Trial Outcomes & Findings for Clinical Performance of Masimo INVSENSOR00057 for Heart Rate Measurements (NCT NCT05462886)
NCT ID: NCT05462886
Last Updated: 2024-09-24
Results Overview
Heart rate accuracy was determined by calculating the accuracy root mean-squared (Arms) difference between the measured values from Masimo INVSENSOR00057 (TestHR\_i) to the reference ECG monitor values (RefHR\_i). Arms= √(∑(i=1 to n) ((TestHR\_i-RefHR\_i )\^2 ))/n
COMPLETED
NA
86 participants
1-3 hours
2024-09-24
Participant Flow
Participant milestones
| Measure |
Masimo INVSENSOR00057
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Masimo INVSENSOR00057 device.
Masimo INVSENSOR00057: noninvasive wearable health monitoring device
|
|---|---|
|
Overall Study
STARTED
|
86
|
|
Overall Study
COMPLETED
|
79
|
|
Overall Study
NOT COMPLETED
|
7
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Clinical Performance of Masimo INVSENSOR00057 for Heart Rate Measurements
Baseline characteristics by cohort
| Measure |
Masimo INVSENSOR00057
n=79 Participants
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Masimo INVSENSOR00057 device.
Masimo INVSENSOR00057: noninvasive wearable health monitoring device
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=93 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
79 Participants
n=93 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=93 Participants
|
|
Sex: Female, Male
Female
|
36 Participants
n=93 Participants
|
|
Sex: Female, Male
Male
|
43 Participants
n=93 Participants
|
|
Race (NIH/OMB)
American Indian or Alaska Native
|
0 Participants
n=93 Participants
|
|
Race (NIH/OMB)
Asian
|
21 Participants
n=93 Participants
|
|
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
|
0 Participants
n=93 Participants
|
|
Race (NIH/OMB)
Black or African American
|
5 Participants
n=93 Participants
|
|
Race (NIH/OMB)
White
|
21 Participants
n=93 Participants
|
|
Race (NIH/OMB)
More than one race
|
8 Participants
n=93 Participants
|
|
Race (NIH/OMB)
Unknown or Not Reported
|
24 Participants
n=93 Participants
|
|
Region of Enrollment
United States
|
79 participants
n=93 Participants
|
PRIMARY outcome
Timeframe: 1-3 hoursPopulation: 17 subjects not included in analysis due to data collection error.
Heart rate accuracy was determined by calculating the accuracy root mean-squared (Arms) difference between the measured values from Masimo INVSENSOR00057 (TestHR\_i) to the reference ECG monitor values (RefHR\_i). Arms= √(∑(i=1 to n) ((TestHR\_i-RefHR\_i )\^2 ))/n
Outcome measures
| Measure |
INVSENSOR00057
n=62 Participants
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive INVSENSOR00057 device.
INVSENSOR00057: noninvasive wearable health monitoring device
|
|---|---|
|
Heart Rate Accuracy of Masimo INVSENSOR00057
|
1.38 bpm
|
Adverse Events
Masimo INVSENSOR00057
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place