Trial Outcomes & Findings for Calibration and Validation of the Masimo Temperature Device in Febrile Patients (NCT NCT05674344)

NCT ID: NCT05674344

Last Updated: 2023-07-25

Results Overview

The accuracy of the Masimo Rad-GT sensor will be determined by calculating the clinical bias (Δcb), limits of agreement (LA), and repeatability (σr). The temperature measurements from Masimo Rad-GT will be compared to the reference temperature measurements (degrees Celsius).

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

170 participants

Primary outcome timeframe

15-30 Minutes

Results posted on

2023-07-25

Participant Flow

Participant milestones

Participant milestones
Measure
Masimo Noninvasive Temperature Device
All subjects who qualify for the trial and participate in data collection will have temperature taken using the Masimo noninvasive temperature device and the reference temperature device. Masimo Temperature device: Non-invasive temperature device
Overall Study
STARTED
170
Overall Study
COMPLETED
146
Overall Study
NOT COMPLETED
24

Reasons for withdrawal

Reasons for withdrawal
Measure
Masimo Noninvasive Temperature Device
All subjects who qualify for the trial and participate in data collection will have temperature taken using the Masimo noninvasive temperature device and the reference temperature device. Masimo Temperature device: Non-invasive temperature device
Overall Study
Screen Fail
21
Overall Study
Incomplete study procedures
3

Baseline Characteristics

Calibration and Validation of the Masimo Temperature Device in Febrile Patients

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Masimo Noninvasive Temperature Device
n=170 Participants
All subjects who qualify for the trial and participate in data collection will have temperature taken using the Masimo noninvasive temperature device and the reference temperature device. Masimo Temperature device: Non-invasive temperature device
Age, Categorical
<=18 years
170 Participants
n=5 Participants
Age, Categorical
Between 18 and 65 years
0 Participants
n=5 Participants
Age, Categorical
>=65 years
0 Participants
n=5 Participants
Sex: Female, Male
Female
90 Participants
n=5 Participants
Sex: Female, Male
Male
80 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
126 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
41 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
3 Participants
n=5 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
n=5 Participants
Race (NIH/OMB)
Asian
12 Participants
n=5 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
1 Participants
n=5 Participants
Race (NIH/OMB)
Black or African American
0 Participants
n=5 Participants
Race (NIH/OMB)
White
92 Participants
n=5 Participants
Race (NIH/OMB)
More than one race
0 Participants
n=5 Participants
Race (NIH/OMB)
Unknown or Not Reported
65 Participants
n=5 Participants
Region of Enrollment
United States
170 participants
n=5 Participants

PRIMARY outcome

Timeframe: 15-30 Minutes

Population: Subject 1 to 91 were part of the calibration phase of the study (91 subjects) and data was excluded from analysis. 14 excluded due to screen fails. Subject 92 to 170 (79 subjects) were part of the validation phase of the study. 10 screen fails. 4: Any subject who does not complete at least one full round of measurements. 26: Any subject whose data suggests existence of external conditions that interfere with RCT (reference clinical thermometer) or Rad-GT measurements.

The accuracy of the Masimo Rad-GT sensor will be determined by calculating the clinical bias (Δcb), limits of agreement (LA), and repeatability (σr). The temperature measurements from Masimo Rad-GT will be compared to the reference temperature measurements (degrees Celsius).

Outcome measures

Outcome measures
Measure
Masimo Noninvasive Temperature Device
n=39 Participants
All subjects who qualify for the trial and participate in data collection will have temperature taken using the Masimo noninvasive temperature device and the reference temperature device. Masimo Temperature device: Non-invasive temperature device
Accuracy of Masimo RAD-GT Temperature Measurments
Clinical Bias
-0.24 Degrees Celsius
Accuracy of Masimo RAD-GT Temperature Measurments
Limits of Agreement
0.85 Degrees Celsius
Accuracy of Masimo RAD-GT Temperature Measurments
Repeatability
0.09 Degrees Celsius

Adverse Events

Masimo Noninvasive Temperature Device

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Chelsea Frank

Masimo Corporation

Phone: (949) 297-7000

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place

Restriction type: GT60