Trial Outcomes & Findings for Early Feasibility Study Using IOPCL MAG to Improve Near Vision in ARMD Subjects (NCT NCT04984590)

NCT ID: NCT04984590

Last Updated: 2025-01-31

Results Overview

Measuring postoperative tilt of the PCIOL and IOPCL MAG complex. Tilt was defined as less than or equal to 10 degrees.

Recruitment status

COMPLETED

Target enrollment

9 participants

Primary outcome timeframe

12 Months

Results posted on

2025-01-31

Participant Flow

Unit of analysis: Eyes

Participant milestones

Participant milestones
Measure
IOPCL AMD MAG
The IOPCL MAG (intraocular pseudophakic capsular lens magnifier) consists of a 4.5 mm diameter optic. The central 1.8mm of the anterior surface has a power of 10.0 diopters. This IOPCL MAG will be implanted over an Alcon Model SN60WF or SA60AT intraocular lens to improve near vision.
Overall Study
STARTED
9 9
Overall Study
COMPLETED
9 9
Overall Study
NOT COMPLETED
0 0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Early Feasibility Study Using IOPCL MAG to Improve Near Vision in ARMD Subjects

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
IOPCL AMD MAG
n=9 Eyes
The IOPCL MAG (intraocular pseudophakic capsular lens magnifier) consists of a 4.5 mm diameter optic. The central 1.8mm of the anterior surface has a power of 10.0 diopters. This IOPCL MAG will be implanted over an Alcon Model SN60WF or SA60AT intraocular lens to improve near vision.
Age, Continuous
83.2 Years
STANDARD_DEVIATION 5.78 • n=5 Participants
Sex: Female, Male
Female
1 Participants
n=5 Participants
Sex: Female, Male
Male
8 Participants
n=5 Participants
Race/Ethnicity, Customized
Not of Hispanic, Latino/a, or Spanish origin
9 Participants
n=5 Participants
Race/Ethnicity, Customized
White
9 Participants
n=5 Participants
Region of Enrollment
United States
9 Participants
n=5 Participants

PRIMARY outcome

Timeframe: 12 Months

Measuring postoperative tilt of the PCIOL and IOPCL MAG complex. Tilt was defined as less than or equal to 10 degrees.

Outcome measures

Outcome measures
Measure
IOPCL AMD MAG
n=9 Eyes
The IOPCL MAG (intraocular pseudophakic capsular lens magnifier) consists of a 4.5 mm diameter optic. The central 1.8mm of the anterior surface has a power of 10.0 diopters. This IOPCL MAG will be implanted over an Alcon Model SN60WF or SA60AT intraocular lens to improve near vision.
Tilt of the PCIOL and IOPCL MAG Complex
PCIOL Tilt <= 10 degrees
9 Eyes
Tilt of the PCIOL and IOPCL MAG Complex
IOPCL MAG Tilt <=10 degrees
9 Eyes

PRIMARY outcome

Timeframe: 12 Months

Measures the decentration of the IOPCL MAG against the PCIOL. Decentration was defined as the horizontal and vertical distances greater than or equal to 0.25mm or less than or equal to 0.75mm.

Outcome measures

Outcome measures
Measure
IOPCL AMD MAG
n=9 Eyes
The IOPCL MAG (intraocular pseudophakic capsular lens magnifier) consists of a 4.5 mm diameter optic. The central 1.8mm of the anterior surface has a power of 10.0 diopters. This IOPCL MAG will be implanted over an Alcon Model SN60WF or SA60AT intraocular lens to improve near vision.
IOPCL MAG Decentration
Number of eyes with IOPCL MAG horizontal and vertical distances >= 0.25mm and <=0.75mm
9 Eyes
IOPCL MAG Decentration
Number of eyes with IOPCL MAG horizontal and vertical distances < 0.25mm and >0.75mm
0 Eyes

PRIMARY outcome

Timeframe: 12 Months

Proportion of eyes (≥ 75%) able to achieve improvement of 10 letters or more of uncorrected near visual acuity (at 14 cm) at 12 months from baseline

Outcome measures

Outcome measures
Measure
IOPCL AMD MAG
n=9 Eyes
The IOPCL MAG (intraocular pseudophakic capsular lens magnifier) consists of a 4.5 mm diameter optic. The central 1.8mm of the anterior surface has a power of 10.0 diopters. This IOPCL MAG will be implanted over an Alcon Model SN60WF or SA60AT intraocular lens to improve near vision.
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)
Gain of >= 20 letters (>= 4 lines)
8 Eyes
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)
Gain of >= 15 letters (>= 3 lines)
9 Eyes
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)
Gain of >= 10 letters (>= 2 lines)
9 Eyes
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)
Gain of >= 5 letters (>= 1 line)
9 Eyes
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)
No change
0 Eyes
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)
Loss of >= 5 letters (>= 1 line)
0 Eyes
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)
Loss of >=10 letters (>= 2 lines)
0 Eyes
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)
Loss of >= 15 letters (>=3 lines)
0 Eyes

SECONDARY outcome

Timeframe: 12 months

Proportion of eyes (≥ 50%) able to achieve improvement of 20 letters or more of uncorrected near visual acuity (at 14 cm) at 12 months from baseline

Outcome measures

Outcome measures
Measure
IOPCL AMD MAG
n=9 Eyes
The IOPCL MAG (intraocular pseudophakic capsular lens magnifier) consists of a 4.5 mm diameter optic. The central 1.8mm of the anterior surface has a power of 10.0 diopters. This IOPCL MAG will be implanted over an Alcon Model SN60WF or SA60AT intraocular lens to improve near vision.
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)
Loss of >= 10 letters (>= 2 lines)
0 Eyes
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)
Loss of >= 15 letters (>= 3 lines)
0 Eyes
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)
Gain of >= 20 letters (>= 4 lines)
8 Eyes
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)
Gain of >= 15 letters (>= 3 lines)
9 Eyes
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)
Gain of >= 10 letters (>= 2 lines)
9 Eyes
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)
Gain of >= 5 letters (>= 1 line)
9 Eyes
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)
No change
0 Eyes
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)
Loss of >= 5 letters (>= 1 line)
0 Eyes

SECONDARY outcome

Timeframe: 12 months

Proportion of eyes (≥ 75%) able to achieve improvement of 10 letters or more when comparing uncorrected near visual acuity (UCNVA) (at 14 cm) at 12 months from baseline BCNVA (at 30 cm, wearing the appropriate preoperative distance refractive correction in addition to +3.0D add)

Outcome measures

Outcome measures
Measure
IOPCL AMD MAG
n=9 Eyes
The IOPCL MAG (intraocular pseudophakic capsular lens magnifier) consists of a 4.5 mm diameter optic. The central 1.8mm of the anterior surface has a power of 10.0 diopters. This IOPCL MAG will be implanted over an Alcon Model SN60WF or SA60AT intraocular lens to improve near vision.
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)
Gain of >= 20 letters (>= 4 lines)
5 Eyes
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)
Gain of >= 15 letters (>= 3 lines)
8 Eyes
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)
Gain of >= 10 letters (>= 2 lines)
9 Eyes
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)
Loss of >= 5 letters (>= 1 line)
0 Eyes
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)
Loss of >= 10 letters (>= 2 lines)
0 Eyes
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)
Loss of >= 15 letters (>= 3 lines)
0 Eyes
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)
Gain of >= 5 letters (>= 1 line)
9 Eyes
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)
No change
0 Eyes

SECONDARY outcome

Timeframe: 12 months

Proportion of eyes (≥ 50%) able to achieve improvement of 20 letters or more when comparing uncorrected near visual acuity (UCNVA) (at 14 cm) at 12 months from baseline BCNVA (at 30 cm, wearing the appropriate preoperative distance refractive correction in addition to +3.0D add)

Outcome measures

Outcome measures
Measure
IOPCL AMD MAG
n=9 Eyes
The IOPCL MAG (intraocular pseudophakic capsular lens magnifier) consists of a 4.5 mm diameter optic. The central 1.8mm of the anterior surface has a power of 10.0 diopters. This IOPCL MAG will be implanted over an Alcon Model SN60WF or SA60AT intraocular lens to improve near vision.
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)
Gain of >= 20 letters (>= 4 lines)
5 Eyes
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)
Gain of >= 15 letters (>= 3 lines)
8 Eyes
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)
Loss of >= 15 letters (>= 3 lines)
0 Eyes
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)
Gain of >= 10 letters (>= 2 lines)
9 Eyes
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)
Gain of >= 5 letters (>= 1 line)
9 Eyes
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)
No change
0 Eyes
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)
Loss of >= 5 letters (>= 1 line)
0 Eyes
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)
Loss of >= 10 letters (>= 2 lines)
0 Eyes

SECONDARY outcome

Timeframe: 12 months

Proportion of eyes (≥ 75%) able to achieve improvement of 10 letters or more when comparing best corrected near visual acuity (BCNVA) (at 14 cm) at 12 months from baseline BCNVA (at 30 cm, wearing the appropriate preoperative distance refractive correction in addition to +3.0D add)

Outcome measures

Outcome measures
Measure
IOPCL AMD MAG
n=9 Eyes
The IOPCL MAG (intraocular pseudophakic capsular lens magnifier) consists of a 4.5 mm diameter optic. The central 1.8mm of the anterior surface has a power of 10.0 diopters. This IOPCL MAG will be implanted over an Alcon Model SN60WF or SA60AT intraocular lens to improve near vision.
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)
Gain of >= 20 letters (>= 4 lines)
5 Eyes
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)
Gain of >= 15 letters (>= 3 lines)
8 Eyes
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)
Gain of >= 10 letters (>= 2 lines)
9 Eyes
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)
Gain of >= 5 letters (>= 1 line)
9 Eyes
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)
No change
0 Eyes
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)
Loss of >= 5 letters (>= 1 line)
0 Eyes
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)
Loss of >= 10 letters (>= 2 lines)
0 Eyes
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)
Loss of >= 15 letters (>= 3 lines)
0 Eyes

SECONDARY outcome

Timeframe: 12 months

Proportion of eyes (≥ 50%) able to achieve improvement of 20 letters or more when comparing best corrected near visual acuity (BCNVA) (at 14 cm) at 12 months from baseline BCNVA (at 30 cm, wearing the appropriate preoperative distance refractive correction in addition to +3.0D add)

Outcome measures

Outcome measures
Measure
IOPCL AMD MAG
n=9 Eyes
The IOPCL MAG (intraocular pseudophakic capsular lens magnifier) consists of a 4.5 mm diameter optic. The central 1.8mm of the anterior surface has a power of 10.0 diopters. This IOPCL MAG will be implanted over an Alcon Model SN60WF or SA60AT intraocular lens to improve near vision.
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)
Gain of >= 20 letters (>= 4 lines)
5 Eyes
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)
Gain of >= 15 letters (>= 3 lines)
8 Eyes
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)
Gain of >= 10 letters (>= 2 lines)
9 Eyes
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)
Gain of >= 5 letters (>= 1 line)
9 Eyes
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)
No change
0 Eyes
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)
Loss of >= 5 letters (>= 1 line)
0 Eyes
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)
Loss of >= 10 letters (>= 2 lines)
0 Eyes
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)
Loss of >= 15 letters (>= 3 lines)
0 Eyes

Adverse Events

IOPCL AMD MAG

Serious events: 0 serious events
Other events: 1 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Other adverse events
Measure
IOPCL AMD MAG
n=9 participants at risk
The IOPCL MAG (intraocular pseudophakic capsular lens magnifier) consists of a 4.5 mm diameter optic. The central 1.8mm of the anterior surface has a power of 10.0 diopters. This IOPCL MAG will be implanted over an Alcon Model SN60WF or SA60AT intraocular lens to improve near vision.
Eye disorders
Bilateral pruritis
11.1%
1/9 • Number of events 1 • 12 Months

Additional Information

Head of Clinical

OnPoint Vision Inc.

Phone: 949-688-0660

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place