Trial Outcomes & Findings for Disposable Powered Articulating Linear Cutter Stapler in Total-Thoracoscopic Anatomic Lobectomy (Segmentectomy) (NCT NCT05329402)

NCT ID: NCT05329402

Last Updated: 2024-05-02

Results Overview

The transection and anastomosis of lung tissues are performed during the operation. The device is withdrawn after successful triggering. The cutting staple line is carefully checked for integrity, air leakage and bleeding. If all staple lines are complete, without air leakage or bleeding, it is judged that the cutting and anastomosis of the device is successful. If there is persistent air leakage and bleeding at the anastomosis site, and conversion to thoracotomy is required, it is judged as device cutting and anastomosis failure.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

164 participants

Primary outcome timeframe

During surgery

Results posted on

2024-05-02

Participant Flow

30 days

Participant milestones

Participant milestones
Measure
Disposable Powered Articulating Endoscopic Linear Cutter Stapler
The Disposable Powered Articulating Endoscopic Linear Cutter Stapler is intended for transection, resection, and/or creation of anastomoses. The instrument has application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures. It can be used with staple line or tissue buttressing materials. The instrument may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen.
ECHELON Flex Powered Articulating Endoscopic Linear Cutters
ECHELON Flex Powered Articulating Endoscopic Linear Cutters has application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures.
Overall Study
STARTED
80
84
Overall Study
COMPLETED
79
83
Overall Study
NOT COMPLETED
1
1

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Race and Ethnicity were not collected from any participant.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Disposable Powered Articulating Endoscopic Linear Cutter Stapler
n=79 Participants
The Disposable Powered Articulating Endoscopic Linear Cutter Stapler is intended for transection, resection, and/or creation of anastomoses. The instrument has application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures. It can be used with staple line or tissue buttressing materials. The instrument may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen. Lung cancer surgery: Only the brand of electric cutting stapler used by different groups was different, and the rest were treated according to routine treatment
ECHELON Flex Powered Articulating Endoscopic Linear Cutters
n=83 Participants
ECHELON Flex Powered Articulating Endoscopic Linear Cutters has application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures. Lung cancer surgery: Only the brand of electric cutting stapler used by different groups was different, and the rest were treated according to routine treatment
Total
n=162 Participants
Total of all reporting groups
Age, Continuous
54.95 years
STANDARD_DEVIATION 9.77 • n=79 Participants
55.33 years
STANDARD_DEVIATION 8.94 • n=83 Participants
55.14 years
STANDARD_DEVIATION 9.33 • n=162 Participants
Sex: Female, Male
Female
40 Participants
n=79 Participants
51 Participants
n=83 Participants
91 Participants
n=162 Participants
Sex: Female, Male
Male
39 Participants
n=79 Participants
32 Participants
n=83 Participants
71 Participants
n=162 Participants
Race and Ethnicity Not Collected
0 Participants
Race and Ethnicity were not collected from any participant.
Region of Enrollment
China
79 participants
n=79 Participants
83 participants
n=83 Participants
162 participants
n=162 Participants
BMI
23.88 kg/m^2
STANDARD_DEVIATION 3.34 • n=79 Participants
24.06 kg/m^2
STANDARD_DEVIATION 3.24 • n=83 Participants
23.97 kg/m^2
STANDARD_DEVIATION 3.28 • n=162 Participants

PRIMARY outcome

Timeframe: During surgery

The transection and anastomosis of lung tissues are performed during the operation. The device is withdrawn after successful triggering. The cutting staple line is carefully checked for integrity, air leakage and bleeding. If all staple lines are complete, without air leakage or bleeding, it is judged that the cutting and anastomosis of the device is successful. If there is persistent air leakage and bleeding at the anastomosis site, and conversion to thoracotomy is required, it is judged as device cutting and anastomosis failure.

Outcome measures

Outcome measures
Measure
Disposable Powered Articulating Endoscopic Linear Cutter Stapler
n=79 Participants
The Disposable Powered Articulating Endoscopic Linear Cutter Stapler is intended for transection, resection, and/or creation of anastomoses. The instrument has application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures. It can be used with staple line or tissue buttressing materials. The instrument may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen. Lung cancer surgery: Only the brand of electric cutting stapler used by different groups was different, and the rest were treated according to routine treatment
ECHELON Flex Powered Articulating Endoscopic Linear Cutters
n=83 Participants
ECHELON Flex Powered Articulating Endoscopic Linear Cutters has application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures. Lung cancer surgery: Only the brand of electric cutting stapler used by different groups was different, and the rest were treated according to routine treatment
Number of Participants With Anastomosis Success
79 Participants
83 Participants

Adverse Events

Disposable Powered Articulating Endoscopic Linear Cutter Stapler

Serious events: 10 serious events
Other events: 44 other events
Deaths: 0 deaths

ECHELON Flex Powered Articulating Endoscopic Linear Cutters

Serious events: 8 serious events
Other events: 54 other events
Deaths: 0 deaths

Serious adverse events

Serious adverse events
Measure
Disposable Powered Articulating Endoscopic Linear Cutter Stapler
n=79 participants at risk
The Disposable Powered Articulating Endoscopic Linear Cutter Stapler is intended for transection, resection, and/or creation of anastomoses. The instrument has application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures. It can be used with staple line or tissue buttressing materials. The instrument may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen.
ECHELON Flex Powered Articulating Endoscopic Linear Cutters
n=83 participants at risk
ECHELON Flex Powered Articulating Endoscopic Linear Cutters has application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures.
Respiratory, thoracic and mediastinal disorders
pneumothorax
2.5%
2/79 • Number of events 2 • 30 days
0.00%
0/83 • 30 days
Respiratory, thoracic and mediastinal disorders
Right pneumothorax
0.00%
0/79 • 30 days
2.4%
2/83 • Number of events 2 • 30 days
Respiratory, thoracic and mediastinal disorders
Pulmonary artery bleeding
1.3%
1/79 • Number of events 1 • 30 days
0.00%
0/83 • 30 days
Respiratory, thoracic and mediastinal disorders
Aerodermectasia
1.3%
1/79 • Number of events 1 • 30 days
0.00%
0/83 • 30 days
Respiratory, thoracic and mediastinal disorders
Thoracic wall lung hernia
1.3%
1/79 • Number of events 1 • 30 days
0.00%
0/83 • 30 days
Cardiac disorders
Atrial fibrillation
0.00%
0/79 • 30 days
1.2%
1/83 • Number of events 1 • 30 days
Respiratory, thoracic and mediastinal disorders
Lung leak gas
1.3%
1/79 • Number of events 1 • 30 days
2.4%
2/83 • Number of events 2 • 30 days
Cardiac disorders
left chest pain
0.00%
0/79 • 30 days
1.2%
1/83 • Number of events 1 • 30 days
Nervous system disorders
Acute cerebral infarction
1.3%
1/79 • Number of events 1 • 30 days
0.00%
0/83 • 30 days
Respiratory, thoracic and mediastinal disorders
Left lower lung with residual pulmonary atelectasis
1.3%
1/79 • Number of events 1 • 30 days
0.00%
0/83 • 30 days
Eye disorders
Glaucoma
1.3%
1/79 • Number of events 1 • 30 days
0.00%
0/83 • 30 days
Respiratory, thoracic and mediastinal disorders
Chylopleura
1.3%
1/79 • Number of events 1 • 30 days
0.00%
0/83 • 30 days
Hepatobiliary disorders
Abnormal liver function
0.00%
0/79 • 30 days
1.2%
1/83 • Number of events 1 • 30 days
Respiratory, thoracic and mediastinal disorders
Asthma
0.00%
0/79 • 30 days
1.2%
1/83 • Number of events 1 • 30 days

Other adverse events

Other adverse events
Measure
Disposable Powered Articulating Endoscopic Linear Cutter Stapler
n=79 participants at risk
The Disposable Powered Articulating Endoscopic Linear Cutter Stapler is intended for transection, resection, and/or creation of anastomoses. The instrument has application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures. It can be used with staple line or tissue buttressing materials. The instrument may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen.
ECHELON Flex Powered Articulating Endoscopic Linear Cutters
n=83 participants at risk
ECHELON Flex Powered Articulating Endoscopic Linear Cutters has application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures.
Gastrointestinal disorders
constipation
13.9%
11/79 • Number of events 11 • 30 days
12.0%
10/83 • Number of events 10 • 30 days
Gastrointestinal disorders
emesis
8.9%
7/79 • Number of events 7 • 30 days
8.4%
7/83 • Number of events 7 • 30 days
Nervous system disorders
aypnia
10.1%
8/79 • Number of events 8 • 30 days
6.0%
5/83 • Number of events 6 • 30 days
General disorders
others
22.8%
18/79 • 30 days
38.6%
32/83 • 30 days

Additional Information

Jiao Liu

Fengh Medical Co., Ltd.

Phone: 13815132106

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place