Trial Outcomes & Findings for The Accuracy of Flap Thickness Versus the Pre-determined Surgical Planning Using WaveLight® FS200 (NCT NCT05501795)
NCT ID: NCT05501795
Last Updated: 2024-07-12
Results Overview
Recruitment status
COMPLETED
Target enrollment
58 participants
Primary outcome timeframe
1 day postoperative
Results posted on
2024-07-12
Participant Flow
Participant milestones
| Measure |
WaveLight® FS200 Femtosecond Laser
Flap creation with the WaveLight® FS200 femtosecond laser
WaveLight® FS200 femtosecond laser: Flap creation with the WaveLight® FS200 femtosecond laser
|
|---|---|
|
Overall Study
STARTED
|
58
|
|
Overall Study
COMPLETED
|
58
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Race and Ethnicity were not collected from any participant.
Baseline characteristics by cohort
| Measure |
WaveLight® FS200 Femtosecond Laser
n=58 Participants
Flap creation with the WaveLight® FS200 femtosecond laser
WaveLight® FS200 femtosecond laser: Flap creation with the WaveLight® FS200 femtosecond laser
|
|---|---|
|
Age, Continuous
|
34.4 years
STANDARD_DEVIATION 7.7 • n=58 Participants
|
|
Sex: Female, Male
Female
|
26 Participants
n=58 Participants
|
|
Sex: Female, Male
Male
|
32 Participants
n=58 Participants
|
PRIMARY outcome
Timeframe: 1 day postoperativePopulation: 116 Eyes of 58 participants
Outcome measures
| Measure |
WaveLight® FS200 Femtosecond Laser
n=116 Eyes
Flap creation with the WaveLight® FS200 femtosecond laser
WaveLight® FS200 femtosecond laser: Flap creation with the WaveLight® FS200 femtosecond laser
|
Pentacam
Pachymetry as measured by the Oculus Pentacam
|
|---|---|---|
|
The Resultant Thickness of the FS200 Created LASIK Flap Compared to Plan at the Surgical Visit.
|
0.64 um
Standard Deviation 9
|
—
|
SECONDARY outcome
Timeframe: 1 day postoperativePopulation: 116 Eyes of 58 participants
Outcome measures
| Measure |
WaveLight® FS200 Femtosecond Laser
n=116 Eyes
Flap creation with the WaveLight® FS200 femtosecond laser
WaveLight® FS200 femtosecond laser: Flap creation with the WaveLight® FS200 femtosecond laser
|
Pentacam
Pachymetry as measured by the Oculus Pentacam
|
|---|---|---|
|
The Resultant Diameter of the FS200 Created LASIK Flap Compared to Plan
|
0.02 mm
Standard Deviation 0.05
|
—
|
SECONDARY outcome
Timeframe: Day 0 (surgical visit)Population: 116 Eyes of 58 participants
Thickness measurements compared between 2 devices: the EX500 and the Oculus Pentacam
Outcome measures
| Measure |
WaveLight® FS200 Femtosecond Laser
n=116 Eyes
Flap creation with the WaveLight® FS200 femtosecond laser
WaveLight® FS200 femtosecond laser: Flap creation with the WaveLight® FS200 femtosecond laser
|
Pentacam
n=116 Eyes
Pachymetry as measured by the Oculus Pentacam
|
|---|---|---|
|
Pre-operative Pachymetry as Measured by the EX500 Versus the Oculus Pentacam
|
555.2 um
Standard Deviation 30.9
|
553.1 um
Standard Deviation 29.4
|
Adverse Events
WaveLight® FS200 Femtosecond Laser
Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place