Trial Outcomes & Findings for The Effects of Filter During CPET on WOB and Aerosol Particle Concentrations (NCT NCT04526925)

NCT ID: NCT04526925

Last Updated: 2023-04-26

Results Overview

The peak aerosol particle concentrations between the situations with and without an in-line filter are reported. The peak concentrations of aerosol particles generated by subjects during CPET, are measured as particles per cubic centimeters of room air. The device measured the concentrations every 6 seconds, it continuously recorded the data for 45 minutes since CPET started and peak concentrations were reported.

Recruitment status

TERMINATED

Study phase

NA

Target enrollment

10 participants

Primary outcome timeframe

45 minutes

Results posted on

2023-04-26

Participant Flow

This is a randomized crossover study, all the 10 participants performed cardiopulmonary exercise test (CPET). During the test, they will be randomly assigned to Standard CPET Without Filter first, then Standard CPET With Filter, or Standard CPET With Filter first, then Standard CPET Without Filter. The time period between two tests were 48 hours.

Participant milestones

Participant milestones
Measure
CPET Without a Filter First Then CPET With a Filter
Standard CPET will be performed without an in-line filter first then CPET will be performed with an in-line filter The filter is connected to the mouthpiece during CPET
CPET With a Filter First Then CPET Without a Filter
Standard CPET will be performed with an in-line filter first then CPET will be performed without an in-line filter
First Intervention
STARTED
4
6
First Intervention
COMPLETED
4
6
First Intervention
NOT COMPLETED
0
0
Second Intervention
STARTED
4
6
Second Intervention
COMPLETED
4
6
Second Intervention
NOT COMPLETED
0
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

The Effects of Filter During CPET on WOB and Aerosol Particle Concentrations

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
CPET Without a Filter First Then CPET With a Filter
n=4 Participants
Standard CPET will be performed without an inline filter first then CPET will be performed with an inline filter.
CPET With a Filter First Then CPET Without a Filter
n=6 Participants
Standard CPET will be performed with an inline filter first then CPET will be performed without an inline filter.
Total
n=10 Participants
Total of all reporting groups
Age, Continuous
31.8 years
STANDARD_DEVIATION 7.4 • n=5 Participants
32.8 years
STANDARD_DEVIATION 9.2 • n=7 Participants
32.4 years
STANDARD_DEVIATION 8.1 • n=5 Participants
Sex: Female, Male
Female
2 Participants
n=5 Participants
3 Participants
n=7 Participants
5 Participants
n=5 Participants
Sex: Female, Male
Male
2 Participants
n=5 Participants
3 Participants
n=7 Participants
5 Participants
n=5 Participants
Race/Ethnicity, Customized
Caucasian
2 Participants
n=5 Participants
3 Participants
n=7 Participants
5 Participants
n=5 Participants
Race/Ethnicity, Customized
Asian
2 Participants
n=5 Participants
2 Participants
n=7 Participants
4 Participants
n=5 Participants
Race/Ethnicity, Customized
Hispanic
0 Participants
n=5 Participants
1 Participants
n=7 Participants
1 Participants
n=5 Participants

PRIMARY outcome

Timeframe: 45 minutes

The peak aerosol particle concentrations between the situations with and without an in-line filter are reported. The peak concentrations of aerosol particles generated by subjects during CPET, are measured as particles per cubic centimeters of room air. The device measured the concentrations every 6 seconds, it continuously recorded the data for 45 minutes since CPET started and peak concentrations were reported.

Outcome measures

Outcome measures
Measure
Standard CPET
n=10 Participants
Standard CPET will be performed without in-line filter Standard mouthpiece: The mouthpiece is the interface that is placed on participant's face in order to measure all the breathing physilogic responses
Standard CPET With In-line Filter
n=10 Participants
An in-line filter will be placed on the mouthpiece during standard CPET Respiratory filter in-line placed with the standard mouthpiece: The respiratory filter is a device that removes solid/large particles from gas
The Peak Aerosol Particle Concentrations During CPET With a Filter and Without a Filter
204.3 particles per cubic centimeters
Standard Deviation 50.6
197.1 particles per cubic centimeters
Standard Deviation 4.6

PRIMARY outcome

Timeframe: 45 minutes

During CPET, participants' breathing volume (minute ventilation) is continuously monitored. The peak minute ventilation with and without a filter is reported.

Outcome measures

Outcome measures
Measure
Standard CPET
n=10 Participants
Standard CPET will be performed without in-line filter Standard mouthpiece: The mouthpiece is the interface that is placed on participant's face in order to measure all the breathing physilogic responses
Standard CPET With In-line Filter
n=10 Participants
An in-line filter will be placed on the mouthpiece during standard CPET Respiratory filter in-line placed with the standard mouthpiece: The respiratory filter is a device that removes solid/large particles from gas
The Peak Minute Ventilation With and Without a Filter
86.1 liter per minute
Standard Deviation 22.7
88.1 liter per minute
Standard Deviation 21.8

PRIMARY outcome

Timeframe: 45 minutes

During CPET, participants' oxygen consumption (namely, oxygen uptake) is continuously monitored. The peak oxygen uptake with and without an inline filter are reported.

Outcome measures

Outcome measures
Measure
Standard CPET
n=10 Participants
Standard CPET will be performed without in-line filter Standard mouthpiece: The mouthpiece is the interface that is placed on participant's face in order to measure all the breathing physilogic responses
Standard CPET With In-line Filter
n=10 Participants
An in-line filter will be placed on the mouthpiece during standard CPET Respiratory filter in-line placed with the standard mouthpiece: The respiratory filter is a device that removes solid/large particles from gas
The Peak Oxygen Uptake With and Without a Filter
2.37 liter per minute
Standard Deviation 0.75
2.42 liter per minute
Standard Deviation 0.75

PRIMARY outcome

Timeframe: 45 minutes

During CPET, participants' exhaled volume of carbon dioxide (CO2) (namely, CO2 output) is continuously monitored. The peak CO2 output with and without an inline filter are reported.

Outcome measures

Outcome measures
Measure
Standard CPET
n=10 Participants
Standard CPET will be performed without in-line filter Standard mouthpiece: The mouthpiece is the interface that is placed on participant's face in order to measure all the breathing physilogic responses
Standard CPET With In-line Filter
n=10 Participants
An in-line filter will be placed on the mouthpiece during standard CPET Respiratory filter in-line placed with the standard mouthpiece: The respiratory filter is a device that removes solid/large particles from gas
The Peak Carbon Dioxide Output With and Without a Filter
2.84 liter per minute
Standard Deviation 0.85
2.80 liter per minute
Standard Deviation 0.80

SECONDARY outcome

Timeframe: 45 minutes

During CPET, patients' heart rate is continuously monitored, and the peak heart rate with and without an inline filter are reported.

Outcome measures

Outcome measures
Measure
Standard CPET
n=10 Participants
Standard CPET will be performed without in-line filter Standard mouthpiece: The mouthpiece is the interface that is placed on participant's face in order to measure all the breathing physilogic responses
Standard CPET With In-line Filter
n=10 Participants
An in-line filter will be placed on the mouthpiece during standard CPET Respiratory filter in-line placed with the standard mouthpiece: The respiratory filter is a device that removes solid/large particles from gas
The Peak Heart Rate With and Without a Filter
164.7 beats per minute
Standard Deviation 11.1
167.1 beats per minute
Standard Deviation 9.2

Adverse Events

Standard CPET

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Standard CPET With In-line Filter

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information