Trial Outcomes & Findings for CP1150 Sound Processor Speech Perception Compared With the Next Generation of Signal Processing Technology (NCT NCT05286385)

NCT ID: NCT05286385

Last Updated: 2025-02-07

Results Overview

Paired difference in percentage CNC Words correct in quiet with the CP1150 Sound Processor and CP1150 Sound Processor with modified firmware.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

10 participants

Primary outcome timeframe

One day (in booth testing)

Results posted on

2025-02-07

Participant Flow

Participant milestones

Participant milestones
Measure
All Study Participants (CP1150 SP Compared With Next Generation Signal Processing Technology)
All study participants attended a single study visit. At the study visit, subjects underwent hearing assessments. Speech perception tests were carried out in all participants in a randomized manner with 4 interventions involving the CP1150 Sound Processor, CP1150 with modified firmware, CP1150 Sound Processor with Forward Focus and the CP1110 Sound Processor
Overall Study
STARTED
10
Overall Study
CP1150 Sound Processor
10
Overall Study
CP1150 Sound Processor With Modified Firmware
10
Overall Study
CP1150 Sound Processor With Forward Focus
10
Overall Study
CP1110 Sound Processor
10
Overall Study
COMPLETED
10
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Race and Ethnicity were not collected from any participant.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
All Study Participants (CP1150 SP Compared With Next Generation Signal Processing Technology)
n=10 Participants
All study participants attended a single study visit. At the study visit, subjects underwent hearing assessments. Speech perception tests were carried out in all study participants in a randomized manner with 4 interventions involving the CP1150 Sound Processor, CP1150 with modified firmware, CP1150 Sound Processor with Forward Focus and the CP1110 Sound Processor.
Age, Continuous
70.6 years
n=10 Participants
Sex: Female, Male
Female
5 Participants
n=10 Participants
Sex: Female, Male
Male
5 Participants
n=10 Participants
Region of Enrollment
Australia
10 participants
n=10 Participants
Age at hearing implant
59.2 years
STANDARD_DEVIATION 7.64 • n=10 Participants

PRIMARY outcome

Timeframe: One day (in booth testing)

Paired difference in percentage CNC Words correct in quiet with the CP1150 Sound Processor and CP1150 Sound Processor with modified firmware.

Outcome measures

Outcome measures
Measure
CP1150 NF
n=10 Participants
All subjects tested with CP1150 Sound Processor with modified firmware.
CP1150
n=10 Participants
All subjects test with CP1150
Speech Perception in Quiet Using an CP1150 Sound Processor
13.82 percentage of correct words
Standard Deviation 10
11.53 percentage of correct words
Standard Deviation 10

SECONDARY outcome

Timeframe: One day (in booth testing)

Paired difference in percentage CNC Words correct in quiet with the CP1150 Sound Processor and CP1150 Sound Processor with forward focus

Outcome measures

Outcome measures
Measure
CP1150 NF
n=10 Participants
All subjects tested with CP1150 Sound Processor with modified firmware.
CP1150
n=10 Participants
All subjects test with CP1150
Speech Perception in Quiet Using CP1150 Sound Processor Using Forward Focus Combined With Standard Microphone Directionality.
11.24 percentage of correct words
Standard Deviation 10
11.53 percentage of correct words
Standard Deviation 10

SECONDARY outcome

Timeframe: One day (in booth testing)

Paired difference in percentage CNC Words correct in quiet with the CP1150 Sound Processor and CP1110 Sound Processor

Outcome measures

Outcome measures
Measure
CP1150 NF
n=10 Participants
All subjects tested with CP1150 Sound Processor with modified firmware.
CP1150
n=10 Participants
All subjects test with CP1150
Speech Perception in Quiet With CP1150 and CP1110 Sound Processors
11.53 percentage of correct words
Standard Deviation 10
11.97 percentage of correct words
Standard Deviation 10

Adverse Events

Test Orders 1, 2, 3 & 4 Combined

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

PRS Specialist, Clinical Affairs

Cochlear Ltd.

Phone: +61294286555

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place