 2008-2010
Riata Leads
n=2148 Participants
Riata leads implanted between 2002-2009
Riata ST
Riata ST leads implanted 2006-2009
QuickSite/QuickFlex
QuickSite/QuickFlex implanted 2006-2010
Riata Leads Lead Age: 6-less Than 7 Years
lead age: 6-less than 7 years
Riata Leads Lead Age: 7 Years or Greater
lead age: 7 years or greater
Riata ST Lead Age: Less Than 5.5 Years
lead age: less than 5.5 years
Riata ST Leads Lead Age: Greater Than or Equal to 5.5 Years
lead age: greater than or equal to 5.5 years
QuickSite/QuickFlex Lead Age: Less Than 5 Years
lead age: less than 5 years
QuickSite/QuickFlex Leads Lead Age: 5 to Less Than 6 Years
lead age: 5 to less than 6 years
QuickSite/QuickFlex Leads Lead Age: Greater Than or Equal to 6 Years
lead age: greater than or equal to 6 years
Comparison of Patients With Electrical Dysfunction to Those Without Electrical Dysfunction (ED).
61.8 years
Standard Deviation 13.4
67.2 years
Standard Deviation 11.9

Adverse Events

Durata Leads

Serious events: 26 serious events
Other events: 20 other events
Deaths: 130 deaths

Riata Leads

Serious events: 28 serious events
Other events: 14 other events
Deaths: 111 deaths

Riata ST Leads

Serious events: 22 serious events
Other events: 9 other events
Deaths: 64 deaths

QuickSite/QuickFlex Leads

Serious events: 7 serious events
Other events: 11 other events
Deaths: 135 deaths

Serious adverse events

Serious adverse events
Measure
Durata Leads
n=982 participants at risk
Subjects implanted with SJM Durata leads
Riata Leads
n=670 participants at risk
Subjects implanted with Riata leads
Riata ST Leads
n=458 participants at risk
Subjects implanted with Riata ST leads
QuickSite/QuickFlex Leads
n=737 participants at risk
Subjects implanted with QuickSite/QuickFlex leads
Product Issues
Abnormal high voltage lead impredance
0.31%
3/982 • Number of events 3 • From enrollment to 3 years of follow-up period
0.75%
5/670 • Number of events 5 • From enrollment to 3 years of follow-up period
1.1%
5/458 • Number of events 5 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period
Product Issues
Abnormal Pacing Impedance
0.81%
8/982 • Number of events 8 • From enrollment to 3 years of follow-up period
0.30%
2/670 • Number of events 2 • From enrollment to 3 years of follow-up period
0.00%
0/458 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period
Product Issues
Decline in R Wave Sensing
0.10%
1/982 • Number of events 1 • From enrollment to 3 years of follow-up period
0.00%
0/670 • From enrollment to 3 years of follow-up period
0.00%
0/458 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period
Product Issues
Elevated pacing thresholds
0.00%
0/982 • From enrollment to 3 years of follow-up period
0.60%
4/670 • Number of events 4 • From enrollment to 3 years of follow-up period
0.22%
1/458 • Number of events 1 • From enrollment to 3 years of follow-up period
0.27%
2/737 • Number of events 2 • From enrollment to 3 years of follow-up period
Product Issues
Externalized Conductor
0.00%
0/982 • From enrollment to 3 years of follow-up period
0.15%
1/670 • Number of events 1 • From enrollment to 3 years of follow-up period
0.00%
0/458 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period
Product Issues
Lead dislodgement or migration
0.10%
1/982 • Number of events 1 • From enrollment to 3 years of follow-up period
0.00%
0/670 • From enrollment to 3 years of follow-up period
0.00%
0/458 • From enrollment to 3 years of follow-up period
0.27%
2/737 • Number of events 2 • From enrollment to 3 years of follow-up period
Product Issues
lead fracture
0.41%
4/982 • Number of events 4 • From enrollment to 3 years of follow-up period
0.30%
2/670 • Number of events 2 • From enrollment to 3 years of follow-up period
0.00%
0/458 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period
Product Issues
lead insulation damage
0.20%
2/982 • Number of events 2 • From enrollment to 3 years of follow-up period
0.15%
1/670 • Number of events 1 • From enrollment to 3 years of follow-up period
0.87%
4/458 • Number of events 4 • From enrollment to 3 years of follow-up period
0.14%
1/737 • Number of events 1 • From enrollment to 3 years of follow-up period
Product Issues
lead noise
0.41%
4/982 • Number of events 4 • From enrollment to 3 years of follow-up period
1.2%
8/670 • Number of events 8 • From enrollment to 3 years of follow-up period
1.7%
8/458 • Number of events 8 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period
Product Issues
oversensing
0.31%
3/982 • Number of events 3 • From enrollment to 3 years of follow-up period
0.45%
3/670 • Number of events 3 • From enrollment to 3 years of follow-up period
0.87%
4/458 • Number of events 4 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period
Product Issues
phrenic nerve/diaphragmatic stimulation
0.00%
0/982 • From enrollment to 3 years of follow-up period
0.00%
0/670 • From enrollment to 3 years of follow-up period
0.00%
0/458 • From enrollment to 3 years of follow-up period
0.27%
2/737 • Number of events 2 • From enrollment to 3 years of follow-up period
Product Issues
shock due to AF with RVR
0.00%
0/982 • From enrollment to 3 years of follow-up period
0.15%
1/670 • Number of events 1 • From enrollment to 3 years of follow-up period
0.00%
0/458 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period
Product Issues
undersensing
0.00%
0/982 • From enrollment to 3 years of follow-up period
0.15%
1/670 • Number of events 1 • From enrollment to 3 years of follow-up period
0.00%
0/458 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period

Other adverse events

Other adverse events
Measure
Durata Leads
n=982 participants at risk
Subjects implanted with SJM Durata leads
Riata Leads
n=670 participants at risk
Subjects implanted with Riata leads
Riata ST Leads
n=458 participants at risk
Subjects implanted with Riata ST leads
QuickSite/QuickFlex Leads
n=737 participants at risk
Subjects implanted with QuickSite/QuickFlex leads
Product Issues
Abnormal High Voltage Lead Impedance
0.20%
2/982 • Number of events 3 • From enrollment to 3 years of follow-up period
0.30%
2/670 • Number of events 2 • From enrollment to 3 years of follow-up period
0.00%
0/458 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period
Product Issues
Abnormal Pacing Impedance
0.51%
5/982 • Number of events 6 • From enrollment to 3 years of follow-up period
0.15%
1/670 • Number of events 1 • From enrollment to 3 years of follow-up period
0.00%
0/458 • From enrollment to 3 years of follow-up period
0.41%
3/737 • Number of events 3 • From enrollment to 3 years of follow-up period
Product Issues
Decrease R Wave
0.10%
1/982 • Number of events 1 • From enrollment to 3 years of follow-up period
0.00%
0/670 • From enrollment to 3 years of follow-up period
0.00%
0/458 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period
Product Issues
Elevated pacing thresholds
0.31%
3/982 • Number of events 3 • From enrollment to 3 years of follow-up period
0.30%
2/670 • Number of events 2 • From enrollment to 3 years of follow-up period
0.22%
1/458 • Number of events 1 • From enrollment to 3 years of follow-up period
0.27%
2/737 • Number of events 2 • From enrollment to 3 years of follow-up period
Product Issues
Lead Insulation Damage
0.00%
0/982 • From enrollment to 3 years of follow-up period
0.15%
1/670 • Number of events 1 • From enrollment to 3 years of follow-up period
0.00%
0/458 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period
Product Issues
Lead Noise
0.41%
4/982 • Number of events 4 • From enrollment to 3 years of follow-up period
0.15%
1/670 • Number of events 1 • From enrollment to 3 years of follow-up period
0.44%
2/458 • Number of events 2 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period
Product Issues
Loss of Capture
0.00%
0/982 • From enrollment to 3 years of follow-up period
0.00%
0/670 • From enrollment to 3 years of follow-up period
0.00%
0/458 • From enrollment to 3 years of follow-up period
0.14%
1/737 • Number of events 1 • From enrollment to 3 years of follow-up period
Product Issues
Loss of Sensing
0.00%
0/982 • From enrollment to 3 years of follow-up period
0.15%
1/670 • Number of events 1 • From enrollment to 3 years of follow-up period
0.00%
0/458 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period
Product Issues
Other Visual Lead Anomalies
0.00%
0/982 • From enrollment to 3 years of follow-up period
0.00%
0/670 • From enrollment to 3 years of follow-up period
0.22%
1/458 • Number of events 1 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period
Product Issues
Oversensing
0.41%
4/982 • Number of events 4 • From enrollment to 3 years of follow-up period
0.30%
2/670 • Number of events 2 • From enrollment to 3 years of follow-up period
0.66%
3/458 • Number of events 3 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period
Product Issues
Oversensing Resulting in Therapy
0.00%
0/982 • From enrollment to 3 years of follow-up period
0.00%
0/670 • From enrollment to 3 years of follow-up period
0.22%
1/458 • Number of events 1 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period
Product Issues
Pectoral Stimulation
0.00%
0/982 • From enrollment to 3 years of follow-up period
0.00%
0/670 • From enrollment to 3 years of follow-up period
0.00%
0/458 • From enrollment to 3 years of follow-up period
0.14%
1/737 • Number of events 1 • From enrollment to 3 years of follow-up period
Product Issues
Phrenic Nerve/Diaphragmatic Stimulation
0.00%
0/982 • From enrollment to 3 years of follow-up period
0.00%
0/670 • From enrollment to 3 years of follow-up period
0.00%
0/458 • From enrollment to 3 years of follow-up period
0.54%
4/737 • Number of events 4 • From enrollment to 3 years of follow-up period
Product Issues
Therapy for Non-Ventricular Rhythm
0.10%
1/982 • Number of events 1 • From enrollment to 3 years of follow-up period
0.45%
3/670 • Number of events 3 • From enrollment to 3 years of follow-up period
0.22%
1/458 • Number of events 1 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period
Product Issues
Undersensing
0.00%
0/982 • From enrollment to 3 years of follow-up period
0.15%
1/670 • Number of events 1 • From enrollment to 3 years of follow-up period
0.00%
0/458 • From enrollment to 3 years of follow-up period
0.00%
0/737 • From enrollment to 3 years of follow-up period

Additional Information

Grant Kim, Program Director - Traditional Cardiac Rhythm Management

Abbott

Phone: 818-493-3147

Results disclosure agreements

  • Principal investigator is a sponsor employee The sponsor can review results communications prior to public release and embargo communications regarding trial results for 12 months from study completion date to allow time for multi-center study publication. After this period, the PI shall submit to sponsor for review a copy of any proposed manuscript at least 30 days prior to the estimated date of submission for publication. Sponsor may require delay of publication not to exceed 60 days for the purpose of filing patent applications.
  • Publication restrictions are in place

Restriction type: OTHER