Trial Outcomes & Findings for Comparison of Non-invasive and Invasive Blood Pressure Monitors (NCT NCT04506775)
NCT ID: NCT04506775
Last Updated: 2025-09-10
Results Overview
255 time points compared across continuous blood pressure (mmHg) As per the FDA standard ANSI/AAMI SP10:2008 or the newly adopted ISO 81060-3:2022 standard, the mean error should be \<5mmHg or \<6mmHg and the standard deviation should be \<8mmHg or \<10mmHg respectively.
COMPLETED
NA
30 participants
Up to 24 Hours
2025-09-10
Participant Flow
Unit of analysis: Wrist
Participant milestones
| Measure |
All Study Participants
Wrist one will have the ViTrack wrist cuff on one wrist. ViTrack: ViTrack is a continuous non-invasive wearable blood pressure cuff.
Wrist two with the radial artery catheter in the opposite wrist. Radial artery catheterization: Radial artery catheterization is the standard of care for measuring blood pressure continuously throughout surgery.
|
|---|---|
|
Overall Study
STARTED
|
30 30
|
|
Overall Study
Wrist 1 (Control)
|
30 30
|
|
Overall Study
Wrist 2 (ViTrack)
|
30 30
|
|
Overall Study
COMPLETED
|
30 30
|
|
Overall Study
NOT COMPLETED
|
0 0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Race and Ethnicity were not collected from any participant.
Baseline characteristics by cohort
| Measure |
All Study Participants
n=30 Participants
Wrist one will have both the ViTrack wrist cuff on one wrist. ViTrack: ViTrack is a continuous non-invasive wearable blood pressure cuff.
Wrist two and the radial artery catheter in the opposite wrist. Radial artery catheterization: Radial artery catheterization is the standard of care for measuring blood pressure continuously throughout surgery.
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=30 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
9 Participants
n=30 Participants
|
|
Age, Categorical
>=65 years
|
21 Participants
n=30 Participants
|
|
Age, Continuous
|
69.4 Years
n=30 Participants
|
|
Sex: Female, Male
Female
|
8 Participants
n=30 Participants
|
|
Sex: Female, Male
Male
|
22 Participants
n=30 Participants
|
|
Region of Enrollment
United States
|
30 participants
n=30 Participants
|
PRIMARY outcome
Timeframe: Up to 24 HoursPopulation: Wrist 1, the ViTrack wrist cuff represents the experimental arm of the study. Wrist 2, the radial arterial catheter, represents the gold standard arterial blood pressure monitor that the ViTrack device is being compared to. The outcome DBP for the ViTrack device reports the mean error and standard deviation for the ViTrack device in comparison to the gold standard in alignment with the FDA standards.
255 time points compared across continuous blood pressure (mmHg) As per the FDA standard ANSI/AAMI SP10:2008 or the newly adopted ISO 81060-3:2022 standard, the mean error should be \<5mmHg or \<6mmHg and the standard deviation should be \<8mmHg or \<10mmHg respectively.
Outcome measures
| Measure |
All Study Participants
n=30 Participants
Wrist one will have both the ViTrack wrist cuff on one wrist. ViTrack: ViTrack is a continuous non-invasive wearable blood pressure cuff.
Wrist two and the radial artery catheter in the opposite wrist.
Radial artery catheterization: Radial artery catheterization is the standard of care for measuring blood pressure continuously throughout surgery.
|
|---|---|
|
Comparison of Diastolic Blood Pressure
|
0.90 mmHg
Standard Deviation 5.97
|
PRIMARY outcome
Timeframe: Up to 24 HoursPopulation: Wrist 1, the ViTrack wrist cuff represents the experimental arm of the study. Wrist 2, the radial arterial catheter, represents the gold standard arterial blood pressure monitor that the ViTrack device is being compared to. The outcome SBP for the ViTrack device reports the mean error and standard deviation for the ViTrack device in comparison to the gold standard in alignment with the FDA standards.
255 time points compared across continuous blood pressure (mmHg) As per the FDA standard ANSI/AAMI SP10:2008 or the newly adopted ISO 81060-3:2022 standard, the mean error should be \<5mmHg or \<6mmHg and the standard deviation should be \<8mmHg or \<10mmHg respectively.
Outcome measures
| Measure |
All Study Participants
n=30 Participants
Wrist one will have both the ViTrack wrist cuff on one wrist. ViTrack: ViTrack is a continuous non-invasive wearable blood pressure cuff.
Wrist two and the radial artery catheter in the opposite wrist.
Radial artery catheterization: Radial artery catheterization is the standard of care for measuring blood pressure continuously throughout surgery.
|
|---|---|
|
Comparison of Systolic Blood Pressure
|
0.19 mmHg
Standard Deviation 5.70
|
SECONDARY outcome
Timeframe: Up to 72 hoursExamining for local skin irritation
Outcome measures
| Measure |
All Study Participants
n=30 Participants
Wrist one will have both the ViTrack wrist cuff on one wrist. ViTrack: ViTrack is a continuous non-invasive wearable blood pressure cuff.
Wrist two and the radial artery catheter in the opposite wrist.
Radial artery catheterization: Radial artery catheterization is the standard of care for measuring blood pressure continuously throughout surgery.
|
|---|---|
|
Skin Irritation
|
0 Participants
|
Adverse Events
All Study Participants
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place