Trial Outcomes & Findings for Standard Versus Flexible Tip Bougie for Videolaryngoscopy (NCT NCT04973176)
NCT ID: NCT04973176
Last Updated: 2025-05-13
Results Overview
Composite score of difficulty of tracheal intubation using the bougie devices minimum score is zero, maximum possible score is 10, higher values represent worse outcome. Total score is sum of sub scales, values are numbers Total 6 sub scale from N1 to N6 ( N1: 0-2, N2: 0-1, N3:0-5, N4: 0-1, N5 0-1, N6 0-1, so total mIDS score is sum of N1 to N 6 sub scales)
COMPLETED
NA
160 participants
average 1 minute
2025-05-13
Participant Flow
Participant milestones
| Measure |
Endotracheal Intubation With C-MAC D Blade Videolaryngoscopy and Standard Bougie
Patients randomised to standard bougie will be intubated using standard bougie (Frova® airway intubation catheter )
Standard Bougie: Standard Bougie for endotracheal intubation
|
Endotracheal Intubation With C-MAC D Blade Videolaryngoscopy and Flexi-tip Bougie
Patients randomised to Flexi-tio bougie will be intubated using Flexi-tip bougie (P3 medical Ltd, Bristol, UK)
Flexi-tip Bougie (P3 Medical Bristol,UK): Flexi-tip Bougie for endotracheal intubation
|
|---|---|---|
|
Overall Study
STARTED
|
80
|
80
|
|
Overall Study
COMPLETED
|
80
|
80
|
|
Overall Study
NOT COMPLETED
|
0
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Race and Ethnicity were not collected from any participant.
Baseline characteristics by cohort
| Measure |
Endotracheal Intubation With C-MAC D Blade Videolaryngoscopy and Standard Bougie
n=80 Participants
Patients randomised to standard bougie will be intubated using standard bougie (Frova® airway intubation catheter )
Standard Bougie: Standard Bougie for endotracheal intubation
|
Endotracheal Intubation With C-MAC D Blade Videolaryngoscopy and Flexi-tip Bougie
n=80 Participants
Patients randomised to Flexi-tio bougie will be intubated using Flexi-tip bougie (P3 medical Ltd, Bristol, UK)
Flexi-tip Bougie (P3 Medical Bristol,UK): Flexi-tip Bougie for endotracheal intubation
|
Total
n=160 Participants
Total of all reporting groups
|
|---|---|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=80 Participants
|
0 Participants
n=80 Participants
|
0 Participants
n=160 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
69 Participants
n=80 Participants
|
70 Participants
n=80 Participants
|
139 Participants
n=160 Participants
|
|
Age, Categorical
>=65 years
|
11 Participants
n=80 Participants
|
10 Participants
n=80 Participants
|
21 Participants
n=160 Participants
|
|
Age, Continuous
|
55.3 year
STANDARD_DEVIATION 16.3 • n=80 Participants
|
55.2 year
STANDARD_DEVIATION 16.1 • n=80 Participants
|
55.2 year
STANDARD_DEVIATION 16.1 • n=160 Participants
|
|
Sex: Female, Male
Female
|
40 Participants
n=80 Participants
|
46 Participants
n=80 Participants
|
86 Participants
n=160 Participants
|
|
Sex: Female, Male
Male
|
40 Participants
n=80 Participants
|
34 Participants
n=80 Participants
|
74 Participants
n=160 Participants
|
|
Race and Ethnicity Not Collected
|
—
|
—
|
0 Participants
Race and Ethnicity were not collected from any participant.
|
|
Region of Enrollment
United Kingdom
|
80 participants
n=80 Participants
|
80 participants
n=80 Participants
|
160 participants
n=160 Participants
|
PRIMARY outcome
Timeframe: average 1 minuteComposite score of difficulty of tracheal intubation using the bougie devices minimum score is zero, maximum possible score is 10, higher values represent worse outcome. Total score is sum of sub scales, values are numbers Total 6 sub scale from N1 to N6 ( N1: 0-2, N2: 0-1, N3:0-5, N4: 0-1, N5 0-1, N6 0-1, so total mIDS score is sum of N1 to N 6 sub scales)
Outcome measures
| Measure |
Endotracheal Intubation With C-MAC D Blade Videolaryngoscopy and Standard Bougie
n=80 Participants
Patients randomised to standard bougie will be intubated using standard bougie (Frova® airway intubation catheter )
Standard Bougie: Standard Bougie for endotracheal intubation
|
Endotracheal Intubation With C-MAC D Blade Videolaryngoscopy and Flexi-tip Bougie
n=80 Participants
Patients randomised to Flexi-tio bougie will be intubated using Flexi-tip bougie (P3 medical Ltd, Bristol, UK)
Flexi-tip Bougie (P3 Medical Bristol,UK): Flexi-tip Bougie for endotracheal intubation
|
|---|---|---|
|
Modified Intubation Difficulty Scale Score (mIDS)
|
1 score on a scale range of 0 to 10 number
Interval 0.0 to 2.0
|
0 score on a scale range of 0 to 10 number
Interval 0.0 to 1.0
|
SECONDARY outcome
Timeframe: less than two minutesTime from when the videolaryngoscope is introduced into the oral cavity until the first capnography waveform is obtained.
Outcome measures
| Measure |
Endotracheal Intubation With C-MAC D Blade Videolaryngoscopy and Standard Bougie
n=80 Participants
Patients randomised to standard bougie will be intubated using standard bougie (Frova® airway intubation catheter )
Standard Bougie: Standard Bougie for endotracheal intubation
|
Endotracheal Intubation With C-MAC D Blade Videolaryngoscopy and Flexi-tip Bougie
n=80 Participants
Patients randomised to Flexi-tio bougie will be intubated using Flexi-tip bougie (P3 medical Ltd, Bristol, UK)
Flexi-tip Bougie (P3 Medical Bristol,UK): Flexi-tip Bougie for endotracheal intubation
|
|---|---|---|
|
Time to Successful Tracheal Intubation
|
42.9 seconds
Standard Deviation 9.1
|
41.0 seconds
Standard Deviation 6.1
|
SECONDARY outcome
Timeframe: less than one minuteTime from when the videolaryngoscope is introduced into the oral cavity to best view of the glottis (vocal cords).
Outcome measures
| Measure |
Endotracheal Intubation With C-MAC D Blade Videolaryngoscopy and Standard Bougie
n=80 Participants
Patients randomised to standard bougie will be intubated using standard bougie (Frova® airway intubation catheter )
Standard Bougie: Standard Bougie for endotracheal intubation
|
Endotracheal Intubation With C-MAC D Blade Videolaryngoscopy and Flexi-tip Bougie
n=80 Participants
Patients randomised to Flexi-tio bougie will be intubated using Flexi-tip bougie (P3 medical Ltd, Bristol, UK)
Flexi-tip Bougie (P3 Medical Bristol,UK): Flexi-tip Bougie for endotracheal intubation
|
|---|---|---|
|
Laryngoscopy Time
|
5 seconds
Standard Deviation 2
|
4.6 seconds
Standard Deviation 1.7
|
SECONDARY outcome
Timeframe: average 1 minuteThis is the percentage of patients being successfully intubated at the first attempt. This will be compared for two bougies
Outcome measures
| Measure |
Endotracheal Intubation With C-MAC D Blade Videolaryngoscopy and Standard Bougie
n=80 Participants
Patients randomised to standard bougie will be intubated using standard bougie (Frova® airway intubation catheter )
Standard Bougie: Standard Bougie for endotracheal intubation
|
Endotracheal Intubation With C-MAC D Blade Videolaryngoscopy and Flexi-tip Bougie
n=80 Participants
Patients randomised to Flexi-tio bougie will be intubated using Flexi-tip bougie (P3 medical Ltd, Bristol, UK)
Flexi-tip Bougie (P3 Medical Bristol,UK): Flexi-tip Bougie for endotracheal intubation
|
|---|---|---|
|
Overall First Attempt Success Rate
|
79 Participants
|
80 Participants
|
SECONDARY outcome
Timeframe: average 1 minutePopulation: this data was not collected and therefore not analysed
This is scored between 0 to 10, 0 being easy to use and 10 being extremely difficult to use
Outcome measures
Outcome data not reported
Adverse Events
Endotracheal Intubation With C-MAC D Blade Videolaryngoscopy and Standard Bougie
Endotracheal Intubation With C-MAC D Blade Videolaryngoscopy and Flexi-tip Bougie
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Dr Cyprian Mendonnca
University Hospitals Coventry and Warwickshire NHS Trust
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place