Trial Outcomes & Findings for Assessment of Measurement Variability Across Automated Biometry Devices (NCT NCT06045325)

NCT ID: NCT06045325

Last Updated: 2025-05-13

Results Overview

The determination of the degree of correlation of white-to-white measurements between various biometric devices and the Orbscan II.

Recruitment status

COMPLETED

Target enrollment

204 participants

Primary outcome timeframe

Day of visit

Results posted on

2025-05-13

Participant Flow

Unit of analysis: eyes

Participant milestones

Participant milestones
Measure
Myopic -1.00D
Myopic patients with a diopter of at least -1.00 Orbscan II: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness. IOL Master 700: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness. IOL Master 500: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness. Atlas 9000: Diagnostic assessment to measure white-to-white. Lenstar 900: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness. Pentacam: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness. iTrace: Diagnostic assessment to measure white-to-white and corneal thickness. Argos: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness. Manual caliper measurement: Manual diagnostic assessment to measure white-to-white.
Overall Study
STARTED
204 408
Overall Study
COMPLETED
204 408
Overall Study
NOT COMPLETED
0 0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Assessment of Measurement Variability Across Automated Biometry Devices

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Myopic -1.00D
n=408 eyes
Myopic patients with a diopter of at least -1.00. All patients will receive scans with each device. Orbscan II: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness. IOL Master 700: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness. IOL Master 500: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness. Atlas 9000: Diagnostic assessment to measure white-to-white. Lenstar 900: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness. Pentacam: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness. iTrace: Diagnostic assessment to measure white-to-white and corneal thickness. Argos: Diagnostic assessment to measure anterior chamber depth, white-to-white, and corneal thickness. Manual caliper measurement: Manual diagnostic assessment to measure white-to-white.
Age, Categorical
<=18 years
0 Participants
n=204 Participants
Age, Categorical
Between 18 and 65 years
204 Participants
n=204 Participants
Age, Categorical
>=65 years
0 Participants
n=204 Participants
Sex/Gender, Customized
Not disclosed
204 Participants
n=204 Participants

PRIMARY outcome

Timeframe: Day of visit

The determination of the degree of correlation of white-to-white measurements between various biometric devices and the Orbscan II.

Outcome measures

Outcome measures
Measure
Argos
n=408 eyes
Argos biometer
Atlas 9000
n=408 eyes
Atlas 9000 biometer
Manual Caliper
n=408 eyes
Manual Caliper
IOLMaster 500
n=408 eyes
IOLMaster 500 biometer
IOLMaster 700
n=408 eyes
IOLMaster 700 biometer
iTrace
n=408 eyes
iTrace device
Lenstar LS900
n=408 eyes
Lenstar LS900 biometer
Pentacam AXL Wave
n=322 eyes
Pentacam AXL Wave device
Pentacam HR
n=408 eyes
Pentacam HR device
The Correlation of White-to-white Measurements Between Biometric Devices.
0.89 Pearson correlation r
Interval 0.87 to 0.91
0.65 Pearson correlation r
Interval 0.59 to 0.71
0.53 Pearson correlation r
Interval 0.46 to 0.6
0.60 Pearson correlation r
Interval 0.54 to 0.66
0.82 Pearson correlation r
Interval 0.78 to 0.85
0.78 Pearson correlation r
Interval 0.74 to 0.82
0.87 Pearson correlation r
Interval 0.84 to 0.89
0.84 Pearson correlation r
Interval 0.81 to 0.87
0.84 Pearson correlation r
Interval 0.81 to 0.87

PRIMARY outcome

Timeframe: Day of visit

The determination of the degree of correlation of anterior chamber depth measurements between various biometric devices and the Orbscan II.

Outcome measures

Outcome measures
Measure
Argos
n=408 eyes
Argos biometer
Atlas 9000
n=408 eyes
Atlas 9000 biometer
Manual Caliper
n=408 eyes
Manual Caliper
IOLMaster 500
n=408 eyes
IOLMaster 500 biometer
IOLMaster 700
n=322 eyes
IOLMaster 700 biometer
iTrace
n=408 eyes
iTrace device
Lenstar LS900
Lenstar LS900 biometer
Pentacam AXL Wave
Pentacam AXL Wave device
Pentacam HR
Pentacam HR device
The Correlation of Anterior Chamber Depth Measurements Between Biometric Devices.
0.91 Pearson correlation r
Interval 0.9 to 0.93
0.39 Pearson correlation r
Interval 0.31 to 0.47
0.33 Pearson correlation r
Interval 0.24 to 0.41
0.41 Pearson correlation r
Interval 0.33 to 0.49
0.95 Pearson correlation r
Interval 0.94 to 0.96
0.96 Pearson correlation r
Interval 0.96 to 0.97

PRIMARY outcome

Timeframe: Day of visit

Population: Note that the iTrace, IOLMaster 500, Atlas 9000, and manual caliper did not measure central corneal thickness, so no central corneal thickness data were collected from these devices.

The determination of the degree of correlation of corneal thickness measurements between various biometric devices and the Orbscan II.

Outcome measures

Outcome measures
Measure
Argos
n=408 eyes
Argos biometer
Atlas 9000
n=408 eyes
Atlas 9000 biometer
Manual Caliper
n=408 eyes
Manual Caliper
IOLMaster 500
n=322 eyes
IOLMaster 500 biometer
IOLMaster 700
n=408 eyes
IOLMaster 700 biometer
iTrace
iTrace device
Lenstar LS900
Lenstar LS900 biometer
Pentacam AXL Wave
Pentacam AXL Wave device
Pentacam HR
Pentacam HR device
The Correlation of Corneal Thickness Measurements Between Biometric Devices.
0.82 Pearson correlation r
Interval 0.78 to 0.85
0.80 Pearson correlation r
Interval 0.76 to 0.83
0.68 Pearson correlation r
Interval 0.63 to 0.73
0.68 Pearson correlation r
Interval 0.62 to 0.73
0.73 Pearson correlation r
Interval 0.68 to 0.77

Adverse Events

Argos

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Atlas 9000

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Manual Caliper

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

IOLMaster 500

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

IOLMaster 700

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

iTrace

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Lenstar LS900

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Pentacam AXL Wave

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Pentacam HR

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Orbscan II

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported