Trial Outcomes & Findings for Noninvasive Hemoglobin Repeatability and Reproducibility in a Blood Donation Setting (NCT NCT04490863)

NCT ID: NCT04490863

Last Updated: 2021-08-31

Results Overview

Coefficient of variation will be computed using Repeatability and Reproducibility variance.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

26 participants

Primary outcome timeframe

Approximately 90 minutes per subject

Results posted on

2021-08-31

Participant Flow

Participant milestones

Participant milestones
Measure
Test Subjects
All subjects are enrolled and receive Rad-67 Pulse oximeter \& DCI Mini sensor for measurement of hemoglobin. Rad-67 Pulse oximeter \& DCI Mini sensor: Noninvasive pulse oximeter that measures hemoglobin
Overall Study
STARTED
26
Overall Study
COMPLETED
26
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Noninvasive Hemoglobin Repeatability and Reproducibility in a Blood Donation Setting

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Test Subjects
n=26 Participants
All subjects are enrolled and receive Rad-67 Pulse oximeter \& DCI Mini sensor for measurement of hemoglobin. Rad-67 Pulse oximeter \& DCI Mini sensor: Noninvasive pulse oximeter that measures hemoglobin
Age, Categorical
<=18 years
1 Participants
n=93 Participants
Age, Categorical
Between 18 and 65 years
23 Participants
n=93 Participants
Age, Categorical
>=65 years
2 Participants
n=93 Participants
Sex: Female, Male
Female
15 Participants
n=93 Participants
Sex: Female, Male
Male
11 Participants
n=93 Participants
Race/Ethnicity, Customized
Race · Asian
2 Participants
n=93 Participants
Race/Ethnicity, Customized
Race · Native Hawaiian or Other Pacific Islander
1 Participants
n=93 Participants
Race/Ethnicity, Customized
Race · Black or African American
1 Participants
n=93 Participants
Race/Ethnicity, Customized
Race · White
12 Participants
n=93 Participants
Race/Ethnicity, Customized
Race · More Than One Race
1 Participants
n=93 Participants
Race/Ethnicity, Customized
Race · Unknown or Not Reported
1 Participants
n=93 Participants
Race/Ethnicity, Customized
Race · Mexican
7 Participants
n=93 Participants
Race/Ethnicity, Customized
Race · Middle Eastern
1 Participants
n=93 Participants
Region of Enrollment
United States
26 participants
n=93 Participants

PRIMARY outcome

Timeframe: Approximately 90 minutes per subject

Coefficient of variation will be computed using Repeatability and Reproducibility variance.

Outcome measures

Outcome measures
Measure
Test Subjects
n=26 Participants
All subjects are enrolled and receive Rad-67 Pulse oximeter \& DCI Mini sensor for measurement of hemoglobin. Rad-67 Pulse oximeter \& DCI Mini sensor: Noninvasive pulse oximeter that measures hemoglobin
Coefficient of Variation of Rad-67
4.50 Coefficient of Variation percentage

Adverse Events

Test Subjects

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Vikram Ramakanth

Masimo Corporation

Phone: 949-297-7000

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place

Restriction type: GT60