Trial Outcomes & Findings for Assessment of Measurement Variability Across Automated Biometry Devices (NCT NCT06045325)
NCT ID: NCT06045325
Last Updated: 2025-05-13
Results Overview
The determination of the degree of correlation of white-to-white measurements between various biometric devices and the Orbscan II.
COMPLETED
204 participants
Day of visit
2025-05-13
Participant Flow
Unit of analysis: eyes
Participant milestones
| Measure |
Myopic -1.00D
Myopic patients with a diopter of at least -1.00
Orbscan II: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness.
IOL Master 700: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness.
IOL Master 500: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness.
Atlas 9000: Diagnostic assessment to measure white-to-white.
Lenstar 900: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness.
Pentacam: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness.
iTrace: Diagnostic assessment to measure white-to-white and corneal thickness.
Argos: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness.
Manual caliper measurement: Manual diagnostic assessment to measure white-to-white.
|
|---|---|
|
Overall Study
STARTED
|
204 408
|
|
Overall Study
COMPLETED
|
204 408
|
|
Overall Study
NOT COMPLETED
|
0 0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Assessment of Measurement Variability Across Automated Biometry Devices
Baseline characteristics by cohort
| Measure |
Myopic -1.00D
n=408 eyes
Myopic patients with a diopter of at least -1.00. All patients will receive scans with each device.
Orbscan II: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness.
IOL Master 700: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness.
IOL Master 500: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness.
Atlas 9000: Diagnostic assessment to measure white-to-white.
Lenstar 900: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness.
Pentacam: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness.
iTrace: Diagnostic assessment to measure white-to-white and corneal thickness.
Argos: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness.
Manual caliper measurement: Manual diagnostic assessment to measure white-to-white.
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=204 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
204 Participants
n=204 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=204 Participants
|
|
Sex/Gender, Customized
Not disclosed
|
204 Participants
n=204 Participants
|
PRIMARY outcome
Timeframe: Day of visitThe determination of the degree of correlation of white-to-white measurements between various biometric devices and the Orbscan II.
Outcome measures
| Measure |
Argos
n=408 eyes
Argos biometer
|
Atlas 9000
n=408 eyes
Atlas 9000 biometer
|
Manual Caliper
n=408 eyes
Manual Caliper
|
IOLMaster 500
n=408 eyes
IOLMaster 500 biometer
|
IOLMaster 700
n=408 eyes
IOLMaster 700 biometer
|
iTrace
n=408 eyes
iTrace device
|
Lenstar LS900
n=408 eyes
Lenstar LS900 biometer
|
Pentacam AXL Wave
n=322 eyes
Pentacam AXL Wave device
|
Pentacam HR
n=408 eyes
Pentacam HR device
|
|---|---|---|---|---|---|---|---|---|---|
|
The Correlation of White-to-white Measurements Between Biometric Devices.
|
0.89 Pearson correlation r
Interval 0.87 to 0.91
|
0.65 Pearson correlation r
Interval 0.59 to 0.71
|
0.53 Pearson correlation r
Interval 0.46 to 0.6
|
0.60 Pearson correlation r
Interval 0.54 to 0.66
|
0.82 Pearson correlation r
Interval 0.78 to 0.85
|
0.78 Pearson correlation r
Interval 0.74 to 0.82
|
0.87 Pearson correlation r
Interval 0.84 to 0.89
|
0.84 Pearson correlation r
Interval 0.81 to 0.87
|
0.84 Pearson correlation r
Interval 0.81 to 0.87
|
PRIMARY outcome
Timeframe: Day of visitThe determination of the degree of correlation of anterior chamber depth measurements between various biometric devices and the Orbscan II.
Outcome measures
| Measure |
Argos
n=408 eyes
Argos biometer
|
Atlas 9000
n=408 eyes
Atlas 9000 biometer
|
Manual Caliper
n=408 eyes
Manual Caliper
|
IOLMaster 500
n=408 eyes
IOLMaster 500 biometer
|
IOLMaster 700
n=322 eyes
IOLMaster 700 biometer
|
iTrace
n=408 eyes
iTrace device
|
Lenstar LS900
Lenstar LS900 biometer
|
Pentacam AXL Wave
Pentacam AXL Wave device
|
Pentacam HR
Pentacam HR device
|
|---|---|---|---|---|---|---|---|---|---|
|
The Correlation of Anterior Chamber Depth Measurements Between Biometric Devices.
|
0.91 Pearson correlation r
Interval 0.9 to 0.93
|
0.39 Pearson correlation r
Interval 0.31 to 0.47
|
0.33 Pearson correlation r
Interval 0.24 to 0.41
|
0.41 Pearson correlation r
Interval 0.33 to 0.49
|
0.95 Pearson correlation r
Interval 0.94 to 0.96
|
0.96 Pearson correlation r
Interval 0.96 to 0.97
|
—
|
—
|
—
|
PRIMARY outcome
Timeframe: Day of visitPopulation: Note that the iTrace, IOLMaster 500, Atlas 9000, and manual caliper did not measure central corneal thickness, so no central corneal thickness data were collected from these devices.
The determination of the degree of correlation of corneal thickness measurements between various biometric devices and the Orbscan II.
Outcome measures
| Measure |
Argos
n=408 eyes
Argos biometer
|
Atlas 9000
n=408 eyes
Atlas 9000 biometer
|
Manual Caliper
n=408 eyes
Manual Caliper
|
IOLMaster 500
n=322 eyes
IOLMaster 500 biometer
|
IOLMaster 700
n=408 eyes
IOLMaster 700 biometer
|
iTrace
iTrace device
|
Lenstar LS900
Lenstar LS900 biometer
|
Pentacam AXL Wave
Pentacam AXL Wave device
|
Pentacam HR
Pentacam HR device
|
|---|---|---|---|---|---|---|---|---|---|
|
The Correlation of Corneal Thickness Measurements Between Biometric Devices.
|
0.82 Pearson correlation r
Interval 0.78 to 0.85
|
0.80 Pearson correlation r
Interval 0.76 to 0.83
|
0.68 Pearson correlation r
Interval 0.63 to 0.73
|
0.68 Pearson correlation r
Interval 0.62 to 0.73
|
0.73 Pearson correlation r
Interval 0.68 to 0.77
|
—
|
—
|
—
|
—
|
Adverse Events
Argos
Atlas 9000
Manual Caliper
IOLMaster 500
IOLMaster 700
iTrace
Lenstar LS900
Pentacam AXL Wave
Pentacam HR
Orbscan II
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place