Trial Outcomes & Findings for Clinical Performance of Masimo Rad-GT (NCT NCT05779397)
NCT ID: NCT05779397
Last Updated: 2023-07-25
Results Overview
The accuracy of the Masimo Rad-GT sensor will be determined by calculating the clinical bias (Δcb), limits of agreement (LA), and repeatability (σr). The temperature measurements from Masimo Rad-GT will be compared to the reference temperature measurements (Degrees Celsius).
COMPLETED
NA
68 participants
1-3 hours
2023-07-25
Participant Flow
Of the 68 enrolled participants, 68 completed the study.
Participant milestones
| Measure |
Rad-GT
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Masimo RAD-GT
Masimo RAD-GT: Non-contact infrared thermometry device
|
|---|---|
|
Overall Study
STARTED
|
68
|
|
Overall Study
COMPLETED
|
68
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Clinical Performance of Masimo Rad-GT
Baseline characteristics by cohort
| Measure |
Rad-GT
n=68 Participants
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Masimo RAD-GT
Masimo RAD-GT: Non-contact infrared thermometry device
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=93 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
68 Participants
n=93 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=93 Participants
|
|
Sex: Female, Male
Female
|
22 Participants
n=93 Participants
|
|
Sex: Female, Male
Male
|
46 Participants
n=93 Participants
|
|
Race/Ethnicity, Customized
Ethnicity · Caucasian
|
32 Participants
n=93 Participants
|
|
Race/Ethnicity, Customized
Ethnicity · Hispanic
|
5 Participants
n=93 Participants
|
|
Race/Ethnicity, Customized
Ethnicity · Asian or Pacific Islander
|
23 Participants
n=93 Participants
|
|
Race/Ethnicity, Customized
Ethnicity · Other
|
6 Participants
n=93 Participants
|
|
Race/Ethnicity, Customized
Ethnicity · More than one
|
2 Participants
n=93 Participants
|
|
Region of Enrollment
United States
|
68 Participants
n=93 Participants
|
PRIMARY outcome
Timeframe: 1-3 hoursPopulation: Data for 7 subjects was excluded as they did not complete at least one full one round of measurements. Data for 37 subjects was excluded as the data suggested existence of external conditions that interfere with RCT (reference clinical thermometer) or Rad-GT measurements.
The accuracy of the Masimo Rad-GT sensor will be determined by calculating the clinical bias (Δcb), limits of agreement (LA), and repeatability (σr). The temperature measurements from Masimo Rad-GT will be compared to the reference temperature measurements (Degrees Celsius).
Outcome measures
| Measure |
Rad-GT
n=24 Participants
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Masimo RAD-GT
Masimo RAD-GT: Non-contact infrared thermometry device
|
|---|---|
|
Accuracy of Masimo Rad-GT Temperature Measurements (Degrees Celsius)
Clinical Bias
|
-0.24 Degrees Celsius
|
|
Accuracy of Masimo Rad-GT Temperature Measurements (Degrees Celsius)
Limits of Agreement
|
0.85 Degrees Celsius
|
|
Accuracy of Masimo Rad-GT Temperature Measurements (Degrees Celsius)
Repeatability
|
0.04 Degrees Celsius
|
Adverse Events
Rad-GT
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place