Trial Outcomes & Findings for Calibration and Validation of the Masimo Temperature Device in Febrile Patients (NCT NCT05674344)
NCT ID: NCT05674344
Last Updated: 2023-07-25
Results Overview
The accuracy of the Masimo Rad-GT sensor will be determined by calculating the clinical bias (Δcb), limits of agreement (LA), and repeatability (σr). The temperature measurements from Masimo Rad-GT will be compared to the reference temperature measurements (degrees Celsius).
COMPLETED
NA
170 participants
15-30 Minutes
2023-07-25
Participant Flow
Participant milestones
| Measure |
Masimo Noninvasive Temperature Device
All subjects who qualify for the trial and participate in data collection will have temperature taken using the Masimo noninvasive temperature device and the reference temperature device.
Masimo Temperature device: Non-invasive temperature device
|
|---|---|
|
Overall Study
STARTED
|
170
|
|
Overall Study
COMPLETED
|
146
|
|
Overall Study
NOT COMPLETED
|
24
|
Reasons for withdrawal
| Measure |
Masimo Noninvasive Temperature Device
All subjects who qualify for the trial and participate in data collection will have temperature taken using the Masimo noninvasive temperature device and the reference temperature device.
Masimo Temperature device: Non-invasive temperature device
|
|---|---|
|
Overall Study
Screen Fail
|
21
|
|
Overall Study
Incomplete study procedures
|
3
|
Baseline Characteristics
Calibration and Validation of the Masimo Temperature Device in Febrile Patients
Baseline characteristics by cohort
| Measure |
Masimo Noninvasive Temperature Device
n=170 Participants
All subjects who qualify for the trial and participate in data collection will have temperature taken using the Masimo noninvasive temperature device and the reference temperature device.
Masimo Temperature device: Non-invasive temperature device
|
|---|---|
|
Age, Categorical
<=18 years
|
170 Participants
n=5 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
0 Participants
n=5 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=5 Participants
|
|
Sex: Female, Male
Female
|
90 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
80 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Hispanic or Latino
|
126 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Not Hispanic or Latino
|
41 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Unknown or Not Reported
|
3 Participants
n=5 Participants
|
|
Race (NIH/OMB)
American Indian or Alaska Native
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Asian
|
12 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
|
1 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Black or African American
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
White
|
92 Participants
n=5 Participants
|
|
Race (NIH/OMB)
More than one race
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Unknown or Not Reported
|
65 Participants
n=5 Participants
|
|
Region of Enrollment
United States
|
170 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 15-30 MinutesPopulation: Subject 1 to 91 were part of the calibration phase of the study (91 subjects) and data was excluded from analysis. 14 excluded due to screen fails. Subject 92 to 170 (79 subjects) were part of the validation phase of the study. 10 screen fails. 4: Any subject who does not complete at least one full round of measurements. 26: Any subject whose data suggests existence of external conditions that interfere with RCT (reference clinical thermometer) or Rad-GT measurements.
The accuracy of the Masimo Rad-GT sensor will be determined by calculating the clinical bias (Δcb), limits of agreement (LA), and repeatability (σr). The temperature measurements from Masimo Rad-GT will be compared to the reference temperature measurements (degrees Celsius).
Outcome measures
| Measure |
Masimo Noninvasive Temperature Device
n=39 Participants
All subjects who qualify for the trial and participate in data collection will have temperature taken using the Masimo noninvasive temperature device and the reference temperature device.
Masimo Temperature device: Non-invasive temperature device
|
|---|---|
|
Accuracy of Masimo RAD-GT Temperature Measurments
Clinical Bias
|
-0.24 Degrees Celsius
|
|
Accuracy of Masimo RAD-GT Temperature Measurments
Limits of Agreement
|
0.85 Degrees Celsius
|
|
Accuracy of Masimo RAD-GT Temperature Measurments
Repeatability
|
0.09 Degrees Celsius
|
Adverse Events
Masimo Noninvasive Temperature Device
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place
Restriction type: GT60