Trial Outcomes & Findings for Evaluation of a Hearing Device for Transmitting Sound to the Inner Ear (NCT NCT05628285)

NCT ID: NCT05628285

Last Updated: 2025-12-03

Results Overview

Functional gain with Sentio 1, i.e., the difference in dB between unaided and aided sound field thresholds, calculated on average for frequencies 500, 1000, 2000 and 4000 Hz (PTA4). In detail, for each patient, sound field thresholds are measured at the frequencies 500, 1000, 2000 and 4000 Hz for which a mean value is calculated and denoted as PTA4. The primary outcome measure is functional gain which is the difference in PTA4 measured unaided (without sound processor) versus aided condition (with sound processor).

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

11 participants

Primary outcome timeframe

1 month

Results posted on

2025-12-03

Participant Flow

Unit of analysis: Ears

Participant milestones

Participant milestones
Measure
Sentio 1 Sound Processor
The sound processor picks up the sound and transfer it to the implant that convert the sound to vibrations that are transmitted to the inner ear. Sentio 1: Sound processor
Overall Study
STARTED
11 11
Overall Study
COMPLETED
10 10
Overall Study
NOT COMPLETED
1 1

Reasons for withdrawal

Reasons for withdrawal
Measure
Sentio 1 Sound Processor
The sound processor picks up the sound and transfer it to the implant that convert the sound to vibrations that are transmitted to the inner ear. Sentio 1: Sound processor
Overall Study
Lost to Follow-up
1

Baseline Characteristics

Race and Ethnicity were not collected from any participant.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Sentio 1 Sound Processor
n=11 Participants
The sound processor picks up the sound and transfer it to the implant that convert the sound to vibrations that are transmitted to the inner ear. Sentio 1: Sound processor
Age, Continuous
51 years
n=11 Participants
Sex: Female, Male
Female
7 Participants
n=11 Participants
Sex: Female, Male
Male
4 Participants
n=11 Participants

PRIMARY outcome

Timeframe: 1 month

Functional gain with Sentio 1, i.e., the difference in dB between unaided and aided sound field thresholds, calculated on average for frequencies 500, 1000, 2000 and 4000 Hz (PTA4). In detail, for each patient, sound field thresholds are measured at the frequencies 500, 1000, 2000 and 4000 Hz for which a mean value is calculated and denoted as PTA4. The primary outcome measure is functional gain which is the difference in PTA4 measured unaided (without sound processor) versus aided condition (with sound processor).

Outcome measures

Outcome measures
Measure
Functional Gain PTA4 (Sentio 1 Sound Processor)
n=10 Ears
The sound processor picks up the sound and transfer it to the implant that convert the sound to vibrations that are transmitted to the inner ear. Sentio 1: Sound processor
Sentio 1 Sound Processor
Measurement at Visit 2 with the Sentio 1 sound processor
Difference Between Visit 2 (Sentio 1) and Visit 1 (Previous Device)
Visit 2 (Sentio 1) - Visit 1 (previous device)
To Assess the Improvement of Hearing With the Sentio1 (PTA4)
32.5 dB
Standard Deviation 6.8

SECONDARY outcome

Timeframe: 1 month

Functional gain with Sentio 1, i.e., the difference in dB between unaided (without sound processor) and aided (with sound processor on) sound field thresholds, for all measured frequencies. In detail, for each patient, sound field thresholds are measured at the frequencies 250, 500, 1000, 2000, 3000, 4000, 6000 and 8000 Hz, with and without the sound processor on.

Outcome measures

Outcome measures
Measure
Functional Gain PTA4 (Sentio 1 Sound Processor)
n=10 Ears
The sound processor picks up the sound and transfer it to the implant that convert the sound to vibrations that are transmitted to the inner ear. Sentio 1: Sound processor
Sentio 1 Sound Processor
Measurement at Visit 2 with the Sentio 1 sound processor
Difference Between Visit 2 (Sentio 1) and Visit 1 (Previous Device)
Visit 2 (Sentio 1) - Visit 1 (previous device)
To Assess the Improvement of Hearing With the Sentio1
4000Hz
32.0 dB
Standard Deviation 8.9
To Assess the Improvement of Hearing With the Sentio1
6000Hz
23.5 dB
Standard Deviation 9.1
To Assess the Improvement of Hearing With the Sentio1
250Hz
18.5 dB
Standard Deviation 11.6
To Assess the Improvement of Hearing With the Sentio1
500Hz
29.5 dB
Standard Deviation 12.1
To Assess the Improvement of Hearing With the Sentio1
1000Hz
35.0 dB
Standard Deviation 6.2
To Assess the Improvement of Hearing With the Sentio1
2000Hz
33.5 dB
Standard Deviation 10.3
To Assess the Improvement of Hearing With the Sentio1
3000Hz
35.0 dB
Standard Deviation 13.1
To Assess the Improvement of Hearing With the Sentio1
8000Hz
17.8 dB
Standard Deviation 12.9

SECONDARY outcome

Timeframe: 1 month

Sentio 1 compared to previous sound processor. 1. Average SSQ12 scores with Sentio 1 for each question, sub-scales and in total. 2. Average SSQ12 scores with current sound processor for each question, sub-scales and in total. 3. Difference in SSQ12 scores between Sentio 1 and current sound processor for each question, sub-scales and in total. Measured by the Speech, Spatial and Quality (SSQ) scale from 0-10 where 0 represents "Not at all" and 10 represents "Perfect". Higher scores mean a better outcome. For each of the scales the item outcomes are averaged. Fot the total score all items are averaged

Outcome measures

Outcome measures
Measure
Functional Gain PTA4 (Sentio 1 Sound Processor)
n=11 Ears
The sound processor picks up the sound and transfer it to the implant that convert the sound to vibrations that are transmitted to the inner ear. Sentio 1: Sound processor
Sentio 1 Sound Processor
n=10 Ears
Measurement at Visit 2 with the Sentio 1 sound processor
Difference Between Visit 2 (Sentio 1) and Visit 1 (Previous Device)
n=10 Ears
Visit 2 (Sentio 1) - Visit 1 (previous device)
Subjective Evaluation of Sound Processor Performance.
SSQ Speech
5.58 score on a scale
Standard Deviation 2.32
6.07 score on a scale
Standard Deviation 2.67
0.579 score on a scale
Standard Deviation 1.537
Subjective Evaluation of Sound Processor Performance.
SSQ Spatial
4.34 score on a scale
Standard Deviation 3.02
4.68 score on a scale
Standard Deviation 2.90
0.353 score on a scale
Standard Deviation 1.590
Subjective Evaluation of Sound Processor Performance.
SSQ Quality
6.21 score on a scale
Standard Deviation 2.12
6.22 score on a scale
Standard Deviation 1.70
0.154 score on a scale
Standard Deviation 0.920
Subjective Evaluation of Sound Processor Performance.
SSQ Total
5.47 score on a scale
Standard Deviation 2.18
5.78 score on a scale
Standard Deviation 2.26
0.402 score on a scale
Standard Deviation 0.937

SECONDARY outcome

Timeframe: 1 month

Percentage (%) of subjects who prefer the Sentio 1 over current sound processor

Outcome measures

Outcome measures
Measure
Functional Gain PTA4 (Sentio 1 Sound Processor)
n=10 Participants
The sound processor picks up the sound and transfer it to the implant that convert the sound to vibrations that are transmitted to the inner ear. Sentio 1: Sound processor
Sentio 1 Sound Processor
Measurement at Visit 2 with the Sentio 1 sound processor
Difference Between Visit 2 (Sentio 1) and Visit 1 (Previous Device)
Visit 2 (Sentio 1) - Visit 1 (previous device)
Subjective Sound Processor Preference
7 Participants

SECONDARY outcome

Timeframe: 1 month

1. Effective gain defined as the difference in dB between aided sound field thresholds with Sentio 1 and unmasked BC thresholds on the aided ear(s). The effective gain is calculated for all measured frequencies. 2. Effective gain with Sentio 1, see definition above, calculated in average for frequencies 500, 1000, 2000 and 4000 Hz (PTA4).

Outcome measures

Outcome measures
Measure
Functional Gain PTA4 (Sentio 1 Sound Processor)
n=10 Participants
The sound processor picks up the sound and transfer it to the implant that convert the sound to vibrations that are transmitted to the inner ear. Sentio 1: Sound processor
Sentio 1 Sound Processor
Measurement at Visit 2 with the Sentio 1 sound processor
Difference Between Visit 2 (Sentio 1) and Visit 1 (Previous Device)
Visit 2 (Sentio 1) - Visit 1 (previous device)
To Assess the Degree to Which Sentio 1 Compensates for the BC Hearing Loss
PTA4
11.9 dB
Standard Deviation 7.7
To Assess the Degree to Which Sentio 1 Compensates for the BC Hearing Loss
3000Hz
12.5 dB
Standard Deviation 9.2
To Assess the Degree to Which Sentio 1 Compensates for the BC Hearing Loss
500Hz
20.5 dB
Standard Deviation 13.6
To Assess the Degree to Which Sentio 1 Compensates for the BC Hearing Loss
1000Hz
11.0 dB
Standard Deviation 7.4
To Assess the Degree to Which Sentio 1 Compensates for the BC Hearing Loss
2000Hz
1.50 dB
Standard Deviation 10.01
To Assess the Degree to Which Sentio 1 Compensates for the BC Hearing Loss
4000Hz
14.5 dB
Standard Deviation 9.0

SECONDARY outcome

Timeframe: 1 month

Difference in dB between Sentio 1 BC in situ thresholds and BC thresholds measured with conventional unmasked BC audiometry for the frequencies 500, 1000, 2000, 3000 and 4000 Hz. The BC in-situ threshold is a measure of the inner ear functionality with the device. This should correspond to the unmasked BC thresholds measured with conventional audiometry. The difference was calculated as the threshold value measured with Sentio 1 minus the threshold value measured with conventional audiometry. Negative values indicate that the thresholds measured with Sentio 1 were lower (better) compared to those measured with conventional audiometry. Conversely, positive values indicate that the thresholds measured with Sentio 1 were higher (worse) compared to conventional audiometry. It's expected that two measurements should be comparable to each with a difference of maximum +/-5 dB.

Outcome measures

Outcome measures
Measure
Functional Gain PTA4 (Sentio 1 Sound Processor)
n=11 Ears
The sound processor picks up the sound and transfer it to the implant that convert the sound to vibrations that are transmitted to the inner ear. Sentio 1: Sound processor
Sentio 1 Sound Processor
Measurement at Visit 2 with the Sentio 1 sound processor
Difference Between Visit 2 (Sentio 1) and Visit 1 (Previous Device)
Visit 2 (Sentio 1) - Visit 1 (previous device)
To Assess the Difference Between BC In-situ Thresholds Measured With Sentio 1 and Conventional Unmasked BC Audiometry.
500Hz
-4.7 dB
Standard Error 11.5
To Assess the Difference Between BC In-situ Thresholds Measured With Sentio 1 and Conventional Unmasked BC Audiometry.
1000Hz
-3.4 dB
Standard Error 9.0
To Assess the Difference Between BC In-situ Thresholds Measured With Sentio 1 and Conventional Unmasked BC Audiometry.
2000Hz
-6.7 dB
Standard Error 7.9
To Assess the Difference Between BC In-situ Thresholds Measured With Sentio 1 and Conventional Unmasked BC Audiometry.
3000Hz
8.2 dB
Standard Error 8.4
To Assess the Difference Between BC In-situ Thresholds Measured With Sentio 1 and Conventional Unmasked BC Audiometry.
4000Hz
9.9 dB
Standard Error 9.2

SECONDARY outcome

Timeframe: 12 months

Tabulation of AEs occurred throughout the investigation and percentage of individuals with reported AEs.

Outcome measures

Outcome data not reported

Adverse Events

Sentio 1 Sound Processor

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Director of Clinical Affairs

Oticon Medical AB

Phone: 0046-31-7486100

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place