Trial Outcomes & Findings for Conduction System Pacing Optimized Therapy (NCT NCT04905290)

NCT ID: NCT04905290

Last Updated: 2025-04-15

Results Overview

Standard Deviation of Activation Times (SDAT) is a measurement of dyssynchrony, taken by the ECG belt. As described in the SDAT Acute Protocol (below) and intervention sections, percent change in SDAT from AV-only pacing was calculated for each patient-pacing configuration combination over the 5 AV delays. Within each patient the difference between CSPOT SDAT and CSP SDAT and the difference between CSPOT SDAT and BiV SDAT were taken. The mean of those differences across patients was calculated along with a 95% confidence interval.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

60 participants

Primary outcome timeframe

At implant during acute protocol

Results posted on

2025-04-15

Participant Flow

Participant milestones

Participant milestones
Measure
Single Arm
Patients will undergo 1) Conduction System Pacing Optimized Therapy (CSPOT) lead placement, then the 2) acute pacing protocol, and then 3) device implant. During the acute pacing protocol (step 2), all patients will undergo three types of pacing configurations and for each pacing configuration several delay settings, as defined in Assigned Interventions. For each intervention, defined as a unique combination of pacing configuration and delay setting, the protocol will alternate for several heartbeats between AV pacing and intervention (5 sections of AV pacing and 4 sections of intervention). For each set of 9, Left Ventricular dP/dt max will be calculated continuously, and Standard Deviation of Activation Time will be selected from one of the sections for atrial-only pacing and one of the intervention sections. Device implant (step 3) will be with either a CRT-D or CRT-P. Each patient's device will be programmed to CSPOT pacing. Patients will be followed for 6 months.
Overall Study
STARTED
60
Overall Study
Implant Attempt
55
Overall Study
COMPLETED
42
Overall Study
NOT COMPLETED
18

Reasons for withdrawal

Reasons for withdrawal
Measure
Single Arm
Patients will undergo 1) Conduction System Pacing Optimized Therapy (CSPOT) lead placement, then the 2) acute pacing protocol, and then 3) device implant. During the acute pacing protocol (step 2), all patients will undergo three types of pacing configurations and for each pacing configuration several delay settings, as defined in Assigned Interventions. For each intervention, defined as a unique combination of pacing configuration and delay setting, the protocol will alternate for several heartbeats between AV pacing and intervention (5 sections of AV pacing and 4 sections of intervention). For each set of 9, Left Ventricular dP/dt max will be calculated continuously, and Standard Deviation of Activation Time will be selected from one of the sections for atrial-only pacing and one of the intervention sections. Device implant (step 3) will be with either a CRT-D or CRT-P. Each patient's device will be programmed to CSPOT pacing. Patients will be followed for 6 months.
Overall Study
Adverse Event
4
Overall Study
Death
2
Overall Study
Withdrawal by Subject
4
Overall Study
Physician Decision
2
Overall Study
Lost to Follow-up
2
Overall Study
Screen Failure
2
Overall Study
Unsuccessful Procedure
2

Baseline Characteristics

Conduction System Pacing Optimized Therapy

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Single Arm
n=60 Participants
Patients will undergo 1) Conduction System Pacing Optimized Therapy (CSPOT) lead placement, then the 2) acute pacing protocol, and then 3) device implant. During the acute pacing protocol (step 2), all patients will undergo three types of pacing configurations and for each pacing configuration several delay settings, as defined in Assigned Interventions. For each intervention, defined as a unique combination of pacing configuration and delay setting, the protocol will alternate for several heartbeats between AV pacing and intervention (5 sections of AV pacing and 4 sections of intervention). For each set of 9, Left Ventricular dP/dt max will be calculated continuously, and Standard Deviation of Activation Time will be selected from one of the sections for atrial-only pacing and one of the intervention sections. Device implant (step 3) will be with either a CRT-D or CRT-P. Each patient's device will be programmed to CSPOT pacing. Patients will be followed for 6 months.
Age, Continuous
68.1 years
STANDARD_DEVIATION 11.1 • n=5 Participants
Sex: Female, Male
Female
24 Participants
n=5 Participants
Sex: Female, Male
Male
36 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
0 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
21 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
39 Participants
n=5 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
n=5 Participants
Race (NIH/OMB)
Asian
0 Participants
n=5 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
n=5 Participants
Race (NIH/OMB)
Black or African American
4 Participants
n=5 Participants
Race (NIH/OMB)
White
18 Participants
n=5 Participants
Race (NIH/OMB)
More than one race
0 Participants
n=5 Participants
Race (NIH/OMB)
Unknown or Not Reported
38 Participants
n=5 Participants
Height
171.1 cm
STANDARD_DEVIATION 10.5 • n=5 Participants
Weight
88.2 kg
STANDARD_DEVIATION 18.9 • n=5 Participants
Systolic Blood Pressure
121.9 mm Hg
STANDARD_DEVIATION 17.5 • n=5 Participants
Diastolic Blood Pressure
70.8 mm Hg
STANDARD_DEVIATION 10.4 • n=5 Participants
NYHA Classification
Class I
5 Participants
n=5 Participants
NYHA Classification
Class II
29 Participants
n=5 Participants
NYHA Classification
Class III
24 Participants
n=5 Participants
NYHA Classification
Class IV
1 Participants
n=5 Participants
NYHA Classification
Not Available
1 Participants
n=5 Participants

PRIMARY outcome

Timeframe: At implant during acute protocol

Population: Subjects with SDAT measurements available for each pacing configuration.

Standard Deviation of Activation Times (SDAT) is a measurement of dyssynchrony, taken by the ECG belt. As described in the SDAT Acute Protocol (below) and intervention sections, percent change in SDAT from AV-only pacing was calculated for each patient-pacing configuration combination over the 5 AV delays. Within each patient the difference between CSPOT SDAT and CSP SDAT and the difference between CSPOT SDAT and BiV SDAT were taken. The mean of those differences across patients was calculated along with a 95% confidence interval.

Outcome measures

Outcome measures
Measure
SDAT Acute Protocol
n=44 Participants
For each intervention, data from 9 sections of atrial only pacing and intervention will be used, beginning with the atrial-only pacing that immediately precedes the first intervention section and ending with the atrial-only pacing immediately following the last intervention pacing section. The SDAT measurement will be taken during one of the 4 intervention pacing sections and one of the 5 atrial-only pacing sections. This will provide 1 pair of SDAT measurements for each patient-intervention combination. For each pair, the percent change from the atrial-only section to the intervention section will be computed. Within each patient, the percent changes calculated from the 5 AV delay settings (default, default + 30ms, default + 60ms, default - 30ms, default - 60ms) will be used to fit a quadratic curve. The minimum value from the fitted quadratic curve within the delay settings will be used as the SDAT measurement for that patient-pacing configuration combination.
QRS Less Than or Equal to 171ms
Subjects who completed acute protocol with QRS width less than or equal to 171ms as measured by 12-lead ECG
Electrical Synchronization Response
CSPOT vs CSP
15.18 Difference in percent change in SDAT
Interval 4.26 to 26.09
Electrical Synchronization Response
CSPOT vs BiV
17.04 Difference in percent change in SDAT
Interval 4.32 to 29.76

PRIMARY outcome

Timeframe: At implant during acute protocol

Population: Subjects with LV dP/dt max measurements available for each pacing configuration

Left Ventricular (LV) dP/dt max, a measurement of the initial velocity of myocardial contraction. As described in the LV dP/dt max (below) and intervention sections, percent change in LV dP/dt max from AV-only pacing was calculated for each patient-pacing configuration combination over the 5 AV delays. Within each patient the difference between CSPOT SDAT and CSP SDAT and the difference between CSPOT SDAT and BiV SDAT were taken. The mean of those differences across patients was calculated along with a 95% confidence interval.

Outcome measures

Outcome measures
Measure
SDAT Acute Protocol
n=41 Participants
For each intervention, data from 9 sections of atrial only pacing and intervention will be used, beginning with the atrial-only pacing that immediately precedes the first intervention section and ending with the atrial-only pacing immediately following the last intervention pacing section. The SDAT measurement will be taken during one of the 4 intervention pacing sections and one of the 5 atrial-only pacing sections. This will provide 1 pair of SDAT measurements for each patient-intervention combination. For each pair, the percent change from the atrial-only section to the intervention section will be computed. Within each patient, the percent changes calculated from the 5 AV delay settings (default, default + 30ms, default + 60ms, default - 30ms, default - 60ms) will be used to fit a quadratic curve. The minimum value from the fitted quadratic curve within the delay settings will be used as the SDAT measurement for that patient-pacing configuration combination.
QRS Less Than or Equal to 171ms
Subjects who completed acute protocol with QRS width less than or equal to 171ms as measured by 12-lead ECG
Hemodynamic Response
CSPOT vs CSP
10.33 Difference in % change LV dP/dt max
Interval 7.15 to 13.51
Hemodynamic Response
CSPOT vs BiV
-1.31 Difference in % change LV dP/dt max
Interval -4.75 to 2.13

SECONDARY outcome

Timeframe: Baseline and 6 months

Population: Subjects with LVEF measurements at both baseline and 6-month follow-up visit

Left ventricular ejection fraction will be measured, based on echocardiography, at the baseline visit and again at the 6-month follow-up visit. Change will be calculated as the value observed at baseline subtracted from the value observed at 6 months.

Outcome measures

Outcome measures
Measure
SDAT Acute Protocol
n=35 Participants
For each intervention, data from 9 sections of atrial only pacing and intervention will be used, beginning with the atrial-only pacing that immediately precedes the first intervention section and ending with the atrial-only pacing immediately following the last intervention pacing section. The SDAT measurement will be taken during one of the 4 intervention pacing sections and one of the 5 atrial-only pacing sections. This will provide 1 pair of SDAT measurements for each patient-intervention combination. For each pair, the percent change from the atrial-only section to the intervention section will be computed. Within each patient, the percent changes calculated from the 5 AV delay settings (default, default + 30ms, default + 60ms, default - 30ms, default - 60ms) will be used to fit a quadratic curve. The minimum value from the fitted quadratic curve within the delay settings will be used as the SDAT measurement for that patient-pacing configuration combination.
QRS Less Than or Equal to 171ms
Subjects who completed acute protocol with QRS width less than or equal to 171ms as measured by 12-lead ECG
Left Ventricular Ejection Fraction (LVEF)
15.24 Change in percentage of LVEF
Interval 10.16 to 20.32

SECONDARY outcome

Timeframe: Baseline and 6 months

Population: Subjects who had LVESV recorded at both baseline and the 6-month follow-up visit

Left ventricular end systolic volume will be measured, based on echocardiography, at the baseline visit and again at the 6-month follow-up visit. The change will be calculated as difference between the 6-month and baseline values, divided by the baseline value.

Outcome measures

Outcome measures
Measure
SDAT Acute Protocol
n=35 Participants
For each intervention, data from 9 sections of atrial only pacing and intervention will be used, beginning with the atrial-only pacing that immediately precedes the first intervention section and ending with the atrial-only pacing immediately following the last intervention pacing section. The SDAT measurement will be taken during one of the 4 intervention pacing sections and one of the 5 atrial-only pacing sections. This will provide 1 pair of SDAT measurements for each patient-intervention combination. For each pair, the percent change from the atrial-only section to the intervention section will be computed. Within each patient, the percent changes calculated from the 5 AV delay settings (default, default + 30ms, default + 60ms, default - 30ms, default - 60ms) will be used to fit a quadratic curve. The minimum value from the fitted quadratic curve within the delay settings will be used as the SDAT measurement for that patient-pacing configuration combination.
QRS Less Than or Equal to 171ms
Subjects who completed acute protocol with QRS width less than or equal to 171ms as measured by 12-lead ECG
Left Ventricular End Systolic Volume (LVESV)
-43.28 percent change lvesv
Interval -51.51 to -35.06

SECONDARY outcome

Timeframe: 6 months

Population: Subjects who had a successful implant procedure.

The Clinical Composite Score (CCS) is a validated 3-level categorical variable that can take the values - Improved, Unchanged, or Worsened - at each follow-up visit. It is based on mortality, HF events, termination of device function, NYHA score, and patient global assessment. Briefly, the scoring system is as follows: * A patient is considered "worsened" if they die, demonstrate a worsened NYHA class, report at least moderately worsened heart-failure symptoms, or are hospitalized or permanently discontinue therapy because of or associated with worsening heart-failure * A patient is considered "improved" if they have not "worsened" and either demonstrate improvement in NYHA class or report at least moderately improved heart-failure symptoms * A patient is considered "stabilized" if they have not "worsened" or "improved" * A patient is considered "unavailable" if they did not complete 6-month visit and was not "worsened"

Outcome measures

Outcome measures
Measure
SDAT Acute Protocol
n=46 Participants
For each intervention, data from 9 sections of atrial only pacing and intervention will be used, beginning with the atrial-only pacing that immediately precedes the first intervention section and ending with the atrial-only pacing immediately following the last intervention pacing section. The SDAT measurement will be taken during one of the 4 intervention pacing sections and one of the 5 atrial-only pacing sections. This will provide 1 pair of SDAT measurements for each patient-intervention combination. For each pair, the percent change from the atrial-only section to the intervention section will be computed. Within each patient, the percent changes calculated from the 5 AV delay settings (default, default + 30ms, default + 60ms, default - 30ms, default - 60ms) will be used to fit a quadratic curve. The minimum value from the fitted quadratic curve within the delay settings will be used as the SDAT measurement for that patient-pacing configuration combination.
QRS Less Than or Equal to 171ms
Subjects who completed acute protocol with QRS width less than or equal to 171ms as measured by 12-lead ECG
Clinical Composite Score (CCS)
Improved
35 Participants
Clinical Composite Score (CCS)
Stabilized
2 Participants
Clinical Composite Score (CCS)
Worsened
6 Participants
Clinical Composite Score (CCS)
Unavailable
3 Participants

SECONDARY outcome

Timeframe: At Implant during acute protocol

Population: Subjects who had SDAT measured during the acute protocol.

Percent change in SDAT (as described in Primary Outcome 1: Arm/Group Description) by baseline QRS subgroup for each pacing configuration CSP, BiV, CSPOT during acute protocol at implant. The QRS subgroups were subjects with QRS width greater than 171ms and QRS width less than or equal to 171ms as measured by 12-lead ECG

Outcome measures

Outcome measures
Measure
SDAT Acute Protocol
n=24 Participants
For each intervention, data from 9 sections of atrial only pacing and intervention will be used, beginning with the atrial-only pacing that immediately precedes the first intervention section and ending with the atrial-only pacing immediately following the last intervention pacing section. The SDAT measurement will be taken during one of the 4 intervention pacing sections and one of the 5 atrial-only pacing sections. This will provide 1 pair of SDAT measurements for each patient-intervention combination. For each pair, the percent change from the atrial-only section to the intervention section will be computed. Within each patient, the percent changes calculated from the 5 AV delay settings (default, default + 30ms, default + 60ms, default - 30ms, default - 60ms) will be used to fit a quadratic curve. The minimum value from the fitted quadratic curve within the delay settings will be used as the SDAT measurement for that patient-pacing configuration combination.
QRS Less Than or Equal to 171ms
n=24 Participants
Subjects who completed acute protocol with QRS width less than or equal to 171ms as measured by 12-lead ECG
Absolute Percent Change in SDAT by QRS Subgroup
CSPOT
42.77 abs percent change in SDAT
Interval 30.21 to 55.32
42.46 abs percent change in SDAT
Interval 31.6 to 53.32
Absolute Percent Change in SDAT by QRS Subgroup
BIV
34.61 abs percent change in SDAT
Interval 23.94 to 45.28
19.72 abs percent change in SDAT
Interval -7.96 to 47.41
Absolute Percent Change in SDAT by QRS Subgroup
CSP
34.26 abs percent change in SDAT
Interval 26.1 to 42.42
22.05 abs percent change in SDAT
Interval -2.18 to 46.28

SECONDARY outcome

Timeframe: At Implant during acute protocol

Population: Subjects who had SDAT measured during the acute protocol.

Percent change in SDAT (as described in Primary Outcome 1: Arm/Group Description) split by subjects with pure Left Bundle Branch Block (LBBB) and with other Conduction Disorders for each pacing configuration CSP, BiV and CSPOT during acute protocol.

Outcome measures

Outcome measures
Measure
SDAT Acute Protocol
n=19 Participants
For each intervention, data from 9 sections of atrial only pacing and intervention will be used, beginning with the atrial-only pacing that immediately precedes the first intervention section and ending with the atrial-only pacing immediately following the last intervention pacing section. The SDAT measurement will be taken during one of the 4 intervention pacing sections and one of the 5 atrial-only pacing sections. This will provide 1 pair of SDAT measurements for each patient-intervention combination. For each pair, the percent change from the atrial-only section to the intervention section will be computed. Within each patient, the percent changes calculated from the 5 AV delay settings (default, default + 30ms, default + 60ms, default - 30ms, default - 60ms) will be used to fit a quadratic curve. The minimum value from the fitted quadratic curve within the delay settings will be used as the SDAT measurement for that patient-pacing configuration combination.
QRS Less Than or Equal to 171ms
n=29 Participants
Subjects who completed acute protocol with QRS width less than or equal to 171ms as measured by 12-lead ECG
Absolute Percent Change in SDAT Split by Subjects With Pure LBBB and With Other Conduction Disorders
CSPOT
51.47 abs percent change in SDAT
Interval 41.44 to 61.5
38.01 abs percent change in SDAT
Interval 27.22 to 48.79
Absolute Percent Change in SDAT Split by Subjects With Pure LBBB and With Other Conduction Disorders
BIV
36.28 abs percent change in SDAT
Interval 28.36 to 44.2
21.31 abs percent change in SDAT
Interval -2.24 to 44.87
Absolute Percent Change in SDAT Split by Subjects With Pure LBBB and With Other Conduction Disorders
CSP
40.18 abs percent change in SDAT
Interval 33.91 to 46.44
20.28 abs percent change in SDAT
Interval 0.14 to 40.42

SECONDARY outcome

Timeframe: At Implant during acute protocol

Population: Subjects who had SDAT measured during the acute protocol.

Absolute Percent change in SDAT (as described in Primary Outcome 1: Arm/Group Description) split by subjects with Ischemic Cardiomyopathy and those with Non-Ischemic Cardiomyopathy for each pacing configuration CSP, BiV and CSPOT during acute protocol.

Outcome measures

Outcome measures
Measure
SDAT Acute Protocol
n=34 Participants
For each intervention, data from 9 sections of atrial only pacing and intervention will be used, beginning with the atrial-only pacing that immediately precedes the first intervention section and ending with the atrial-only pacing immediately following the last intervention pacing section. The SDAT measurement will be taken during one of the 4 intervention pacing sections and one of the 5 atrial-only pacing sections. This will provide 1 pair of SDAT measurements for each patient-intervention combination. For each pair, the percent change from the atrial-only section to the intervention section will be computed. Within each patient, the percent changes calculated from the 5 AV delay settings (default, default + 30ms, default + 60ms, default - 30ms, default - 60ms) will be used to fit a quadratic curve. The minimum value from the fitted quadratic curve within the delay settings will be used as the SDAT measurement for that patient-pacing configuration combination.
QRS Less Than or Equal to 171ms
n=14 Participants
Subjects who completed acute protocol with QRS width less than or equal to 171ms as measured by 12-lead ECG
Absolute Percent Change in SDAT Split by Subjects With Ischemic Cardiomyopathy and Those With Non-Ischemic Cardiomyopathy
CSPOT
48.96 abs percent change in SDAT
Interval 41.32 to 56.61
28.95 abs percent change in SDAT
Interval 10.75 to 47.15
Absolute Percent Change in SDAT Split by Subjects With Ischemic Cardiomyopathy and Those With Non-Ischemic Cardiomyopathy
BIV
36.95 abs percent change in SDAT
Interval 29.71 to 44.2
3.74 abs percent change in SDAT
Interval -44.37 to 51.85
Absolute Percent Change in SDAT Split by Subjects With Ischemic Cardiomyopathy and Those With Non-Ischemic Cardiomyopathy
CSP
38.83 abs percent change in SDAT
Interval 32.17 to 45.49
2.23 abs percent change in SDAT
Interval -37.24 to 41.71

SECONDARY outcome

Timeframe: At Implant during acute protocol

Population: Subjects who had LV dP/dt max measured during acute protocol

Percent change in LV dP/dt max (as described in Primary Outcome 2: Arm/Group Description)split by subjects with Ischemic Cardiomyopathy and those with Non-Ischemic Cardiomyopathy for each pacing configuration CSP, BiV and CSPOT during acute protocol.

Outcome measures

Outcome measures
Measure
SDAT Acute Protocol
n=31 Participants
For each intervention, data from 9 sections of atrial only pacing and intervention will be used, beginning with the atrial-only pacing that immediately precedes the first intervention section and ending with the atrial-only pacing immediately following the last intervention pacing section. The SDAT measurement will be taken during one of the 4 intervention pacing sections and one of the 5 atrial-only pacing sections. This will provide 1 pair of SDAT measurements for each patient-intervention combination. For each pair, the percent change from the atrial-only section to the intervention section will be computed. Within each patient, the percent changes calculated from the 5 AV delay settings (default, default + 30ms, default + 60ms, default - 30ms, default - 60ms) will be used to fit a quadratic curve. The minimum value from the fitted quadratic curve within the delay settings will be used as the SDAT measurement for that patient-pacing configuration combination.
QRS Less Than or Equal to 171ms
n=14 Participants
Subjects who completed acute protocol with QRS width less than or equal to 171ms as measured by 12-lead ECG
Percent Change in LV dP/dt Max Split by Subjects With Ischemic Cardiomyopathy and Those With Non-Ischemic Cardiomyopathy
CSPOT
30.05 percent change in LV dP/dt max
Interval 23.81 to 36.29
17.56 percent change in LV dP/dt max
Interval 10.82 to 24.31
Percent Change in LV dP/dt Max Split by Subjects With Ischemic Cardiomyopathy and Those With Non-Ischemic Cardiomyopathy
BIV
30.64 percent change in LV dP/dt max
Interval 22.15 to 39.12
17.62 percent change in LV dP/dt max
Interval 11.56 to 23.69
Percent Change in LV dP/dt Max Split by Subjects With Ischemic Cardiomyopathy and Those With Non-Ischemic Cardiomyopathy
CSP
20.03 percent change in LV dP/dt max
Interval 15.96 to 24.1
8.32 percent change in LV dP/dt max
Interval 2.13 to 14.51

SECONDARY outcome

Timeframe: At Implant during acute protocol

Population: Subjects who had LV dP/dt max measured during acute protocol

Percent change in LV dP/dt max (as described in Primary Outcome 2: Arm/Group Description) split by subjects with pure Left Bundle Branch Block (LBBB) and with other Conduction Disorders (NIVCD) for each pacing configuration CSP, BiV and CSPOT during acute protocol

Outcome measures

Outcome measures
Measure
SDAT Acute Protocol
n=18 Participants
For each intervention, data from 9 sections of atrial only pacing and intervention will be used, beginning with the atrial-only pacing that immediately precedes the first intervention section and ending with the atrial-only pacing immediately following the last intervention pacing section. The SDAT measurement will be taken during one of the 4 intervention pacing sections and one of the 5 atrial-only pacing sections. This will provide 1 pair of SDAT measurements for each patient-intervention combination. For each pair, the percent change from the atrial-only section to the intervention section will be computed. Within each patient, the percent changes calculated from the 5 AV delay settings (default, default + 30ms, default + 60ms, default - 30ms, default - 60ms) will be used to fit a quadratic curve. The minimum value from the fitted quadratic curve within the delay settings will be used as the SDAT measurement for that patient-pacing configuration combination.
QRS Less Than or Equal to 171ms
n=27 Participants
Subjects who completed acute protocol with QRS width less than or equal to 171ms as measured by 12-lead ECG
Percent Change in LV dP/dt Max Split by Subjects With Pure Left Bundle Branch Block (LBBB) and With Other Conduction Disorders (NIVCD)
CSPOT
28.08 percent change in LV dP/dt max
Interval 20.4 to 35.76
24.59 percent change in LV dP/dt max
Interval 17.98 to 31.21
Percent Change in LV dP/dt Max Split by Subjects With Pure Left Bundle Branch Block (LBBB) and With Other Conduction Disorders (NIVCD)
BIV
29.35 percent change in LV dP/dt max
Interval 17.73 to 40.98
24.65 percent change in LV dP/dt max
Interval 17.05 to 32.25
Percent Change in LV dP/dt Max Split by Subjects With Pure Left Bundle Branch Block (LBBB) and With Other Conduction Disorders (NIVCD)
CSP
18.02 percent change in LV dP/dt max
Interval 12.65 to 23.39
15.29 percent change in LV dP/dt max
Interval 10.12 to 20.47

SECONDARY outcome

Timeframe: At Implant during acute protocol

Population: Subjects who had LV dP/dt max measured during acute protocol

Percent change in LV dP/dt max (as described in Primary Outcome 2: Arm/Group Description) by QRS subgroup for each pacing configuration CSP, BiV and CSPOT during acute protocol. The QRS subgroups were subjects with QRS width greater than 171ms and QRS width less than or equal to 171ms as measured by 12-lead ECG

Outcome measures

Outcome measures
Measure
SDAT Acute Protocol
n=23 Participants
For each intervention, data from 9 sections of atrial only pacing and intervention will be used, beginning with the atrial-only pacing that immediately precedes the first intervention section and ending with the atrial-only pacing immediately following the last intervention pacing section. The SDAT measurement will be taken during one of the 4 intervention pacing sections and one of the 5 atrial-only pacing sections. This will provide 1 pair of SDAT measurements for each patient-intervention combination. For each pair, the percent change from the atrial-only section to the intervention section will be computed. Within each patient, the percent changes calculated from the 5 AV delay settings (default, default + 30ms, default + 60ms, default - 30ms, default - 60ms) will be used to fit a quadratic curve. The minimum value from the fitted quadratic curve within the delay settings will be used as the SDAT measurement for that patient-pacing configuration combination.
QRS Less Than or Equal to 171ms
n=22 Participants
Subjects who completed acute protocol with QRS width less than or equal to 171ms as measured by 12-lead ECG
Improvement in LV dP/dt Max by QRS Subgroup
CSPOT
29.28 percent change in LV dP/dt max
Interval 21.32 to 37.24
22.77 percent change in LV dP/dt max
Interval 16.39 to 29.14
Improvement in LV dP/dt Max by QRS Subgroup
BIV
32.18 percent change in LV dP/dt max
Interval 20.68 to 43.68
20.88 percent change in LV dP/dt max
Interval 15.74 to 26.02
Improvement in LV dP/dt Max by QRS Subgroup
CSP
18.2 percent change in LV dP/dt max
Interval 12.88 to 23.51
14.49 percent change in LV dP/dt max
Interval 9.16 to 19.82

Adverse Events

Single Arm

Serious events: 20 serious events
Other events: 14 other events
Deaths: 2 deaths

Serious adverse events

Serious adverse events
Measure
Single Arm
n=60 participants at risk
Patients will undergo 1) Conduction System Pacing Optimized Therapy (CSPOT) lead placement, then the 2) acute pacing protocol, and then 3) device implant. During the acute pacing protocol (step 2), all patients will undergo three types of pacing configurations and for each pacing configuration several delay settings, as defined in Assigned Interventions. For each intervention, defined as a unique combination of pacing configuration and delay setting, the protocol will alternate for several heartbeats between AV pacing and intervention (5 sections of AV pacing and 4 sections of intervention). For each set of 9, Left Ventricular dP/dt max will be calculated continuously, and Standard Deviation of Activation Time will be selected from one of the sections for atrial-only pacing and one of the intervention sections. Device implant (step 3) will be with either a CRT-D or CRT-P. Each patient's device will be programmed to CSPOT pacing. Patients will be followed for 6 months.
Blood and lymphatic system disorders
Thrombocytopenia
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Cardiac disorders
Acute myocardial infarction
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Cardiac disorders
Arrhythmia
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Cardiac disorders
Atrioventricular block complete
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Cardiac disorders
Cardiac perforation
3.3%
2/60 • Number of events 2 • From enrollment through 6 month follow-up
Cardiac disorders
Pericarditis
3.3%
2/60 • Number of events 2 • From enrollment through 6 month follow-up
Cardiac disorders
Supraventricular tachycardia
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Cardiac disorders
Ventricular fibrillation
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
General disorders
Adverse drug reaction
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Infections and infestations
Endocarditis
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Injury, poisoning and procedural complications
Vascular pseudoaneurysm
3.3%
2/60 • Number of events 2 • From enrollment through 6 month follow-up
Metabolism and nutrition disorders
Failure to thrive
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Neoplasms benign, malignant and unspecified (incl cysts and polyps)
Myelodysplastic syndrome
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Nervous system disorders
Transient ischaemic attack
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Product Issues
Device fastener issue
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Product Issues
Device lead issue
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Renal and urinary disorders
Acute kidney injury
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Respiratory, thoracic and mediastinal disorders
Dyspnoea
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up

Other adverse events

Other adverse events
Measure
Single Arm
n=60 participants at risk
Patients will undergo 1) Conduction System Pacing Optimized Therapy (CSPOT) lead placement, then the 2) acute pacing protocol, and then 3) device implant. During the acute pacing protocol (step 2), all patients will undergo three types of pacing configurations and for each pacing configuration several delay settings, as defined in Assigned Interventions. For each intervention, defined as a unique combination of pacing configuration and delay setting, the protocol will alternate for several heartbeats between AV pacing and intervention (5 sections of AV pacing and 4 sections of intervention). For each set of 9, Left Ventricular dP/dt max will be calculated continuously, and Standard Deviation of Activation Time will be selected from one of the sections for atrial-only pacing and one of the intervention sections. Device implant (step 3) will be with either a CRT-D or CRT-P. Each patient's device will be programmed to CSPOT pacing. Patients will be followed for 6 months.
Cardiac disorders
Atrial fibrillation
8.3%
5/60 • Number of events 5 • From enrollment through 6 month follow-up
Cardiac disorders
Atrioventricular block complete
5.0%
3/60 • Number of events 4 • From enrollment through 6 month follow-up
Cardiac disorders
Atrioventricular block second degree
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Cardiac disorders
Cardiac failure
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Cardiac disorders
Ventricular tachycardia
3.3%
2/60 • Number of events 2 • From enrollment through 6 month follow-up
Endocrine disorders
Adrenal mass
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Gastrointestinal disorders
Gastrointestinal disorder
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
General disorders
Chest pain
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
General disorders
Implant site haematoma
3.3%
2/60 • Number of events 2 • From enrollment through 6 month follow-up
General disorders
Non-cardiac chest pain
3.3%
2/60 • Number of events 2 • From enrollment through 6 month follow-up
General disorders
Peripheral swelling
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Injury, poisoning and procedural complications
Device placement issue
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Injury, poisoning and procedural complications
Vascular access site swelling
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Nervous system disorders
Headache
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Product Issues
Device stimulation issue
3.3%
2/60 • Number of events 2 • From enrollment through 6 month follow-up
Product Issues
Lead dislodgement
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Product Issues
Oversensing
3.3%
2/60 • Number of events 2 • From enrollment through 6 month follow-up
Vascular disorders
Haematoma
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Vascular disorders
Hypertension
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up
Vascular disorders
Venous thrombosis limb
1.7%
1/60 • Number of events 1 • From enrollment through 6 month follow-up

Additional Information

James Burrell, Senior Statistician

Medtronic

Phone: 18006338766

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place