Trial Outcomes & Findings for The Effects of Filter During CPET on WOB and Aerosol Particle Concentrations (NCT NCT04526925)
NCT ID: NCT04526925
Last Updated: 2023-04-26
Results Overview
The peak aerosol particle concentrations between the situations with and without an in-line filter are reported. The peak concentrations of aerosol particles generated by subjects during CPET, are measured as particles per cubic centimeters of room air. The device measured the concentrations every 6 seconds, it continuously recorded the data for 45 minutes since CPET started and peak concentrations were reported.
TERMINATED
NA
10 participants
45 minutes
2023-04-26
Participant Flow
This is a randomized crossover study, all the 10 participants performed cardiopulmonary exercise test (CPET). During the test, they will be randomly assigned to Standard CPET Without Filter first, then Standard CPET With Filter, or Standard CPET With Filter first, then Standard CPET Without Filter. The time period between two tests were 48 hours.
Participant milestones
| Measure |
CPET Without a Filter First Then CPET With a Filter
Standard CPET will be performed without an in-line filter first then CPET will be performed with an in-line filter
The filter is connected to the mouthpiece during CPET
|
CPET With a Filter First Then CPET Without a Filter
Standard CPET will be performed with an in-line filter first then CPET will be performed without an in-line filter
|
|---|---|---|
|
First Intervention
STARTED
|
4
|
6
|
|
First Intervention
COMPLETED
|
4
|
6
|
|
First Intervention
NOT COMPLETED
|
0
|
0
|
|
Second Intervention
STARTED
|
4
|
6
|
|
Second Intervention
COMPLETED
|
4
|
6
|
|
Second Intervention
NOT COMPLETED
|
0
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
The Effects of Filter During CPET on WOB and Aerosol Particle Concentrations
Baseline characteristics by cohort
| Measure |
CPET Without a Filter First Then CPET With a Filter
n=4 Participants
Standard CPET will be performed without an inline filter first then CPET will be performed with an inline filter.
|
CPET With a Filter First Then CPET Without a Filter
n=6 Participants
Standard CPET will be performed with an inline filter first then CPET will be performed without an inline filter.
|
Total
n=10 Participants
Total of all reporting groups
|
|---|---|---|---|
|
Age, Continuous
|
31.8 years
STANDARD_DEVIATION 7.4 • n=5 Participants
|
32.8 years
STANDARD_DEVIATION 9.2 • n=7 Participants
|
32.4 years
STANDARD_DEVIATION 8.1 • n=5 Participants
|
|
Sex: Female, Male
Female
|
2 Participants
n=5 Participants
|
3 Participants
n=7 Participants
|
5 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
2 Participants
n=5 Participants
|
3 Participants
n=7 Participants
|
5 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Caucasian
|
2 Participants
n=5 Participants
|
3 Participants
n=7 Participants
|
5 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Asian
|
2 Participants
n=5 Participants
|
2 Participants
n=7 Participants
|
4 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Hispanic
|
0 Participants
n=5 Participants
|
1 Participants
n=7 Participants
|
1 Participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 45 minutesThe peak aerosol particle concentrations between the situations with and without an in-line filter are reported. The peak concentrations of aerosol particles generated by subjects during CPET, are measured as particles per cubic centimeters of room air. The device measured the concentrations every 6 seconds, it continuously recorded the data for 45 minutes since CPET started and peak concentrations were reported.
Outcome measures
| Measure |
Standard CPET
n=10 Participants
Standard CPET will be performed without in-line filter
Standard mouthpiece: The mouthpiece is the interface that is placed on participant's face in order to measure all the breathing physilogic responses
|
Standard CPET With In-line Filter
n=10 Participants
An in-line filter will be placed on the mouthpiece during standard CPET
Respiratory filter in-line placed with the standard mouthpiece: The respiratory filter is a device that removes solid/large particles from gas
|
|---|---|---|
|
The Peak Aerosol Particle Concentrations During CPET With a Filter and Without a Filter
|
204.3 particles per cubic centimeters
Standard Deviation 50.6
|
197.1 particles per cubic centimeters
Standard Deviation 4.6
|
PRIMARY outcome
Timeframe: 45 minutesDuring CPET, participants' breathing volume (minute ventilation) is continuously monitored. The peak minute ventilation with and without a filter is reported.
Outcome measures
| Measure |
Standard CPET
n=10 Participants
Standard CPET will be performed without in-line filter
Standard mouthpiece: The mouthpiece is the interface that is placed on participant's face in order to measure all the breathing physilogic responses
|
Standard CPET With In-line Filter
n=10 Participants
An in-line filter will be placed on the mouthpiece during standard CPET
Respiratory filter in-line placed with the standard mouthpiece: The respiratory filter is a device that removes solid/large particles from gas
|
|---|---|---|
|
The Peak Minute Ventilation With and Without a Filter
|
86.1 liter per minute
Standard Deviation 22.7
|
88.1 liter per minute
Standard Deviation 21.8
|
PRIMARY outcome
Timeframe: 45 minutesDuring CPET, participants' oxygen consumption (namely, oxygen uptake) is continuously monitored. The peak oxygen uptake with and without an inline filter are reported.
Outcome measures
| Measure |
Standard CPET
n=10 Participants
Standard CPET will be performed without in-line filter
Standard mouthpiece: The mouthpiece is the interface that is placed on participant's face in order to measure all the breathing physilogic responses
|
Standard CPET With In-line Filter
n=10 Participants
An in-line filter will be placed on the mouthpiece during standard CPET
Respiratory filter in-line placed with the standard mouthpiece: The respiratory filter is a device that removes solid/large particles from gas
|
|---|---|---|
|
The Peak Oxygen Uptake With and Without a Filter
|
2.37 liter per minute
Standard Deviation 0.75
|
2.42 liter per minute
Standard Deviation 0.75
|
PRIMARY outcome
Timeframe: 45 minutesDuring CPET, participants' exhaled volume of carbon dioxide (CO2) (namely, CO2 output) is continuously monitored. The peak CO2 output with and without an inline filter are reported.
Outcome measures
| Measure |
Standard CPET
n=10 Participants
Standard CPET will be performed without in-line filter
Standard mouthpiece: The mouthpiece is the interface that is placed on participant's face in order to measure all the breathing physilogic responses
|
Standard CPET With In-line Filter
n=10 Participants
An in-line filter will be placed on the mouthpiece during standard CPET
Respiratory filter in-line placed with the standard mouthpiece: The respiratory filter is a device that removes solid/large particles from gas
|
|---|---|---|
|
The Peak Carbon Dioxide Output With and Without a Filter
|
2.84 liter per minute
Standard Deviation 0.85
|
2.80 liter per minute
Standard Deviation 0.80
|
SECONDARY outcome
Timeframe: 45 minutesDuring CPET, patients' heart rate is continuously monitored, and the peak heart rate with and without an inline filter are reported.
Outcome measures
| Measure |
Standard CPET
n=10 Participants
Standard CPET will be performed without in-line filter
Standard mouthpiece: The mouthpiece is the interface that is placed on participant's face in order to measure all the breathing physilogic responses
|
Standard CPET With In-line Filter
n=10 Participants
An in-line filter will be placed on the mouthpiece during standard CPET
Respiratory filter in-line placed with the standard mouthpiece: The respiratory filter is a device that removes solid/large particles from gas
|
|---|---|---|
|
The Peak Heart Rate With and Without a Filter
|
164.7 beats per minute
Standard Deviation 11.1
|
167.1 beats per minute
Standard Deviation 9.2
|
Adverse Events
Standard CPET
Standard CPET With In-line Filter
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place