Trial Outcomes & Findings for Real World Perspectives on Advancements in Hearing Aid Technology (NCT NCT05468918)

NCT ID: NCT05468918

Last Updated: 2023-08-22

Results Overview

A speech in noise test in which participants repeat words which are presented in the presence of background noise. In a 12-speaker array, speech is presented from speakers either on the left side of the participant (-90 degrees) or from behind the participant (180 degrees) while background noise is presented from remaining speakers. Scores are calculated as the number of words correctly repeated and expressed as a percentage. A higher percentage or score is better.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

14 participants

Primary outcome timeframe

Day 1 of study

Results posted on

2023-08-22

Participant Flow

Participant milestones

Participant milestones
Measure
Participants Tested With Comparator Devices First, Than Audeo Lumity Devices
All participants who completed speech in noise testing in the lab with the comparator devices first, and then the Audeo Lumity devices. Phonak Audeo previous generation hearing aids: Receiver-in-canal (RIC) hearing aids with universal earpieces Phonak Audeo Lumity hearing aids: Receiver-in-canal (RIC) hearing aids with universal earpieces
Participants Tested With Audeo Lumity First, Than Comparator Devices
All participants who completed speech in noise testing in the lab with the Audeo Lumity devices first, and then the comparator devices. Phonak Audeo previous generation hearing aids: Receiver-in-canal (RIC) hearing aids with universal earpieces Phonak Audeo Lumity hearing aids: Receiver-in-canal (RIC) hearing aids with universal earpieces
Overall Study
STARTED
7
7
Overall Study
COMPLETED
7
7
Overall Study
NOT COMPLETED
0
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Race and Ethnicity were not collected from any participant.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
All Participants Tested With Both Devices.
n=14 Participants
Participants who complete lab testing and home trials with both the comparator devices and the Audeo Lumity devices.
Age, Continuous
75 years
n=14 Participants
Sex: Female, Male
Female
8 Participants
n=14 Participants
Sex: Female, Male
Male
6 Participants
n=14 Participants
Region of Enrollment
United States
14 Participants
n=14 Participants

PRIMARY outcome

Timeframe: Day 1 of study

A speech in noise test in which participants repeat words which are presented in the presence of background noise. In a 12-speaker array, speech is presented from speakers either on the left side of the participant (-90 degrees) or from behind the participant (180 degrees) while background noise is presented from remaining speakers. Scores are calculated as the number of words correctly repeated and expressed as a percentage. A higher percentage or score is better.

Outcome measures

Outcome measures
Measure
All Participants Tested With Both Devices.
n=14 Participants
Participants who complete lab testing and home trials with both the comparator devices and the Audeo Lumity devices.
Speech in Noise Test
Percent correct with Lumity devices
84.84 percentage of correct responses
Interval 52.5 to 99.0
Speech in Noise Test
Percent correct with comparator devices
77.16 percentage of correct responses
Interval 24.0 to 98.5

PRIMARY outcome

Timeframe: Day 1 of study

Following the speech in noise testing in the lab, participants rated their listening effort on a scale from 1 to 10, with 1 = no effort and 10 = a lot of effort. In this case, a lower rating is better as it indicates less perceived listening effort.

Outcome measures

Outcome measures
Measure
All Participants Tested With Both Devices.
n=14 Participants
Participants who complete lab testing and home trials with both the comparator devices and the Audeo Lumity devices.
Subjective Perception of Listening Effort
Listening effort rating with Lumity devices
5.82 score on a scale
Interval 1.0 to 10.0
Subjective Perception of Listening Effort
Listening effort rating with comparator devices
6.82 score on a scale
Interval 3.0 to 10.0

Adverse Events

All Participants Who Wore Comparator Devices

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

All Participants Who Wore Audeo Lumity Devices

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Director of Research

Sonova USCS

Phone: 630-821-5000

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place