Trial Outcomes & Findings for Dexcom G6 for Total Pancreatectomy With Islet Autotransplantation (TPIAT) Patients (NCT NCT05607576)
NCT ID: NCT05607576
Last Updated: 2025-05-06
Results Overview
Blood Glucose levels (mg/dL) obtained by the Dexcom G6 sensor and transmitter compared to blood glucose levels (mg/dL) obtained by a handheld glucometer until the transmitter is changed. The Clarke Error Grid Analysis method was used to compare absolute values. Accuracy reported as percentage of comparators in Zone A and B.
COMPLETED
NA
18 participants
3 months
2025-05-06
Participant Flow
Participant milestones
| Measure |
Uncovered
Dexcom G6 CGM will be uncovered during standard of care radiologic procedures.
Uncovered CGM during radiologic procedures: Dexcom G6 CGM will not be covered during radiologic procedures.
|
|---|---|
|
Overall Study
STARTED
|
18
|
|
Overall Study
COMPLETED
|
14
|
|
Overall Study
NOT COMPLETED
|
4
|
Reasons for withdrawal
| Measure |
Uncovered
Dexcom G6 CGM will be uncovered during standard of care radiologic procedures.
Uncovered CGM during radiologic procedures: Dexcom G6 CGM will not be covered during radiologic procedures.
|
|---|---|
|
Overall Study
immediate sensor error
|
2
|
|
Overall Study
Physician Decision
|
2
|
Baseline Characteristics
Dexcom G6 for Total Pancreatectomy With Islet Autotransplantation (TPIAT) Patients
Baseline characteristics by cohort
| Measure |
Uncovered
n=14 Participants
Dexcom G6 CGM will be uncovered during standard of care radiologic procedures.
Uncovered CGM during radiologic procedures: Dexcom G6 CGM will not be covered during radiologic procedures.
|
|---|---|
|
Age, Categorical
<=18 years
|
14 Participants
n=5 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
0 Participants
n=5 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=5 Participants
|
|
Sex: Female, Male
Female
|
9 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
5 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Hispanic or Latino
|
3 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Not Hispanic or Latino
|
10 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Unknown or Not Reported
|
1 Participants
n=5 Participants
|
|
Race (NIH/OMB)
American Indian or Alaska Native
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Asian
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Black or African American
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
White
|
12 Participants
n=5 Participants
|
|
Race (NIH/OMB)
More than one race
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Unknown or Not Reported
|
2 Participants
n=5 Participants
|
|
BMI percentile
|
79.4 Percentile
n=5 Participants
|
|
Insurance
Private
|
8 Participants
n=5 Participants
|
|
Insurance
Medicaid
|
5 Participants
n=5 Participants
|
|
Insurance
Private and Medicaid
|
1 Participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 3 monthsPopulation: 4168 CGM-glucometer matched data points analyzed, 67% of which were collected from the HG (ACCU-CHEK®). A total of 1375 matched data points were collected from the participants in the outpatient setting using the SIG (Contour®). There were 1435 matched pairs before radiation exposure and 2,733 matched data points after radiation exposure. Absolute blood glucose readings were plotted and analyzed using the Clarke Error Grid Analysis. Data points are matched blood glucose readings.
Blood Glucose levels (mg/dL) obtained by the Dexcom G6 sensor and transmitter compared to blood glucose levels (mg/dL) obtained by a handheld glucometer until the transmitter is changed. The Clarke Error Grid Analysis method was used to compare absolute values. Accuracy reported as percentage of comparators in Zone A and B.
Outcome measures
| Measure |
CGM Uncovered -Pre Radiation Exposures
n=1435 Matched blood glucose readings
Uncovered CGM during radiologic procedures: Dexcom G6 CGM will not be covered during radiologic procedures, This data grouping provides baseline data.
|
CGM Uncovered - Post Radiation Exposure
n=1832 Matched blood glucose readings
blood glucose readings after transmitter was exposed to scatter radiation.
|
CGM Uncovered - Post Radiation Exposure >500 µGy
n=901 Matched blood glucose readings
blood glucose readings after transmitter was exposed to scatter radiation greater than 500 µGy
|
Handheld Clinical Glucometer
n=2793 Matched blood glucose readings
ACCU-CHEK handheld glucometer used in clinical setting
|
Handheld Home Glucometer.
n=1375 Matched blood glucose readings
Contour glucometer used at home.
|
|---|---|---|---|---|---|
|
Dexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)
error grid section A
|
806 Matched blood glucose readings
|
1226 Matched blood glucose readings
|
704 Matched blood glucose readings
|
1715 Matched blood glucose readings
|
1020 Matched blood glucose readings
|
|
Dexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)
error grid section B
|
606 Matched blood glucose readings
|
570 Matched blood glucose readings
|
187 Matched blood glucose readings
|
1033 Matched blood glucose readings
|
329 Matched blood glucose readings
|
|
Dexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)
error grid section C
|
6 Matched blood glucose readings
|
7 Matched blood glucose readings
|
5 Matched blood glucose readings
|
56 Matched blood glucose readings
|
69 Matched blood glucose readings
|
|
Dexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)
error grid section D
|
17 Matched blood glucose readings
|
31 Matched blood glucose readings
|
5 Matched blood glucose readings
|
47 Matched blood glucose readings
|
15 Matched blood glucose readings
|
PRIMARY outcome
Timeframe: 3 monthsPopulation: Data collected on functionality of transmitter over time (blood glucose (BG) mg/dL/months from TPIAT). Data points analyzed using the Clarke Error Grid analysis of matched blood glucose readings.
Transmitter functional duration is determined by time from TPIAT procedure and CGM placement (baseline) to time for Transmitter changed (approximately 3 months), Functionality is determined by analysis of Blood Glucose (mg/dL) in month groupings (0-1month, 1-2months, 2+ months from baseline. Blood Glucose levels (mg/dL) obtained by the Dexcom G6 sensor and transmitter compared to blood glucose levels (mg/dL) obtained by a handheld glucometer until the transmitter is changed. The Clarke Error Grid Analysis method was used to compare absolute values. Accuracy reported as percentage of comparators in Zone A and B.
Outcome measures
| Measure |
CGM Uncovered -Pre Radiation Exposures
n=3294 Matched blood glucose readings
Uncovered CGM during radiologic procedures: Dexcom G6 CGM will not be covered during radiologic procedures, This data grouping provides baseline data.
|
CGM Uncovered - Post Radiation Exposure
n=586 Matched blood glucose readings
blood glucose readings after transmitter was exposed to scatter radiation.
|
CGM Uncovered - Post Radiation Exposure >500 µGy
n=288 Matched blood glucose readings
blood glucose readings after transmitter was exposed to scatter radiation greater than 500 µGy
|
Handheld Clinical Glucometer
ACCU-CHEK handheld glucometer used in clinical setting
|
Handheld Home Glucometer.
Contour glucometer used at home.
|
|---|---|---|---|---|---|
|
Post Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)
error grid Zone A
|
2065 Matched blood glucose readings
|
446 Matched blood glucose readings
|
226 Matched blood glucose readings
|
—
|
—
|
|
Post Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)
error grid Zone B
|
1173 Matched blood glucose readings
|
128 Matched blood glucose readings
|
61 Matched blood glucose readings
|
—
|
—
|
|
Post Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)
error grid Zone C
|
16 Matched blood glucose readings
|
2 Matched blood glucose readings
|
0 Matched blood glucose readings
|
—
|
—
|
|
Post Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)
error grid Zone D
|
43 Matched blood glucose readings
|
10 Matched blood glucose readings
|
1 Matched blood glucose readings
|
—
|
—
|
Adverse Events
Uncovered
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Deborah Elder; Associate Professor
Cincinnati Children's Hospital Medical Center
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place