Trial Outcomes & Findings for Efficacy of the Erchonia Corporation THL™ for Providing Relief of Tinnitus Symptoms (NCT NCT05374421)

NCT ID: NCT05374421

Last Updated: 2024-03-12

Results Overview

The Tinnitus Functional Index (TFI) has been highly recommended to be used in both clinical and research settings for tinnitus measurement (Meikle et al., 2011). The self-report questionnaire, consists of 25 items with a response option on an 11-point Likert scale from 0-10. The calculation of overall TFI score is as follows: (1) Sum all valid answers. (2) Divide by the number of questions for which the sum of the valid answers was based to yield the respondent's mean item score. (3) Multiply by 10 to attain the respondent's overall TFI score (0-100 range). Total TFI score can be categorized as Level of Tinnitus Severity as follows: 0 to 17: Not a problem,18 to 31: Small problem, 32 to 53: Moderate problem, 54 to 72: Big Problem, 73 to 100: Very big problem. A negative (-) change indicates the negative impact of tinnitus on the individual's everyday life has improved (lessened).

Recruitment status

TERMINATED

Study phase

NA

Target enrollment

23 participants

Primary outcome timeframe

Baseline and 4 weeks

Results posted on

2024-03-12

Participant Flow

The study procedure started with attainment of subject consent followed by subject randomization into a procedure group. Succeeding procedure group assignment the investigator performed the study qualification evaluation to determine if the subject was eligible. The "Protocol enrollment" number includes all subjects that signed the consent form, while the number of subjects that "Started" reflects the total number of subjects that satisfied the qualification criteria

Participant milestones

Participant milestones
Measure
Erchonia® THL™
The Erchonia® EVRL™ is made up of (2) 405-nanometer violet laser diodes mounted in a portable handheld device. Erchonia® THL™: The Erchonia® THL™ is made up of (2) 405-nanometer violet laser diodes mounted in a portable handheld device. The diodes are positioned approximately 4 inches from the skin and applied in a continuous back and forth sweeping motion from the top of the ear down the side of the neck. The treatments will be self-administered by the subject at his or her's place of residence, once per day for 28 days. Each procedure administration lasts 5 minutes per ear, for a total of 10 minutes.
Placebo Laser
The Placebo Laser has the same appearance as the Erchonia® THL™ but does not emit any therapeutic light. Placebo Laser: The Placebo Laser has the same appearance as the Erchonia® THL™ but does not emit any therapeutic light. The diodes are positioned approximately 4 inches from the skin and applied in a continuous back and forth sweeping motion from the top of the ear down the side of the neck. The treatments will be self-administered by the subject at his or her's place of residence, once per day for 28 days. Each procedure administration lasts 5 minutes per ear, for a total of 10 minutes.
Overall Study
STARTED
8
7
Overall Study
COMPLETED
5
3
Overall Study
NOT COMPLETED
3
4

Reasons for withdrawal

Reasons for withdrawal
Measure
Erchonia® THL™
The Erchonia® EVRL™ is made up of (2) 405-nanometer violet laser diodes mounted in a portable handheld device. Erchonia® THL™: The Erchonia® THL™ is made up of (2) 405-nanometer violet laser diodes mounted in a portable handheld device. The diodes are positioned approximately 4 inches from the skin and applied in a continuous back and forth sweeping motion from the top of the ear down the side of the neck. The treatments will be self-administered by the subject at his or her's place of residence, once per day for 28 days. Each procedure administration lasts 5 minutes per ear, for a total of 10 minutes.
Placebo Laser
The Placebo Laser has the same appearance as the Erchonia® THL™ but does not emit any therapeutic light. Placebo Laser: The Placebo Laser has the same appearance as the Erchonia® THL™ but does not emit any therapeutic light. The diodes are positioned approximately 4 inches from the skin and applied in a continuous back and forth sweeping motion from the top of the ear down the side of the neck. The treatments will be self-administered by the subject at his or her's place of residence, once per day for 28 days. Each procedure administration lasts 5 minutes per ear, for a total of 10 minutes.
Overall Study
Withdrawal by Subject
3
4

Baseline Characteristics

Efficacy of the Erchonia Corporation THL™ for Providing Relief of Tinnitus Symptoms

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Erchonia® THL™
n=8 Participants
The Erchonia® EVRL™ is made up of (2) 405-nanometer violet laser diodes mounted in a portable handheld device. Erchonia® THL™: The Erchonia® THL™ is made up of (2) 405-nanometer violet laser diodes mounted in a portable handheld device. The diodes are positioned approximately 4 inches from the skin and applied in a continuous back and forth sweeping motion from the top of the ear down the side of the neck. The treatments will be self-administered by the subject at his or her's place of residence, once per day for 28 days. Each procedure administration lasts 5 minutes per ear, for a total of 10 minutes.
Placebo Laser
n=7 Participants
The Placebo Laser has the same appearance as the Erchonia® THL™ but does not emit any therapeutic light. Placebo Laser: The Placebo Laser has the same appearance as the Erchonia® THL™ but does not emit any therapeutic light. The diodes are positioned approximately 4 inches from the skin and applied in a continuous back and forth sweeping motion from the top of the ear down the side of the neck. The treatments will be self-administered by the subject at his or her's place of residence, once per day for 28 days. Each procedure administration lasts 5 minutes per ear, for a total of 10 minutes.
Total
n=15 Participants
Total of all reporting groups
Race (NIH/OMB)
Asian
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
Race (NIH/OMB)
Black or African American
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
Race (NIH/OMB)
White
8 Participants
n=5 Participants
7 Participants
n=7 Participants
15 Participants
n=5 Participants
Race (NIH/OMB)
More than one race
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
Region of Enrollment
United States
8 participants
n=5 Participants
7 participants
n=7 Participants
15 participants
n=5 Participants
Age, Categorical
<=18 years
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
Age, Categorical
Between 18 and 65 years
2 Participants
n=5 Participants
2 Participants
n=7 Participants
4 Participants
n=5 Participants
Age, Categorical
>=65 years
6 Participants
n=5 Participants
5 Participants
n=7 Participants
11 Participants
n=5 Participants
Age, Continuous
70.4 years
STANDARD_DEVIATION 8.33 • n=5 Participants
55.6 years
STANDARD_DEVIATION 12.25 • n=7 Participants
64.8 years
STANDARD_DEVIATION 12.27 • n=5 Participants
Sex: Female, Male
Female
6 Participants
n=5 Participants
5 Participants
n=7 Participants
11 Participants
n=5 Participants
Sex: Female, Male
Male
2 Participants
n=5 Participants
2 Participants
n=7 Participants
4 Participants
n=5 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants

PRIMARY outcome

Timeframe: Baseline and 4 weeks

The Tinnitus Functional Index (TFI) has been highly recommended to be used in both clinical and research settings for tinnitus measurement (Meikle et al., 2011). The self-report questionnaire, consists of 25 items with a response option on an 11-point Likert scale from 0-10. The calculation of overall TFI score is as follows: (1) Sum all valid answers. (2) Divide by the number of questions for which the sum of the valid answers was based to yield the respondent's mean item score. (3) Multiply by 10 to attain the respondent's overall TFI score (0-100 range). Total TFI score can be categorized as Level of Tinnitus Severity as follows: 0 to 17: Not a problem,18 to 31: Small problem, 32 to 53: Moderate problem, 54 to 72: Big Problem, 73 to 100: Very big problem. A negative (-) change indicates the negative impact of tinnitus on the individual's everyday life has improved (lessened).

Outcome measures

Outcome measures
Measure
Erchonia® THL™
n=5 Participants
The Erchonia® EVRL™ is made up of (2) 405-nanometer violet laser diodes mounted in a portable handheld device. Erchonia® THL™: The Erchonia® THL™ is made up of (2) 405-nanometer violet laser diodes mounted in a portable handheld device. The diodes are positioned approximately 4 inches from the skin and applied in a continuous back and forth sweeping motion from the top of the ear down the side of the neck. The treatments will be self-administered by the subject at his or her's place of residence, once per day for 28 days. Each procedure administration lasts 5 minutes per ear, for a total of 10 minutes.
Placebo Laser
n=3 Participants
The Placebo Laser has the same appearance as the Erchonia® THL™ but does not emit any therapeutic light. Placebo Laser: The Placebo Laser has the same appearance as the Erchonia® THL™ but does not emit any therapeutic light. The diodes are positioned approximately 4 inches from the skin and applied in a continuous back and forth sweeping motion from the top of the ear down the side of the neck. The treatments will be self-administered by the subject at his or her's place of residence, once per day for 28 days. Each procedure administration lasts 5 minutes per ear, for a total of 10 minutes.
Mean Change From Baseline to Study Endpoint in the Tinnitus Functional Index (TFI)
-8.96 score on a scale
Standard Deviation 15.53
-2.8 score on a scale
Standard Deviation 5.7

Adverse Events

Erchonia® THL™

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Placebo Laser

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Clinical Affairs Manager

Erchonia Corporation

Phone: 864-531-0064

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place