Trial Outcomes & Findings for Electroconvulsive Therapy Amplitude Titration (NCT NCT04621786)
NCT ID: NCT04621786
Last Updated: 2024-01-11
Results Overview
Electric field modeling is used to calculate Ebrain from the pre-ECT structural MRI. Ebrain is an individual's electric field strength (Volts/meter) per unit current (milliampere). To avoid confounds from the tissue boundary effects, Ebrain is calculated as the 90th percentile of maximal. Higher values (0.2 Volts/meter per milliampere) indicate that an individual receives a higher electric field strength per unit current. The Ebrain in this sample ranges from 0.1 to 0.19 Volts/meter per milliampere. Linear regression model assessed the relationship between amplitude-determined seizure (independent variable) and Ebrain (dependent variable) with beta coefficient and 95% confidence intervals.
COMPLETED
NA
41 participants
First treatment
2024-01-11
Participant Flow
Participant milestones
| Measure |
Experimental Arm
All subjects enrolled will receive amplitude titration for their first treatment. The remainder of the ECT series will be completed with traditional (800mA) pulse amplitude with right unilateral electrode placement. This investigation only includes the single open-label arm.
Mecta Spectrum 5000Q paired with Soterix Medical 4X1 HD - ECT Multi-Channel Stimulation Interface: The Mecta Spectrum 5000Q paired with Soterix Medical 4X1 HD-ECT Multi-Channel Stimulation Interface will reduce ECT current amplitude for amplitude-seizure titration.
|
|---|---|
|
Overall Study
STARTED
|
41
|
|
Overall Study
COMPLETED
|
29
|
|
Overall Study
NOT COMPLETED
|
12
|
Reasons for withdrawal
| Measure |
Experimental Arm
All subjects enrolled will receive amplitude titration for their first treatment. The remainder of the ECT series will be completed with traditional (800mA) pulse amplitude with right unilateral electrode placement. This investigation only includes the single open-label arm.
Mecta Spectrum 5000Q paired with Soterix Medical 4X1 HD - ECT Multi-Channel Stimulation Interface: The Mecta Spectrum 5000Q paired with Soterix Medical 4X1 HD-ECT Multi-Channel Stimulation Interface will reduce ECT current amplitude for amplitude-seizure titration.
|
|---|---|
|
Overall Study
Lost to Follow-up
|
9
|
|
Overall Study
Protocol Violation
|
3
|
Baseline Characteristics
Electroconvulsive Therapy Amplitude Titration
Baseline characteristics by cohort
| Measure |
Experimental Arm
n=29 Participants
All subjects enrolled will receive amplitude titration for their first treatment. The remainder of the ECT series will be completed with traditional (800mA) pulse amplitude with right unilateral electrode placement. This investigation only includes the single open-label arm.
Mecta Spectrum 5000Q paired with Soterix Medical 4X1 HD - ECT Multi-Channel Stimulation Interface: The Mecta Spectrum 5000Q paired with Soterix Medical 4X1 HD-ECT Multi-Channel Stimulation Interface will reduce ECT current amplitude for amplitude-seizure titration.
|
|---|---|
|
Race (NIH/OMB)
Asian
|
1 Participants
n=5 Participants
|
|
Age, Continuous
|
64.2 years
STANDARD_DEVIATION 7.9 • n=5 Participants
|
|
Sex: Female, Male
Female
|
17 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
12 Participants
n=5 Participants
|
|
Race (NIH/OMB)
American Indian or Alaska Native
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Black or African American
|
1 Participants
n=5 Participants
|
|
Race (NIH/OMB)
White
|
27 Participants
n=5 Participants
|
|
Race (NIH/OMB)
More than one race
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Unknown or Not Reported
|
0 Participants
n=5 Participants
|
|
Region of Enrollment
United States
|
29 participants
n=5 Participants
|
|
Amplitude determined seizure
|
312 milliamperes
STANDARD_DEVIATION 113 • n=5 Participants
|
PRIMARY outcome
Timeframe: First treatmentPopulation: Amplitude determined seizure (miliamperes) predicting Ebrain (Volts/meter per milliampere)
Electric field modeling is used to calculate Ebrain from the pre-ECT structural MRI. Ebrain is an individual's electric field strength (Volts/meter) per unit current (milliampere). To avoid confounds from the tissue boundary effects, Ebrain is calculated as the 90th percentile of maximal. Higher values (0.2 Volts/meter per milliampere) indicate that an individual receives a higher electric field strength per unit current. The Ebrain in this sample ranges from 0.1 to 0.19 Volts/meter per milliampere. Linear regression model assessed the relationship between amplitude-determined seizure (independent variable) and Ebrain (dependent variable) with beta coefficient and 95% confidence intervals.
Outcome measures
| Measure |
Experimental Arm
n=29 Participants
All subjects enrolled will receive amplitude titration for their first treatment. The remainder of the ECT series will be completed with traditional (800mA) pulse amplitude with right unilateral electrode placement. This investigation only includes the single open-label arm.
Mecta Spectrum 5000Q paired with Soterix Medical 4X1 HD - ECT Multi-Channel Stimulation Interface: The Mecta Spectrum 5000Q paired with Soterix Medical 4X1 HD-ECT Multi-Channel Stimulation Interface will reduce ECT current amplitude for amplitude-seizure titration.
|
|---|---|
|
Beta Coefficient From Linear Regression of Amplitude-determined Seizure (Independent Variable) and Ebrain (Dependent Variable)
|
-.0001054 beta coefficient from linear regression
Interval -0.0001767 to -0.0000341
|
Adverse Events
Experimental Arm
Serious adverse events
| Measure |
Experimental Arm
n=29 participants at risk
All subjects enrolled will receive amplitude titration for their first treatment. The remainder of the ECT series will be completed with traditional (800mA) pulse amplitude with right unilateral electrode placement. This investigation only includes the single open-label arm.
Mecta Spectrum 5000Q paired with Soterix Medical 4X1 HD - ECT Multi-Channel Stimulation Interface: The Mecta Spectrum 5000Q paired with Soterix Medical 4X1 HD-ECT Multi-Channel Stimulation Interface will reduce ECT current amplitude for amplitude-seizure titration.
|
|---|---|
|
Psychiatric disorders
Attempted self-harm
|
3.4%
1/29 • Number of events 1 • 3 months
|
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place