Trial Outcomes & Findings for SpO2 Accuracy of Noninvasive Disposable Pulse Oximeter Sensor With Motion Conditions (NCT NCT05264116)
NCT ID: NCT05264116
Last Updated: 2025-05-20
Results Overview
Percentage of SpO2 (oxygen saturation by pulse oximetry) measured by the Reprocessed Oximeter pulse oximetry sensors during motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. Accuracy root mean square (ARMS) between measured SpO2 and reference SaO2 (arterial oxygen saturation) must meet the 3% specification for each reprocessed pulse oximetry sensor style. Accuracy will be determined by comparing the noninvasive blood oxygen saturation measurement of the pulse oximeter to that obtained from a blood sample and calculating the arithmetic root mean square error (Arms) value. Standard deviation of the differences is computed as the precision. Square root of the sum of the squares of bias and precision is computed as the Arms Error value.
COMPLETED
NA
12 participants
1-2 hours
2025-05-20
Participant Flow
Participant milestones
| Measure |
Healthy Adult Participants
All participants are enrolled in the test group and receive the noninvasive adhesive reprocessed pulse oximeter sensors.
STRYKER SUSTAINABILITY SOLUTIONS PULSE OXIMETERS: Reprocessed noninvasive pulse oximeter sensors.
|
|---|---|
|
Overall Study
STARTED
|
12
|
|
Overall Study
COMPLETED
|
12
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
SpO2 Accuracy of Noninvasive Disposable Pulse Oximeter Sensor With Motion Conditions
Baseline characteristics by cohort
| Measure |
Healthy Adult Participants
n=12 Participants
All participants are enrolled in the test group and receive the noninvasive adhesive reprocessed pulse oximeter sensors.
STRYKER SUSTAINABILITY SOLUTIONS PULSE OXIMETERS: Reprocessed noninvasive pulse oximeter sensors.
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=5 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
12 Participants
n=5 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=5 Participants
|
|
Sex: Female, Male
Female
|
6 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
6 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Caucasian
|
5 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Black or African American
|
2 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Asian
|
2 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Hispanic
|
2 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Multiethnic
|
1 Participants
n=5 Participants
|
|
Region of Enrollment
United States
|
12 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 1-2 hoursPercentage of SpO2 (oxygen saturation by pulse oximetry) measured by the Reprocessed Oximeter pulse oximetry sensors during motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. Accuracy root mean square (ARMS) between measured SpO2 and reference SaO2 (arterial oxygen saturation) must meet the 3% specification for each reprocessed pulse oximetry sensor style. Accuracy will be determined by comparing the noninvasive blood oxygen saturation measurement of the pulse oximeter to that obtained from a blood sample and calculating the arithmetic root mean square error (Arms) value. Standard deviation of the differences is computed as the precision. Square root of the sum of the squares of bias and precision is computed as the Arms Error value.
Outcome measures
| Measure |
Device 1
n=12 Participants
All participants are enrolled in the test group and receive the noninvasive adhesive reprocessed pulse oximeter sensors. Device 1 consists of non-woven butterfly style tape.
STRYKER SUSTAINABILITY SOLUTIONS PULSE OXIMETERS: Reprocessed noninvasive pulse oximeter sensors.
|
Device 2
n=12 Participants
All participants are enrolled in the test group and receive the noninvasive adhesive reprocessed pulse oximeter sensors. Device 2 consists of a woven style tape.
STRYKER SUSTAINABILITY SOLUTIONS PULSE OXIMETERS: Reprocessed noninvasive pulse oximeter sensors.
|
|---|---|---|
|
Accuracy of Sensor by Arms Calculation Under Motion Conditions
|
2.80 Root Mean Square Error (%)
|
2.66 Root Mean Square Error (%)
|
Adverse Events
Device 1
Device 2
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place