EARSATS-19: In-ear Measurement of Blood Oxygen Saturation in COVID-19 Follow up
NCT ID: NCT04529408
Last Updated: 2023-10-17
Study Results
The study team has not published outcome measurements, participant flow, or safety data for this trial yet. Check back later for updates.
Basic Information
Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.
COMPLETED
60 participants
OBSERVATIONAL
2020-08-26
2022-08-10
Brief Summary
Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.
DESIGN Non-inferiority study
AIMS To evaluate qualitative and quantitative performance of in-ear SpO2 monitoring against the gold standard right finger-clip pulse oximeter -- towards validation for use in COVID-19 in the acute ambulatory and long-term monitoring setting
OUTCOME MEASURES
In-ear SpO2 compared with gold-standard finger-clip pulse oximeter:
Correlation between SpO2 measurements at rest Correlation between SpO2 measurements during 6 minute walk test Signal quality during 6 minute walk test Qualitative evaluation of clinical and patient user acceptability using questionnaires
POPULATION 30 patients attending COVID-19 follow-up clinic and 30 patients with chronic lung disease attending routine outpatient investigations
ELIGIBILITY Aged 18 and above, no upper age limit Able to give informed consent No abnormal ear anatomy.
DURATION 12 months
Related Clinical Trials
Explore similar clinical trials based on study characteristics and research focus.
Feasibility of Remote Evaluation and Monitoring of Acoustic Pathophysiological Signals With External Sensor Technology in Covid-19
NCT04695821
COVID-19 Surveillance Based on Smart Wearable Device
NCT04459637
COntinuous Signs Monitoring In Covid-19 Patients
NCT04581031
imPulse⢠Una Full-spectrum, Over Clothing E-stethoscope
NCT04556149
Prospective Study to Validate the Clinical Accuracy of the Norbert Device to Measure Oxygen Saturation
NCT05685654
Detailed Description
Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.
The COVID-19 pandemic places substantial pressure on clinical resources with ambulatory patients who were often discharged home to be cared for on virtual wards with daily telephone consults. Now the initial surge has settled, we are beginning to understand the challenge of following up patients who have had COVID-19 at volumes that mean doing this in a timely fashion is not possible (BTS Guidance 6 weeks).
There is mounting concern that the huge number of COVID-19 survivors will be at increased risk of chronic lung diseases such as pulmonary fibrosis, where early capture of patterns of low SpO2 could unearth desperately needed insights for preventing disease progression. The inconveniences, impracticalities and inaccuracies of conventional measurement of SpO2 (the percentage unit of blood oxygen) using a conventional finger probe for discrete readings are well established. A user-friendly, remote continuous SpO2 monitoring tool would therefore allow the rapid identification of any deterioration post discharge and also track longer-term recovery, allowing the accurate triage of patients to limited follow up resources.
Hearables and In-Ear Photoplethysmography
Reflective photoplethysmography (PPG), which uses a changes in the ratio between red and infrared light to estimate SpO2, can be measured from any site with skin that has vasculature, but for SpO2 measurement has conventionally been from fingers or sometimes the earlobe. The recent interest in the development of Hearables9 has promoted the ear canal as a preferred site for the measurement of vital signs in digital health technology. Indeed, the ear canal, which acts as a shield from external electrical noise (resembling a nature-built Faraday cage), represents a unique opportunity for physiological measurements, due to its general fixed position relative to the heart (ear-electrocardiogram) and proximity to the brain (ear-electroencephalogram). Relative to limbs, the head moves less in daily life and this is one property that has helped establish the feasibility of wearables to detect Ear-ECG and Ear-EEG.
Unlike the finger and ear lobe PPG signal, in-ear PPG has shown that the ear canal, owing to being a narrow cavity with consistent blood supply, offers a signal which is stable and resistant to changes in blood volume occurring during hypothermia . This attribute has established the ear canal as a preferred site for accurate core body temperature measurement. In-ear PPG has also been shown to be far more sensitive than earlobe and finger PPG to amplitude variations that arise from respiration, thus allowing for a better measurement of respiratory rate. Furthermore, a significant delay has been evidenced between earlobe pulse oximetry and pulse oximetry on the hand or the foot for detection of hypoxemia (low levels of blood oxygen)..
Previous work led by Professor Mandic's group in the Faculty of Engineering showed that in a study of 14 healthy volunteers in-ear PPG compared with the current standard of care - the finger clip -- the in-ear sensor shows non-inferiority and potential superiority with regard to detection of a mean blood oxygen delay (the time it took from detecting minimal blood oxygen change in the ear to detecting minimal blood oxygen in the finger) reduction of 12.4 seconds, thus being faster at identifying acute desaturation.
RATIONALE FOR CURRENT STUDY
This project sets out to establish the feasibility of SpO2 measurement from the ear canal as a convenient site for monitoring in the post COVID-19 follow up patient population, and perform a comprehensive comparison with the right index finger - the conventional clinical measurement site. The initial phase of the study will start with a total of n = 60 patients; 30 patients, who will be attending hospital as part of their routine clinical post COVID-19 follow up, which already includes SpO2 monitoring and a six minute walk test (6MWT). This study will also recruit a further 30 patients with chronic lung disease also attending for these routine outpatient investigations.
The continuous monitoring and rapid detection of desaturation can provide earlier warning of deterioration or detailed tracking of an improving trajectory. Current alternative monitoring strategies are reactive and all revolve around the classical finger or earlobe clip, which are less sensitive and present a number of challenges for continual monitoring. Living daily life with a finger clip, or even an earlobe clip, is problematic with much higher signal noise from movement artefacts with the finger clip compared to the ear bud placed within a stable head position.
From a patient perspective, the ear bud form factor that will be used in this study is familiar to them, being similar to an in-ear headphone, and will be much less intrusive to daily activities than a finger clip. Therefore the potential for continuous monitoring during the active day and the night time is much more user-friendly, and vital for the younger and potentially more active COVID-19 patients. Continuous monitoring would allow the development of protocolised management pathways to provide increasing levels of supported self-care to patients, freeing up vital resources during future pandemics.
The data generated in this pilot feasibility study would allow scaling to prepare for larger research projects in any subsequent surge discharge and follow up. The next stage would move to outpatient monitoring, with ambulatory patients going home with the device and the incorporation of further COVID patient populations such as inpatients being monitored for deterioration, where these devices could provide accurate continuous monitoring of SpO2, detecting any deterioration before they develop breathlessness and their condition becomes life threatening. This project therefore directly complements the NHS' strategic priority for early supported discharge monitoring for COVID-19 positive patients and follow up for long-term complications.
The COVID-19 follow up clinic cohort (n = 30) is ideally placed to evaluate this technology; from a quantitative point of view, a significant number of these patients will likely experience desaturation during the 6MWT, serving as an opportunity to test the device's ability to detect this. Qualitatively, since many of these patients will have either had hospital admissions or home monitoring -- and therefore lived experience of SpO2 monitoring -- they are uniquely qualified to offer valuable input on user experience. Similarly, the chronic lung disease cohort (n = 30), attending for routine OP baseline SpO2 measurement and 6MWT are also well-suited to this study.
In the long term this technology will also prove an invaluable device for the home monitoring of respiratory patients. Beyond the current pandemic there exists a general quest for more personalised health data, such that pulse oximetry measurement of capillary oxygen saturation (SpO2) will likely expand into both the clinical and consumer market of wearable health technology in the near future.
Conditions
See the medical conditions and disease areas that this research is targeting or investigating.
Study Design
Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.
COHORT
PROSPECTIVE
Study Groups
Review each arm or cohort in the study, along with the interventions and objectives associated with them.
COVID-19 Follow up clinic
Patients attending routine post-COVID-19 follow up clinic
EarSats Pulse Oximeter Probe
Off-the-shelf pulse oximeter embedded into a memory foam ear plug for measuring blood oxygen saturation (SpO2) from the ear canal; in-ear SpO2 measured at-rest and during 6 minute walk test
Pulmonary Disease clinic
Patients attending routine outpatient appointment for pulmonary disease
EarSats Pulse Oximeter Probe
Off-the-shelf pulse oximeter embedded into a memory foam ear plug for measuring blood oxygen saturation (SpO2) from the ear canal; in-ear SpO2 measured at-rest and during 6 minute walk test
Interventions
Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.
EarSats Pulse Oximeter Probe
Off-the-shelf pulse oximeter embedded into a memory foam ear plug for measuring blood oxygen saturation (SpO2) from the ear canal; in-ear SpO2 measured at-rest and during 6 minute walk test
Eligibility Criteria
Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.
Inclusion Criteria
* Able to give informed consent (where not able to use CIE, paper consent form will be made available)
Exclusion Criteria
18 Years
ALL
No
Sponsors
Meet the organizations funding or collaborating on the study and learn about their roles.
Imperial College London
OTHER
Responsible Party
Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.
Principal Investigators
Learn about the lead researchers overseeing the trial and their institutional affiliations.
Nicholas Peters, MD
Role: PRINCIPAL_INVESTIGATOR
Imperial College London
Locations
Explore where the study is taking place and check the recruitment status at each participating site.
Imperial College Healthcare NHS Trust
London, Non-US/Non-Canadian, United Kingdom
Countries
Review the countries where the study has at least one active or historical site.
References
Explore related publications, articles, or registry entries linked to this study.
Goverdovsky V, von Rosenberg W, Nakamura T, Looney D, Sharp DJ, Papavassiliou C, Morrell MJ, Mandic DP. Hearables: Multimodal physiological in-ear sensing. Sci Rep. 2017 Jul 31;7(1):6948. doi: 10.1038/s41598-017-06925-2.
Other Identifiers
Review additional registry numbers or institutional identifiers associated with this trial.
285603
Identifier Type: -
Identifier Source: org_study_id
More Related Trials
Additional clinical trials that may be relevant based on similarity analysis.